The Summer Camp Study: Blood Glucose Control With a Bi-Hormonal Bionic Endocrine Pancreas

The Summer Camp Study: Feasibility of Outpatient Automated Blood Glucose Control With a Bi-Hormonal Bionic Endocrine Pancreas in a Pediatric Population at the Clara Barton Diabetes Camps

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01833988
Enrollment
32
Registered
2013-04-17
Start date
2013-04-30
Completion date
2014-12-31
Last updated
2017-09-08

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Type 1 Diabetes

Keywords

bionic pancreas, artificial pancreas, insulin, glucagon, continuous glucose monitoring (CGM), outpatient, insulin pump, pediatrics, children, camp, summer camp

Brief summary

This study will test the hypothesis that a wearable automated bionic pancreas system that automatically delivers both insulin and glucagon can improved glycemic control vs. usual care for young people with type 1 diabetes 12-20 in a diabetes camp environment.

Detailed description

The bionic pancreas will be compared to usual care in a crossover design in which each volunteer will serve as his or her own control. Each volunteer will be under closed-loop glucose control for five days and usual camp level of diabetes care for five days in random order with a one day washout period in between.

Interventions

DEVICEBi-hormonal Bionic Pancreas

Automated blood glucose control via a closed-loop bionic pancreas device.

OTHERUsual Care

Comparator week to closed-loop control, utilizing usual camp care and the subject's own insulin pump.

Sponsors

Boston University
CollaboratorOTHER
Massachusetts General Hospital
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
12 Years to 20 Years
Healthy volunteers
No

Inclusion criteria

* Age 12-20 years with type 1 diabetes for at least one year. * Diabetes managed using an insulin infusion pump and rapid- or very-rapid-acting insulins including insulin aspart (NovoLog), insulin lispro (Humalog), and insulin glulisine (Apidra) for at least three months prior to enrollment. * Otherwise healthy (mild chronic disease such as asthma will be allowed if well controlled that do not require medications that result in exclusion).

Exclusion criteria

* Unable to provide informed assent * Unable to comply with study procedures. * Current participation in another diabetes-related clinical trial other than one that is primarily observational in nature. * Total daily dose (TDD) of insulin that is \> 2 units/kg. * Pregnancy (positive urine HCG), plan to become pregnant in the immediate future, or sexually active without use of contraception * Hypoglycemia unawareness (self-reported or legal guardian report of consistent lack of hypoglycemia symptoms when BG is \< 50 mg/dl) * End stage renal disease on dialysis (hemodialysis or peritoneal dialysis). * History of prolonged QT or arrhythmia * History of congenital heart disease or current known cardiac disease * Acute illness (other than non-vomiting viral illness) or exacerbation of chronic illness other than type 1 diabetes at the time of the study. * Seizure disorder or history of hypoglycemic seizures or coma in the last five years * Untreated or inadequately treated mental illness (indicators would include symptoms such as psychosis, hallucinations, mania, and any psychiatric hospitalization in the last year), or treatment with second generation anti-psychotic medications, which are known to affect glucose regulation. * Electrically powered implants (e.g. cochlear implants, neurostimulators) that might be susceptible to radiofrequency interference. * Use non-insulin, injectable (e.g. exenatide, pramlintide) or oral (e.g. thiazolidinediones, biguanides, sulfonylureas, meglitinides, dipeptidyl peptidase-4 inhibitors, acarbose)anti-diabetic medications. * History of adverse reaction to glucagon (including allergy) besides nausea and vomiting. * Unwilling or unable to completely avoid acetaminophen during the usual care and closed-loop BG control portions of the study. * History of eating disorder such as anorexia, bulimia, diabulemia or omission of insulin to manipulate weight * History of intentional, inappropriate administration of insulin leading to severe hypoglycemia requiring treatment * Any factors that, in the opinion of the principal investigator, would interfere with the safe completion of the study procedures.

Design outcomes

Primary

MeasureTime frame
Percentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm1 week
Difference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours.1 week

Secondary

MeasureTime frameDescription
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dlDay 2-5
Difference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care PeriodDay 2-5
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue MeasurementsDay 1-5
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean BG During Exercise1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise1 week
Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm)1 week
Difference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG)Day 2-5Day 2-5
Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure)1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure)1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl1 week
Difference Between Closed-loop and Open-loop in Mean CGMG in the Four Hour Period Following Meals1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.1 weekDifference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in standard deviation of CGMG values (glycemic variability) in different BG ranges. %\<70 70-120 70-180 %\>180 %\>250
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)1 week
Difference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring.1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night1 week
Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at NightDay 2-5Day 2-5
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night1 week
Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure) at Night1 week
Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure) at Night1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG During Exercise1 week
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Average BG as Determined From All HemoCue Measurements Taken During the Day/Nighttime Including All Extra Measurements.1 weekDifference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in average BG as determined from all HemoCue measurements taken during the day/nighttime including all extra measurements taken before meals, taken during exercise, and taken for hypoglycemia monitoring.

