Attain Performa(TM) Quadripolar Lead Study

Attain Performa(TM) Quadripolar Lead Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01751022
Enrollment
1202
Registered
2012-12-17
Start date
2012-12-31
Completion date
2015-02-28
Last updated
2016-12-05

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Heart Failure

Keywords

heart failure, cardiac pacing, cardiac resynchronization therapy, left ventricular lead

Brief summary

The purpose of the study is to evaluate the safety and efficacy of the Medtronic Attain Performa Quadripolar Leads (Model 4298, 4398, and 4598) during and post the implant procedure. This study will also assess the interactions of the Attain Performa leads with the entire Medtronic CRT-D system.

Interventions

DEVICEDevice: Attain Performa Pacing Lead (along with a standard CRT-D system)

Implant and follow-up of study lead, separate analysis for each lead model

Sponsors

Medtronic Cardiac Rhythm and Heart Failure
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patient is indicated for implant of a CRT-D device and left-heart lead per local indications (In US only this is based on Class I and II indications for CRT-D implant per HRS/ACC/AHA guidelines) * Patient (or legally authorized representative) has signed and dated the study-specific Consent Form * Patient is 18 years of age or older, or is of legal age to give informed consent per local and national law * Patient is expected to remain available for follow-up visits * Patient understands the study and agrees to comply with study protocol

Exclusion criteria

* Patient has a previous LV lead implanted or previous implant attempt within 30 days of enrollment or has ongoing AEs from a previous unsuccessful implant attempt * Patient has contraindications for standard transvenous cardiac pacing (e.g., mechanical right heart valve) * Patient has had a heart transplant (Note: Patients waiting for heart transplants are allowed in the study) * Patient is contraindicated for \< 1 mg dexamethasone acetate * Patient is currently enrolled or planning to participate in a potentially confounding drug or device study during the course of this study. (Note: Co-enrollment in concurrent studies may be allowed provided that documented pre-approval is obtained from Medtronic's study manager) * Patient has a life expectancy less than 180 days * Patient with

Design outcomes

Primary

MeasureTime frameDescription
Lead Complication-free Rate at 6 MonthsImplant to 6 months post-implantThe three Attain Performa LV leads models are evaluated separately. The primary safety objective is listed as following: \- Model 4298/4398: The Attain Performa Model 4298/4398 lead will be considered safe if the probability of subjects freed of Attain Performa lead-related complications at 6 months post-implant is greater than 87% (i.e., the one-sided 97.5% lower confidence bound must be greater than 87%). \- Model 4598: The safety performance of the Attain Performa Model 4598 lead will be characterized by summarizing the probability of subjects who are free from Attain Performa LV lead related complications at 6 months. The lower boundaries of the 97.5% confidence intervals for the all lead models are greater than the pacing threshold of 87%, thus concluding that the crtiera was met for all lead models.
LV Pacing Capture Thresholds Per Attain Performa Lead Model6 months post-implant

Secondary

MeasureTime frameDescription
Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelImplant up to 1-month post implant
Pacing Capture Thresholds at the Final Programmed Pacing Polarity6 months post-implant
Percentage of Subjects With Presence of PNS in All LV Lead Pacing Polarities6 months post-implantPercentage of patients with presence of PNS in all LV lead pacing polarities at 8.0 V at 0.5ms performed at 6-month visit.
Pacing Impedance at the Final Programmed Pacing Polarity6 month post-implantPacing impedance for each LV pacing polarity. Noticed that pacing impedance values are not recorded for reversed LV pacing polarities, since impedance from LV1 to LV2 is the same as from LV2 to LV1. Impedance is a measurement of current/resistance between the pacing lead and the cardiac tissue (measured in Ohms).
Complication Rate for Individual Attain Performa Lead Related Events6 month post-Implant
Implant Related Times Per Attain Performa Lead ModelImplant up to 1-month post implant
Percentage of Subjects With Successful Implant Per Attain Performa Lead ModelImplant up to 1-month post implant

