Study Comparing Flexible Spectral-Imaging Color Enhancement (FICE) With Olympus Narrow Band Imaging (NBI)

A Concurrence Study of Diagnostic Visualization Comparing Flexible Spectral-Imaging Color Enhancement (FICE) With Olympus Narrow Band Imaging (NBI)

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02005393
Acronym
FICE
Enrollment
20
Registered
2013-12-09
Start date
2013-10-31
Completion date
2013-11-30
Last updated
2017-03-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Routine Endoscopy

Keywords

FICE, NBI

Brief summary

The study is intended to test the image quality of the FICE system to the predicate device, Narrow Band Imaging (NBI). The hypothesis is the images are of equal quality.

Detailed description

No additional description required.

Interventions

DEVICEFICE Image Acquisition followed by NBI Image Acquisition

Images Captured with FICE Device, followed by same locations imaged with NBI Image Acquisition System

Sponsors

Fujifilm Medical Systems USA, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
21 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Referred for diagnostic endoscopy or colonoscopy * Ability to comply and willingness to partake in study

Exclusion criteria

* less than 21 years of age * is an inmate

Design outcomes

Primary

MeasureTime frameDescription
Image Quality Scores Using Likert ScaleOne dayLikert score ratings of image quality for FICE and NBI images. Each FICE setting was compared to NBI, to determine which FICE settings were equivalent to NBI. Likert Scores: 1. Not seen; 2. poor but usable, characteristic features are detectable but details are not fully reproduced; features just visible; 3. good; allows an adequate assessment, details of anatomical structures are visible but not necessarily clearly defined; details emerging; 4. very good; allows an excellent assesssment, anatomical details are clearly defined; details clear

Countries

United States

Participant flow

Recruitment details

Twenty subjects were recruited at Mayo Clinic, Scottsdale between 23Oct2013 and 08Nov2013 to have two endoscopies completed - one using Fujifilm's endoscope and processor equipped with FICE, and the second immediately following with Olympus' endoscope and processor equipped with NBI.

Participants by arm

ArmCount
FICE Image Acquisition Followed by NBI Image Acquisition
Images Captured with FICE Image Acquisition System immediately followed by Image Acquisition with NBI FICE Image Acquisition System: Images Captured with FICE Device
20
Total20

Baseline characteristics

CharacteristicFICE Image Acquisition Followed by NBI Image Acquisition
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
20 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
20 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race/Ethnicity, Customized
American Indian or Alaska Native
0 participants
Race/Ethnicity, Customized
Asian
3 participants
Race/Ethnicity, Customized
Black or African American
1 participants
Race/Ethnicity, Customized
More than one race
0 participants
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
0 participants
Race/Ethnicity, Customized
Unknown or Not Reported
0 participants
Race/Ethnicity, Customized
White
16 participants
Region of Enrollment
United States
20 participants
Sex: Female, Male
Female
12 Participants
Sex: Female, Male
Male
8 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 20
serious
Total, serious adverse events
0 / 20

Outcome results

Primary

Image Quality Scores Using Likert Scale

Likert score ratings of image quality for FICE and NBI images. Each FICE setting was compared to NBI, to determine which FICE settings were equivalent to NBI. Likert Scores: 1. Not seen; 2. poor but usable, characteristic features are detectable but details are not fully reproduced; features just visible; 3. good; allows an adequate assessment, details of anatomical structures are visible but not necessarily clearly defined; details emerging; 4. very good; allows an excellent assesssment, anatomical details are clearly defined; details clear

Time frame: One day

Population: All patients had imaging done with both FICE, immediately followed by NBI. Images for one subject were not included in the Reader Study due to an error on the Compact Flash card which resulted in three (3) corrupt FICE images (FICE settings 0, 1 \& 2); therefore making 19 out of 20 participants analyzed.

ArmMeasureGroupValue (MEAN)Dispersion
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 03.2 units on a scaleStandard Deviation 0.7
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 13.3 units on a scaleStandard Deviation 0.7
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 23.4 units on a scaleStandard Deviation 0.6
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 33.3 units on a scaleStandard Deviation 0.7
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 43.3 units on a scaleStandard Deviation 0.8
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 53.1 units on a scaleStandard Deviation 0.8
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 93.2 units on a scaleStandard Deviation 0.8
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 63.4 units on a scaleStandard Deviation 0.7
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 73.0 units on a scaleStandard Deviation 0.8
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleFICE 83.1 units on a scaleStandard Deviation 0.7
FICE Image Acquisition Followed by NBI Image AcquisitionImage Quality Scores Using Likert ScaleNBI3.1 units on a scaleStandard Deviation 1.1
Comparison: All 20 subjects enrolled in the study, including Subject 219 that was dropped from the Reader assessments due to corrupted images, were all rated as having similar white light images between the FICE and NBI procedures for each location used in the concurrence study. This assessment was intended to assure that no procedural sequence or visualization bias was introduced into the image acquisition process. The overall Reader mean (SD) for both FICE and NBI were equal to or greater than 3.0 (0.8).