Contrast Enhanced Computed Tomography
Conditions
Keywords
CECT, catheter, contrast
Brief summary
The purpose of this study is to compare multidetector computed tomography (CT) image quality following peripheral venous high pressure injection of iodinated intravenous contrast media between a 20 gauge (GA) fenestrated, intravenous catheter (BD Nexiva™ Diffusics™) and an 18 GA non-fenestrated, intravenous catheter. In the contrast enhanced CT procedure (CECT), a machine (power injector) is used to rapidly deliver contrast agent to an individual's circulatory system. The contrast allows for greater differentiation of tissues in images of the area of interest, affording a better opportunity to visually assess the area and hence make a more accurate diagnosis. CECT often requires a high rate of contrast flow (≥ 5 mL/sec) through the IV catheter to provide sufficient contrast for an acceptable image. Typically an 18 GA IV catheter is required to achieve these high flow rates; however, often the patient does not have a vein that will accommodate this large size catheter. The BD Nexiva™ Diffusics™ IV catheter is designed to deliver a higher flow rate than a conventional catheter of the same gauge by incorporating fenestrations (small openings) near the tip of the catheter that allow fluid to exit the catheter in addition to that which normally exits through the main opening at the tip of the catheter.
Detailed description
This is a prospective study performed in adult human subjects who have been referred for a CECT. The study will occur during a single visit. Once consented and enrolled, subjects whose antecubital veins can accommodate an 18 GA IV catheter (as assessed by the Inserter or designee of the PI), will be randomized to receive either a conventional 18 GA IV catheter (reference, control) or a 20 GA BD Nexiva™ Diffusics™ IV catheter (test). Subjects whose veins are not considered suitable for an 18 GA IV catheter will be placed in a non-randomized cohort for placement of a 20 GA BD Nexiva™ Diffusics™ IV catheter. Subjects will receive one (1) catheter for the study. The primary objective of this study is to compare multidetector computed tomography (CT) image quality following peripheral venous high pressure injection of iodinated intravenous contrast media between a 20 GA fenestrated, intravenous catheter (BD Nexiva™ Diffusics™) and an 18 GA non-fenestrated, intravenous catheter. The 20 GA BD Nexiva™ Diffusics™ IV catheter will be considered non-inferior to conventional 18 GA IV catheters if the 95% upper bound for the difference in percentage of images of acceptable quality between the images acquired using the 18 GA IV catheters and the 20 GA BD Nexiva™ Diffusics™ IV catheters is less than 15%. Study catheters will be inserted by appropriately licensed healthcare professionals (Inserters). Prior to enrollment, potential Inserters will be trained and qualified by the Sponsor on proper use of the BD Nexiva™ Diffusics™ IV catheter in both an artificial hand model and then in human patients. Inserters will then be expected to use the test product for a period of two (2) weeks, in order to gain familiarity with the product. Because BD Nexiva™ Diffusics™ IV catheters are not currently used in the Johns Hopkins Outpatient Center, the Sponsor will provide BD Nexiva™ Diffusics™ IV catheters for use during this training and familiarization period.
Interventions
The Inserter will have a single attempt on each arm (antecubital area) in which to insert the study catheter. If the catheter is not successfully inserted on the first attempt, the Inserter may make a second attempt in the contralateral arm. If the Inserter is unsuccessful in the second attempt, the subject is to be discontinued from the Study, and additional (non-Study) insertion attempts may be made according to facility policies.
The Inserter will have a single attempt on each arm (antecubital area) in which to insert the study catheter. If the catheter is not successfully inserted on the first attempt, the Inserter may make a second attempt in the contralateral arm. If after the first attempt, there is no vein in the contralateral arm that can accommodate an 18 gauge catheter, the subject is to be discontinued from the Study, and additional (non-Study) insertion attempts may be made according to facility policies.