Countries

Australia, Austria, Belgium, Canada, Chile, Denmark, Finland, France, Germany, Greece, Hungary, India, Israel, Italy, Malaysia, Netherlands, Norway, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, United States

Participant flow

Pre-assignment details

Out of the 1202 Intial enrollments; 77 were excluded due to to the following pre-implant events: 1. unsuccessful cannulation (25) 2. death (3) 3. Exit prior to implant attempt (43) 4. Attain Performa Lead implant not attempted (6) Resulting in 1125 subjects with Attain Performa Lead Implant Attempts

Participants by arm

ArmCount
Attain Performa LV Lead Model 4298
Subjects with an attempted Attain Performa LV Lead Model 4298 implant
499
Attain Performa LV Lead Model 4398
Subjects with an attempted Attain Performa LV Lead Model 4398 implant
326
Attain Performa LV Lead Model 4598
Subjects with an attempted Attain Performa LV Lead Model 4598 implant
300
Total1,125

Baseline characteristics

CharacteristicAttain Performa LV Lead Model 4598TotalAttain Performa LV Lead Model 4298Attain Performa LV Lead Model 4398
Age, Continuous67 years
STANDARD_DEVIATION 11
67 years
STANDARD_DEVIATION 11
68 years
STANDARD_DEVIATION 11
66 years
STANDARD_DEVIATION 11
Left Ventricular Ejection Fraction (LVEF)27 %
STANDARD_DEVIATION 7
26 %
STANDARD_DEVIATION 7
25 %
STANDARD_DEVIATION 7
26 %
STANDARD_DEVIATION 7
New York Heart Association Classification (NYHA)
I
4 participants16 participants8 participants4 participants
New York Heart Association Classification (NYHA)
II
89 participants308 participants125 participants94 participants
New York Heart Association Classification (NYHA)
III
198 participants766 participants352 participants216 participants
New York Heart Association Classification (NYHA)
IV
9 participants35 participants14 participants12 participants
QRS (Electrocardiogram Complex)156 ms
STANDARD_DEVIATION 23
156 ms
STANDARD_DEVIATION 23
155 ms
STANDARD_DEVIATION 23
156 ms
STANDARD_DEVIATION 24
Race/Ethnicity, Customized
American Indian or Alaska Native
5 participants9 participants4 participants0 participants
Race/Ethnicity, Customized
Asian
17 participants32 participants7 participants8 participants
Race/Ethnicity, Customized
Black or African American
17 participants76 participants38 participants21 participants
Race/Ethnicity, Customized
Hispanic or Latino
11 participants40 participants18 participants11 participants
Race/Ethnicity, Customized
Native Hawaiian or Pacific Islander
0 participants1 participants0 participants1 participants
Race/Ethnicity, Customized
Not reportable per local laws or regulations
48 participants99 participants30 participants21 participants
Race/Ethnicity, Customized
Subject/physician chose not to provide information
5 participants28 participants14 participants9 participants
Race/Ethnicity, Customized
Two or more races
0 participants4 participants1 participants3 participants
Race/Ethnicity, Customized
White or Caucasian
197 participants836 participants387 participants252 participants
Region of Enrollment
Australia
5 participants23 participants17 participants1 participants
Region of Enrollment
Austria
2 participants10 participants8 participants0 participants
Region of Enrollment
Belgium
5 participants9 participants3 participants1 participants
Region of Enrollment
Canada
21 participants60 participants17 participants22 participants
Region of Enrollment
Chile
0 participants12 participants12 participants0 participants
Region of Enrollment
Denmark
8 participants10 participants1 participants1 participants
Region of Enrollment
Finland
1 participants4 participants0 participants3 participants
Region of Enrollment
France
8 participants10 participants0 participants2 participants
Region of Enrollment
Germany
19 participants29 participants7 participants3 participants
Region of Enrollment
Greece
0 participants1 participants0 participants1 participants
Region of Enrollment
Hungary
4 participants7 participants0 participants3 participants
Region of Enrollment
India
15 participants21 participants3 participants3 participants
Region of Enrollment
Israel
2 participants4 participants0 participants2 participants
Region of Enrollment
Italy
10 participants28 participants8 participants10 participants
Region of Enrollment
Malaysia
1 participants2 participants0 participants1 participants
Region of Enrollment
Netherlands
9 participants23 participants13 participants1 participants
Region of Enrollment
Norway
0 participants5 participants5 participants0 participants
Region of Enrollment
Romania
1 participants12 participants11 participants0 participants
Region of Enrollment
Saudi Arabia
2 participants17 participants9 participants6 participants
Region of Enrollment
Serbia
8 participants10 participants0 participants2 participants
Region of Enrollment
Slovakia
7 participants13 participants4 participants2 participants
Region of Enrollment
Slovenia
3 participants8 participants0 participants5 participants
Region of Enrollment
South Africa
2 participants7 participants3 participants2 participants
Region of Enrollment
Spain
0 participants4 participants3 participants1 participants
Region of Enrollment
Switzerland
1 participants4 participants2 participants1 participants
Region of Enrollment
United Kingdom
14 participants31 participants13 participants4 participants
Region of Enrollment
United States
152 participants761 participants360 participants249 participants
Sex: Female, Male
Female
93 Participants329 Participants143 Participants93 Participants
Sex: Female, Male
Male
207 Participants796 Participants356 Participants233 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
deaths
Total, all-cause mortality
— / —— / —— / —
other
Total, other adverse events
155 / 49984 / 32684 / 300
serious
Total, serious adverse events
218 / 499134 / 326117 / 300

