Post-market Clinical Follow-up of a Magnetic Bone Conduction Implant (Cochlear Baha Attract System)

Post-market Clinical Follow-up of a Magnetic Bone Conduction Implant (Cochlear Baha Attract System)

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02022085
Enrollment
54
Registered
2013-12-27
Start date
2014-06-30
Completion date
2017-11-30
Last updated
2018-10-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Deafness, Hearing Loss, Hearing Loss, Conductive

Brief summary

The rationale behind this post-market clinical follow-up investigation is to collect data regarding the usability and clinical performance of the Baha Attract System in subjects with hearing impairment that are candidates for Baha surgery: * to evaluate the efficacy of the Baha Attract System in terms of hearing performance compared to the unaided situation and compared to a pre-operative test situation using the sound processor on a Baha Softband; * to evaluate the mid- and long-term safety of the Baha Attract System.

Detailed description

This investigation was designed to collect data regarding the usability and clinical performance of the Baha Attract System in subjects with hearing impairment that were candidates for Baha surgery. The investigation was performed in an open design since it was not possible to perform the investigation in a blinded fashion. The main evaluations of the investigation, i.e. free-field hearing tests, are standard audiological procedures that are routinely used in daily clinical practice at hospitals worldwide for evaluation of hearing performance in hearing impaired patients. The primary efficacy evaluation, audiometric thresholds, renders comparable results across different sites and countries as it is not a language-specific test. Language-specific audiological tests were performed as secondary efficacy evaluations; to enable comparison of data across sites, validated word lists were used at all sites and equivalent speaker configurations (speech from front, noise from behind) and test procedures (same noise and adaptive speech levels, etc) were used. For the primary evaluation in the investigation, the Baha Attract System was compared to the pre-operative unaided situation, as the intended use of the system is to improve hearing performance in patients with conductive or mixed hearing loss or single-sided sensorineural deafness. Audiological test results with the Baha Attract System were also compared against results obtained with the same Sound Processor on a Baha Softband. Use of the Sound Processor on a Softband is a standard procedure that is routinely used as a pre-operative test prior to bone conduction hearing implant surgery. The Softband test allows the patient to pre-operatively experience hearing through the Sound Processor to get an indication of the post-operative hearing outcome and to choose a suitable Sound Processor. The generic and hearing-specific quality of life questionnaires used in the investigation (HUI3, APHAB, SSQ) are validated scales that have been frequently used and reported in the literature.

Interventions

The Cochlear Baha Attract System incorporates both implantable and external parts. The implantable parts include the BI300 Implant and the BIM400 Implant Magnet, which is intended to be fixated to the BI300 Implant. The external parts include the Sound Processor Magnet (SP Magnet), which, together with the Implant Magnet, constitute the transcutaneous coupling. The SP Magnet incorporates a soft material (Soft pad) at the tissue facing surface that is designed to distribute the pressure over the skin. The Baha Sound Processor attaches to the SP Magnet via a snap coupling.

Sponsors

Cochlear Bone Anchored Solutions
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

Adult subject, i.e. ≥ 18 years of age Conductive or mixed hearing loss in the ear to be implanted: Bone conduction thresholds with a pure tone average PTA4 of \< 30 dB hearing level (mean of 500, 1000, 2000 and 4000 Hz). OR Single-sided sensorineural deafness (SSD): European sites: Bone conduction thresholds with a pure tone average PTA4 of \< 30dB hearing level (mean of 500, 1000, 2000 and 4000 Hz) in the good ear. US sites: Air conduction thresholds with a pure tone average PTA4 of 20 dB hearing level (mean of 500, 1000, 2000 and 3000 Hz) in the good ear OR Subject is indicated for an AC CROS but-for some reason-cannot or will not use an AC CROS. No previous bone conduction implant on the side of the skull to be implanted. Signed informed consent.

Exclusion criteria

Subjects that are scheduled for simultaneous bilateral implant surgery. The investigation is limited to subjects with unilateral use of the Baha Attract System (however, bilateral hearing loss is not an exclusion criterion). Suitable implant position for the BI300 Implant (4 mm or 3 mm) not found during surgery due to insufficient bone quality and/or bone thickness. Less than 3 mm soft tissue thickness at the planned implant site. Subjects that have received radiation therapy at the same side of the skull where the Baha Attract System will be positioned. Condition that could jeopardise osseointegration and/or wound healing as judged by the investigator (e.g. osteoporosis, psoriasis, use of corticosteroids). Uncontrolled diabetes as judged by the investigator. Condition that may have an impact on the outcome of the investigation as judged by the investigator. Unable to follow investigational procedures (e.g. to complete quality of life scales). Participation in another investigation with pharmaceuticals and/or medical device.

Design outcomes

Primary

MeasureTime frameDescription
Hearing Performance: Threshold Audiometry PTA4: Unaided Versus Baha AttractBaseline (unaided) before surgery, 6, 12 and 24 months after surgeryThe change of hearing performance with the Baha Attract System (aided) at 6, 12 and 24 months from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz). The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractBaseline (unaided) before surgery, 24 months after surgeryThe change of hearing performance with the Baha Attract System (aided) from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz). The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Secondary

