Skip to content

Galilei Lens Professional vs. Predicate Devices: a Comparison Study

GALILEI Lens Professional vs. Predicate Devices: a Comparison Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01961089
Enrollment
40
Registered
2013-10-11
Start date
2013-10-31
Completion date
2013-12-31
Last updated
2019-10-08

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cataract

Brief summary

Study question: How does the new GALILEI Lens Professional compare to predicate devices? Study hypothesis: The GALILEI Lens Professional achieves a better precision in the postoperative, calculated target refraction with cataract surgeries than predicate devices, and it is more versatile in the measurable parameters as well as more user-friendly.

Detailed description

Background: An accessory to the commercially available GALILEI Dual-Scheimpflug Corneal Topographer/Keratometer with the name EBR Accessory was developed by the company Ziemer Ophthalmic Systems AG for the measurement of intraocular distances. The combination is called GALILEI Lens Professional. The measurement principle is based on on short coherence interferometry/-reflectometry, which has gained wide-spread application and has been applied clinically, including one of the members of this study group, for the precise measurement of axial, intraocular distances (12.1.). Such EBR Accessory measurements, in combination with measurements by the GALILEI, may be applied to given intra-ocular lens (IOL) types and IOL equations, an permit the calculation of suggested IOLs that are implanted during cataract surgery, in order to achieve the desired vision correction. To register this accessory for sale, the precision and agreement to existing devices with similar applications (predicate devices) must be demonstrated. One complete measurement consists of a cornea-topography/keratometry scan, followed immediately by three consecutive, axial biometry scans of the anterior segment (cornea and crystalline lens) and - after a short re-alignment - three consecutive, axial biometry scans of the retina. The measurement process is continuous, but divided in two steps. Blinking by the patient is possible in between the two steps, which increases patient comfort and, in certain cases, measurement success. Precision is evaluated by three repeated, complete measurements with the GALILEI Lens Professional and each predicate device. Two standard devices with comparable optical technology that are used in routine cataract surgery were declared as predicate devices. These two devise are typically used in combination with standard keratometers, as which the GALILEI Lens Professional may be employed as well. Goal: The comparison of the GALILEI Lens Professional to predicate devices with respect to precision in measuring postoperative, calculated target refractions in cataract patients, and versatility of measurable quantities.

Interventions

DEVICEGalilei Lens Professional

Measurement device

DEVICEIOLMaster

Measurement device

DEVICELenstar

Measurement device

Sponsors

Ziemer Ophthalmic Systems AG
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
SINGLE (Subject)

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* spherical equivalent between -10 D and +10 D * best-corrected VA of 20/100 or better.

Exclusion criteria

* strabism * blepharitis * nystagmus * amblyopia * anisometropia (spherical equivalent \> 1D) * angle closure glaucoma * seizure disorder * brain damages * Down syndrome * trisomy 13 or 18 * cerebral palsy * other serious disorders of the eye or central nervous systems

Design outcomes

Primary

MeasureTime frameDescription
Agreement With Predicate Devices in Terms of Agreement 13 monthsAssessed parameters: Axial Length (AL; G6, IOLM, LS) Central Corneal Thickness (CCT; G6, LS)) Anterior Chamber Depth (ACD; G6, IOLM, LS) Lens Thickness (LT; G6, LS) Corneal Curvature (SimK; G6, IOLM, LS) White-to-White (WtW; G6, IOL, LS)
Agreement With Predicate Devices in Terms of Agreement 23 monthsAssessed Parameters: Simulated Corneal Curvature (SimK; G6, IOLM, LS)

Secondary

MeasureTime frameDescription
Agreement With Devices of the Same Type in Terms of Repeatability3 monthsAssessed parameters: Axial Length (AL) Central Corneal Thickness (CCT) Anterior Chamber Depth (ACD) Lens Thickness (LT) Corneal Curvature (SimK) White-to-White (WtW)

Countries

Switzerland

Participant flow

Recruitment details

40 adult subjects were recruited from among patients seen at the ORASIS eye clinics, including 20 people with cataract in at least one eye and 20 subjects without cataract. The subjects included 5 people with prior refractive surgery, 5 people with high myopia, 5 people with high hyperopia, and 3 persons with dilated pupils.

Participants by arm

ArmCount
Device
Galilei Lens Professional versus predicate devices Galilei G6 Lens Professional (Ziemer Ophthalmic Systems AG) IOLMaster (Carl Zeiss Meditech) Lenstar 900 (Haag-Streit AG)
40
Total40

Baseline characteristics

CharacteristicDevice
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
19 Participants
Age, Categorical
Between 18 and 65 years
21 Participants
Age, Continuous58.8 years
STANDARD_DEVIATION 18.9
Region of Enrollment
Switzerland
40 participants
Sex: Female, Male
Female
29 Participants
Sex: Female, Male
Male
11 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 40
serious
Total, serious adverse events
0 / 40

Outcome results

Primary

Agreement With Predicate Devices in Terms of Agreement 1

Assessed parameters: Axial Length (AL; G6, IOLM, LS) Central Corneal Thickness (CCT; G6, LS)) Anterior Chamber Depth (ACD; G6, IOLM, LS) Lens Thickness (LT; G6, LS) Corneal Curvature (SimK; G6, IOLM, LS) White-to-White (WtW; G6, IOL, LS)

Time frame: 3 months

Population: The measurement of Central Corneal Thickness (CCT) and Lens Thickness (LT) is not possible with the IOLM: the IOLM cannot measure these two parameters and does therefore not display results for CCT and LT. As a consequence, the comparison in CCT and LT is only possible between the G6 and the LS, but not with respect to the IOLM.