Outcome results

Primary

Lead Complication-free Rate at 6 Months

The three Attain Performa LV leads models are evaluated separately. The primary safety objective is listed as following: \- Model 4298/4398: The Attain Performa Model 4298/4398 lead will be considered safe if the probability of subjects freed of Attain Performa lead-related complications at 6 months post-implant is greater than 87% (i.e., the one-sided 97.5% lower confidence bound must be greater than 87%). \- Model 4598: The safety performance of the Attain Performa Model 4598 lead will be characterized by summarizing the probability of subjects who are free from Attain Performa LV lead related complications at 6 months. The lower boundaries of the 97.5% confidence intervals for the all lead models are greater than the pacing threshold of 87%, thus concluding that the crtiera was met for all lead models.

Time frame: Implant to 6 months post-implant

Population: Subjects with an attempted Attain Performa LV Lead Model. Results for lead model 4298 comes from its PMA-S Clinical Report Version 1, 27MAR2014; results for lead model 4398 comes from its PMA-S Clinical Report Version 3, 03SEP2014; results for lead model 4598 comes from its PMA-S Clinical Report Version 1, 29AUG2014.

ArmMeasureValue (NUMBER)
Attain Performa LV Lead Model 4298Lead Complication-free Rate at 6 Months96.04 Survival Probability (%)
Attain Performa LV Lead Model 4398Lead Complication-free Rate at 6 Months98.75 Survival Probability (%)
Attain Performa LV Lead Model 4598Lead Complication-free Rate at 6 Months96.21 Survival Probability (%)
Primary

LV Pacing Capture Thresholds Per Attain Performa Lead Model

Time frame: 6 months post-implant

Population: Subjects with Attain Performa LV Lead Model implanted and valid pacing thresholds measured at the 6 month follow-up visit. Results for model 4298 comes from its PMA-S Clinical Report Version 1, 27MAR2014; for model 4398 comes from its PMA-S Clinical Report Version 3, 03SEP2014; for model 4598 comes from its PMA-S Clinical Report V. 1, 29AUG2014.