Sponsors
Study design
Eligibility
Inclusion criteria
* Must be at least 18 years old at the time of enrollment, * Must meet the investigational site's standard criteria for multi-detector Computed Tomography (CT), * Per institutional protocol, the ordered scan must require a flow rate of 5.0-7.0 mL/sec (inclusive) for Visipaque 320 or 5.0-7.5 mL/sec (inclusive) for Omnipaque350 warmed to 35°C, * The planned catheter insertion site must be located in the antecubital area, * Must require any of the following multi-detector CT procedures: CT angiogram,CT pancreas,CT liver, CT kidney, Cardiac CT, * Must be willing and able to provide informed consent, * Must be able to read, write, and follow instructions in English, * Must be able to accommodate a 20 G x 1.00 inch IV catheter,
Exclusion criteria
* Preexisting IV catheter or contrast-compatible port in place suitable for power injection with the intent to be used for contrast enhanced CT, * Subjects who have reported an adverse reaction that precludes use of iodine-based contrast media, * Subjects who have had a prior extravasation event involving solutions considered irritants or vesicants.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Acceptable Image Quality | at the time of image assessment | Study images were assessed by a US board-certified radiologist to determine whether the image is of acceptable quality. The radiologist was not informed of the study device used for the injection that produced the image under evaluation. Subjective image quality assessment for acceptability was determined by: * The report of the reading radiologist in the section of the report where the radiologist indicates if the image is acceptable or not. The absence of a comment in this section will be interpreted as acceptable, and, * the assessment of an independent single second reader (such as the subinvestigator or research radiologist) blinded to the reading radiologist's report and the infusion catheter type. In the case of a non-concurrence, the Principal Investigator assessed the image and his or her assessment had final authority. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Chest CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects) | at the time of image assessment | Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified. |
| Complete Chest and Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects) | at the time of image assessment | Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified. |
| Maximum Flow Rate | immediately after power injection | Maximum flow rate guidelines provided in the BD Nexiva Diffusics Instructions for Use were to be followed. The maximum flow rate (milliliters per second, or mL/sec) utilized was recorded by the Radiology Technician immediately following the power injection of iodinated intravenous contrast media. Results are descriptively summarized; no formal acceptance criteria were specified by the protocol. |
| Catheter Insertion Success | immediately after catheter insertion | Insertion is successful when the catheter can be flushed or infused to demonstrate patency, and there is no inadvertent administration of a solution or medication into the tissue surrounding the IV catheter. The number of participants with successfully inserted catheters after the first or second insertion attempt is reported. The clinician inserting the IV catheter made a clinical judgement as to whether the catheter was successfully placed; the protocol did not define more specific criteria. Catheter insertion success rates were determined from each IV insertion attempted in the Study. |
| Extravasation of Contrast Media | upon contrast injection | The number of subjects who experienced injections with extravasation of contrast media is reported. |
| Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Randomized Subjects | at the time of image assessment | Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified. |
| High Pressure Alarm | immediately after contrast injection | The number of subjects who experienced injections with activation of the high pressure alarm is reported. Immediately following the power injection, the clinician performing the power injection recorded whether or not the high pressure alarm sounded. |
| Catheter Dislodgement | Immediately following power injection | The number of subjects who experienced partial or complete dislodgement of the catheter from the subject prior to power injection procedure completion is reported. The information about the IV catheter was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection. |
| Catheter Transfixation | immediately after power injection | The number of subjects who experienced catheter transfixation (the IV penetrating the opposite wall of the vein) is reported. This information was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection. |
| Catheter Integrity Failure | immediately after power injection | Catheter integrity failure generally refers to any portion of the device breaking or malfunctioning so that its proper function is no longer assured. In this study, the most relevant catheter integrity failures to be assessed were fluid leakage, tubing rupture, or tubing separation from the hub. The number of subjects who experienced injections with a catheter integrity failure during injection is reported. The information about the IV catheter was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection. |
| Automatic Injection Shutoff | immediately after contrast injection | The number of subjects who experienced injections with automatic injection shutoff is reported. Immediately after the power injection, the injection technician recorded whether or not an automatic injection shutoff occurred. |
Countries
United States
Participant flow
Recruitment details
Subjects were recruited from those referred to the Johns Hopkins Outpatient Center for contrast enhanced computed tomography, or CECT. Subjects were recruited in January and February of 2013. The Overall Study table summarizes the number of subjects that Started and Completed in each Reporting Group, including the type of CECT procedure.
Pre-assignment details
Subjects whose veins were determined to be unable to accommodate an 18 GA IV catheter, but that could accommodate a 20 GA IV catheter, were not randomized but were assigned to receive the 20 GA BD Nexiva™ Diffusics™ IV catheter (non-randomized cohort).