ArmMeasureGroupValue (MEAN)Dispersion
DeviceAgreement With Predicate Devices in Terms of Agreement 1AL G6-IOLM-0.010 Units are mmStandard Deviation 0.049
DeviceAgreement With Predicate Devices in Terms of Agreement 1AL G6-LS-0.023 Units are mmStandard Deviation 0.048
DeviceAgreement With Predicate Devices in Terms of Agreement 1AL IOLM-LS-0.014 Units are mmStandard Deviation 0.02
DeviceAgreement With Predicate Devices in Terms of Agreement 1CCT G6-LS0.005 Units are mmStandard Deviation 0.008
DeviceAgreement With Predicate Devices in Terms of Agreement 1ACD G6-IOLM0.137 Units are mmStandard Deviation 0.104
DeviceAgreement With Predicate Devices in Terms of Agreement 1ACD G6-LS0.052 Units are mmStandard Deviation 0.168
DeviceAgreement With Predicate Devices in Terms of Agreement 1ACD IOLM-LS-0.087 Units are mmStandard Deviation 0.22
DeviceAgreement With Predicate Devices in Terms of Agreement 1LT G6-LS0.070 Units are mmStandard Deviation 0.256
DeviceAgreement With Predicate Devices in Terms of Agreement 1WtW G6-IOLM0.150 Units are mmStandard Deviation 0.199
DeviceAgreement With Predicate Devices in Terms of Agreement 1WtW G6-LS-0.121 Units are mmStandard Deviation 0.173
DeviceAgreement With Predicate Devices in Terms of Agreement 1WtW IOLM-LS-0.276 Units are mmStandard Deviation 0.191
Comparison: P-value comparing AL between G6 and IOLMp-value: 0.09Paired t-test
Comparison: P-value comparing AL between G6 and LSp-value: <0.001Paired t-test
Comparison: P-value comparing AL between IOLM and LSp-value: <0.0001Paired t-test
Comparison: P-value comparing CCT between G6 and LSp-value: <0.0001Paired t-test
Comparison: P-value comparing ACD between G6 and IOLMp-value: <0.0001Paired t-test
Comparison: P-value comparing ACD between G6 and LSp-value: <0.01Paired t-test
Comparison: P-value comparing ACD between IOLM and LSp-value: 0.02Paired t-test
Comparison: P-value comparing LT between G6 and LSp-value: <0.01Paired t-test
Comparison: P-value comparing WtW between G6 and IOLMp-value: <0.0001Paired t-test
Comparison: P-value comparing WtW between G6 and LSp-value: <0.0001Paired t-test
Comparison: P-value comparing WtW between IOLM and LSp-value: <0.001Paired t-test
Primary

Agreement With Predicate Devices in Terms of Agreement 2

Assessed Parameters: Simulated Corneal Curvature (SimK; G6, IOLM, LS)

Time frame: 3 months

Population: SimK

ArmMeasureGroupValue (MEAN)Dispersion
DeviceAgreement With Predicate Devices in Terms of Agreement 2SimK G6-IOLM-0.031 DioptersStandard Deviation 0.247
DeviceAgreement With Predicate Devices in Terms of Agreement 2SimK G6-LS-0.033 DioptersStandard Deviation 0.326
DeviceAgreement With Predicate Devices in Terms of Agreement 2SimK IOLM-LS0.008 DioptersStandard Deviation 0.198
Comparison: P-value comparing SimK between G6 and IOLMp-value: 0.13Paired t-test
Comparison: P-value comparing SimK between G6 and LSp-value: 0.25Paired t-test
Comparison: P-value comparing SimK between IOLM and LSp-value: 0.21Paired t-test
Secondary

Agreement With Devices of the Same Type in Terms of Repeatability

Assessed parameters: Axial Length (AL) Central Corneal Thickness (CCT) Anterior Chamber Depth (ACD) Lens Thickness (LT) Corneal Curvature (SimK) White-to-White (WtW)

Time frame: 3 months

Population: The IOLM does not measure CCT and LT. Thus, repeatability results are available with the G6 and the LS, but not with the IOLM. The IOL Master displays one average value only for WtW, which does not permit the calculation of repeatability under the present study design.

ArmMeasureGroupValue (NUMBER)
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityWtW LS0.0042 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityAL G60.0018 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityAL IOLM0.0010 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityAL LS0.0008 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityCCT G60.0040 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityCCT LS0.0082 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityACD G60.0259 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityACD IOLM0.0084 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityACD LS0.0391 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityLT G60.0394 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityLT LS0.0262 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilitySimK G60.0028 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilitySimK IOLM0.0024 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilitySimK LS0.0050 Coefficient of variation
DeviceAgreement With Devices of the Same Type in Terms of RepeatabilityWtW G60.0149 Coefficient of variation
Comparison: Repeatability was determined with random effect ANOVAs (estimates of random effects). They were calculated as the square root of the sum of the (Device x EyeID) interaction component plus (EyeID) and (Device) variance components plus the residual variance component. A (Device x Operator) interaction component was not assessed because each device was consistently operated by the same operator such that there was no Device x Operator interaction component. CV = Coefficient of Variation = SD/mean.p-value: 0.05ANOVA

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026