Skip to content

Clinical Trial of Transtek Body Fat Analyzer (GBF-1251-B & Other 3 Models)

Verify the Functions and Efficiency of Transtek (Trade Mark) Body Fat Analyzer, GBF-1251-B, BF-1255-B, BF-1256-B, and GBF-1257-B

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01763775
Acronym
BS-BT
Enrollment
50
Registered
2013-01-09
Start date
2012-10-31
Completion date
2012-10-31
Last updated
2013-01-09

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Body Fat Disorder, Bone Mass, Weight

Keywords

weight, total body water, body fat, muscle mass, bone mass

Brief summary

The clinical protocol of the clinical testing of this device: 1. Objective of the test: To verify the functions and efficiency of devices. 2. Test methods and procedures: Comparison Test. 3. Device Under Test (DUT): Transtek Body Fat Analyzer, Model: GBF-1251-B, BF-1255-B, BF-1256-B, and GBF-1257-B. 4. Comparison device: Transtek, Glass Body Fat Analyzer, GBF-950-D. 5. Study endpoints: DUT and the comparison device are substantial equivalence. 6. Statistical methodology used: Description of statistical methods.

Detailed description

1. Test Purpose: The aim of clinical test is to collect weight, body fat, total body water, bone mass, and muscle mass reading value which measured by DUT and reference device. 2. Target Subject: The subject database shall contain at least 25 male and 25 female subjects and all above 18 years old. 3. Test procedures: 1. Record the name, age, gender, and height of each subject. 2. Let subject seated calmly for two minutes before test, dry the feet. 3. Use device to measure weight, body fat, total body water, bone mass, and muscle mass; record reading value. 4. For each subject, repeat 3) to get 3 pairs measurements. 5. Repeat step 1) to 4) for every device (DUT: GBF-1251-B, BF-1255-B, BF-1256-B, and GBF-1257-B; and reference device: GBF-950-D). 6. Repeat step 1) to 5) for each subject. 4. Note: 1. No motion and speaking are allowed during the measurement. 2. All DUT use the same algorithms and contact the patient at the same body locations (feet), and operate on the same frequency. 3. Test environment: Temperature: 22±1℃; Relative humidity: 40\ 70%.

Interventions

DEVICETranstek 125X

Measuring weight, total body water, body fat, muscle mass, and bone mass. Measure each participants by Transtek 125X (GBF-1251-B, BF-1255-B, BF-1256-B, GBF-1257-B).

DEVICETranstek 950D

Measuring weight, total body water, body fat, muscle mass, and bone mass. Measure each participants by Transtek 950D (GBF-950-D).

Sponsors

West China Hospital
CollaboratorOTHER
Leo Wang
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
FACTORIAL
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* male,female

Exclusion criteria

* below 18 years old

Design outcomes

Primary

MeasureTime frameDescription
BMI (Body Mass Index)5 DayBody Fat, Total Body Water, Muscle Mass, Bone Mass

Secondary

MeasureTime frameDescription
Weight5 DaysBody Weight

Countries

China

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026