ProspeCtive, nOn-randoMized, MulticENter Clinical Evaluation of Edwards Pericardial Bioprostheses With a New Tissue Treatment Platform (COMMENCE)

ProspeCtive, nOn-randoMized, MulticENter Clinical Evaluation of Edwards Pericardial Aortic and Mitral Bioprostheses (Models 11000A and 11000M) With a New Tissue Treatment Platform (COMMENCE)

Status
Completed
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01757665
Acronym
COMMENCE
Enrollment
777
Registered
2012-12-31
Start date
2012-12-11
Completion date
2026-03-23
Last updated
2026-05-07

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Aortic Stenosis, Aortic Valve Insufficiency, Heart Failure, Mitral Stenosis, Mitral Valve Insufficiency

Keywords

Aortic valve replacement, Mitral valve replacement

Brief summary

The objective of this trial is to confirm that the modifications to tissue processing, valve sterilization and packaging do not raise any new questions of safety and effectiveness in subjects who require replacement of their native or prosthetic aortic or mitral valve.

Detailed description

Multicenter, prospective, single arm trial - Up to seven hundred (700) aortic valve replacement (AVR) subjects and up to one hundred seventy-five (175) mitral valve replacement (MVR) subjects at up to forty (40) clinical sites will be enrolled. The trial will include male and female patients, 18 years or older, requiring replacement for a diseased, damaged, or malfunctioning native or prosthetic aortic or mitral valve. Patients will be followed and assessed after implant for up to 5 years.

Interventions

DEVICEEdwards Aortic and Mitral Bioprostheses Models 11000A and 11000M

Implant of an aortic valve, Model 11000A or mitral valve, Model 11000M

Sponsors

Edwards Lifesciences
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

Eligibility Criteria: Inclusion Criteria: 1. Is 18 years or older 2. Provides written informed consent prior to trial procedures 3. Agrees to attend all follow-up assessments for up to 5 years and is willing to comply with specified follow-up evaluations at clinical investigational sites that are participating in the COMMENCE trial and/or obtain the protocol-specified diagnostic tests at centers that are under the same IRB or the same healthcare system 4. Diagnosed with aortic or mitral valve disease requiring valve replacement based on pre- operative evaluation 5. Scheduled to undergo planned aortic or mitral valve replacement with or without concomitant bypass surgery 6. Scheduled to undergo planned aortic valve replacement with or without resection and replacement of the ascending aorta from the sinotubular junction and without the need for circulatory arrest for hemi arch or arch replacement

Exclusion criteria

A subject meeting any of the following criteria shall be excluded: 1. Requires emergency surgery 2. Requires planned multiple valve replacement/ repair (with the exception of mitral valve replacement with tricuspid valve repair) 3. Has prior valve surgery, which included implant of a bioprosthetic valve, mechanical valve, or annuloplasty ring that will remain in situ 4. Requires a surgical procedure outside of the cardiac area (e.g. vascular bypass) 5. Requires surgical replacement of the aortic root 6. Has active endocarditis/myocarditis or endocarditis/myocarditis within 3 months to the scheduled aortic or mitral valve replacement surgery 7. Has renal insufficiency as determined by creatinine (S-Cr) level ≥ 2.5 mg/dL or end -stage renal disease requiring chronic dialysis at screening visit 8. Has MRI or CT scan confirmed stroke, cerebrovascular accident (CVA) or transient ischemic attack (TIA) within 6 months (180 days) prior to planned valve surgery 9. Has acute myocardial infarction (MI) within 30 days prior to planned valve surgery 10. Has presence of non-cardiac disease limiting life expectancy to less than 12 months 11. Diagnosed with hypertrophic obstructive cardiomyopathy (HOCM) 12. Diagnosed with abnormal calcium metabolism and hyperparathyroidism 13. Exhibits left ventricular ejection fraction ≤ 20% as validated by diagnostic procedure prior to planned valve surgery 14. Echocardiographic evidence of an intra-cardiac mass, thrombus, or vegetation 15. Hemodynamic or respiratory instability requiring inotropic support, mechanical circulatory support, or mechanical ventilation within 30 days prior to planned valve surgery 16. Documented leukopenia (WBC \< 3.5x 10³/μL), acute anemia (Hgb \< 10.0 gm/dL or 6 mmol/L) or thrombocytopenia (platelet count \< 50x 10³/μL) accompanied by history of bleeding diathesis and coagulopathy 17. Has prior organ transplant or is currently an organ transplant candidate 18. Current or recent participation (within 6 weeks prior to surgery) in another drug or device trial 19. Was previously implanted with trial device (Model 11000A or Model 11000M) 20. Pregnant (female subject of childbearing potential only), lactating or planning to become pregnant during the duration of participation in trial 21. Currently incarcerated or unable to give voluntary informed consent 22. Documented history of substance (drug or alcohol) abuse within the last 5 years prior to implant 23. Requires concomitant left ventricular assist device (LVAD) placement

Design outcomes

Primary

MeasureTime frameDescription
Subjects With Structural Valve Deterioration1 Year Post ImplantThe rate of implanted subjects that experience structural valve deterioration (SVD) of the trial valve by the time of the post operative day (POD) 390 follow-up visit. Structural valve deterioration includes dysfunction or deterioration intrinsic to the valve. Examples of SVD includes complications such as wear, fracture, calcification, leaflet tear.

Secondary

MeasureTime frameDescription
Percentage of Subjects With Early Adverse EventsEvents occuring within 30 days of procedureNumber of subjects with early adverse events occurring within 30 days of procedure divided by the number of enrolled subjects times 100
Percentage of Late Adverse Events Divided by Late Patient YearsEvents occurring >= 31 days and up through 3 years post-implantNumber of late events divided by the total number of late patient years times 100. Late patient years are calculated from 31 days post-implant to the date of the last contact (follow up or adverse event).

Countries

Canada, Poland, United States

Contacts

PRINCIPAL_INVESTIGATORJohn Puskas, MD

Icahn School of Medicine at Mount Sinai

Participant flow

Participants by arm

ArmCount
Edwards Pericardial Aortic Bioprosthesis, Model 11000A
The Edwards Pericardial Aortic Bioprosthesis, Model 11000A, is indicated for patients who require replacement of their native or prosthetic aortic valve.
694
Edwards Pericardial Mitral Bioprosthesis, Model 11000M
The Edwards Pericardial Mitral Bioprosthesis, Model 11000M, is indicated for patients who require replacement of their native or prosthetic mitral valve.
83
Total777