MeasureTime frameDescription
Hearing Performance: Threshold Audiometry PTA4: Sound Processor on Softband Versus Baha AttractBaseline before surgery, 6, 12 and 24 months after surgeryThe change of hearing performance with the Baha Attract System (aided) from the aided hearing performance, sound processor on a softband, before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz) The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on SoftbandBaseline (aided) before surgery, 6 months after surgeryThe change of hearing performance with the Baha Attract System (aided) from the aided hearing performance with Sound processor on a Softband before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies 250, 500, 1000, 2000, 3000, 4000, 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
Adaptive Speech Recognition in Noise Ratio: Baha Attract Versus Sound Processor on SoftbandBaseline (aided) before surgery, 6, 12 and 24 months after surgeryThe change of hearing performance with the Baha Attract System compared to the pre-operative aided situation with Sound Processor on a softband; Adaptive Speech recognition in Noise measured as signal to noise ratio. A ratio of 1 reflects the ability to correctly hear sentences at 65 dB, in the presence of 65 dB background noise. A lower ratio than 1 reflects the ability to correctly hear sentences below 65 dB. A ratio higher than 1 reflects the ability to correctly hear sentences presented above 65 dB.
Speech in Quiet: Baha Attract Versus Sound Processor on SoftbandBaseline (aided) before surgery, 6, 12 and 24 months after surgeryThe change of hearing performance with the Baha Attract System at 6 months compared to the pre-operative aided situation with the Sound Processor on a softband; Speech in quiet at 50, 65 and 80dB. Speech in Quiet is a measure of percentage of correct words. A higher score reflects a higher percentage of correct words.
Health Utility Index (HUI)Baseline (unaided) before surgery, 6 and 24 months after surgeryChange of health status and health related quality of life using the generic quality of life scale Health Utilities Index (HUI3) when wearing Baha Attract System compared to the pre-operative unaided situation. A health utility value of 1.00 indicates perfect health while a score of 0.00 indicates death. The change from unaided to aided hearing is presented. A positive value indicates an improved quality of life, a negative value indicates impaired quality of life. HUI score of 1 (maximum) describes a state of perfect health. HUI score of 0 describes a state of dead. A negative score of Comprehensive Health State describes a state worse than dead.
Abbreviated Profile of Hearing Aid Benefit (APHAB)Baseline before surgery, 6 and 24 months after surgeryMeasuring change of Ease of communication, Reverberation, Background noise, Aversiveness and a Global score with the Baha Attract System from the pre-operative unaided situation. The absolute APHAB scale is between 0 and 100%, where 0% indicates no problems and 100% indicates always problem. The change from unaided to aided hearing is presented. A positive value indicates an improvement, a negative value an impairment.
Speech, Spatial and Qualities of Hearing Scale (SSQ)Baseline before surgery, 6 and 24 months after surgeryMeasuring change of speech, spatial and hearing experiences with the Baha Attract System from the pre-operative unaided situation. A scale from 0 to 10, where 0 represents can not hear at all, and 10 hear perfectly. The change from unaided to aided hearing is presented. A positive value indicates improved hearing, a negative value indicates impaired hearing.
Adaptive Speech Recognition in Noise Ratio: Unaided Versus Baha AttractBaseline (unaided) before surgery, 6, 12 and 24 months after surgeryThe change of hearing performance with the Baha Attract System compared to the pre-operative unaided situation; Adaptive speech recognition in noise measured as signal to noise ratio The adaptive speech test in noise was conducted using validated lists of phonetically balanced sentences, with speech presented from the front (0 degrees azimuth) and noise from the back (180 degrees azimuth). The noise was kept constant at 65 dB SPL, and the speech was adapted in 2 dB steps to establish the speech-to-noise ratio (SNR) providing a 50% level of understanding. A ratio of 1 reflects the ability to correctly hear sentences at 65 dB, in the presence of 65 dB background noise. A lower ratio than 1 reflects the ability to correctly hear sentences below 65 dB. A ratio higher than 1 reflects the ability to correctly hear sentences presented above 65 dB. Manchester and Birmingham are excluded from this analysis due to incorrect set up.
Tissue Reduction Performed During SurgeryVisit 2 (surgery)Surgical thinning of the soft tissue flap was advocated when the soft tissue thickness exceeded 6 mm
Implant StabilityVisit 2 (surgery)Implant Stability Quotient - ISQ, a scale from 1 to 100, where 100 represent the highest stability
Magnetic Force4, 6, 12 weeks, 6, 12 and 24 monthsTo investigate if the magnetic force required for sound processor magnet retention will change over time
Sound Processor Magnet Choice4, 6, 12 weeks, 6, 12 and 24 monthsTo investigate how sound processor magnet choice will change over time. Six different magnetic strength could be chosen; SPM 1 had the lowest strength and SPM 6 the the highest.
Pain & DiscomfortWeek 6, Week 12, Month 6, Month 12, Month 24Degree of pain and discomfort.
Numbness When Tested With a PinDay 10, Week 4, Week 6, Week 12, Month 6, Month 12, Month 24Degree of numbness when tested with a pin
Numbness When Tested With a Cotton SwabDay 10, Week 4, Week 6, Week 12, Month 6, Month 12, Month 24Degree of numbness when tested with a Cotton Swab
Time to Perform SurgeryVisit 2 (Surgery)Time of first incision to time of last suture
Speech in Quiet, Baha Attract Versus UnaidedBaseline before surgery, 6, 12 and 24 months after surgeryThe change of hearing performance with the Baha Attract System at 6, 12 and 24 months compared to the pre-operative unaided situation; Speech in quiet at 50, 65 and 80dB. Speech in Quiet is a measure of percentage of correct words. A higher score reflects a higher percentage of correct words.

Countries

Netherlands, Poland, United Kingdom, United States

Participant flow

Participants by arm

ArmCount
Baha Attract System
This transcutaneous solution is based on a magnet coupling using magnets on both side of the skin; * One implant magnet * One external magnet on which a sound processor is attached, i.e. the Sound Processor magnet (SP magnet) Baha Attract System: The Cochlear Baha Attract System incorporates both implantable and external parts. The implantable parts include the BI300 Implant and the BIM400 Implant Magnet, which is intended to be fixated to the BI300 Implant. The external parts include the Sound Processor Magnet (SP Magnet), which, together with the Implant Magnet, constitute the transcutaneous coupling. The SP Magnet incorporates a soft material (Soft pad) at the tissue facing surface that is designed to distribute the pressure over the skin. The Baha Sound Processor attaches to the SP Magnet via a snap coupling.
54
Total54

Baseline characteristics

CharacteristicBaha Attract System
Age, Continuous42.1 years
STANDARD_DEVIATION 13.6
Region of Enrollment
Netherlands
23 participants
Region of Enrollment
Poland
14 participants
Region of Enrollment
United Kingdom
16 participants
Region of Enrollment
United States
1 participants
Sex: Female, Male
Female
33 Participants
Sex: Female, Male
Male
21 Participants
Type of hearing loss
Conductive hearing loss
39 participants
Type of hearing loss
Single-sided sensorineural deafness
15 participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
1 / 54
other
Total, other adverse events
29 / 54
serious
Total, serious adverse events
13 / 54

Outcome results

Primary

Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract

The change of hearing performance with the Baha Attract System (aided) from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry at frequencies of 250, 500, 1000, 2000, 3000, 4000 and 6000 Hz). The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline (unaided) before surgery, 12 months after surgery

