Evaluation of Blood Glucose Monitoring Systems With Blood Samples From Neonates

Performance of Three Bayer Blood Glucose Monitoring Systems With Blood Samples From Neonates

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01756274
Enrollment
162
Registered
2012-12-25
Start date
2012-12-31
Completion date
2013-02-28
Last updated
2017-02-15

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study was to evaluate the performance of three Bayer Blood Glucose Monitoring Systems (BGMS) with neonatal blood. Blood samples used in this study were 'left-over samples'. The blood samples were from heel sticks of neonates, collected (into a tube) and sent to the laboratory. All meter Blood Glucose (BG) results were compared with capillary plasma results obtained with a reference laboratory glucose method (Cobas® 6000 instrument).

Interventions

DEVICEContour® NEXT BGMS

Lab professionals tested the BG concentration using Contour® NEXT BGMS (Blood Glucose Monitoring System). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method, Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).

Lab professionals tested the BG concentration using Contour® PLUS BGMS (Blood Glucose Monitoring System). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method, Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).

Lab professionals tested the BG concentration using Contour® Next EZ BGMS (Blood Glucose Monitoring System). All meter BG results were compared with plasma results obtained with a reference laboratory glucose method, Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
No minimum to 30 Days
Healthy volunteers
Yes

Inclusion criteria

* Routine blood samples (heelstick) collected from neonates (less than 30 days of age). * A portion of the samples (approximately 10%) will be from subjects \<24 hours old. * A portion of the samples (at least 10%) will be from subjects in the NICU. * Sample blood volume must be sufficient to complete investigational testing in addition to prescribed testing.

Exclusion criteria

* Samples from babies \>=30 days of age

Design outcomes

Primary

MeasureTime frameDescription
Percent of Blood Glucose Results Within +/-15 mg/dL(<75 mg/dL) and Within +/-20% (>=75 mg/dL) of the Reference Instrument BG Value30 minutesLaboratory professionals tested the BG concentration of 'left-over samples' using plasma referenced Blood Glucose Monitoring Systems. BGMS results were compared with capillary plasma BG results obtained with a Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).

Secondary

MeasureTime frameDescription
Percent of BG Results (Per Population) Within +/-15 mg/dL (<100 mg/dL)and Within +/-15% (>=100 mg/dL) of the Reference Instrument BG Value30 minutesLaboratory professionals tested the BG concentration of 'left-over samples' using plasma referenced Blood Glucose Monitoring Systems. BGMS results were compared with capillary plasma BG results obtained with a Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).

Other

MeasureTime frameDescription
Number of Blood Samples From Babies Less Than 24 Hours Old That Produced Meter Results Beyond What Random Chance Would Predict (i.e. Outside 95% Limits for Studentized Residuals)30 minutesTo evaluate the effect of neonatal age on the meter systems' performances, the number of blood samples that produced unusual meter results out of the total number of blood samples that came from babies less than 24 hours old was reported. Studentized residuals were used to measure the degree to which meter BG results departed from what would be expected using a linear model. (This analysis is not related to BGM accuracy status, which has already been reported.)
Number of Blood Samples From Babies in the Neonatal Intensive Care Unit (NICU) That Produced Meter Results Beyond What Random Chance Would Predict (i.e., Outside 95% Limits for Studentized Residuals)30 minutesTo evaluate how the meter systems perform with blood samples drawn in the Neonatal Intensive Care Unit, the number of blood samples that produced unusual meter results out of the total number of blood samples that came from babies in Neonatal Intensive Care was reported. Studentized residuals were used to measure the degree to which meter BG results departed from what would be expected using a linear model. (This analysis is not related to BGM accuracy status, which has already been reported.)