Baseline characteristics

CharacteristicEdwards Pericardial Mitral Bioprosthesis, Model 11000MTotalEdwards Pericardial Aortic Bioprosthesis, Model 11000A
Age, Continuous68.9 years
STANDARD_DEVIATION 9.3
67.2 years
STANDARD_DEVIATION 11.4
67 years
STANDARD_DEVIATION 11.6
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants2 Participants2 Participants
Race (NIH/OMB)
Asian
2 Participants8 Participants6 Participants
Race (NIH/OMB)
Black or African American
2 Participants25 Participants23 Participants
Race (NIH/OMB)
More than one race
0 Participants2 Participants2 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants2 Participants2 Participants
Race (NIH/OMB)
Unknown or Not Reported
6 Participants43 Participants37 Participants
Race (NIH/OMB)
White
73 Participants695 Participants622 Participants
Sex: Female, Male
Female
49 Participants245 Participants196 Participants
Sex: Female, Male
Male
34 Participants532 Participants498 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
26 / 6898 / 82
other
Total, other adverse events
542 / 58966 / 82
serious
Total, serious adverse events
317 / 68958 / 82

Outcome results

Primary

Subjects With Structural Valve Deterioration

The rate of implanted subjects that experience structural valve deterioration (SVD) of the trial valve by the time of the post operative day (POD) 390 follow-up visit. Structural valve deterioration includes dysfunction or deterioration intrinsic to the valve. Examples of SVD includes complications such as wear, fracture, calcification, leaflet tear.

Time frame: 1 Year Post Implant

Population: The outcome is reported for subjects who received the Model 11000A or Model 11000M device where data is available. The SVD result combines the aortic and mitral implanted subjects. The study was powered for the combined cohort for this endpoint as there would be too few mitral subjects to report the arms separately.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubjects With Structural Valve Deterioration1 Participants
Comparison: The null hypothesis is that the rate of structural valve deterioration at one year is greater than 1%. The alternative hypothesis is that this rate is less than 1%.
Secondary

Percentage of Late Adverse Events Divided by Late Patient Years

Number of late events divided by the total number of late patient years times 100. Late patient years are calculated from 31 days post-implant to the date of the last contact (follow up or adverse event).

Time frame: Events occurring >= 31 days and up through 3 years post-implant

Population: The outcome is reported for subjects who received the Model 11000A or Model 11000M device where data is available.

ArmMeasureGroupValue (NUMBER)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsThromboembolism2.1 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsHemolysis0.0 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsAll Bleeding2.6 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsAll Paravalvular Leak (OPC)0.2 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsTrial Valve Related Mortality0.7 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsMajor Paravalvular Leak (OPC)0.1 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsMajor Bleeding1.4 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsNon-structural Valve Dysfunction0.0 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsValve Thrombosis0.0 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsReoperation0.2 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsEndocarditis0.6 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsExplant0.2 percentage of events/late patient years
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Late Adverse Events Divided by Late Patient YearsAll Cause Mortality2.2 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsExplant1.7 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsAll Cause Mortality6.1 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsTrial Valve Related Mortality0.9 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsThromboembolism1.7 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsValve Thrombosis0.0 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsAll Bleeding8.7 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsMajor Bleeding5.2 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsEndocarditis0.9 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsHemolysis0.0 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsAll Paravalvular Leak (OPC)0.0 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsMajor Paravalvular Leak (OPC)0.0 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsNon-structural Valve Dysfunction0.9 percentage of events/late patient years
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Late Adverse Events Divided by Late Patient YearsReoperation2.6 percentage of events/late patient years
Secondary

Percentage of Subjects With Early Adverse Events

Number of subjects with early adverse events occurring within 30 days of procedure divided by the number of enrolled subjects times 100

Time frame: Events occuring within 30 days of procedure

Population: The outcome is reported for subjects who received the Model 11000A or Model 11000M device where data is available.

ArmMeasureGroupValue (NUMBER)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsValve Thrombosis0.0 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsHemolysis0.0 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsAll Cause Mortality1.2 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsAll Paravalvular Leak (OPC)0.3 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsAll Bleeding0.9 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsMajor Paravalvular Leak (OPC)0.1 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsThromboembolism2.2 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsNon-structural Valve Dysfunction0.0 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsMajor Bleeding0.7 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsReoperation0.1 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsTrial Valve Related Mortality0.4 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsExplant0.0 percentage of subjects
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MPercentage of Subjects With Early Adverse EventsEndocarditis0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsExplant0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsTrial Valve Related Mortality1.2 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsThromboembolism2.4 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsValve Thrombosis0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsAll Bleeding1.2 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsMajor Bleeding1.2 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsEndocarditis0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsHemolysis0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsAll Paravalvular Leak (OPC)0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsMajor Paravalvular Leak (OPC)0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsNon-structural Valve Dysfunction0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsReoperation0.0 percentage of subjects
Edwards Pericardial Mitral Bioprosthesis, Model 11000MPercentage of Subjects With Early Adverse EventsAll Cause Mortality1.2 percentage of subjects
Other Pre-specified

Subject's Amount of Paravalvular Leak Over Time - 11000A

Paravalvular leak refers to blood flowing through a channel between the implanted artificial valve and the cardiac tissue as a result of inappropriate sealing. Paravalvular leak is evaluated by echocardiography over time. It is assessed on a scale from minimum of 0 to maximum of 4, where 0 = no leak, 1 = a trace leak, 2 = a mild leak, 3 = a moderate leak, and 4 = a severe leak. Higher numbers on the scale show a worsening outcome.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+2 Mild1 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year0 None12 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+2 Mild2 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months0 None20 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+2 Mild0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year0 None6 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+1 Trivial/Trace1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year0 None92 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+1 Trivial/Trace1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+2 Mild3 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months0 None118 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year0 None37 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+1 Trivial/Trace1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+2 Mild0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+2 Mild4 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year0 None61 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+1 Trivial/Trace3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+2 Mild2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year0 None152 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+1 Trivial/Trace3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+2 Mild3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months0 None191 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year0 None127 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+2 Mild2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+2 Mild1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+2 Mild1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+1 Trivial/Trace1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year0 None46 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months0 None182 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+1 Trivial/Trace1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months0 None87 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+2 Mild1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year0 None67 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+2 Mild2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year0 None16 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months0 None12 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+1 Trivial/Trace6 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+2 Mild3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+1 Trivial/Trace7 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year+2 Mild9 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months0 None610 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+1 Trivial/Trace8 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year+2 Mild10 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A2 Year0 None166 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A1 Year0 None450 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Paravalvular Leak Over Time - 11000A3 Months+3 Moderate1 Participants
Other Pre-specified

Subject's Amount of Paravalvular Leak Over Time - 11000M

Paravalvular leak refers to blood flowing through a channel between the implanted artificial valve and the cardiac tissue as a result of inappropriate sealing. Paravalvular leak is evaluated by echocardiography over time. It is assessed on a scale from minimum of 0 to maximum of 4, where 0 = no leak, 1 = a trace leak, 2 = a mild leak, 3 = a moderate leak, and 4 = a severe leak. Higher numbers on the scale show a worsening outcome.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months0 None5 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year0 None4 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months0 None26 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year0 None13 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year0 None25 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year0 None21 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months0 None23 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year0 None7 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months0 None15 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year0 None11 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year0 None2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year0 None2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months0 None6 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year0 None6 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year0 None24 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months0 None75 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M3 Months+1 Trivial/Trace0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Paravalvular Leak Over Time - 11000M1 Year0 None67 Participants
Other Pre-specified

