Skip to content

Comparative Evaluation of Handling Characteristics of Two Adrenaline Autoinjectors,EpiPen in Comparison With Jext

Comparative Evaluation of Handling Characteristics of Two Adrenaline Autoinjectors,EpiPen in Comparison With Jext; the HEJ-study

Status
Completed
Phases
Phase 4
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01887405
Acronym
HEJ
Enrollment
91
Registered
2013-06-26
Start date
2012-01-31
Completion date
2012-09-30
Last updated
2013-06-26

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Handling of Auto-injectors

Keywords

Adrenaline, Auto-injector, Jext, Epipen

Brief summary

The purpose of the study is to demonstrate non-inferiority of Jext compared with EpiPen for the proportion of subjects with a successful self-injection. And to evaluate and compare the handling characteristics of two auto-injectors (Jext and EpiPen).

Interventions

DRUGAdrenaline auto-injector

Each subject will be randomly assigned to use either the Jext or EpiPen first. The alternative auto-injector will then be used by the same subject in the second study session. Valid for both Jext and Epipen: Active ingredients: Adrenaline tartrate Dosage form: Prefilled auto-injector for single use, delivering one intramuscular injection Dose/strength: 150 μg Excipients: Sodium chloride, sodium metabisulphite (E223), hydrochloric acid (for pH adjustment), water

Sponsors

ALK Nordic A/S, Danmark Filial
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
SUPPORTIVE_CARE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
16 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Written informed consent obtained before entering the study 1 * Prescription of EpiPen or AnaPen for at least 2 months * Age ≥16 years * Body weight \>50 kg * Subject willing and able to comply with study protocol

Exclusion criteria

* Any of the following concomitant diseases: cardiovascular disease, e.g. arrhythmias or signs of coronary atherosclerosis, heart failure, uncontrolled hypertension, uncontrolled diabetes mellitus, cancer, hyperthyroidism, severe renal impairment * Known blood-born infection, e.g. hepatitis and/or HIV * Concomitant treatment with medication which may potentiate the effect of adrenaline, e.g. catechol-O-methyl transferase (COMT) inhibitors, monoamine oxidase (MAO) inhibitors and tricyclic antidepressants * Pregnancy * Being immediate family of the investigator or study staff, defined as the investigator's/staff's spouse, parent, child, grandparent or grandchild

Design outcomes

Primary

MeasureTime frameDescription
Non-inferiority of Jext compared with EpiPen for the proportion of subjects with a successful self-injection.1 dayThe following steps were recorded as correctly done or not by the observer in a check list approach (yes/no): 1. The subject identified the safety cap. 2. The subject removed the safety cap. 3. The subject identified the end of the auto-injector where the needle presents itself 4. The subject used sufficient force to actuate the injection. 5. The subject injected the auto-injector into the anterolateral thigh at a right angle (approximately 90°) to the thigh. 6. The time spent for the injection of adrenaline was more than 10 seconds, i.e. with the needle in situ. If all 6 steps were performed correctly, the injection was considered as successful. These data form the basis for evaluation of the binary primary endpoint (successful injection or not).

Secondary

MeasureTime frameDescription
To evaluate and compare the handling characteristics of two auto-injectors (Jext and EpiPen) regarding subject preference for one of the two auto-injectors over the other, the time to perform the injection and any hesitance to perform the injection1 dayFor evaluation of the secondary endpoints, the observer assessed the following: The time used for the actuation of the injection, i.e. the time from starting to read the instructions to having injected the adrenaline as recorded by a stop watch. This time was divided into two parts: * The time for reading the information * The time used to perform the injection Any hesitation from the subject to actuate the injection on a 4-step scale: None, Minor, Severe or Completely Inhibited (i.e. no injection) At the end of the second session the subject was asked the following question: Which auto-injector do you prefer? The following responses were possible: Jext / EpiPen / I don't have any preference for any of the auto-injectors. The preference is a secondary endpoint.

Other

MeasureTime frameDescription
Evaluation of instructions and AEs1 dayAfter the injection the subject evaluated the package of instructions (leaflet as well as the information on the auto-injector) for the following (yes/no): Were the instructions concise? Were the instructions comprehensible? Do you have any further questions concerning the instructions or do you see any lack of clarity? These have been assessed as other endpoints. Safety assessments included recording of all AEs and serious adverse events (SAEs) were done.

Countries

Sweden

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026