Countries

United States

Participant flow

Participants by arm

ArmCount
Neonates 'Left-over' Blood Samples
Blood samples used in this study are 'left-over samples' from heel sticks of neonates, collected (into a tube) and sent to the laboratory. Lab professionals tested the BG concentration using 3 Bayer Blood Glucose Monitoring Systems: Contour® NEXT, Contour® PLUS, and Contour® Next EZ BGMS. Each subject could contribute up to 2 blood samples. Baseline Characteristics for Participant Flow based on number of blood samples, not number of subjects.
162
Neonates 'Left-over' Blood Samples
Blood samples used in this study are 'left-over samples' from heel sticks of neonates, collected (into a tube) and sent to the laboratory. Lab professionals tested the BG concentration using 3 Bayer Blood Glucose Monitoring Systems: Contour® NEXT, Contour® PLUS, and Contour® Next EZ BGMS. Each subject could contribute up to 2 blood samples. Baseline Characteristics for Participant Flow based on number of blood samples, not number of subjects.
211
Total373

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyLaboratory error1
Overall StudyProtocol Violation2

Baseline characteristics

CharacteristicNeonates 'Left-over' Blood Samples
Age, Customized
1-29 days
186 blood samples
Age, Customized
< 24 hours
25 blood samples
Nursery Location
NICU (Neonatal Intensive Care Unit)
30 blood samples
Nursery Location
Other
22 blood samples
Nursery Location
Special Care
101 blood samples
Nursery Location
Well Baby
58 blood samples
Region of Enrollment
United States
211 blood samples
Sex/Gender, Customized
Female
100 blood samples
Sex/Gender, Customized
Male
111 blood samples

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 162
serious
Total, serious adverse events
0 / 162

Outcome results

Primary

Percent of Blood Glucose Results Within +/-15 mg/dL(<75 mg/dL) and Within +/-20% (>=75 mg/dL) of the Reference Instrument BG Value

Laboratory professionals tested the BG concentration of 'left-over samples' using plasma referenced Blood Glucose Monitoring Systems. BGMS results were compared with capillary plasma BG results obtained with a Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).

Time frame: 30 minutes

Population: Each subject could contribute up to 2 blood samples. Of 217 blood samples 211(217-6) were included in data analysis. Two samples could not be used because they had been taken from sources that had not been specified in the protocol. Four samples had no reference method results due to laboratory errors.

ArmMeasureGroupValue (NUMBER)
Neonates 'Left-over' Blood SamplesPercent of Blood Glucose Results Within +/-15 mg/dL(<75 mg/dL) and Within +/-20% (>=75 mg/dL) of the Reference Instrument BG ValueContour® NEXT BGMS100.0 percentage of Blood Glucose Results
Neonates 'Left-over' Blood SamplesPercent of Blood Glucose Results Within +/-15 mg/dL(<75 mg/dL) and Within +/-20% (>=75 mg/dL) of the Reference Instrument BG ValueContour® PLUS BGMS99.5 percentage of Blood Glucose Results
Neonates 'Left-over' Blood SamplesPercent of Blood Glucose Results Within +/-15 mg/dL(<75 mg/dL) and Within +/-20% (>=75 mg/dL) of the Reference Instrument BG ValueContour® Next EZ BGMS99.5 percentage of Blood Glucose Results
Secondary

Percent of BG Results (Per Population) Within +/-15 mg/dL (<100 mg/dL)and Within +/-15% (>=100 mg/dL) of the Reference Instrument BG Value

Laboratory professionals tested the BG concentration of 'left-over samples' using plasma referenced Blood Glucose Monitoring Systems. BGMS results were compared with capillary plasma BG results obtained with a Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).

Time frame: 30 minutes

Population: Each subject could contribute up to 2 blood samples. Of 217 blood samples 211(217-6) were included in data analysis. Two samples could not be used because they had been taken from sources that had not been specified in the protocol. Four samples had no reference method results due to laboratory errors.