ArmMeasureValue (MEAN)
Attain Performa LV Lead Model 4298LV Pacing Capture Thresholds Per Attain Performa Lead Model93.92 percentage of participants
Attain Performa LV Lead Model 4398LV Pacing Capture Thresholds Per Attain Performa Lead Model96.24 percentage of participants
Attain Performa LV Lead Model 4598LV Pacing Capture Thresholds Per Attain Performa Lead Model95.86 percentage of participants
Secondary

Complication Rate for Individual Attain Performa Lead Related Events

Time frame: 6 month post-Implant

Population: Subjects with Attain Performa LV Lead Model successfully implanted. Results for model 4298 comes from its PMA-S Clinical Report Version 1, 27MAR2014; for model 4398 comes from its PMA-S Clinical Report Version 3, 03SEP2014; for model 4598 comes from its PMA-S Clinical Report Version 1, 29AUG2014.

ArmMeasureValue (NUMBER)
Attain Performa LV Lead Model 4298Complication Rate for Individual Attain Performa Lead Related Events3.70 percentage of participants
Attain Performa LV Lead Model 4398Complication Rate for Individual Attain Performa Lead Related Events1.58 percentage of participants
Attain Performa LV Lead Model 4598Complication Rate for Individual Attain Performa Lead Related Events3.74 percentage of participants
Secondary

Implant Related Times Per Attain Performa Lead Model

Time frame: Implant up to 1-month post implant

Population: Subjects with Attain Performa LV Lead Model successfully implanted. Results for model 4298 comes from its PMA-S Clinical Report Version 1, 27MAR2014; for model 4398 comes from its PMA-S Clinical Report Version 3, 03SEP2014; for model 4598 comes from its PMA-S Clinical Report Version 1, 29AUG2014.

ArmMeasureGroupValue (MEAN)Dispersion
Attain Performa LV Lead Model 4298Implant Related Times Per Attain Performa Lead ModelCannulation6.3 minutesStandard Deviation 9.5
Attain Performa LV Lead Model 4298Implant Related Times Per Attain Performa Lead ModelFluoroscopy21.7 minutesStandard Deviation 15.4
Attain Performa LV Lead Model 4298Implant Related Times Per Attain Performa Lead ModelLV Lead Placement13.0 minutesStandard Deviation 15.9
Attain Performa LV Lead Model 4298Implant Related Times Per Attain Performa Lead ModelTotal Operation109.6 minutesStandard Deviation 49
Attain Performa LV Lead Model 4398Implant Related Times Per Attain Performa Lead ModelTotal Operation107.96 minutesStandard Deviation 53.46
Attain Performa LV Lead Model 4398Implant Related Times Per Attain Performa Lead ModelCannulation8.45 minutesStandard Deviation 14.47
Attain Performa LV Lead Model 4398Implant Related Times Per Attain Performa Lead ModelLV Lead Placement17.26 minutesStandard Deviation 26.21
Attain Performa LV Lead Model 4398Implant Related Times Per Attain Performa Lead ModelFluoroscopy23.00 minutesStandard Deviation 17.29
Attain Performa LV Lead Model 4598Implant Related Times Per Attain Performa Lead ModelTotal Operation95.09 minutesStandard Deviation 43.38
Attain Performa LV Lead Model 4598Implant Related Times Per Attain Performa Lead ModelFluoroscopy18.77 minutesStandard Deviation 12.66
Attain Performa LV Lead Model 4598Implant Related Times Per Attain Performa Lead ModelLV Lead Placement11.82 minutesStandard Deviation 14.17
Attain Performa LV Lead Model 4598Implant Related Times Per Attain Performa Lead ModelCannulation6.87 minutesStandard Deviation 11.19
Secondary

Pacing Capture Thresholds at the Final Programmed Pacing Polarity

Time frame: 6 months post-implant

Population: Subjects with Attain Performa LV Lead Model implanted and valid pacing thresholds measured at the 6 month follow-up visit. Results for model 4298 comes from its PMA-S Clinical Report Version 1, 27MAR2014; for model 4398 comes from its PMA-S Clinical Report Version 3, 03SEP2014; for model 4598 comes from its PMA-S Clinical Report V. 1, 29AUG2014.