Other

MeasureTime frame
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose1 week

Countries

United States

Participant flow

Participants by arm

ArmCount
All Study Participants
The bionic pancreas (closed loop) will be compared to usual care (subject's own insulin pump) in a crossover design in which each volunteer will serve as his or her own control. Each volunteer will be under closed-loop glucose control for five days and usual camp level of diabetes care for five days in random order with a two day washout period in between.
32
Total32

Baseline characteristics

CharacteristicAll Study Participants
Age, Continuous16 years
STANDARD_DEVIATION 3
Region of Enrollment
United States
32 participants
Sex: Female, Male
All study participants
Female
16 Participants
Sex: Female, Male
All study participants
Male
16 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
— / —— / —
other
Total, other adverse events
6 / 324 / 32
serious
Total, serious adverse events
0 / 320 / 32

Outcome results

Primary

Difference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours.

Time frame: 1 week

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasDifference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours.138 mg/dLStandard Deviation 18
ControlDifference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours.157 mg/dLStandard Deviation 27
Primary

Percentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm

Time frame: 1 week

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasPercentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm6.1 percentage of timeStandard Deviation 4.5
ControlPercentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm7.6 percentage of timeStandard Deviation 7.3
Secondary

Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure) at Night

Time frame: 1 week

Population: This outcome was not analyzed as we feel that the outcome percentage of time \<50mg/dl at nightime is an equivalent indicator of severe overnight hypoglycemia exposure instead of area over the curve, and can be used to compare outcomes with other studies.

Secondary

Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure)

Time frame: 1 week

Population: This outcome was not analyzed as we feel that the outcome percentage of time \<70mg/dl is an equivalent indicator of hypoglycemia exposure instead of area over the curve, and can be used to compare outcomes with other studies.

Secondary

Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure) at Night

Time frame: 1 week

Population: This outcome was not analyzed as we feel that the outcome percentage of time \<70mg/dl at nightime is an equivalent indicator of overnight hypoglycemia exposure instead of area over the curve, and can be used to compare outcomes with other studies.

Secondary

Difference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring.

Time frame: 1 week

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring.141 mg/dlStandard Deviation 20
ControlDifference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring.162 mg/dlStandard Deviation 37
Secondary

Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)

Time frame: 1 week

ArmMeasureGroupValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-12051.5 percentage of timeStandard Deviation 12.3
Bionic PancreasDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>18013.0 percentage of timeStandard Deviation 7.6
Bionic PancreasDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-18084.6 percentage of timeStandard Deviation 8.5
Bionic PancreasDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>2502.8 percentage of timeStandard Deviation 3.4
Bionic PancreasDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)<702.4 percentage of timeStandard Deviation 2.1
ControlDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>2508.6 percentage of timeStandard Deviation 10.9
ControlDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)<705.0 percentage of timeStandard Deviation 5.4
ControlDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-12027.4 percentage of timeStandard Deviation 17.1
ControlDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-18065.9 percentage of timeStandard Deviation 16.9
ControlDifference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>18029.8 percentage of timeStandard Deviation 19.8
Secondary

Difference Between Closed-loop and Open-loop in Mean CGMG in the Four Hour Period Following Meals

Time frame: 1 week

Population: Data was not collected to measure this outcome

Secondary

Difference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG)

Day 2-5

Time frame: Day 2-5

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG)142 mg/dlStandard Deviation 12
ControlDifference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG)158 mg/dlStandard Deviation 27
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure)

Time frame: 1 week

Population: This outcome was not analyzed as we feel that the outcome percentage of time \<50mg/dl is a n equivalent indicator of severe hypoglycemia instead of area over the curve, and can be used to compare outcomes with other studies.

Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Average BG as Determined From All HemoCue Measurements Taken During the Day/Nighttime Including All Extra Measurements.

Difference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in average BG as determined from all HemoCue measurements taken during the day/nighttime including all extra measurements taken before meals, taken during exercise, and taken for hypoglycemia monitoring.

Time frame: 1 week

Population: This data was not analysed to prevent bias. This information would have been misleading (unscheduled BG checks were performed to confirm a low BG mostly). Overall CGM trend, incidence/duration of hypoglycemia and BG trend during and after exercise is separately analyzed under other secondary outcomes.

Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)

Time frame: 1 week

Population: Fraction of time \< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, \> 180 mg/dl, \> 250 mg/dl

ArmMeasureGroupValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-120 mg/dl39.4 percentage of timeStandard Deviation 8.2
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>180 mg/dl24.0 percentage of timeStandard Deviation 8.5
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-180 mg/dl73.3 percentage of timeStandard Deviation 9
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>250 mg/dl7.6 percentage of timeStandard Deviation 5.6
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)<70 mg/dl2.7 percentage of timeStandard Deviation 2.2
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>250 mg/dl10.8 percentage of timeStandard Deviation 8.7
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)<70 mg/dl5.3 percentage of timeStandard Deviation 4.6
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-120 mg/dl29.9 percentage of timeStandard Deviation 12.3
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)70-180 mg/dl63.8 percentage of timeStandard Deviation 13.3
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)>180 mg/dl30.9 percentage of timeStandard Deviation 15
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night

Time frame: 1 week

ArmMeasureGroupValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night70-120 mg/dl55.3 percentage of timeStandard Deviation 13.7
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night>180 mg/dl10.5 percentage of timeStandard Deviation 7
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night70-180 mg/dl86.9 percentage of timeStandard Deviation 8.1
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night>250 mg/dl1.8 percentage of timeStandard Deviation 2.5
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night<70 mg/dl2.6 percentage of timeStandard Deviation 2.5
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night>250 mg/dl9.5 percentage of timeStandard Deviation 12.5
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night<70 mg/dl4.0 percentage of timeStandard Deviation 5.3
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night70-120 mg/dl28.3 percentage of timeStandard Deviation 17.3
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night70-180 mg/dl66.7 percentage of timeStandard Deviation 19.9
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night>180 mg/dl29.3 percentage of timeStandard Deviation 22
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean BG During Exercise

Time frame: 1 week

Population: Data on exercise not collected

Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at Night

Day 2-5

Time frame: Day 2-5

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at Night124 mg/dlStandard Deviation 11
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at Night157 mg/dlStandard Deviation 36
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG During Exercise

Time frame: 1 week

Population: Data not obtained for exercise period

Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia

Time frame: 1 week

ArmMeasureValue (NUMBER)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia97 number of interventions
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia210 number of interventions
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night

Time frame: 1 week

ArmMeasureValue (NUMBER)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night29 interventions
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night48 interventions
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue Measurements

Time frame: Day 1-5

ArmMeasureValue (NUMBER)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue Measurements97 events
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue Measurements210 events
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise

Time frame: 1 week

Population: Individual data during exercise was not analyzed. No episodes of severe hypoglycemia were observed on bionic pancreas and only 1 episode of severe hypoglycemia occurred in usual care arm (insulin pump)

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise0 Participants
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise1 Participants
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dl

Time frame: Day 2-5

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dl29 Participants
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dl14 Participants
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl

Time frame: 1 week

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl26 Participants
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl17 Participants
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)

Time frame: 1 week

ArmMeasureGroupValue (NUMBER)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)70-12012.3 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)>1807.6 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)70-1808.5 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)>2503.4 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)<702.1 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)>25010.9 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)<705.4 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)70-12017.1 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)70-18016.9 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)>18019.8 Standard deviation of mean values
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.

Difference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in standard deviation of CGMG values (glycemic variability) in different BG ranges. %\<70 70-120 70-180 %\>180 %\>250

Time frame: 1 week

ArmMeasureGroupValue (NUMBER)
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.70-1208.2 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.>1808.5 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.70-1809.0 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.>2505.6 Standard deviation of mean values
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.<702.2 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.>2508.7 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.<704.6 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.70-12012.3 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.70-18013.3 Standard deviation of mean values
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.>18015.0 Standard deviation of mean values
Secondary

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night

Time frame: 1 week

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night4.1 percentage of timeStandard Deviation 6.1
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night4.4 percentage of timeStandard Deviation 6.7
Secondary

Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm)

Time frame: 1 week

ArmMeasureGroupValue (MEAN)Dispersion
Bionic PancreasDifference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm)Mean CGMG on Day 1169.2 mg/dlStandard Deviation 31.3
Bionic PancreasDifference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm)Mean CGMG day 2-5142 mg/dlStandard Deviation 12
ControlDifference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm)Mean CGMG on Day 1156 mg/dlStandard Deviation 37
ControlDifference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm)Mean CGMG day 2-5158 mg/dlStandard Deviation 27
Secondary

Difference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care Period

Time frame: Day 2-5

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasDifference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care Period75 percentage of daysStandard Deviation 21.5
ControlDifference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care Period49.3 percentage of daysStandard Deviation 36.2
Other Pre-specified

Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose

Time frame: 1 week

ArmMeasureValue (MEAN)Dispersion
Bionic PancreasDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose0.82 unit/kg/dayStandard Deviation 0.16
ControlDifference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose0.79 unit/kg/dayStandard Deviation 0.17