Population: Intention-to-Treat population (ITT); includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 250 Hz-18.5 dBStandard Deviation 14.3
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 500 Hz-26.0 dBStandard Deviation 13
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 1000 Hz-26.5 dBStandard Deviation 14.2
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 4000 Hz-12.5 dBStandard Deviation 12.2
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 2000 Hz-18.5 dBStandard Deviation 13.5
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 3000 Hz-16.3 dBStandard Deviation 13.2
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 6000 Hz-7.25 dBStandard Deviation 11.19
Comparison: Baha Attract after 12 months vs Unaided situation Pre-Opp-value: <0.0001Fisher's non-parametric permutation test
Primary

Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract

The change of hearing performance with the Baha Attract System (aided) from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz). The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline (unaided) before surgery, 24 months after surgery

Population: Intention-to-Treat population (ITT); includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 250 Hz-19.9 dBStandard Deviation 11
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 2000 Hz-19.6 dBStandard Deviation 11.8
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 3000 Hz-18.4 dBStandard Deviation 10.9
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 500 Hz-25.3 dBStandard Deviation 9.6
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 1000 Hz-28.0 dBStandard Deviation 12.4
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 4000 Hz13.2 dBStandard Deviation 12
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 6000 Hz-8.67 dBStandard Deviation 8.88
Comparison: Baha Attract after 24 months vs Unaided situation Pre-Opp-value: <0.0001Fisher's non-parametric permutation test
Primary

Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract

The change of hearing performance with the Baha Attract System (aided) from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies 250, 500, 1000, 2000, 3000, 4000 and 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline (unaided) before surgery, 6 months after surgery

Population: Intention-to-Treat population (ITT); includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 500 Hz-26.6 dBStandard Deviation 12.8
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 6000 Hz-7.04 dBStandard Deviation 8.61
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 250 Hz-16.9 dBStandard Deviation 12.9
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 1000 Hz-26.4 dBStandard Deviation 14.1
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 2000 Hz-18.7 dBStandard Deviation 12.1
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 3000 Hz-16.9 dBStandard Deviation 12.7
Baha Attract SystemHearing Performance: Change in Threshold Audiometry: Unaided Versus Baha AttractChange of Threshold value at 4000 Hz-12.3 dBStandard Deviation 10.6
Comparison: Baha Attract after 6 months vs Unaided situation Pre-Opp-value: <0.0001Fisher's non-parametric permutation test
Primary

Hearing Performance: Threshold Audiometry PTA4: Unaided Versus Baha Attract

The change of hearing performance with the Baha Attract System (aided) at 6, 12 and 24 months from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz). The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline (unaided) before surgery, 6, 12 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Threshold Audiometry PTA4: Unaided Versus Baha AttractPTA4 6 months, Baha Attract vs Unaided-21.0 dBStandard Deviation 10.4
Baha Attract SystemHearing Performance: Threshold Audiometry PTA4: Unaided Versus Baha AttractPTA4 24 months, Baha Attract vs Unaided-21.5 dBStandard Deviation 8.3
Baha Attract SystemHearing Performance: Threshold Audiometry PTA4: Unaided Versus Baha AttractPTA4 12 months, Baha Attract vs Baha Attract-20.9 dBStandard Deviation 10.4
Comparison: PTA4 6 months with Baha Attract vs Unaidedp-value: <0.0001Fisher's non-parametric permutation test
Comparison: PTA4 12 months with Baha Attract vs Unaidedp-value: <0.0001Fisher's non-parametric permutation test
Comparison: PTA4 24 months with Baha Attract vs Unaidedp-value: <0.0001Fisher's non-parametric permutation test
Secondary

Abbreviated Profile of Hearing Aid Benefit (APHAB)

Measuring change of Ease of communication, Reverberation, Background noise, Aversiveness and a Global score with the Baha Attract System from the pre-operative unaided situation. The absolute APHAB scale is between 0 and 100%, where 0% indicates no problems and 100% indicates always problem. The change from unaided to aided hearing is presented. A positive value indicates an improvement, a negative value an impairment.

Time frame: Baseline before surgery, 6 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all patients who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)6 months: Background noise26.3 units on scaleStandard Deviation 20.6
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)6 months: Reverberation22.0 units on scaleStandard Deviation 19
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)24 months: Reverberation20.8 units on scaleStandard Deviation 19.9
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)6 months: Ease of communication20.4 units on scaleStandard Deviation 24.9
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)6 months: Aversiveness1.32 units on scaleStandard Deviation 23.45
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)6 months: Global22.9 units on scaleStandard Deviation 18.1
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)24 months: Ease of communication18.7 units on scaleStandard Deviation 26.3
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)24 months: Background noise23.7 units on scaleStandard Deviation 20.4
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)24 months: Aversiveness-0.596 units on scaleStandard Deviation 19.646
Baha Attract SystemAbbreviated Profile of Hearing Aid Benefit (APHAB)24 months: Global21.1 units on scaleStandard Deviation 18.9
Comparison: 6 months: Ease of communicationp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 6 months: Background noisep-value: <0.0001Fisher's non-parametric permutation test
Comparison: 6 months: Reverberationp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 6 months: Aversivenessp-value: 0.69Fisher's non-parametric permutation test
Comparison: 6 months: Globalp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 24 months: Ease of communicationp-value: <0.001Fisher's non-parametric permutation test
Comparison: 24 months: Background noisep-value: <0.001Fisher's non-parametric permutation test
Comparison: 24 months: Reverberationp-value: <0.001Fisher's non-parametric permutation test
Comparison: 24 months: Aversivenessp-value: 0.84Fisher's non-parametric permutation test
Comparison: 24 months: Globalp-value: <0.001Fisher's non-parametric permutation test
Secondary

Adaptive Speech Recognition in Noise Ratio: Baha Attract Versus Sound Processor on Softband

The change of hearing performance with the Baha Attract System compared to the pre-operative aided situation with Sound Processor on a softband; Adaptive Speech recognition in Noise measured as signal to noise ratio. A ratio of 1 reflects the ability to correctly hear sentences at 65 dB, in the presence of 65 dB background noise. A lower ratio than 1 reflects the ability to correctly hear sentences below 65 dB. A ratio higher than 1 reflects the ability to correctly hear sentences presented above 65 dB.

