Patent Ductus Arteriosus
Conditions
Keywords
PDA, Echo, PES, Handheld echo system
Brief summary
This is a single-center, feasibility study involving all neonates admitted to the neonatal intensive care unit with orders for a cardiac echo evaluation. Parents will be approached and provided a description of the study. No written consent will be requested, but parents have the option of opting out. Each patient will have a 5 minute Pocket echocardiography system scan followed by a full echo performed on a traditional full featured echo system. The objective is to assess if the current Food and Drug administration (FDA) approved Pocket echocardiography system (PES) can detect patent ductus arteriosus (PDA) in neonates as comparable to traditional full featured echo systems (FFES) and/or physical exam alone.
Detailed description
All newborns admitted to the neonatal intensive care unit with orders for a cardiac echo evaluation will be included in this study. Parents will be approached and provided a description of the study. No written consent will be requested, but parents have the option of opting out. If parents agree to be part of this study, a 5-minute PES scan will be performed first so as to not bias the sonographer. Then, a full echocardiogram study will be performed on a traditional FFES. The scans will be performed at the bedside. The images will be reviewed and compared by fellow and/or attending and by experienced sonographer. The scans will include short axis view, ductal view and arch view. The objective of the study is to assess if the current FDA approved PES can detect PDA in neonates as comparable to traditional full featured echo systems and/or physical exam alone. Using PES for PDA diagnosis will be an accessible, economic and effective way of complementing the findings in our physical exam by increasing diagnostic accuracy, and detecting PDA's at the bedside comparable to a full size traditional echocardiographic system.
Interventions
A 5-minute Pocket echocardiography system scan (PES) scan will be performed to detect PDA on neonates. This scan will be followed by a Full Featured Echocardiography System Scan (FFES) and scan results will be compared.
Sponsors
Study design
Eligibility
Inclusion criteria
* All newborn infants with orders for cardiac echocardiography evaluation in clinically stable condition
Exclusion criteria
* Known congenital heart disease. * Clinically unstable (Meaning by unstable patients: Those patients that need cardiopulmonary resuscitation, or in maximum ventilatory or inotrope support and in immediate need for Extracorporeal Membrane Oxygenation or major surgery or whenever the physician attending responsible for the clinical care of the patient considers that a traditional echo needs to be done urgently)
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| False Positives Results of Pocket Echocardiography System (PES) Detection of Patent Ductus Arteriosus (PDA) to Full Featured Echo System (FFES) | at baseline | Evaluate the usefulness of the currently FDA approved Pocket Echocardiography System (PES) in PDA detection as compared to Full Featured Echo System (FFES) by looking at false positives between the two systems. |
| False Negative Results of Pocket Echocardiography System (PES) Detection of Patent Ductus Arteriosus (PDA) to Full Featured Echo System (FFES) | at baseline | Evaluate the usefulness of the currently FDA approved Pocket Echocardiography System (PES) in PDA detection as compared to Full Featured Echo System (FFES) by looking at false negatives between the two systems. |
Countries
United States
Participant flow
Recruitment details
Patients recruited in the NICU.
Participants by arm
| Arm | Count |
|---|---|
| PES First, Then FFES A 5-minute Pocket echocardiography system scan (PES) scan will be performed to detect PDA on neonates. This scan will be followed by a Full Featured Echocardiography System Scan (FFES) and scan results will be compared. | 50 |
| Total | 50 |
Baseline characteristics
| Characteristic | PES First, Then FFES |
|---|---|
| Age, Categorical <=18 years | 50 Participants |
| Age, Categorical >=65 years | 0 Participants |
| Age, Categorical Between 18 and 65 years | 0 Participants |
| Age, Continuous | 34.8 weeks STANDARD_DEVIATION 4.98 |
| Body Weight | 2.470 kg STANDARD_DEVIATION 1.157 |
| Region of Enrollment United States | 50 participants |
| Sex: Female, Male Female | 17 Participants |
| Sex: Female, Male Male | 33 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 0 / 50 |
| serious Total, serious adverse events | 0 / 50 |
Outcome results
False Negative Results of Pocket Echocardiography System (PES) Detection of Patent Ductus Arteriosus (PDA) to Full Featured Echo System (FFES)
Evaluate the usefulness of the currently FDA approved Pocket Echocardiography System (PES) in PDA detection as compared to Full Featured Echo System (FFES) by looking at false negatives between the two systems.
Time frame: at baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| PES First, Then FFES | False Negative Results of Pocket Echocardiography System (PES) Detection of Patent Ductus Arteriosus (PDA) to Full Featured Echo System (FFES) | 3.75 percentage of false negative | Standard Deviation 4.63 |
False Positives Results of Pocket Echocardiography System (PES) Detection of Patent Ductus Arteriosus (PDA) to Full Featured Echo System (FFES)
Evaluate the usefulness of the currently FDA approved Pocket Echocardiography System (PES) in PDA detection as compared to Full Featured Echo System (FFES) by looking at false positives between the two systems.
Time frame: at baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| PES First, Then FFES | False Positives Results of Pocket Echocardiography System (PES) Detection of Patent Ductus Arteriosus (PDA) to Full Featured Echo System (FFES) | 10.8 percentage of false positives | Standard Deviation 1.2 |