Subject's Amount of Total Valvular Regurgitation Over Time - 11000A

Valvular regurgitation occurs when the valve in the heart does not close tightly allowing some of the blood that was pumped out of the heart to leak back into it. Valvular regurgitation is evaluated by echocardiography over time. It is assessed on a scale from 0 to 4, where 0 represents no regurgitation and 4 represents severe regurgitation.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year0 None11 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+2 Mild2 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months0 None17 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+1 Trivial/Trace3 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year0 None4 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+2 Mild2 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+1 Trivial/Trace3 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+2 Mild1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+1 Trivial/Trace5 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months0 None106 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+1 Trivial/Trace11 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+2 Mild1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year0 None88 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+1 Trivial/Trace6 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+2 Mild2 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+3 Moderate1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year0 None32 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+2 Mild5 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+1 Trivial/Trace25 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months0 None166 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+2 Mild8 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+2 Mild3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year0 None125 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+4 Severe1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+1 Trivial/Trace11 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year0 None51 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+1 Trivial/Trace26 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+2 Mild3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year0 None112 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+1 Trivial/Trace12 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+2 Mild7 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+1 Trivial/Trace20 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year0 None38 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+1 Trivial/Trace9 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+2 Mild2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months0 None162 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+2 Mild4 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+1 Trivial/Trace17 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year0 None11 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+1 Trivial/Trace5 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months0 None69 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year0 None49 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+2 Mild2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+2 Mild2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+1 Trivial/Trace20 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+2 Mild1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+1 Trivial/Trace1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months0 None11 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+1 Trivial/Trace64 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+2 Mild14 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+1 Trivial/Trace80 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+2 Mild10 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year0 None136 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months0 None531 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year0 None385 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+2 Mild21 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A1 Year+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+1 Trivial/Trace32 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A2 Year+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Amount of Total Valvular Regurgitation Over Time - 11000A3 Months+4 Severe1 Participants
Other Pre-specified

Subject's Amount of Total Valvular Regurgitation Over Time - 11000M

Mitral valvular regurgitation occurs when the mitral valve in the heart does not close tightly allowing some of the blood that was pumped out of the heart to leak back into it. Mitral valvular regurgitation is evaluated by echocardiography over time. It is assessed on a scale from 0 to 4, where 0 represents no regurgitation and 4 represents severe regurgitation.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+1 Trivial/Trace1 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year0 None3 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+1 Trivial/Trace1 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months0 None4 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year0 None6 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+3 Moderate1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+2 Mild3 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+1 Trivial/Trace5 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+3 Moderate1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+3 Moderate1 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year0 None15 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+2 Mild4 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+2 Mild4 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+1 Trivial/Trace7 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months0 None15 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months0 None18 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+1 Trivial/Trace5 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year0 None13 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+1 Trivial/Trace7 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year0 None6 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+1 Trivial/Trace1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+1 Trivial/Trace1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+4 Severe1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+1 Trivial/Trace4 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+1 Trivial/Trace3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months0 None10 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year0 None1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year0 None8 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+2 Mild2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year0 None3 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+2 Mild1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+3 Moderate0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year0 None0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+2 Mild0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months0 None4 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+1 Trivial/Trace2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months0 None51 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+1 Trivial/Trace18 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+3 Moderate2 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year0 None42 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+2 Mild4 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+2 Mild5 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M3 Months+2 Mild5 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year0 None13 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+4 Severe1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+4 Severe0 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+3 Moderate1 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M1 Year+1 Trivial/Trace19 Participants
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Amount of Total Valvular Regurgitation Over Time - 11000M2 Year+1 Trivial/Trace6 Participants
Other Pre-specified

Subject's Average Cardiac Index Over Time- 11000A

Cardiac index is an assessment that divides the cardiac output from left ventricle in one minute by the person's body surface area (BSA), thus relating heart performance to the size of the individual.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Index Over Time- 11000A1 Year2.6 L/min/m^2Standard Deviation 0.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Index Over Time- 11000A2 Year2.2 L/min/m^2Standard Deviation 0.8
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Index Over Time- 11000A3 Months3.0 L/min/m^2Standard Deviation 0.8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Index Over Time- 11000A1 Year2.5 L/min/m^2Standard Deviation 1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Index Over Time- 11000A3 Months2.4 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Index Over Time- 11000A2 Year2.3 L/min/m^2Standard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Index Over Time- 11000A1 Year2.2 L/min/m^2Standard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Index Over Time- 11000A3 Months2.4 L/min/m^2Standard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Index Over Time- 11000A2 Year2.2 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Index Over Time- 11000A2 Year2.3 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Index Over Time- 11000A1 Year2.2 L/min/m^2Standard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Index Over Time- 11000A3 Months2.3 L/min/m^2Standard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Index Over Time- 11000A1 Year2.4 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Index Over Time- 11000A2 Year2.0 L/min/m^2Standard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Index Over Time- 11000A3 Months2.4 L/min/m^2Standard Deviation 0.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Index Over Time- 11000A3 Months2.0 L/min/m^2Standard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Cardiac Index Over Time- 11000A2 Year2.2 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Cardiac Index Over Time- 11000A1 Year2.3 L/min/m^2Standard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Cardiac Index Over Time- 11000A3 Months2.4 L/min/m^2Standard Deviation 0.7
Other Pre-specified

Subject's Average Cardiac Index Over Time - 11000M

Cardiac index is an assessment that divides the cardiac output from left ventricle in one minute by the person's body surface area(BSA), thus relating heart performance to the size of the individual.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Index Over Time - 11000M1 Year2.0 L/min/m^2Standard Deviation 0.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Index Over Time - 11000M3 Months2.2 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Index Over Time - 11000M1 Year1.9 L/min/m^2Standard Deviation 0.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Index Over Time - 11000M3 Months2.1 L/min/m^2Standard Deviation 0.6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Index Over Time - 11000M2 Year1.9 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Index Over Time - 11000M1 Year2.2 L/min/m^2Standard Deviation 0.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Index Over Time - 11000M3 Months2.2 L/min/m^2Standard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Index Over Time - 11000M2 Year2.1 L/min/m^2Standard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Index Over Time - 11000M2 Year1.9 L/min/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Index Over Time - 11000M3 Months1.9 L/min/m^2Standard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Index Over Time - 11000M1 Year1.9 L/min/m^2Standard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Index Over Time - 11000M3 Months2.1 L/min/m^2Standard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Index Over Time - 11000M1 Year1.8 L/min/m^2Standard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Index Over Time - 11000M2 Year1.6 L/min/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Index Over Time - 11000M1 Year2.0 L/min/m^2Standard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Index Over Time - 11000M3 Months2.1 L/min/m^2Standard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Index Over Time - 11000M2 Year1.9 L/min/m^2Standard Deviation 0.6
Other Pre-specified