Time frame: Baseline (aided) before surgery, 6, 12 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all patient who received the surgical intervention.~Speech in Noise analysis excludes patients from Birmingham and Manchester due to incorrect set up.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemAdaptive Speech Recognition in Noise Ratio: Baha Attract Versus Sound Processor on Softband6 months: Speech in Noise-0.508 change in SNR dBStandard Deviation 4.173
Baha Attract SystemAdaptive Speech Recognition in Noise Ratio: Baha Attract Versus Sound Processor on Softband12 months: Speech in Noise-1.05 change in SNR dBStandard Deviation 4.75
Baha Attract SystemAdaptive Speech Recognition in Noise Ratio: Baha Attract Versus Sound Processor on Softband24 months: Speech in Noise-0.739 change in SNR dBStandard Deviation 4.106
Comparison: 6 months: Speech in Noise with Baha Attract vs Softbandp-value: 0.46Fisher's non-parametric permutation test
Comparison: 12 months: Speech in Noise with Baha Attract vs Softbandp-value: 0.19Fisher's non-parametric permutation test
Comparison: 24 months: Speech in Noise with Baha Attract vs Softbandp-value: 0.31Fisher's non-parametric permutation test
Secondary

Adaptive Speech Recognition in Noise Ratio: Unaided Versus Baha Attract

The change of hearing performance with the Baha Attract System compared to the pre-operative unaided situation; Adaptive speech recognition in noise measured as signal to noise ratio The adaptive speech test in noise was conducted using validated lists of phonetically balanced sentences, with speech presented from the front (0 degrees azimuth) and noise from the back (180 degrees azimuth). The noise was kept constant at 65 dB SPL, and the speech was adapted in 2 dB steps to establish the speech-to-noise ratio (SNR) providing a 50% level of understanding. A ratio of 1 reflects the ability to correctly hear sentences at 65 dB, in the presence of 65 dB background noise. A lower ratio than 1 reflects the ability to correctly hear sentences below 65 dB. A ratio higher than 1 reflects the ability to correctly hear sentences presented above 65 dB. Manchester and Birmingham are excluded from this analysis due to incorrect set up.

Time frame: Baseline (unaided) before surgery, 6, 12 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all patient who received the surgical intervention.~Adaptive Speech in Noise analysis excluded subjects from Birmingham and Manchester clinics due to incorrect set up.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemAdaptive Speech Recognition in Noise Ratio: Unaided Versus Baha AttractSNR 6 months, Baha Attract vs Unaided-4.30 dBStandard Deviation 5.64
Baha Attract SystemAdaptive Speech Recognition in Noise Ratio: Unaided Versus Baha AttractSNR 24 months, Baha Attract vs Unaided-4.47 dBStandard Deviation 6.11
Baha Attract SystemAdaptive Speech Recognition in Noise Ratio: Unaided Versus Baha AttractSNR 12 months, Baha Attract vs Unaided4.79 dBStandard Deviation 6.17
Comparison: Speech in Noise, 6 months Baha Attract vs Unaidedp-value: <0.0001Fisher's non-parametric permutation test
Comparison: Speech in Noise, 12 months Baha Attract vs Unaidedp-value: <0.0001Fisher's non-parametric permutation test
Comparison: Speech in Noise, 24 months Baha Attract vs Unaidedp-value: <0.0001Fisher's non-parametric permutation test
Secondary

Health Utility Index (HUI)

Change of health status and health related quality of life using the generic quality of life scale Health Utilities Index (HUI3) when wearing Baha Attract System compared to the pre-operative unaided situation. A health utility value of 1.00 indicates perfect health while a score of 0.00 indicates death. The change from unaided to aided hearing is presented. A positive value indicates an improved quality of life, a negative value indicates impaired quality of life. HUI score of 1 (maximum) describes a state of perfect health. HUI score of 0 describes a state of dead. A negative score of Comprehensive Health State describes a state worse than dead.

Time frame: Baseline (unaided) before surgery, 6 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all patients who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHealth Utility Index (HUI)6 months: Comprehensive health state0.063 units on a scaleStandard Deviation 0.249
Baha Attract SystemHealth Utility Index (HUI)6 months: Vision-0.003 units on a scaleStandard Deviation 0.033
Baha Attract SystemHealth Utility Index (HUI)6 months: Cognition0.005 units on a scaleStandard Deviation 0.158
Baha Attract SystemHealth Utility Index (HUI)6 months: Pain0.020 units on a scaleStandard Deviation 0.123
Baha Attract SystemHealth Utility Index (HUI)24 months: Comprehensive health state0.069 units on a scaleStandard Deviation 0.24
Baha Attract SystemHealth Utility Index (HUI)24 months: Vision0.002 units on a scaleStandard Deviation 0.015
Baha Attract SystemHealth Utility Index (HUI)24 months: Dexterity-0.025 units on a scaleStandard Deviation 0.151
Baha Attract SystemHealth Utility Index (HUI)24 months: Emotion0.014 units on a scaleStandard Deviation 0.089
Baha Attract SystemHealth Utility Index (HUI)24 months: Cognition0.003 units on a scaleStandard Deviation 0.134
Baha Attract SystemHealth Utility Index (HUI)24 months: Pain0.040 units on a scaleStandard Deviation 0.117
Baha Attract SystemHealth Utility Index (HUI)6 months: Hearing0.111 units on a scaleStandard Deviation 0.321
Baha Attract SystemHealth Utility Index (HUI)6 months: Speech0.045 units on a scaleStandard Deviation 0.153
Baha Attract SystemHealth Utility Index (HUI)6 months: Ambulation-0.016 units on a scaleStandard Deviation 0.067
Baha Attract SystemHealth Utility Index (HUI)6 months: Dexterity-0.021 units on a scaleStandard Deviation 0.108
Baha Attract SystemHealth Utility Index (HUI)6 months: Emotion0.017 units on a scaleStandard Deviation 0.152
Baha Attract SystemHealth Utility Index (HUI)24 months: Hearing0.119 units on a scaleStandard Deviation 0.358
Baha Attract SystemHealth Utility Index (HUI)24 months: Speech0.051 units on a scaleStandard Deviation 0.135
Baha Attract SystemHealth Utility Index (HUI)24 months: Ambulation-0.015 units on a scaleStandard Deviation 0.076
Comparison: 6 months change in Comprehensive health statep-value: 0.088Fisher's non-parametric permutation test
Comparison: 6 months change in Visionp-value: 0.88Fisher's non-parametric permutation test
Comparison: 6 months change in Hearingp-value: 0.02Fisher's non-parametric permutation test
Comparison: 6 months change in Speechp-value: 0.039Fisher's non-parametric permutation test
Comparison: 6 months change in Ambulationp-value: 0.25Fisher's non-parametric permutation test
Comparison: 6 months change in Dexterityp-value: 0.38Fisher's non-parametric permutation test
Comparison: 6 months change in Emotionp-value: 0.43Fisher's non-parametric permutation test
Comparison: 6 months change in Cognitionp-value: 0.85Fisher's non-parametric permutation test
Comparison: 6 months change in Painp-value: 0.25Fisher's non-parametric permutation test
Comparison: 24 months change in Comprehensive health statep-value: 0.088Fisher's non-parametric permutation test
Comparison: 24 months change in Visionp-value: 0.63Fisher's non-parametric permutation test
Comparison: 24 months change in Hearingp-value: 0.045Fisher's non-parametric permutation test
Comparison: 24 months change in Speechp-value: 0.016Fisher's non-parametric permutation test
Comparison: 24 months change in Ambulationp-value: 0.25Fisher's non-parametric permutation test
Comparison: 24 months change in Dexterityp-value: 0.5Wilcoxan Signed Rank Test
Comparison: 24 months change in Emotionp-value: 0.29Fisher's non-parametric permutation test
Comparison: 24 months change in Cognitionp-value: 0.92Fisher's non-parametric permutation test
Comparison: 24 months change in Painp-value: 0.02Fisher's non-parametric permutation test
Secondary

Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband

The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance with Sound processor on a Softband before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies: 250, 500, 1000, 2000, 3000, 4000, 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline (aided) before surgery, 12 months after surgery

Population: Intention-to-Treat population (ITT); includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband12 months: Change of Threshold value at 250 Hz-2.55 dBStandard Deviation 11.15
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband12 months: Change of Threshold value at 500 Hz-2.84 dBStandard Deviation 10.74
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband12 months: Change of Threshold value at 3000 Hz4.02 dBStandard Deviation 11.53
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband12 months: Change of Threshold value at 1000 Hz-0.392 dBStandard Deviation 6.989
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband12 months: Change of Threshold value at 2000 Hz0.392 dBStandard Deviation 9.265
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband12 months: Change of Threshold value at 4000 Hz3.63 dBStandard Deviation 11.84
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband12 months: Change of Threshold value at 6000 Hz5.59 dBStandard Deviation 11.52
Comparison: 250Hz: 12 months with Baha Attract vs Softbandp-value: 0.11Fisher's non-parametric permutation test
Comparison: 500Hz: 12 months with Baha Attract vs Softbandp-value: 0.066Fisher's non-parametric permutation test
Comparison: 1000Hz: 12 months with Baha Attract vs Softbandp-value: 0.71Fisher's non-parametric permutation test
Comparison: 2000Hz: 12 months with Baha Attract vs Softbandp-value: 0.78Fisher's non-parametric permutation test
Comparison: 3000Hz: 12 months with Baha Attract vs Softbandp-value: 0.015Fisher's non-parametric permutation test
Comparison: 4000Hz: 12 months with Baha Attract vs Softbandp-value: 0.035Fisher's non-parametric permutation test
Comparison: 6000Hz: 12 months with Baha Attract vs Softbandp-value: 0.0013Fisher's non-parametric permutation test
Secondary

Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband

The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance with Sound processor on a Softband before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies: 250, 500, 1000, 2000, 3000, 4000, 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline (aided) before surgery, 24 months after surgery

Population: Intention-to-Treat population (ITT); includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband24 months: Change of Threshold value at 1000 Hz-1.20 dBStandard Deviation 6.34
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband24 months: Change of Threshold value at 2000 Hz-0.761 dBStandard Deviation 8.69
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband24 months: Change of Threshold value at 6000 Hz5.22 dBStandard Deviation 11.72
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband24 months: Change of Threshold value at 250 Hz-3.67 dBStandard Deviation 8.35
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband24 months: Change of Threshold value at 500 Hz-1.74 dBStandard Deviation 10.23
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband24 months: Change of Threshold value at 3000 Hz2.61 dBStandard Deviation 9.23
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband24 months: Change of Threshold value at 4000 Hz3.15 dBStandard Deviation 11.17
Comparison: 250Hz: 24 months with Baha Attract vs Softbandp-value: 0.0051Fisher's non-parametric permutation test
Comparison: 500Hz: 24 months with Baha Attract vs Softbandp-value: 0.26Fisher's non-parametric permutation test
Comparison: 1000Hz: 24 months with Baha Attract vs Softbandp-value: 0.22Fisher's non-parametric permutation test
Comparison: 2000Hz: 24 months with Baha Attract vs Softbandp-value: 0.57Fisher's non-parametric permutation test
Comparison: 3000Hz: 24 months with Baha Attract vs Softbandp-value: 0.064Fisher's non-parametric permutation test
Comparison: 4000Hz: 24 months with Baha Attract vs Softbandp-value: 0.064Fisher's non-parametric permutation test
Comparison: 6000Hz: 24 months with Baha Attract vs Softbandp-value: 0.0051Fisher's non-parametric permutation test
Secondary

Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband

The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance with Sound processor on a Softband before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies 250, 500, 1000, 2000, 3000, 4000, 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline (aided) before surgery, 6 months after surgery

Population: Intention-to-Treat population (ITT); includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband6 months: Change of Threshold value at 250 Hz-1.30 dBStandard Deviation 9.77
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband6 months: Change of Threshold value at 500 Hz-4.17 dBStandard Deviation 7.99
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband6 months: Change of Threshold value at 1000 Hz-1.11 dBStandard Deviation 6.49
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband6 months: Change of Threshold value at 2000 Hz-0.463 dBStandard Deviation 7.022
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband6 months: Change of Threshold value at 3000 Hz2.69 dBStandard Deviation 9.4
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband6 months: Change of Threshold value at 4000 Hz3.24 dBStandard Deviation 9.07
Baha Attract SystemHearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband6 months: Change of Threshold value at 6000 Hz5.56 dBStandard Deviation 9.6
Comparison: 250Hz: 6 months with Baha Attract vs Softbandp-value: 0.34Fisher's non-parametric permutation test
Comparison: 500Hz: 6 months with Baha Attract vs Softbandp-value: 0.0004Fisher's non-parametric permutation test
Comparison: 1000Hz: 6 months with Baha Attract vs Softbandp-value: 0.22Fisher's non-parametric permutation test
Comparison: 2000Hz: 6 months with Baha Attract vs Softbandp-value: 0.64Fisher's non-parametric permutation test
Comparison: 3000Hz: 6 months with Baha Attract vs Softbandp-value: 0.039Fisher's non-parametric permutation test
Comparison: 4000Hz: 6 months with Baha Attract vs Softbandp-value: 0.012Fisher's non-parametric permutation test
Comparison: 6000Hz: 6 months with Baha Attract vs Softbandp-value: <0.0001Fisher's non-parametric permutation test
Secondary