Participants by arm
| Arm | Count |
|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | 102 |
| 18 GA Conventional Catheter - Randomized | 103 |
| 20 GA BD Nexiva Diffusics - Nonrandomized | 33 |
| Total | 238 |
Withdrawals & dropouts
| Period | Reason | FG000 | FG001 | FG002 |
|---|---|---|---|---|
| Overall Study | Adverse Event | 0 | 0 | 1 |
| Overall Study | Unable to insert IV in antecubital vein | 1 | 1 | 5 |
Baseline characteristics
| Characteristic | 20 GA BD Nexiva Diffusics - Randomized | 18 GA Conventional Catheter - Randomized | 20 GA BD Nexiva Diffusics - Nonrandomized | Total |
|---|---|---|---|---|
| Age, Continuous | 61.6 years STANDARD_DEVIATION 12.9 | 60.0 years STANDARD_DEVIATION 13.6 | 58.8 years STANDARD_DEVIATION 12.3 | 60.5 years STANDARD_DEVIATION 13.1 |
| Sex: Female, Male Female | 45 Participants | 37 Participants | 20 Participants | 102 Participants |
| Sex: Female, Male Male | 57 Participants | 66 Participants | 13 Participants | 136 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk | EG002 affected / at risk |
|---|---|---|---|
| deaths Total, all-cause mortality | — / — | — / — | — / — |
| other Total, other adverse events | 0 / 102 | 0 / 103 | 1 / 33 |
| serious Total, serious adverse events | 0 / 102 | 0 / 103 | 0 / 33 |
Outcome results
Acceptable Image Quality
Study images were assessed by a US board-certified radiologist to determine whether the image is of acceptable quality. The radiologist was not informed of the study device used for the injection that produced the image under evaluation. Subjective image quality assessment for acceptability was determined by: * The report of the reading radiologist in the section of the report where the radiologist indicates if the image is acceptable or not. The absence of a comment in this section will be interpreted as acceptable, and, * the assessment of an independent single second reader (such as the subinvestigator or research radiologist) blinded to the reading radiologist's report and the infusion catheter type. In the case of a non-concurrence, the Principal Investigator assessed the image and his or her assessment had final authority.
Time frame: at the time of image assessment
Population: One subject randomized to the 18 GA conventional catheter was excluded from the per-protocol analysis. Due to catheter placement, the required minimum flow rate of 5 mL/sec could not be achieved; contrast was administered at 4.5 mL/sec.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Acceptable Image Quality | 100 percentage of participants |
| 18 GA Conventional Catheter - Randomized | Acceptable Image Quality | 100 percentage of participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Acceptable Image Quality | 100 percentage of participants |
Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Randomized Subjects
Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified.
Time frame: at the time of image assessment
Population: 126 subjects had HU measurements available from abdominal CT imaging.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Randomized Subjects | 400.9 Hounsfield Units | Standard Deviation 69.8 |
| 18 GA Conventional Catheter - Randomized | Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Randomized Subjects | 402.5 Hounsfield Units | Standard Deviation 91.5 |
Automatic Injection Shutoff
The number of subjects who experienced injections with automatic injection shutoff is reported. Immediately after the power injection, the injection technician recorded whether or not an automatic injection shutoff occurred.
Time frame: immediately after contrast injection
Population: One subject randomized to the 18 GA conventional IV was excluded from the analysis as the required minimum flow rate could not be achieved due to IV placement. Non-randomized group analysis includes 1 subject with media extravasation with first study IV (adverse event). Subject was discontinued and not included in primary outcomes analysis.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Automatic Injection Shutoff | 0 participants |
| 18 GA Conventional Catheter - Randomized | Automatic Injection Shutoff | 0 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Automatic Injection Shutoff | 0 participants |
Catheter Dislodgement
The number of subjects who experienced partial or complete dislodgement of the catheter from the subject prior to power injection procedure completion is reported. The information about the IV catheter was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection.
Time frame: Immediately following power injection
Population: One subject randomized to the 18 GA conventional IV was excluded from the analysis as the required minimum flow rate could not be achieved due to IV placement. Non-randomized group analysis includes 1 subject with media extravasation with first study IV (adverse event). Subject was discontinued and not included in primary outcomes analysis.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Catheter Dislodgement | 0 participants |
| 18 GA Conventional Catheter - Randomized | Catheter Dislodgement | 0 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Catheter Dislodgement | 0 participants |
Catheter Insertion Success
Insertion is successful when the catheter can be flushed or infused to demonstrate patency, and there is no inadvertent administration of a solution or medication into the tissue surrounding the IV catheter. The number of participants with successfully inserted catheters after the first or second insertion attempt is reported. The clinician inserting the IV catheter made a clinical judgement as to whether the catheter was successfully placed; the protocol did not define more specific criteria. Catheter insertion success rates were determined from each IV insertion attempted in the Study.
Time frame: immediately after catheter insertion
Population: All subjects for whom an insertion was attempted are included in the analysis for insertion success. This includes some subjects that did not go on to complete the study.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Catheter Insertion Success | Success with second attempt | 5 participants |
| 20 GA BD Nexiva Diffusics - Randomized | Catheter Insertion Success | Unsuccessful after second attempt | 1 participants |
| 20 GA BD Nexiva Diffusics - Randomized | Catheter Insertion Success | Success with first insertion attempt | 96 participants |
| 18 GA Conventional Catheter - Randomized | Catheter Insertion Success | Success with second attempt | 2 participants |
| 18 GA Conventional Catheter - Randomized | Catheter Insertion Success | Unsuccessful after second attempt | 1 participants |
| 18 GA Conventional Catheter - Randomized | Catheter Insertion Success | Success with first insertion attempt | 100 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Catheter Insertion Success | Unsuccessful after second attempt | 5 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Catheter Insertion Success | Success with first insertion attempt | 20 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Catheter Insertion Success | Success with second attempt | 8 participants |
Catheter Integrity Failure
Catheter integrity failure generally refers to any portion of the device breaking or malfunctioning so that its proper function is no longer assured. In this study, the most relevant catheter integrity failures to be assessed were fluid leakage, tubing rupture, or tubing separation from the hub. The number of subjects who experienced injections with a catheter integrity failure during injection is reported. The information about the IV catheter was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection.