Subject's Average Cardiac Output Over Time - 11000A

The amount of blood the heart pumps through the circulatory system in a minute.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Output Over Time - 11000A3 Months5.1 Liters/minuteStandard Deviation 1.3
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Output Over Time - 11000A1 Year4.5 Liters/minuteStandard Deviation 0.9
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Output Over Time - 11000A2 Year3.7 Liters/minuteStandard Deviation 1.1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Output Over Time - 11000A1 Year4.7 Liters/minuteStandard Deviation 1.8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Output Over Time - 11000A3 Months4.5 Liters/minuteStandard Deviation 1.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Output Over Time - 11000A2 Year4.3 Liters/minuteStandard Deviation 1.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Output Over Time - 11000A2 Year4.4 Liters/minuteStandard Deviation 1.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Output Over Time - 11000A3 Months4.7 Liters/minuteStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Output Over Time - 11000A1 Year4.4 Liters/minuteStandard Deviation 1.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Output Over Time - 11000A3 Months4.8 Liters/minuteStandard Deviation 1.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Output Over Time - 11000A2 Year4.9 Liters/minuteStandard Deviation 1.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Output Over Time - 11000A1 Year4.8 Liters/minuteStandard Deviation 1.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Output Over Time - 11000A3 Months5.2 Liters/minuteStandard Deviation 1.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Output Over Time - 11000A1 Year5.4 Liters/minuteStandard Deviation 1.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Output Over Time - 11000A2 Year4.4 Liters/minuteStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Output Over Time - 11000A3 Months4.4 Liters/minuteStandard Deviation 1.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Cardiac Output Over Time - 11000A2 Year4.5 Liters/minuteStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Cardiac Output Over Time - 11000A3 Months4.8 Liters/minuteStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Cardiac Output Over Time - 11000A1 Year4.7 Liters/minuteStandard Deviation 1.6
Other Pre-specified

Subject's Average Cardiac Output Over Time - 11000M

The amount of blood the heart pumps through the circulatory system in a minute.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Output Over Time - 11000M1 Year3.7 Liters/MinuteStandard Deviation 1
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Cardiac Output Over Time - 11000M3 Months4.0 Liters/MinuteStandard Deviation 1.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Output Over Time - 11000M1 Year3.3 Liters/MinuteStandard Deviation 1.1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Output Over Time - 11000M3 Months3.6 Liters/MinuteStandard Deviation 0.9
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Cardiac Output Over Time - 11000M2 Year3.3 Liters/MinuteStandard Deviation 1.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Output Over Time - 11000M1 Year4.4 Liters/MinuteStandard Deviation 1.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Output Over Time - 11000M3 Months4.3 Liters/MinuteStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Cardiac Output Over Time - 11000M2 Year4.1 Liters/MinuteStandard Deviation 1.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Output Over Time - 11000M2 Year3.6 Liters/MinuteStandard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Output Over Time - 11000M3 Months3.9 Liters/MinuteStandard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Cardiac Output Over Time - 11000M1 Year3.8 Liters/MinuteStandard Deviation 1.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Output Over Time - 11000M3 Months4.6 Liters/MinuteStandard Deviation 1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Output Over Time - 11000M1 Year3.8 Liters/MinuteStandard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Cardiac Output Over Time - 11000M2 Year3.3 Liters/MinuteStandard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Output Over Time - 11000M1 Year3.8 Liters/MinuteStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Output Over Time - 11000M3 Months4.0 Liters/MinuteStandard Deviation 1.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Cardiac Output Over Time - 11000M2 Year3.6 Liters/MinuteStandard Deviation 1.1
Other Pre-specified

Subject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A

Effective orifice area index represents the minimal cross-sectional area of the blood flow downstream of the aortic valve divided by the person's body surface area. Effective orifice area index is evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A1 Year0.6 cm^2/m^2Standard Deviation 0.1
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A2 Year0.6 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A3 Months0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A1 Year0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A3 Months0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A2 Year0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A1 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A2 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A2 Year0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A1 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A1 Year1.0 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A2 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A3 Months1.1 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A3 Months1.1 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A2 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A1 Year0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.3
Other Pre-specified

Subject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M

Effective orifice area index represents the minimal cross-sectional area of the blood flow downstream of the mitral valve divided by the person's body surface area. Effective orifice area index is evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M1 Year0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M3 Months0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M1 Year0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M3 Months0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M2 Year0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M1 Year0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M3 Months0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M2 Year0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M2 Year0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M3 Months0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M1 Year0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M3 Months0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M1 Year0.7 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M2 Year0.6 cm^2/m^2Standard Deviation 0.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M1 Year0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M3 Months0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Index (EOAI) Measurements - 11000M2 Year0.7 cm^2/m^2Standard Deviation 0.3
Other Pre-specified

Subject's Average Effective Orifice Area Measurements - 11000A

Effective orifice area represents the cross-sectional area of the blood flow downstream of the aortic valve. Effective orifice area is evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Measurements - 11000A1 Year1.1 Centimeters SquaredStandard Deviation 0.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Measurements - 11000A2 Year1.0 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Measurements - 11000A3 Months1.4 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Measurements - 11000A1 Year1.3 Centimeters SquaredStandard Deviation 0.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Measurements - 11000A3 Months1.5 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Measurements - 11000A2 Year1.3 Centimeters SquaredStandard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Measurements - 11000A1 Year1.6 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Measurements - 11000A3 Months1.7 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Measurements - 11000A2 Year1.6 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Measurements - 11000A2 Year1.9 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Measurements - 11000A1 Year1.8 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Measurements - 11000A3 Months1.8 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Measurements - 11000A1 Year2.2 Centimeters SquaredStandard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Measurements - 11000A2 Year1.8 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Measurements - 11000A3 Months2.3 Centimeters SquaredStandard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Measurements - 11000A3 Months2.5 Centimeters SquaredStandard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Effective Orifice Area Measurements - 11000A2 Year1.6 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Effective Orifice Area Measurements - 11000A1 Year1.7 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Effective Orifice Area Measurements - 11000A3 Months1.8 Centimeters SquaredStandard Deviation 0.6
Other Pre-specified

