Type 1 Diabetes
Conditions
Keywords
bionic pancreas, artificial pancreas, insulin, glucagon, continuous glucose monitoring (CGM), outpatient, insulin pump, pediatrics, children, camp, summer camp
Brief summary
This study will test the hypothesis that a wearable automated bionic pancreas system that automatically delivers both insulin and glucagon can improved glycemic control vs. usual care for young people with type 1 diabetes 12-20 in a diabetes camp environment.
Detailed description
The bionic pancreas will be compared to usual care in a crossover design in which each volunteer will serve as his or her own control. Each volunteer will be under closed-loop glucose control for five days and usual camp level of diabetes care for five days in random order with a one day washout period in between.
Interventions
Automated blood glucose control via a closed-loop bionic pancreas device.
Comparator week to closed-loop control, utilizing usual camp care and the subject's own insulin pump.
Sponsors
Study design
Eligibility
Inclusion criteria
* Age 12-20 years with type 1 diabetes for at least one year. * Diabetes managed using an insulin infusion pump and rapid- or very-rapid-acting insulins including insulin aspart (NovoLog), insulin lispro (Humalog), and insulin glulisine (Apidra) for at least three months prior to enrollment. * Otherwise healthy (mild chronic disease such as asthma will be allowed if well controlled that do not require medications that result in exclusion).
Exclusion criteria
* Unable to provide informed assent * Unable to comply with study procedures. * Current participation in another diabetes-related clinical trial other than one that is primarily observational in nature. * Total daily dose (TDD) of insulin that is \> 2 units/kg. * Pregnancy (positive urine HCG), plan to become pregnant in the immediate future, or sexually active without use of contraception * Hypoglycemia unawareness (self-reported or legal guardian report of consistent lack of hypoglycemia symptoms when BG is \< 50 mg/dl) * End stage renal disease on dialysis (hemodialysis or peritoneal dialysis). * History of prolonged QT or arrhythmia * History of congenital heart disease or current known cardiac disease * Acute illness (other than non-vomiting viral illness) or exacerbation of chronic illness other than type 1 diabetes at the time of the study. * Seizure disorder or history of hypoglycemic seizures or coma in the last five years * Untreated or inadequately treated mental illness (indicators would include symptoms such as psychosis, hallucinations, mania, and any psychiatric hospitalization in the last year), or treatment with second generation anti-psychotic medications, which are known to affect glucose regulation. * Electrically powered implants (e.g. cochlear implants, neurostimulators) that might be susceptible to radiofrequency interference. * Use non-insulin, injectable (e.g. exenatide, pramlintide) or oral (e.g. thiazolidinediones, biguanides, sulfonylureas, meglitinides, dipeptidyl peptidase-4 inhibitors, acarbose)anti-diabetic medications. * History of adverse reaction to glucagon (including allergy) besides nausea and vomiting. * Unwilling or unable to completely avoid acetaminophen during the usual care and closed-loop BG control portions of the study. * History of eating disorder such as anorexia, bulimia, diabulemia or omission of insulin to manipulate weight * History of intentional, inappropriate administration of insulin leading to severe hypoglycemia requiring treatment * Any factors that, in the opinion of the principal investigator, would interfere with the safe completion of the study procedures.
