Skip to content

CEP249 Pediatric Accuracy Study

A Performance Evaluation of the Enlite™ and Enlite 3 Glucose Sensor to Support Use in Children

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01991470
Enrollment
186
Registered
2013-11-25
Start date
2015-03-31
Completion date
2016-12-31
Last updated
2018-09-27

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Insulin Requiring Diabetes

Keywords

Insulin requiring diabetes, Children, Sensor, Accuracy

Brief summary

The purpose of this study is to evaluate the Enlite Sensor™ and Enlite 3 Sensor performance in pediatric patients with insulin requiring diabetes at the beginning, middle and end of sensor wear.

Detailed description

This study is a prospective, single arm, multi-center, in-clinic study.. There will be two consecutive phases of the study: A total of up to 260 subjects (2-18 years old) will be enrolled at up to 15 investigational centers Subjects will come for clinic visit for YSI reference glucose testing and for Enlite Sensor™ and Enlite 3 Sensor™ testing. SMBG (Self-Monitoring of Blood Glucose) reference measurement is for subjects 2-6 years of age and cannot tolerate YSI. Study has a run-in period and a study period with a total of 10 visits

Interventions

DEVICEEnlite sensors

Subject wears enlite sensor first and followed by enlite 3 sensor.

Sponsors

Medtronic Diabetes
Lead SponsorINDUSTRY

Study design

Intervention model
SEQUENTIAL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
2 Years to 18 Years
Healthy volunteers
No

Inclusion criteria

1. Subject is 2-18 years of age at time of screening 2. Subject has been diagnosed with insulin requiring diabetes mellitus for at least one year. 3. Subject is willing to perform greater than or equal to 4 finger stick blood glucose measurements daily 4. Subject is willing to perform required sensor calibrations 5. Subject is willing to wear the system (Guardian Mobile application, pumps, sensors, meter) continuously throughout the study 6. Adequate venous access as assessed by investigator or appropriate staff

Exclusion criteria

1. Subject is unable to tolerate tape adhesive in the area of sensor placement 2. Subject has any unresolved adverse skin condition in the area of sensor or device placement (e.g., psoriasis, rash, Staphylococcus infection) 3. Subject is actively participating in an investigational study (drug or device) wherein they have received treatment from an investigational study (drug or device) in the last 2 weeks 4. Females of child-bearing potential who have a positive pregnancy test at screening or plans to become pregnant during the course of the study 5. Subjects with hematocrit lower than the normal age specific reference range per central or local lab testing

Design outcomes

Primary

MeasureTime frameDescription
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump6 daysEnlite Sensor accuracy relative to the YSI (Yellow Spring Instruments) reference. Each subject wore two 530G pumps. The results for the lower serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the primary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System7 daysEnlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG (Self-Monitoring of Blood Glucose) reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

Secondary

MeasureTime frameDescription
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump, Using One Additional Calibration During FST6 DaysEnlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump, Using One Additional Calibration During FST6 DaysEnlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the GST3C Transmitter7 daysEnlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With GST3C Transmitter, Using One Additional Calibration During FST7 DaysEnlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System, Using One Additional Calibration During FST7 DaysEnlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump6 DaysEnlite Sensor accuracy relative to the YSI reference. Each subject wore two 530G pumps. The results for the higher serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the secondary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

Countries

United States

Participant flow

Pre-assignment details

Two phase study: Enlite and Enlite 3. Some participated only Enlite phase and some only participated in Enlite 3 phase.

Participants by arm

ArmCount
2-18 Year Olds Wearing Enlite Sensor
2-18 years old wearing Enlite and/or Enlite3 sensors.
186
Total186

Withdrawals & dropouts

PeriodReasonFG000
Enlite 1 PeriodAdverse Event1
Enlite 1 PeriodInclusion/Exclusion not met1
Enlite 1 PeriodWithdrawal by Subject18
Enlite 3 PeriodDUE TO MULTIPLE SENSOR ISSUES1
Enlite 3 PeriodLost to Follow-up1
Enlite 3 PeriodPhysician Decision2

Baseline characteristics

Characteristic2-18 Year Olds Wearing Enlite Sensor
Age, Continuous13.2 years
STANDARD_DEVIATION 3.79
Region of Enrollment
United States
186 participants
Sex: Female, Male
Female
83 Participants
Sex: Female, Male
Male
103 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 1610 / 158
other
Total, other adverse events
25 / 16123 / 158
serious
Total, serious adverse events
2 / 1611 / 158

Outcome results

Primary

Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System

Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG (Self-Monitoring of Blood Glucose) reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

Time frame: 7 days

Population: a total of 25 subjects do not have device data

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System10.88 percentage of differenceStandard Deviation 10.65
Primary

Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump

Enlite Sensor accuracy relative to the YSI (Yellow Spring Instruments) reference. Each subject wore two 530G pumps. The results for the lower serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the primary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

Time frame: 6 days

Population: All subjects wearing Enlite sensor with lower serial number of 530G pump. A total of 30 subjects did not have device data.

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump15.02 percentage of differenceStandard Deviation 13.37
Secondary

Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With GST3C Transmitter, Using One Additional Calibration During FST

Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

Time frame: 7 Days

Population: A total of 41 subjects did not have device data.

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With GST3C Transmitter, Using One Additional Calibration During FST10.85 percentage of differenceStandard Deviation 10.35
Secondary

Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the GST3C Transmitter

Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

Time frame: 7 days

Population: A total of 41 subjects did not have device data.

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the GST3C Transmitter11.08 percentage of differenceStandard Deviation 10.58
Secondary

Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System, Using One Additional Calibration During FST

Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

Time frame: 7 Days

Population: A total of 34 subjects did not have device data.

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System, Using One Additional Calibration During FST10.05 percentage of differenceStandard Deviation 9.26
Secondary

Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump

Enlite Sensor accuracy relative to the YSI reference. Each subject wore two 530G pumps. The results for the higher serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the secondary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

Time frame: 6 Days

Population: All subjects wearing Enlite sensors with high serial number 530G pump. A total of 35 subjects did not have device data.

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump17.50 percentage of differenceStandard Deviation 19.7
Secondary

Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump, Using One Additional Calibration During FST

Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

Time frame: 6 Days

Population: A total of 34 subjects did not have device data.

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump, Using One Additional Calibration During FST16.94 percentage of differenceStandard Deviation 18.74
Secondary

Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump, Using One Additional Calibration During FST

Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

Time frame: 6 Days

Population: A total of 32 subjects did not have device data.

ArmMeasureValue (MEAN)Dispersion
Enlite Sensor Accuracy MARD, Lower Serial NumberEnlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump, Using One Additional Calibration During FST14.37 percentage of differenceStandard Deviation 12.99

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026