ArmMeasureValue (MEAN)Dispersion
Attain Performa LV Lead Model 4298Pacing Capture Thresholds at the Final Programmed Pacing Polarity1.06 VoltsStandard Deviation 0.74
Attain Performa LV Lead Model 4398Pacing Capture Thresholds at the Final Programmed Pacing Polarity1.15 VoltsStandard Deviation 0.76
Attain Performa LV Lead Model 4598Pacing Capture Thresholds at the Final Programmed Pacing Polarity1.10 VoltsStandard Deviation 0.79
Secondary

Pacing Impedance at the Final Programmed Pacing Polarity

Pacing impedance for each LV pacing polarity. Noticed that pacing impedance values are not recorded for reversed LV pacing polarities, since impedance from LV1 to LV2 is the same as from LV2 to LV1. Impedance is a measurement of current/resistance between the pacing lead and the cardiac tissue (measured in Ohms).

Time frame: 6 month post-implant

Population: Subjects with Attain Performa LV lead implanted and complete Medtronic Quad CRT-D system and valid measures of lead impedance at 6-month visit. Results for model 4298 comes from its PMA-S Clinical Report V1, 27MAR14; for model 4398 comes from its PMA-S Clinical Report V3, 03SEP14; for model 4598 comes from its PMA-S Clinical Report V1, 29AUG14.

ArmMeasureGroupValue (MEAN)Dispersion
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV1 to RV Coil460.56 OhmsStandard Deviation 124.26
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV2780.49 OhmsStandard Deviation 228
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV3776.64 OhmsStandard Deviation 192.41
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV4750.14 OhmsStandard Deviation 171.98
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV2 to RV Coil462.63 OhmsStandard Deviation 168.82
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV2 to LV3562.40 OhmsStandard Deviation 179.24
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV2 to LV4730.26 OhmsStandard Deviation 196.11
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV3 to RV Coil450.29 OhmsStandard Deviation 107.29
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV3 to LV4715.20 OhmsStandard Deviation 159.58
Attain Performa LV Lead Model 4298Pacing Impedance at the Final Programmed Pacing PolarityLV4 to RV Coil406.24 OhmsStandard Deviation 87.12
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV3 to RV Coil458.86 OhmsStandard Deviation 110.62
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV2 to RV Coil476.71 OhmsStandard Deviation 193.46
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV4 to RV Coil407.17 OhmsStandard Deviation 87.09
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV2 to LV3568.36 OhmsStandard Deviation 197.87
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV3 to LV4733.30 OhmsStandard Deviation 164.21
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV2 to LV4755.85 OhmsStandard Deviation 214.04
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV1 to RV Coil466.24 OhmsStandard Deviation 123.63
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV2802.52 OhmsStandard Deviation 240.3
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV3792.62 OhmsStandard Deviation 195.78
Attain Performa LV Lead Model 4398Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV4761.79 OhmsStandard Deviation 175.08
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV2826.69 OhmsStandard Deviation 275.82
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV4799.70 OhmsStandard Deviation 225.09
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV3 to RV Coil468.90 OhmsStandard Deviation 247.12
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV4 to RV Coil421.69 OhmsStandard Deviation 180.51
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV2 to RV Coil483.99 OhmsStandard Deviation 247.31
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV1 to RV Coil496.45 OhmsStandard Deviation 133.92
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV2 to LV4756.48 OhmsStandard Deviation 257.78
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV2 to LV3580.02 OhmsStandard Deviation 247.99
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV3 to LV4740.99 OhmsStandard Deviation 254.57
Attain Performa LV Lead Model 4598Pacing Impedance at the Final Programmed Pacing PolarityLV1 to LV3826.15 OhmsStandard Deviation 278.05
Secondary

Percentage of Subjects With Presence of PNS in All LV Lead Pacing Polarities

Percentage of patients with presence of PNS in all LV lead pacing polarities at 8.0 V at 0.5ms performed at 6-month visit.