ArmMeasureGroupValue (NUMBER)
Neonates 'Left-over' Blood SamplesPercent of BG Results (Per Population) Within +/-15 mg/dL (<100 mg/dL)and Within +/-15% (>=100 mg/dL) of the Reference Instrument BG ValueContour® NEXT BGMS99.5 percentage of BLOOD GLUCOSE RESULTS
Neonates 'Left-over' Blood SamplesPercent of BG Results (Per Population) Within +/-15 mg/dL (<100 mg/dL)and Within +/-15% (>=100 mg/dL) of the Reference Instrument BG ValueContour® PLUS BGMS99.1 percentage of BLOOD GLUCOSE RESULTS
Neonates 'Left-over' Blood SamplesPercent of BG Results (Per Population) Within +/-15 mg/dL (<100 mg/dL)and Within +/-15% (>=100 mg/dL) of the Reference Instrument BG ValueContour® Next EZ BGMS99.1 percentage of BLOOD GLUCOSE RESULTS
Other Pre-specified

Number of Blood Samples From Babies in the Neonatal Intensive Care Unit (NICU) That Produced Meter Results Beyond What Random Chance Would Predict (i.e., Outside 95% Limits for Studentized Residuals)

To evaluate how the meter systems perform with blood samples drawn in the Neonatal Intensive Care Unit, the number of blood samples that produced unusual meter results out of the total number of blood samples that came from babies in Neonatal Intensive Care was reported. Studentized residuals were used to measure the degree to which meter BG results departed from what would be expected using a linear model. (This analysis is not related to BGM accuracy status, which has already been reported.)

Time frame: 30 minutes

Population: Of 217 'left-over' blood samples, 211(217-6) were included in data analysis and, of these, thirty (30) blood samples were obtained from babies in the Neonatal Intensive Care Unit (NICU).

ArmMeasureGroupValue (NUMBER)
Neonates 'Left-over' Blood SamplesNumber of Blood Samples From Babies in the Neonatal Intensive Care Unit (NICU) That Produced Meter Results Beyond What Random Chance Would Predict (i.e., Outside 95% Limits for Studentized Residuals)Contour® NEXT4 BLOOD SAMPLES
Neonates 'Left-over' Blood SamplesNumber of Blood Samples From Babies in the Neonatal Intensive Care Unit (NICU) That Produced Meter Results Beyond What Random Chance Would Predict (i.e., Outside 95% Limits for Studentized Residuals)Contour® PLUS2 BLOOD SAMPLES
Neonates 'Left-over' Blood SamplesNumber of Blood Samples From Babies in the Neonatal Intensive Care Unit (NICU) That Produced Meter Results Beyond What Random Chance Would Predict (i.e., Outside 95% Limits for Studentized Residuals)Contour® Next EZ3 BLOOD SAMPLES
Other Pre-specified

Number of Blood Samples From Babies Less Than 24 Hours Old That Produced Meter Results Beyond What Random Chance Would Predict (i.e. Outside 95% Limits for Studentized Residuals)

To evaluate the effect of neonatal age on the meter systems' performances, the number of blood samples that produced unusual meter results out of the total number of blood samples that came from babies less than 24 hours old was reported. Studentized residuals were used to measure the degree to which meter BG results departed from what would be expected using a linear model. (This analysis is not related to BGM accuracy status, which has already been reported.)

Time frame: 30 minutes

Population: Of 217 'left-over' blood samples, 211(217-6) were included in data analysis. Of these, twenty five (25) blood samples were obtained from babies that were less than 24 hours old.

ArmMeasureGroupValue (NUMBER)
Neonates 'Left-over' Blood SamplesNumber of Blood Samples From Babies Less Than 24 Hours Old That Produced Meter Results Beyond What Random Chance Would Predict (i.e. Outside 95% Limits for Studentized Residuals)Contour® NEXT1 BLOOD SAMPLES
Neonates 'Left-over' Blood SamplesNumber of Blood Samples From Babies Less Than 24 Hours Old That Produced Meter Results Beyond What Random Chance Would Predict (i.e. Outside 95% Limits for Studentized Residuals)Contour® PLUS2 BLOOD SAMPLES
Neonates 'Left-over' Blood SamplesNumber of Blood Samples From Babies Less Than 24 Hours Old That Produced Meter Results Beyond What Random Chance Would Predict (i.e. Outside 95% Limits for Studentized Residuals)Contour® Next EZ1 BLOOD SAMPLES

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026