Subject's Average Effective Orifice Area Measurements - 11000M

Effective orifice area represents the cross-sectional area of the blood flow downstream of the mitral valve. Effective orifice area is evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Measurements - 11000M1 Year1.1 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Effective Orifice Area Measurements - 11000M3 Months1.2 Centimeters SquaredStandard Deviation 0.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Measurements - 11000M1 Year1.2 Centimeters SquaredStandard Deviation 0.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Measurements - 11000M3 Months1.4 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Effective Orifice Area Measurements - 11000M2 Year1.3 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Measurements - 11000M1 Year1.5 Centimeters SquaredStandard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Measurements - 11000M3 Months1.6 Centimeters SquaredStandard Deviation 0.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Effective Orifice Area Measurements - 11000M2 Year1.4 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Measurements - 11000M2 Year1.6 Centimeters SquaredStandard Deviation 0
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Measurements - 11000M3 Months1.6 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Effective Orifice Area Measurements - 11000M1 Year1.4 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Measurements - 11000M3 Months1.8 Centimeters SquaredStandard Deviation 0.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Measurements - 11000M1 Year1.5 Centimeters SquaredStandard Deviation 0.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Effective Orifice Area Measurements - 11000M2 Year1.2 Centimeters SquaredStandard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Measurements - 11000M1 Year1.4 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Measurements - 11000M3 Months1.5 Centimeters SquaredStandard Deviation 0.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Effective Orifice Area Measurements - 11000M2 Year1.4 Centimeters SquaredStandard Deviation 0.4
Other Pre-specified

Subject's Average Hematocrit Percentage

Laboratory Analysis of Hematocrit Percentage on blood drawn from subjects. Hematocrit is the proportion of red blood cells to the plasma (liquid portion of the blood).

Time frame: Baseline, Discharge, 3 Months, 1 Year and 2 Year

Population: The outcome is reported for subjects who received the Model 11000A or 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hematocrit PercentageDischarge30.1 percentage of red blood cellsStandard Deviation 4.1
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hematocrit Percentage1 Year41.9 percentage of red blood cellsStandard Deviation 4.1
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hematocrit Percentage3 Months41.0 percentage of red blood cellsStandard Deviation 4.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hematocrit Percentage2 Year42.0 percentage of red blood cellsStandard Deviation 3.9
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hematocrit PercentageBaseline41.0 percentage of red blood cellsStandard Deviation 4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hematocrit Percentage2 Year42.4 percentage of red blood cellsStandard Deviation 4.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hematocrit PercentageBaseline40.3 percentage of red blood cellsStandard Deviation 4.6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hematocrit PercentageDischarge30.1 percentage of red blood cellsStandard Deviation 3.9
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hematocrit Percentage3 Months39.3 percentage of red blood cellsStandard Deviation 5.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hematocrit Percentage1 Year41.3 percentage of red blood cellsStandard Deviation 4.8
Other Pre-specified

Subject's Average Hemoglobin Count

Laboratory Analysis of Hemoglobin Count on blood drawn from subjects. Hemoglobin is an oxygen-carrying protein in red blood cells.

Time frame: Baseline, Discharge, 3 Months, 1 Year and 2 Year

Population: The outcome is reported for subjects who received the Model 11000A or 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hemoglobin CountDischarge10.0 g/dlStandard Deviation 1.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hemoglobin Count1 Year14.0 g/dlStandard Deviation 1.7
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hemoglobin Count3 Months13.5 g/dlStandard Deviation 1.6
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hemoglobin Count2 Year14.0 g/dlStandard Deviation 1.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Hemoglobin CountBaseline13.8 g/dlStandard Deviation 1.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hemoglobin Count2 Year14.0 g/dlStandard Deviation 1.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hemoglobin CountBaseline13.3 g/dlStandard Deviation 1.6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hemoglobin CountDischarge9.9 g/dlStandard Deviation 1.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hemoglobin Count3 Months12.7 g/dlStandard Deviation 1.9
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Hemoglobin Count1 Year13.6 g/dlStandard Deviation 1.6
Other Pre-specified

Subject's Average International Normalized Ratio

Laboratory Analysis of International Normalized Ratio (INR) on blood drawn from subjects. The INR is a calculation based on results of a prothrombin time (PT). The PT is a blood test that measures the time it takes for the plasma (liquid portion of the blood) to clot. INR results will vary according to a person's age, the medicines they take, and any health problems they may have. In general, the higher the INR number, the longer it takes for the blood to clot. In healthy people an INR of 1.1 or below is considered normal. An INR range of 2.0 to 3.0 is generally an effective therapeutic range for people taking blood thinner medication.

Time frame: Baseline, Discharge, 3 Months, 1 Year and 2 Year

Population: The outcome is reported for subjects who received the Model 11000A or 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average International Normalized RatioDischarge1.3 RatioStandard Deviation 0.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average International Normalized Ratio1 Year1.2 RatioStandard Deviation 0.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average International Normalized Ratio3 Months1.3 RatioStandard Deviation 0.7
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average International Normalized Ratio2 Year1.2 RatioStandard Deviation 0.6
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average International Normalized RatioBaseline1.1 RatioStandard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average International Normalized Ratio2 Year1.6 RatioStandard Deviation 0.8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average International Normalized RatioBaseline1.4 RatioStandard Deviation 0.7
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average International Normalized RatioDischarge1.7 RatioStandard Deviation 0.6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average International Normalized Ratio3 Months1.9 RatioStandard Deviation 0.8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average International Normalized Ratio1 Year1.7 RatioStandard Deviation 0.8
Other Pre-specified

Subject's Average Mean Gradient Measurements - 11000A

Mean gradient is the average flow of blood through the aortic valve measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. Mean gradient values depend on the size and type of valve.

Time frame: 3 Months, 1 Year, and 2 Year Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Mean Gradient Measurements - 11000A1 Year17.6 mmHgStandard Deviation 7.8
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Mean Gradient Measurements - 11000A2 Year16.1 mmHgStandard Deviation 6.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Mean Gradient Measurements - 11000A3 Months13.8 mmHgStandard Deviation 6.1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Mean Gradient Measurements - 11000A1 Year12.6 mmHgStandard Deviation 4.7
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Mean Gradient Measurements - 11000A3 Months10.7 mmHgStandard Deviation 4.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Mean Gradient Measurements - 11000A2 Year11.2 mmHgStandard Deviation 4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Mean Gradient Measurements - 11000A1 Year10.1 mmHgStandard Deviation 3.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Mean Gradient Measurements - 11000A3 Months9.4 mmHgStandard Deviation 3.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Mean Gradient Measurements - 11000A2 Year9.9 mmHgStandard Deviation 4.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Mean Gradient Measurements - 11000A2 Year9.2 mmHgStandard Deviation 3.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Mean Gradient Measurements - 11000A1 Year9.6 mmHgStandard Deviation 5.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Mean Gradient Measurements - 11000A3 Months8.7 mmHgStandard Deviation 3.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Mean Gradient Measurements - 11000A1 Year8.2 mmHgStandard Deviation 3.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Mean Gradient Measurements - 11000A2 Year8.3 mmHgStandard Deviation 3.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Mean Gradient Measurements - 11000A3 Months6.8 mmHgStandard Deviation 2.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Mean Gradient Measurements - 11000A3 Months4.7 mmHgStandard Deviation 1.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Mean Gradient Measurements - 11000A2 Year10.1 mmHgStandard Deviation 4.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Mean Gradient Measurements - 11000A1 Year10.4 mmHgStandard Deviation 4.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Mean Gradient Measurements - 11000A3 Months9.1 mmHgStandard Deviation 4
Other Pre-specified