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| Percentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm | 1 week |
| Difference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours. | 1 week |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dl | Day 2-5 | — |
| Difference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care Period | Day 2-5 | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue Measurements | Day 1-5 | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean BG During Exercise | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise | 1 week | — |
| Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm) | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG) | Day 2-5 | Day 2-5 |
| Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure) | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure) | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl | 1 week | — |
| Difference Between Closed-loop and Open-loop in Mean CGMG in the Four Hour Period Following Meals | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | 1 week | Difference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in standard deviation of CGMG values (glycemic variability) in different BG ranges. %\<70 70-120 70-180 %\>180 %\>250 |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | 1 week | — |
| Difference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring. | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night | 1 week | — |
| Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at Night | Day 2-5 | Day 2-5 |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | 1 week | — |
| Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure) at Night | 1 week | — |
| Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure) at Night | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG During Exercise | 1 week | — |
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Average BG as Determined From All HemoCue Measurements Taken During the Day/Nighttime Including All Extra Measurements. | 1 week | Difference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in average BG as determined from all HemoCue measurements taken during the day/nighttime including all extra measurements taken before meals, taken during exercise, and taken for hypoglycemia monitoring. |
Other
| Measure | Time frame |
|---|---|
| Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose | 1 week |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| All Study Participants The bionic pancreas (closed loop) will be compared to usual care (subject's own insulin pump) in a crossover design in which each volunteer will serve as his or her own control. Each volunteer will be under closed-loop glucose control for five days and usual camp level of diabetes care for five days in random order with a two day washout period in between. | 32 |
| Total | 32 |
Baseline characteristics
| Characteristic | All Study Participants |
|---|---|
| Age, Continuous | 16 years STANDARD_DEVIATION 3 |
| Region of Enrollment United States | 32 participants |
| Sex: Female, Male All study participants Female | 16 Participants |
| Sex: Female, Male All study participants Male | 16 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | — / — | — / — |
| other Total, other adverse events | 6 / 32 | 4 / 32 |
| serious Total, serious adverse events | 0 / 32 | 0 / 32 |
Outcome results
Difference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours.
Time frame: 1 week
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Difference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours. | 138 mg/dL | Standard Deviation 18 |
| Control | Difference in Average Blood Glucose (BG) Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) Periods as Determined From All Scheduled HemoCue Measurements With Mean Evenly Weighted Across the Daytime and Nighttime Hours. | 157 mg/dL | Standard Deviation 27 |
Percentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm
Time frame: 1 week
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Percentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm | 6.1 percentage of time | Standard Deviation 4.5 |
| Control | Percentage of Time With a Low Plasma Glucose Reading (Less Than 70mg/dl) in the Bionic Pancreas Arm as Compared to Insulin Pump Arm | 7.6 percentage of time | Standard Deviation 7.3 |
Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure) at Night
Time frame: 1 week
Population: This outcome was not analyzed as we feel that the outcome percentage of time \<50mg/dl at nightime is an equivalent indicator of severe overnight hypoglycemia exposure instead of area over the curve, and can be used to compare outcomes with other studies.
Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure)
Time frame: 1 week
Population: This outcome was not analyzed as we feel that the outcome percentage of time \<70mg/dl is an equivalent indicator of hypoglycemia exposure instead of area over the curve, and can be used to compare outcomes with other studies.
Difference Between Closed-loop and Open-loop in Area Over the Curve and Below 70 mg/dl (Measure of Total Hypoglycemia Exposure) at Night
Time frame: 1 week
Population: This outcome was not analyzed as we feel that the outcome percentage of time \<70mg/dl at nightime is an equivalent indicator of overnight hypoglycemia exposure instead of area over the curve, and can be used to compare outcomes with other studies.
Difference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring.