Hearing Performance: Threshold Audiometry PTA4: Sound Processor on Softband Versus Baha Attract

The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance, sound processor on a softband, before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz) The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Time frame: Baseline before surgery, 6, 12 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all subjects who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemHearing Performance: Threshold Audiometry PTA4: Sound Processor on Softband Versus Baha Attract12 months vs Softband: PTA40.196 dBStandard Deviation 6.719
Baha Attract SystemHearing Performance: Threshold Audiometry PTA4: Sound Processor on Softband Versus Baha Attract24 months vs Softband: PTA4-0.136 dBStandard Deviation 6.472
Baha Attract SystemHearing Performance: Threshold Audiometry PTA4: Sound Processor on Softband Versus Baha Attract6 months vs Softband: PTA4-0.62 dBStandard Deviation 5.138
Comparison: 6 months with Baha Attract vs Softband: change in PTA4p-value: 0.38Fisher's non-parametric permutation test
Comparison: 12 months with Baha Attract vs Softband: change in PTA4p-value: 0.84Fisher's non-parametric permutation test
Comparison: 24 months with Baha Attract vs Softband: change in PTA4p-value: 0.89Fisher's non-parametric permutation test
Secondary

Implant Stability

Implant Stability Quotient - ISQ, a scale from 1 to 100, where 100 represent the highest stability

Time frame: Visit 2 (surgery)

Population: Intention-to-treat (ITT) population; includes all patients who received surgical intervention.

ArmMeasureValue (MEAN)Dispersion
Baha Attract SystemImplant Stability77.3 units on a scaleStandard Deviation 9
Secondary

Magnetic Force

To investigate if the magnetic force required for sound processor magnet retention will change over time

Time frame: 4, 6, 12 weeks, 6, 12 and 24 months

Population: Intention-to-Treat (ITT) population; includes all patients who received surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemMagnetic Force4 weeks1.01 NewtonStandard Deviation 0.4
Baha Attract SystemMagnetic Force6 weeks1.07 NewtonStandard Deviation 0.33
Baha Attract SystemMagnetic Force12 weeks1.15 NewtonStandard Deviation 0.41
Baha Attract SystemMagnetic Force6 months1.16 NewtonStandard Deviation 0.42
Baha Attract SystemMagnetic Force12 months1.15 NewtonStandard Deviation 0.38
Baha Attract SystemMagnetic Force24 months1.09 NewtonStandard Deviation 0.41
Secondary

Numbness When Tested With a Cotton Swab

Degree of numbness when tested with a Cotton Swab

Time frame: Day 10, Week 4, Week 6, Week 12, Month 6, Month 12, Month 24

Population: Safety population; includes all patients who received the surgical intervention.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Baha Attract SystemNumbness When Tested With a Cotton SwabDay 10: Numbness tested with a Cotton SwabNo Numbness13 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabDay 10: Numbness tested with a Cotton SwabNumbness within 2 cm from implant9 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabDay 10: Numbness tested with a Cotton SwabNumbness within & beyond 2cm from Implant32 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 4: Numbness tested with a Cotton SwabNo Numbness18 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 4: Numbness tested with a Cotton SwabNumbness within 2 cm from implant6 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 4: Numbness tested with a Cotton SwabNumbness within & beyond 2cm from Implant30 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 6: Numbness tested with a Cotton SwabNo Numbness23 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 6: Numbness tested with a Cotton SwabNumbness within 2 cm from implant2 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 6: Numbness tested with a Cotton SwabNumbness within & beyond 2cm from Implant28 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 12: Numbness tested with a Cotton SwabNo Numbness31 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 12: Numbness tested with a Cotton SwabNumbness within 2 cm from implant6 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabWeek 12: Numbness tested with a Cotton SwabNumbness within & beyond 2cm from Implant16 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 6: Numbness tested with a Cotton SwabNo Numbness43 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 6: Numbness tested with a Cotton SwabNumbness within 2 cm from implant2 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 6: Numbness tested with a Cotton SwabNumbness within & beyond 2cm from Implant7 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 12: Numbness tested with a Cotton SwabNo Numbness44 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 12: Numbness tested with a Cotton SwabNumbness within 2 cm from implant3 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 12: Numbness tested with a Cotton SwabNumbness within & beyond 2cm from Implant5 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 24: Numbness tested with a Cotton SwabNo Numbness42 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 24: Numbness tested with a Cotton SwabNumbness within 2 cm from implant3 Participants
Baha Attract SystemNumbness When Tested With a Cotton SwabMonth 24: Numbness tested with a Cotton SwabNumbness within & beyond 2cm from Implant1 Participants
Secondary