Time frame: immediately after power injection
Population: One subject randomized to the 18 GA conventional IV was excluded from the analysis as the required minimum flow rate could not be achieved due to IV placement. Non-randomized group analysis includes 1 subject with media extravasation with first study IV (adverse event). Subject was discontinued and not included in primary outcomes analysis.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Catheter Integrity Failure | 0 participants |
| 18 GA Conventional Catheter - Randomized | Catheter Integrity Failure | 0 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Catheter Integrity Failure | 0 participants |
Catheter Transfixation
The number of subjects who experienced catheter transfixation (the IV penetrating the opposite wall of the vein) is reported. This information was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection.
Time frame: immediately after power injection
Population: One subject randomized to the 18 GA conventional IV was excluded from the analysis as the required minimum flow rate could not be achieved due to IV placement. Non-randomized group analysis includes 1 subject with media extravasation with first study IV (adverse event). Subject was discontinued and not included in primary outcomes analysis.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Catheter Transfixation | 0 participants |
| 18 GA Conventional Catheter - Randomized | Catheter Transfixation | 0 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Catheter Transfixation | 0 participants |
Chest CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects)
Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified.
Time frame: at the time of image assessment
Population: 20 subjects had HU measurements available from thoracic CT imaging.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Chest CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects) | 428.9 Hounsfield Units | Standard Deviation 76.9 |
| 18 GA Conventional Catheter - Randomized | Chest CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects) | 402.2 Hounsfield Units | Standard Deviation 87.2 |
Complete Chest and Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects)
Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified.
Time frame: at the time of image assessment
Population: 54 subjects had HU measurements available from complete chest and abdomen CT imaging.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Complete Chest and Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects) | 373.2 Hounsfield Units | Standard Deviation 68.9 |
| 18 GA Conventional Catheter - Randomized | Complete Chest and Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects) | 387.7 Hounsfield Units | Standard Deviation 71.7 |
Extravasation of Contrast Media
The number of subjects who experienced injections with extravasation of contrast media is reported.
Time frame: upon contrast injection
Population: One subject randomized to the 18 GA IV was excluded from the per-protocol analysis as minimum flow rate could not be achieved. Non-randomized group analysis includes 1 subject with media extravasation with first study IV (adverse event). Subject was discontinued and not included in primary outcomes analysis.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Extravasation of Contrast Media | 0 participants |
| 18 GA Conventional Catheter - Randomized | Extravasation of Contrast Media | 0 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Extravasation of Contrast Media | 1 participants |
High Pressure Alarm
The number of subjects who experienced injections with activation of the high pressure alarm is reported. Immediately following the power injection, the clinician performing the power injection recorded whether or not the high pressure alarm sounded.
Time frame: immediately after contrast injection
Population: One subject randomized to the 18 GA conventional IV was excluded from the analysis as the required minimum flow rate could not be achieved due to IV placement. Non-randomized group analysis includes 1 subject with media extravasation with first study IV (adverse event). Subject was discontinued and not included in primary outcomes analysis.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | High Pressure Alarm | 0 participants |
| 18 GA Conventional Catheter - Randomized | High Pressure Alarm | 1 participants |
| 20 GA BD Nexiva Diffusics - Nonrandomized | High Pressure Alarm | 1 participants |
Maximum Flow Rate
Maximum flow rate guidelines provided in the BD Nexiva Diffusics Instructions for Use were to be followed. The maximum flow rate (milliliters per second, or mL/sec) utilized was recorded by the Radiology Technician immediately following the power injection of iodinated intravenous contrast media. Results are descriptively summarized; no formal acceptance criteria were specified by the protocol.
Time frame: immediately after power injection
Population: One subject randomized to the 18 GA conventional IV was excluded from the analysis as the required minimum flow rate could not be achieved due to IV placement. Non-randomized flow rate analysis only includes subjects who completed successful contrast delivery; one subject with extravasation upon delivery was discontinued and is not included.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| 20 GA BD Nexiva Diffusics - Randomized | Maximum Flow Rate | 5.58 mL/second | Standard Deviation 0.49 |
| 18 GA Conventional Catheter - Randomized | Maximum Flow Rate | 5.74 mL/second | Standard Deviation 0.66 |
| 20 GA BD Nexiva Diffusics - Nonrandomized | Maximum Flow Rate | 5.46 mL/second | Standard Deviation 0.44 |