Subject's Average Mean Gradient Measurements - 11000M

Mean gradient is the average flow of blood through the mitral valve measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. Mean gradient values depend on the size and type of valve.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Mean Gradient Measurements - 11000M1 Year4.9 mmHgStandard Deviation 1.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Mean Gradient Measurements - 11000M3 Months5.6 mmHgStandard Deviation 0.7
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Mean Gradient Measurements - 11000M1 Year4.1 mmHgStandard Deviation 1.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Mean Gradient Measurements - 11000M3 Months4.1 mmHgStandard Deviation 1.7
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Mean Gradient Measurements - 11000M2 Year3.9 mmHgStandard Deviation 1.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Mean Gradient Measurements - 11000M1 Year4.1 mmHgStandard Deviation 1.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Mean Gradient Measurements - 11000M3 Months4.0 mmHgStandard Deviation 1.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Mean Gradient Measurements - 11000M2 Year3.8 mmHgStandard Deviation 1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Mean Gradient Measurements - 11000M2 Year3.5 mmHgStandard Deviation 0.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Mean Gradient Measurements - 11000M3 Months3.4 mmHgStandard Deviation 1.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Mean Gradient Measurements - 11000M1 Year3.9 mmHgStandard Deviation 2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Mean Gradient Measurements - 11000M3 Months3.3 mmHgStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Mean Gradient Measurements - 11000M1 Year3.3 mmHgStandard Deviation 1.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Mean Gradient Measurements - 11000M2 Year3.4 mmHgStandard Deviation 2.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Mean Gradient Measurements - 11000M1 Year4.0 mmHgStandard Deviation 1.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Mean Gradient Measurements - 11000M3 Months4.0 mmHgStandard Deviation 1.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Mean Gradient Measurements - 11000M2 Year3.8 mmHgStandard Deviation 1.5
Other Pre-specified

Subject's Average Partial Thromboplastin Time

Laboratory Analysis of partial thromboplastin time (PTT) on blood drawn from subjects. PTT is a blood test that looks at how long it takes for the blood to clot.

Time frame: Baseline, Discharge, 3 Months, 1 Year and 2 Year

Population: The outcome is reported for subjects who received the Model 11000A or 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Partial Thromboplastin TimeDischarge32.1 SecondsStandard Deviation 8.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Partial Thromboplastin Time1 Year29.4 SecondsStandard Deviation 5.7
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Partial Thromboplastin Time3 Months31.2 SecondsStandard Deviation 15.8
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Partial Thromboplastin Time2 Year31.4 SecondsStandard Deviation 14.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Partial Thromboplastin TimeBaseline30.0 SecondsStandard Deviation 6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Partial Thromboplastin Time2 Year33.7 SecondsStandard Deviation 9.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Partial Thromboplastin TimeBaseline35.1 SecondsStandard Deviation 14.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Partial Thromboplastin TimeDischarge37.2 SecondsStandard Deviation 14.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Partial Thromboplastin Time3 Months36.0 SecondsStandard Deviation 8.1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Partial Thromboplastin Time1 Year35.3 SecondsStandard Deviation 14.3
Other Pre-specified

Subject's Average Peak Gradients Measurements Over Time - 11000A

Peak gradient is the maximum value measured of flow of blood through the aortic valve as measured in millimeters of mercury. Gradients are evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Peak Gradients Measurements Over Time - 11000A1 Year35.0 mmHgStandard Deviation 16.3
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Peak Gradients Measurements Over Time - 11000A2 Year31.7 mmHgStandard Deviation 12
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Peak Gradients Measurements Over Time - 11000A3 Month27.5 mmHgStandard Deviation 12.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Peak Gradients Measurements Over Time - 11000A1 Year25.2 mmHgStandard Deviation 8.9
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Peak Gradients Measurements Over Time - 11000A3 Month21.7 mmHgStandard Deviation 8.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Peak Gradients Measurements Over Time - 11000A2 Year22.0 mmHgStandard Deviation 7.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Peak Gradients Measurements Over Time - 11000A1 Year20.1 mmHgStandard Deviation 7.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Peak Gradients Measurements Over Time - 11000A3 Month18.8 mmHgStandard Deviation 6.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Peak Gradients Measurements Over Time - 11000A2 Year19.8 mmHgStandard Deviation 8.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Peak Gradients Measurements Over Time - 11000A2 Year18.2 mmHgStandard Deviation 6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Peak Gradients Measurements Over Time - 11000A1 Year19.2 mmHgStandard Deviation 8.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Peak Gradients Measurements Over Time - 11000A3 Month17.5 mmHgStandard Deviation 6.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Peak Gradients Measurements Over Time - 11000A1 Year16.7 mmHgStandard Deviation 6.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Peak Gradients Measurements Over Time - 11000A2 Year16.4 mmHgStandard Deviation 6.6
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Peak Gradients Measurements Over Time - 11000A3 Month13.8 mmHgStandard Deviation 4.9
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Peak Gradients Measurements Over Time - 11000A3 Month10.1 mmHgStandard Deviation 3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Peak Gradients Measurements Over Time - 11000A2 Year20.1 mmHgStandard Deviation 8.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Peak Gradients Measurements Over Time - 11000A1 Year20.9 mmHgStandard Deviation 9.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Peak Gradients Measurements Over Time - 11000A3 Month18.4 mmHgStandard Deviation 7.6
Other Pre-specified

Subject's Average Peak Gradients Measurements Over Time - 11000M

Peak gradient is the maximum value measured of flow of blood through the mitral valve as measured in millimeters of mercury. Gradients are evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Peak Gradients Measurements Over Time - 11000M1 Year10.6 mmHgStandard Deviation 3
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Peak Gradients Measurements Over Time - 11000M3 Months12.0 mmHgStandard Deviation 1.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Peak Gradients Measurements Over Time - 11000M1 Year9.9 mmHgStandard Deviation 3.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Peak Gradients Measurements Over Time - 11000M3 Months9.8 mmHgStandard Deviation 4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Peak Gradients Measurements Over Time - 11000M2 Year9.5 mmHgStandard Deviation 4.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Peak Gradients Measurements Over Time - 11000M1 Year10.4 mmHgStandard Deviation 3.7
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Peak Gradients Measurements Over Time - 11000M3 Months10.2 mmHgStandard Deviation 4.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Peak Gradients Measurements Over Time - 11000M2 Year10.5 mmHgStandard Deviation 3.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Peak Gradients Measurements Over Time - 11000M2 Year6.6 mmHgStandard Deviation 1.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Peak Gradients Measurements Over Time - 11000M3 Months9.0 mmHgStandard Deviation 2.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Peak Gradients Measurements Over Time - 11000M1 Year10.2 mmHgStandard Deviation 5.5
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Peak Gradients Measurements Over Time - 11000M3 Months17.4 mmHgStandard Deviation 22.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Peak Gradients Measurements Over Time - 11000M1 Year9.0 mmHgStandard Deviation 4.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Peak Gradients Measurements Over Time - 11000M2 Year8.3 mmHgStandard Deviation 3.4
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Peak Gradients Measurements Over Time - 11000M1 Year10.1 mmHgStandard Deviation 3.8
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Peak Gradients Measurements Over Time - 11000M3 Months10.5 mmHgStandard Deviation 7.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Peak Gradients Measurements Over Time - 11000M2 Year9.4 mmHgStandard Deviation 4
Other Pre-specified