Time frame: 1 week
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring. | 141 mg/dl | Standard Deviation 20 |
| Control | Difference Between Closed-loop and Open-loop in Average BG as Determined From All HemoCue Measurements Taken During the Nighttime Including All Extra Measurements Taken for Hypoglycemia Monitoring. | 162 mg/dl | Standard Deviation 37 |
Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)
Time frame: 1 week
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-120 | 51.5 percentage of time | Standard Deviation 12.3 |
| Bionic Pancreas | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >180 | 13.0 percentage of time | Standard Deviation 7.6 |
| Bionic Pancreas | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-180 | 84.6 percentage of time | Standard Deviation 8.5 |
| Bionic Pancreas | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >250 | 2.8 percentage of time | Standard Deviation 3.4 |
| Bionic Pancreas | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | <70 | 2.4 percentage of time | Standard Deviation 2.1 |
| Control | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >250 | 8.6 percentage of time | Standard Deviation 10.9 |
| Control | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | <70 | 5.0 percentage of time | Standard Deviation 5.4 |
| Control | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-120 | 27.4 percentage of time | Standard Deviation 17.1 |
| Control | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-180 | 65.9 percentage of time | Standard Deviation 16.9 |
| Control | Difference Between Closed-loop and Open-loop in Fraction of Time at Night Spent Within Glucose Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >180 | 29.8 percentage of time | Standard Deviation 19.8 |
Difference Between Closed-loop and Open-loop in Mean CGMG in the Four Hour Period Following Meals
Time frame: 1 week
Population: Data was not collected to measure this outcome
Difference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG)
Day 2-5
Time frame: Day 2-5
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG) | 142 mg/dl | Standard Deviation 12 |
| Control | Difference Between Closed-loop (Bionic Pancreas) and Open-loop (Insulin Pump) in Mean Continuous Glucose Monitoring Glucose (CGMG) | 158 mg/dl | Standard Deviation 27 |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Area Over the Curve and Below 50 mg/dl (Measure of Total Hypoglycemia Exposure)
Time frame: 1 week
Population: This outcome was not analyzed as we feel that the outcome percentage of time \<50mg/dl is a n equivalent indicator of severe hypoglycemia instead of area over the curve, and can be used to compare outcomes with other studies.
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Average BG as Determined From All HemoCue Measurements Taken During the Day/Nighttime Including All Extra Measurements.
Difference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in average BG as determined from all HemoCue measurements taken during the day/nighttime including all extra measurements taken before meals, taken during exercise, and taken for hypoglycemia monitoring.
Time frame: 1 week
Population: This data was not analysed to prevent bias. This information would have been misleading (unscheduled BG checks were performed to confirm a low BG mostly). Overall CGM trend, incidence/duration of hypoglycemia and BG trend during and after exercise is separately analyzed under other secondary outcomes.
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl)
Time frame: 1 week
Population: Fraction of time \< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, \> 180 mg/dl, \> 250 mg/dl
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-120 mg/dl | 39.4 percentage of time | Standard Deviation 8.2 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >180 mg/dl | 24.0 percentage of time | Standard Deviation 8.5 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-180 mg/dl | 73.3 percentage of time | Standard Deviation 9 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >250 mg/dl | 7.6 percentage of time | Standard Deviation 5.6 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | <70 mg/dl | 2.7 percentage of time | Standard Deviation 2.2 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >250 mg/dl | 10.8 percentage of time | Standard Deviation 8.7 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | <70 mg/dl | 5.3 percentage of time | Standard Deviation 4.6 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-120 mg/dl | 29.9 percentage of time | Standard Deviation 12.3 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | 70-180 mg/dl | 63.8 percentage of time | Standard Deviation 13.3 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG (Continuous Glucose Monitor) Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) | >180 mg/dl | 30.9 percentage of time | Standard Deviation 15 |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night
Time frame: 1 week
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | 70-120 mg/dl | 55.3 percentage of time | Standard Deviation 13.7 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | >180 mg/dl | 10.5 percentage of time | Standard Deviation 7 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | 70-180 mg/dl | 86.9 percentage of time | Standard Deviation 8.1 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | >250 mg/dl | 1.8 percentage of time | Standard Deviation 2.5 |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | <70 mg/dl | 2.6 percentage of time | Standard Deviation 2.5 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | >250 mg/dl | 9.5 percentage of time | Standard Deviation 12.5 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | <70 mg/dl | 4.0 percentage of time | Standard Deviation 5.3 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | 70-120 mg/dl | 28.3 percentage of time | Standard Deviation 17.3 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | 70-180 mg/dl | 66.7 percentage of time | Standard Deviation 19.9 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Fraction of Time Spent Within CGMG Ranges (< 70 mg/dl, 70-120 mg/dl, 70-180 mg/dl, > 180 mg/dl, > 250 mg/dl) at Night | >180 mg/dl | 29.3 percentage of time | Standard Deviation 22 |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean BG During Exercise