Numbness When Tested With a Pin

Degree of numbness when tested with a pin

Time frame: Day 10, Week 4, Week 6, Week 12, Month 6, Month 12, Month 24

Population: Safety population; includes all patients who received the surgical intervention.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Baha Attract SystemNumbness When Tested With a PinWeek 12: Numbness tested with a pinNo Numbness33 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 12: Numbness tested with a pinNumbness within 2cm of Implant5 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 12: Numbness tested with a pinNumbness within & beyond 2cm from Implant15 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 6: Numbness tested with a PinNo Numbness42 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 6: Numbness tested with a PinNumbness within 2cm of Implant3 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 6: Numbness tested with a PinNumbness within & beyond 2cm from Implant7 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 12: Numbness tested with a PinNo Numbness45 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 12: Numbness tested with a PinNumbness within 2cm of Implant3 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 12: Numbness tested with a PinNumbness within & beyond 2cm from Implant4 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 24: Numbness tested with a PinNo Numbness45 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 24: Numbness tested with a PinNumbness within 2cm of Implant0 Participants
Baha Attract SystemNumbness When Tested With a PinMonth 24: Numbness tested with a PinNumbness within & beyond 2cm from Implant1 Participants
Baha Attract SystemNumbness When Tested With a PinDay 10: Numbness tested with a PinNo Numbness12 Participants
Baha Attract SystemNumbness When Tested With a PinDay 10: Numbness tested with a PinNumbness within 2cm of Implant9 Participants
Baha Attract SystemNumbness When Tested With a PinDay 10: Numbness tested with a PinNumbness within & beyond 2cm from Implant33 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 4: Numbness tested with a pinNo Numbness15 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 4: Numbness tested with a pinNumbness within 2cm of Implant7 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 4: Numbness tested with a pinNumbness within & beyond 2cm from Implant32 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 6: Numbness tested with a pinNo Numbness21 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 6: Numbness tested with a pinNumbness within 2cm of Implant3 Participants
Baha Attract SystemNumbness When Tested With a PinWeek 6: Numbness tested with a pinNumbness within & beyond 2cm from Implant29 Participants
Secondary

Pain & Discomfort

Degree of pain and discomfort.

Time frame: Week 6, Week 12, Month 6, Month 12, Month 24

Population: Safety population; includes all patients who received the surgical intervention.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Baha Attract SystemPain & DiscomfortWeek 6No Discomfort or Pain21 Participants
Baha Attract SystemPain & DiscomfortWeek 6Slight Discomfort or Pain15 Participants
Baha Attract SystemPain & DiscomfortWeek 6Discomfort or Pain15 Participants
Baha Attract SystemPain & DiscomfortWeek 6Excessive Discomfort or Pain2 Participants
Baha Attract SystemPain & DiscomfortWeek 12No Discomfort or Pain24 Participants
Baha Attract SystemPain & DiscomfortWeek 12Slight Discomfort or Pain16 Participants
Baha Attract SystemPain & DiscomfortWeek 12Discomfort or Pain11 Participants
Baha Attract SystemPain & DiscomfortWeek 12Excessive Discomfort or Pain2 Participants
Baha Attract SystemPain & DiscomfortMonth 6No Discomfort or Pain33 Participants
Baha Attract SystemPain & DiscomfortMonth 6Slight Discomfort or Pain15 Participants
Baha Attract SystemPain & DiscomfortMonth 6Discomfort or Pain5 Participants
Baha Attract SystemPain & DiscomfortMonth 6Excessive Discomfort or Pain0 Participants
Baha Attract SystemPain & DiscomfortMonth 12No Discomfort or Pain37 Participants
Baha Attract SystemPain & DiscomfortMonth 12Slight Discomfort or Pain11 Participants
Baha Attract SystemPain & DiscomfortMonth 12Discomfort or Pain4 Participants
Baha Attract SystemPain & DiscomfortMonth 12Excessive Discomfort or Pain0 Participants
Baha Attract SystemPain & DiscomfortMonth 24No Discomfort or Pain40 Participants
Baha Attract SystemPain & DiscomfortMonth 24Slight Discomfort or Pain4 Participants
Baha Attract SystemPain & DiscomfortMonth 24Discomfort or Pain3 Participants
Baha Attract SystemPain & DiscomfortMonth 24Excessive Discomfort or Pain0 Participants
Secondary

Sound Processor Magnet Choice

To investigate how sound processor magnet choice will change over time. Six different magnetic strength could be chosen; SPM 1 had the lowest strength and SPM 6 the the highest.

Time frame: 4, 6, 12 weeks, 6, 12 and 24 months

Population: Intention-to-Treat (ITT) population; includes all patients who received surgical intervention at week 4. At week 6, 12 and month 6, 12 and 24 one patient had left the study.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Baha Attract SystemSound Processor Magnet ChoiceMonth 24SPM 45 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 24SPM 317 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 24SPM 58 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 4SPM 619 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 4SPM 519 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 4SPM 49 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 4SPM 36 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 4SPM 21 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 4SPM 10 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 6SPM 611 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 6SPM 510 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 6SPM 416 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 6SPM 313 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 6SPM 22 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 6SPM 11 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 12SPM 66 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 12SPM 512 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 12SPM 411 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 12SPM 315 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 12SPM 29 Participants
Baha Attract SystemSound Processor Magnet ChoiceWeek 12SPM 10 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 6SPM 65 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 6SPM 59 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 6SPM 412 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 6SPM 318 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 6SPM 29 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 6SPM 10 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 12SPM 66 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 12SPM 511 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 12SPM 412 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 12SPM 314 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 12SPM 28 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 12SPM 11 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 24SPM 65 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 24SPM 29 Participants
Baha Attract SystemSound Processor Magnet ChoiceMonth 24SPM 12 Participants
Secondary

Speech in Quiet: Baha Attract Versus Sound Processor on Softband

The change of hearing performance with the Baha Attract System at 6 months compared to the pre-operative aided situation with the Sound Processor on a softband; Speech in quiet at 50, 65 and 80dB. Speech in Quiet is a measure of percentage of correct words. A higher score reflects a higher percentage of correct words.

Time frame: Baseline (aided) before surgery, 6, 12 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all patients who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband6 months: Speech in quiet at 50dB2.74 Change of % correct words at the dB leveStandard Deviation 29.94
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband6 months: Speech in quiet at 65dB-2.11 Change of % correct words at the dB leveStandard Deviation 10.62
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband6 months: Speech in quiet at 80dB-0.33 Change of % correct words at the dB leveStandard Deviation 5
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband12 months: Speech in quiet at 65dB-2.69 Change of % correct words at the dB leveStandard Deviation 11.16
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband24 months: Speech in quiet at 65dB-5.67 Change of % correct words at the dB leveStandard Deviation 16.48
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband24 months: Speech in quiet at 80dB-0.696 Change of % correct words at the dB leveStandard Deviation 8.178
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband12 months: Speech in quiet at 50dB0.373 Change of % correct words at the dB leveStandard Deviation 27.201
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband12 months: Speech in quiet at 80dB-0.961 Change of % correct words at the dB leveStandard Deviation 8.457
Baha Attract SystemSpeech in Quiet: Baha Attract Versus Sound Processor on Softband24 months: Speech in quiet at 50dB-11.6 Change of % correct words at the dB leveStandard Deviation 32.6
Comparison: 6 months Speech in Quiet at 50dBp-value: 0.43Fisher's non-parametric permutation test
Comparison: 6 months Speech in Quiet at 65dBp-value: 0.16Fisher's non-parametric permutation test
Comparison: 6 months Speech in Quiet at 80dBp-value: 0.65Fisher's non-parametric permutation test
Comparison: 12 months Speech in Quiet at 50dBp-value: 0.93Fisher's non-parametric permutation test
Comparison: 12 months Speech in Quiet at 65dBp-value: 0.094Fisher's non-parametric permutation test
Comparison: 12 months Speech in Quiet at 80dBp-value: 0.44Fisher's non-parametric permutation test
Comparison: 24 months Speech in Quiet at 50dBp-value: 0.021Fisher's non-parametric permutation test
Comparison: 24 months Speech in Quiet at 65dBp-value: 0.024Fisher's non-parametric permutation test
Comparison: 24 months Speech in Quiet at 80dBp-value: 0.59Fisher's non-parametric permutation test
Secondary