Subject's Average Performance Index Measurements - 11000A

Performance index is defined as the subject's effective orifice area (the cross sectional area of the blood flow downstream of the aortic valve) divided by the subject's native orifice area. Effective orifice area is evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Performance Index Measurements - 11000A1 Year0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Performance Index Measurements - 11000A2 Year0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Performance Index Measurements - 11000A3 Months1.1 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Performance Index Measurements - 11000A1 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Performance Index Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Performance Index Measurements - 11000A2 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Performance Index Measurements - 11000A1 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Performance Index Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Performance Index Measurements - 11000A2 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Performance Index Measurements - 11000A2 Year0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Performance Index Measurements - 11000A1 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Performance Index Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Performance Index Measurements - 11000A1 Year0.9 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Performance Index Measurements - 11000A2 Year0.7 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Performance Index Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Performance Index Measurements - 11000A3 Months0.8 cm^2/m^2Standard Deviation 0.3
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Performance Index Measurements - 11000A2 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Performance Index Measurements - 11000A1 Year0.8 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - TotalSubject's Average Performance Index Measurements - 11000A3 Months0.9 cm^2/m^2Standard Deviation 0.2
Other Pre-specified

Subject's Average Performance Index Measurements - 11000M

Performance index is defined as the subject's effective orifice area (the crosssectional area of the blood flow downstream of the mitral valve) divided by the subject's native orifice area. Effective orifice area is evaluated by echocardiography over time.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Performance Index Measurements - 11000M1 Year0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Performance Index Measurements - 11000M3 Months0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Performance Index Measurements - 11000M2 Year0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Performance Index Measurements - 11000M1 Year0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Performance Index Measurements - 11000M3 Months0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Performance Index Measurements - 11000M2 Year0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Performance Index Measurements - 11000M3 Months0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 23mmSubject's Average Performance Index Measurements - 11000M1 Year0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Performance Index Measurements - 11000M3 Months0.5 cm^2/m^2Standard Deviation 0.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Performance Index Measurements - 11000M1 Year0.5 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 25mmSubject's Average Performance Index Measurements - 11000M2 Year0.6 cm^2/m^2Standard Deviation 0
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Performance Index Measurements - 11000M3 Months0.6 cm^2/m^2Standard Deviation 0.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Performance Index Measurements - 11000M2 Year0.4 cm^2/m^2Standard Deviation 0.1
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 27mmSubject's Average Performance Index Measurements - 11000M1 Year0.5 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Performance Index Measurements - 11000M1 Year0.5 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Performance Index Measurements - 11000M3 Months0.6 cm^2/m^2Standard Deviation 0.2
Edwards Pericardial Aortic Bioprosthesis, Model 11000A - 29mmSubject's Average Performance Index Measurements - 11000M2 Year0.6 cm^2/m^2Standard Deviation 0.2
Other Pre-specified

Subject's Average Plasma Free Hemoglobin

Laboratory Analysis of Plasma Free Hemoglobin on blood drawn from subjects. This blood test measures the level of free hemoglobin in the plasma (liquid portion of the blood).

Time frame: Baseline, Discharge, 3 Months, 1 Year and 2 Year

Population: The outcome is reported for subjects who received the Model 11000A or 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Plasma Free HemoglobinDischarge7.3 mg/dlStandard Deviation 12.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Plasma Free Hemoglobin1 Year13.8 mg/dlStandard Deviation 25.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Plasma Free Hemoglobin3 Months12.4 mg/dlStandard Deviation 22.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Plasma Free Hemoglobin2 Year17.3 mg/dlStandard Deviation 33.9
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Plasma Free HemoglobinBaseline11.6 mg/dlStandard Deviation 18.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Plasma Free Hemoglobin2 Year20.5 mg/dlStandard Deviation 33.9
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Plasma Free HemoglobinBaseline19.8 mg/dlStandard Deviation 32.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Plasma Free HemoglobinDischarge7.2 mg/dlStandard Deviation 6.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Plasma Free Hemoglobin3 Months10.6 mg/dlStandard Deviation 12.1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Plasma Free Hemoglobin1 Year8.6 mg/dlStandard Deviation 12.3
Other Pre-specified

Subject's Average Platelet Count

Laboratory Analysis of Platelet Count on blood drawn from subjects; platelets help with blood clotting.

Time frame: Baseline, Discharge, 3 Months, 1 Year, and 2 Years

Population: The outcome is reported for subjects who received the Model 11000A or 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Platelet CountDischarge217.7 10^3 platelets per microliterStandard Deviation 90.1
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Platelet Count1 Year207.7 10^3 platelets per microliterStandard Deviation 65.8
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Platelet Count3 Months215.3 10^3 platelets per microliterStandard Deviation 62.6
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Platelet Count2 Year200.6 10^3 platelets per microliterStandard Deviation 54.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Platelet CountBaseline208.6 10^3 platelets per microliterStandard Deviation 59
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Platelet Count2 Year204.0 10^3 platelets per microliterStandard Deviation 54.1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Platelet CountBaseline203.6 10^3 platelets per microliterStandard Deviation 52.1
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Platelet CountDischarge196.0 10^3 platelets per microliterStandard Deviation 75.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Platelet Count3 Months221.0 10^3 platelets per microliterStandard Deviation 71.8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Platelet Count1 Year207.8 10^3 platelets per microliterStandard Deviation 51.7
Other Pre-specified

Subject's Average Prothrombin Time

Laboratory Analysis of Prothrombin Time (PT) on blood drawn from subjects. The PT is a blood test that measures the time it takes for the plasma (liquid portion of the blood) to clot.