Time frame: 1 week
Population: Data on exercise not collected
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at Night
Day 2-5
Time frame: Day 2-5
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at Night | 124 mg/dl | Standard Deviation 11 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG at Night | 157 mg/dl | Standard Deviation 36 |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean CGMG During Exercise
Time frame: 1 week
Population: Data not obtained for exercise period
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia
Time frame: 1 week
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia | 97 number of interventions |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia | 210 number of interventions |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night
Time frame: 1 week
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night | 29 interventions |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Carbohydrate Interventions for Hypoglycemia at Night | 48 interventions |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue Measurements
Time frame: Day 1-5
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue Measurements | 97 events |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Hypoglycemic Events (BG <70mg/dl) as Determined From HemoCue Measurements | 210 events |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise
Time frame: 1 week
Population: Individual data during exercise was not analyzed. No episodes of severe hypoglycemia were observed on bionic pancreas and only 1 episode of severe hypoglycemia occurred in usual care arm (insulin pump)
| Arm | Measure | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise | 0 Participants |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Severe Hypoglycemic Episodes and Nadir BG During Exercise | 1 Participants |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dl
Time frame: Day 2-5
| Arm | Measure | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dl | 29 Participants |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean BG < 154 mg/dl | 14 Participants |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl
Time frame: 1 week
| Arm | Measure | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl | 26 Participants |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Number of Subjects With Mean CGMG < 154 mg/dl | 17 Participants |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM)
Time frame: 1 week
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | 70-120 | 12.3 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | >180 | 7.6 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | 70-180 | 8.5 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | >250 | 3.4 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | <70 | 2.1 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | >250 | 10.9 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | <70 | 5.4 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | 70-120 | 17.1 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | 70-180 | 16.9 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values at Night (11:00 PM to 7:00 AM) | >180 | 19.8 Standard deviation of mean values |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges.
Difference between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in standard deviation of CGMG values (glycemic variability) in different BG ranges. %\<70 70-120 70-180 %\>180 %\>250
Time frame: 1 week
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | 70-120 | 8.2 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | >180 | 8.5 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | 70-180 | 9.0 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | >250 | 5.6 Standard deviation of mean values |
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | <70 | 2.2 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | >250 | 8.7 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | <70 | 4.6 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | 70-120 | 12.3 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | 70-180 | 13.3 Standard deviation of mean values |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Standard Deviation of CGMG Values (Glycemic Variability) in Different BG Ranges. | >180 | 15.0 Standard deviation of mean values |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night
Time frame: 1 week
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night | 4.1 percentage of time | Standard Deviation 6.1 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Time Spent in Hypoglycemia (Plasma BG <Than 70 mg/dl) at Night | 4.4 percentage of time | Standard Deviation 6.7 |
Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm)
Time frame: 1 week
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| Bionic Pancreas | Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm) | Mean CGMG on Day 1 | 169.2 mg/dl | Standard Deviation 31.3 |
| Bionic Pancreas | Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm) | Mean CGMG day 2-5 | 142 mg/dl | Standard Deviation 12 |
| Control | Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm) | Mean CGMG on Day 1 | 156 mg/dl | Standard Deviation 37 |
| Control | Difference in Mean CGMG on Day 1 vs. Remaining Days (Days 2-5) Between Closed Loop (Bionic Pancreas Arm) and Usual Care (Insulin Pump Arm) | Mean CGMG day 2-5 | 158 mg/dl | Standard Deviation 27 |
Difference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care Period
Time frame: Day 2-5
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Difference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care Period | 75 percentage of days | Standard Deviation 21.5 |
| Control | Difference in the Percentage of Study Days With Mean CGM BG </= 154 mg/dl Over the Duration of the Closed-loop Period vs. the Usual Care Period | 49.3 percentage of days | Standard Deviation 36.2 |
Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose
Time frame: 1 week
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Bionic Pancreas | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose | 0.82 unit/kg/day | Standard Deviation 0.16 |
| Control | Difference Between Closed-loop (Bionic Pancreas Arm) and Open-loop (Insulin Pump Arm) in Mean Insulin Total Daily Dose | 0.79 unit/kg/day | Standard Deviation 0.17 |