Speech in Quiet, Baha Attract Versus Unaided

The change of hearing performance with the Baha Attract System at 6, 12 and 24 months compared to the pre-operative unaided situation; Speech in quiet at 50, 65 and 80dB. Speech in Quiet is a measure of percentage of correct words. A higher score reflects a higher percentage of correct words.

Time frame: Baseline before surgery, 6, 12 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all patients who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided6 months: Speech in quiet at 50dB52.1 Change of % correct words at the dB leveStandard Deviation 29.2
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided6 months: Speech in quiet at 65dB43.4 Change of % correct words at the dB leveStandard Deviation 31.5
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided6 months: Speech in quiet at 80dB13.3 Change of % correct words at the dB leveStandard Deviation 20.6
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided12 months: Speech in quiet at 50dB50.7 Change of % correct words at the dB leveStandard Deviation 30.2
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided12 months: Speech in quiet at 65dB45.5 Change of % correct words at the dB leveStandard Deviation 31.4
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided12 months: Speech in quiet at 80dB13.9 Change of % correct words at the dB leveStandard Deviation 21.2
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided24 months: Speech in quiet at 50dB40.4 Change of % correct words at the dB leveStandard Deviation 33.3
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided24 months: Speech in quiet at 65dB43.3 Change of % correct words at the dB leveStandard Deviation 29.9
Baha Attract SystemSpeech in Quiet, Baha Attract Versus Unaided24 months: Speech in quiet at 80dB14.0 Change of % correct words at the dB leveStandard Deviation 20.7
Comparison: 6 months: Speech in quiet at 50dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 6 months: Speech in quiet at 65dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 6 months: Speech in quiet at 80dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 12 months: Speech in quiet at 50dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 12 months: Speech in quiet at 65dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 12 months: Speech in quiet at 80dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 24 months: Speech in quiet at 50dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 24 months: Speech in quiet at 65dBp-value: <0.0001Fisher's non-parametric permutation test
Comparison: 24 months: Speech in quiet at 80dBp-value: <0.0001Fisher's non-parametric permutation test
Secondary

Speech, Spatial and Qualities of Hearing Scale (SSQ)

Measuring change of speech, spatial and hearing experiences with the Baha Attract System from the pre-operative unaided situation. A scale from 0 to 10, where 0 represents can not hear at all, and 10 hear perfectly. The change from unaided to aided hearing is presented. A positive value indicates improved hearing, a negative value indicates impaired hearing.

Time frame: Baseline before surgery, 6 and 24 months after surgery

Population: Intention-to-Treat (ITT) population; includes all patient who received the surgical intervention.

ArmMeasureGroupValue (MEAN)Dispersion
Baha Attract SystemSpeech, Spatial and Qualities of Hearing Scale (SSQ)Qualities: Change from pre-op to 24 months1.30 units on scaleStandard Deviation 1.39
Baha Attract SystemSpeech, Spatial and Qualities of Hearing Scale (SSQ)Speech: Change from pre-op to 6 months2.49 units on scaleStandard Deviation 1.67
Baha Attract SystemSpeech, Spatial and Qualities of Hearing Scale (SSQ)Spatial: Change from pre-op to 6 months1.87 units on scaleStandard Deviation 1.9
Baha Attract SystemSpeech, Spatial and Qualities of Hearing Scale (SSQ)Qualities: Change from pre-op to 6 months1.68 units on scaleStandard Deviation 1.36
Baha Attract SystemSpeech, Spatial and Qualities of Hearing Scale (SSQ)Speech: Change from pre-op to 24 months2.07 units on scaleStandard Deviation 1.57
Baha Attract SystemSpeech, Spatial and Qualities of Hearing Scale (SSQ)Spatial: Change from pre-op to 24 months1.36 units on scaleStandard Deviation 1.74
Comparison: Speech: Change from pre-op to 6 monthsp-value: <0.0001Fisher's non-parametric permutation test
Comparison: Spatial: Change from pre-op to 6 monthsp-value: <0.0001Fisher's non-parametric permutation test
Comparison: Qualities: Change from pre-op to 6 monthsp-value: <0.0001Fisher's non-parametric permutation test
Comparison: Speech: Change from pre-op to 24 monthsp-value: <0.0001Fisher's non-parametric permutation test
Comparison: Spatial: Change from pre-op to 24 monthsp-value: <0.0001Fisher's non-parametric permutation test
Comparison: Qualities: Change from pre-op to 24 monthsp-value: <0.0001Fisher's non-parametric permutation test
Secondary

Time to Perform Surgery

Time of first incision to time of last suture

Time frame: Visit 2 (Surgery)

Population: Intention-to-Treat (ITT) population; includes all patients who received the surgical intervention.

ArmMeasureValue (MEAN)Dispersion
Baha Attract SystemTime to Perform Surgery38.7 MinutesStandard Deviation 10.7
Secondary

Tissue Reduction Performed During Surgery

Surgical thinning of the soft tissue flap was advocated when the soft tissue thickness exceeded 6 mm

Time frame: Visit 2 (surgery)

Population: Intention-to-Treat (ITT) population; all patients who received surgical intervention.

ArmMeasureGroupValue (NUMBER)
Baha Attract SystemTissue Reduction Performed During SurgeryTissue thinning12 participants
Baha Attract SystemTissue Reduction Performed During SurgeryNo tissue thinning42 participants