Time frame: Baseline, Discharge, 3 Months, 1 Year, and 2 Years

Population: The outcome is reported for subjects who received the Model 11000A or 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Prothrombin Time3 Months15.2 SecondsStandard Deviation 7.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Prothrombin TimeDischarge15.0 SecondsStandard Deviation 5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Prothrombin Time1 Year13.8 SecondsStandard Deviation 5.7
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Prothrombin Time2 Year14.1 SecondsStandard Deviation 6.4
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Prothrombin TimeBaseline12.7 SecondsStandard Deviation 2.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Prothrombin Time2 Year17.3 SecondsStandard Deviation 7.9
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Prothrombin TimeBaseline16.2 SecondsStandard Deviation 8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Prothrombin TimeDischarge20.1 SecondsStandard Deviation 7.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Prothrombin Time3 Months21.0 SecondsStandard Deviation 8.4
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Prothrombin Time1 Year19.1 SecondsStandard Deviation 8.8
Other Pre-specified

Subject's Average Red Blood Cells Count

Laboratory Analysis of Red Blood Cell (RBC) Count on blood drawn from subjects; RBC carry oxygen.

Time frame: Baseline, Discharge, 3 Months, 1 Year and 2 Year

Population: The outcome is reported for subjects who received the Model 11000A or Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Red Blood Cells Count2 Year4.6 10^6 cells/microlitersStandard Deviation 0.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Red Blood Cells CountBaseline4.5 10^6 cells/microlitersStandard Deviation 0.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Red Blood Cells CountDischarge3.3 10^6 cells/microlitersStandard Deviation 0.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Red Blood Cells Count3 Months4.7 10^6 cells/microlitersStandard Deviation 0.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Red Blood Cells Count1 Year4.7 10^6 cells/microlitersStandard Deviation 0.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Red Blood Cells Count1 Year4.5 10^6 cells/microlitersStandard Deviation 0.6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Red Blood Cells Count3 Months4.5 10^6 cells/microlitersStandard Deviation 0.7
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Red Blood Cells CountBaseline4.4 10^6 cells/microlitersStandard Deviation 0.6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Red Blood Cells Count2 Year4.6 10^6 cells/microlitersStandard Deviation 0.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Red Blood Cells CountDischarge3.3 10^6 cells/microlitersStandard Deviation 0.4
Other Pre-specified

Subject's Average Score at Baseline and 1 Year on the Quality of Life Survey

The Medical Outcomes Study Short-Form 12 (SF-12) contains two components - the Physical Component Summary (PCS) and the Mental Component Summary (MCS). The SF-12 Physical Component Summary questionnaire scale ranges from a maximum of 100, which reflects the best health status to a minimum of 0, which reflects the worst health status. The SF-12 Mental Component Summary scale ranges from a maximum of 100, which reflects the best health status to a minimum of 0, which reflects the worst health status.

Time frame: Baseline and one year post-implant

Population: The outcome is reported for subjects who received the Model 11000A or Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyPhysical Health-Baseline42.9 units on a scaleStandard Deviation 9.7
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyPhysical Health-1 Year49.1 units on a scaleStandard Deviation 9.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyMental Health-Baseline50.9 units on a scaleStandard Deviation 9.5
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyMental Health-1 Year53.8 units on a scaleStandard Deviation 8.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyMental Health-1 Year52.3 units on a scaleStandard Deviation 9.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyPhysical Health-Baseline40.1 units on a scaleStandard Deviation 8.9
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyMental Health-Baseline49.2 units on a scaleStandard Deviation 11.5
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyPhysical Health-1 Year48.8 units on a scaleStandard Deviation 7
Other Pre-specified

Subjects Average Serum Glycerol Levels

Laboratory analysis of serum glycerol in blood drawn from subjects. This blood test measures the amount of glycerol (a naturally occurring carbohydrate, that can be used as a fuel source by the body) in the serum (liquid portion of the blood).

Time frame: Pre-Implant (post-heparinization) and Post-Implant (between 60 and 120 minutes after heart was restarted)

Population: The outcome is reported for subjects who received the Model 11000A device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubjects Average Serum Glycerol LevelsPre-Implant405.4 micromol/LStandard Deviation 270.2
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubjects Average Serum Glycerol LevelsPost-Implant245.6 micromol/LStandard Deviation 345.8
Other Pre-specified

Subject's Average White Blood Cell Count

Laboratory analysis of White Blood Cell (WBC) count on blood drawn from subject; WBC fight infection.

Time frame: Baseline, Discharge, 3 Months, 1 Year and 2 Year

Population: The outcome is reported for subjects who received the Model 11000A or Model 11000M device where data is available.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average White Blood Cell CountBaseline7.2 10^3 cells /microliterStandard Deviation 1.9
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average White Blood Cell Count1 Year7.3 10^3 cells /microliterStandard Deviation 1.9
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average White Blood Cell Count3 Months7.1 10^3 cells /microliterStandard Deviation 1.7
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average White Blood Cell Count2 Year7.3 10^3 cells /microliterStandard Deviation 1.8
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's Average White Blood Cell CountDischarge9.1 10^3 cells /microliterStandard Deviation 2.8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average White Blood Cell Count2 Year7.4 10^3 cells /microliterStandard Deviation 1.8
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average White Blood Cell CountBaseline7.4 10^3 cells /microliterStandard Deviation 2.6
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average White Blood Cell CountDischarge10.4 10^3 cells /microliterStandard Deviation 5.3
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average White Blood Cell Count3 Months7.1 10^3 cells /microliterStandard Deviation 2
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's Average White Blood Cell Count1 Year7.1 10^3 cells /microliterStandard Deviation 2.3
Other Pre-specified

Subject's New York Heart Association (NYHA) Functional Class Compared to Baseline

The New York Heart Association (NYHA) functional classification system relates symptoms to everyday activities and the patient's quality of life. Class I. No limitation of physical activity. Ordinary physical activity does not cause undue fatigue, palpitation, dyspnea (shortness of breath). Class II. Slight limitation of physical activity. Comfortable at rest. Ordinary physical activity results in fatigue, palpitation, dyspnea (shortness of breath). Class III. Marked limitation of physical activity. Comfortable at rest. Less than ordinary activity causes fatigue, palpitation, or dyspnea. Class IV. Unable to carry on any physical activity without discomfort. Symptoms of heart failure at rest. If any physical activity is undertaken, discomfort increases.

Time frame: 3 Months, 1 Year, and 2 Years Post Implant

Population: The outcome is reported for subjects who received the Model 11000A or Model 11000M device where data is available.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline3 MonthsSame200 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline1 YearWorsened17 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline1 YearImproved331 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline2 YearImproved157 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline3 MonthsWorsened14 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline2 YearSame75 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline1 YearSame161 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline2 YearWorsened8 Participants
Edwards Pericardial Bioprosthesis, Models 11000A & 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline3 MonthsImproved431 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline2 YearWorsened0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline3 MonthsImproved66 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline3 MonthsSame7 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline3 MonthsWorsened2 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline1 YearImproved66 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline1 YearSame7 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline1 YearWorsened0 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline2 YearImproved26 Participants
Edwards Pericardial Mitral Bioprosthesis, Model 11000MSubject's New York Heart Association (NYHA) Functional Class Compared to Baseline2 YearSame3 Participants