Time frame: 6 months post-implant

Population: Subjects with Attain Performa LV Lead implanted and at least 1 valid pacing threshold at any LV lead pacing polarity measured at the 6 month visit. Results for model 4298 comes from its PMA-S Clinical Report V1, 27MAR14; for model 4398 comes from its PMA-S Clinical Report V3, 03SEP14; for model 4598 comes from its PMA-S Clinical Report V1, 29AUG14.

ArmMeasureGroupValue (NUMBER)
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV212.91 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV434.18 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV426.58 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-RV COIL21.52 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV238.48 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV126.58 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-RV COIL33.92 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV221.27 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2 -LV134.94 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-LV312.66 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV338.99 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-LV427.85 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-RV COIL31.39 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-RV COIL42.03 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV133.42 percentage of participants
Attain Performa LV Lead Model 4298Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV320.76 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV340.45 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2 -LV132.21 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-RV COIL44.19 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV423.60 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV434.83 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV215.36 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-RV COIL17.98 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-LV318.35 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-RV COIL31.84 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV124.34 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV131.46 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-LV425.47 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV220.22 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-RV COIL33.71 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV319.10 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV240.45 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV131.84 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-RV COIL44.57 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV235.58 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV335.58 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV1-LV437.83 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-RV COIL27.72 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2 -LV127.34 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-LV310.86 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV2-LV422.47 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-RV COIL24.72 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV28.61 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV3-LV418.73 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-RV COIL19.10 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV124.72 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV217.60 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Presence of PNS in All LV Lead Pacing PolaritiesLV4-LV316.10 percentage of participants
Secondary

Percentage of Subjects With Successful Implant Per Attain Performa Lead Model

Time frame: Implant up to 1-month post implant

Population: Subjects with an attempted Attain Performa LV Lead Model. Results for lead model 4298 comes from its PMA-S Clinical Report Version 1, 27MAR2014; results for lead model 4398 comes from its PMA-S Clinical Report Version 3, 03SEP2014; results for lead model 4598 comes from its PMA-S Clinical Report Version 1, 29AUG2014.

ArmMeasureValue (NUMBER)
Attain Performa LV Lead Model 4298Percentage of Subjects With Successful Implant Per Attain Performa Lead Model97.60 percentage of participants
Attain Performa LV Lead Model 4398Percentage of Subjects With Successful Implant Per Attain Performa Lead Model97.55 percentage of participants
Attain Performa LV Lead Model 4598Percentage of Subjects With Successful Implant Per Attain Performa Lead Model98.00 percentage of participants
Secondary

Rate of Overall Acceptable Lead Handling Per Attain Performa Lead Model

Time frame: Implant up to 1-month post implant

Population: Subjects with an attempted Attain Performa LV Lead Model. Results for lead model 4298 comes from its PMA-S Clinical Report Version 1, 27MAR2014; results for lead model 4398 comes from its PMA-S Clinical Report Version 3, 03SEP2014; results for lead model 4598 comes from its PMA-S Clinical Report Version 1, 29AUG2014.

ArmMeasureGroupValue (NUMBER)
Attain Performa LV Lead Model 4298Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelUnacceptable0.20 percentage of participants
Attain Performa LV Lead Model 4298Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelAcceptable98.00 percentage of participants
Attain Performa LV Lead Model 4298Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelNo Response1.80 percentage of participants
Attain Performa LV Lead Model 4398Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelUnacceptable0.31 percentage of participants
Attain Performa LV Lead Model 4398Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelAcceptable98.47 percentage of participants
Attain Performa LV Lead Model 4398Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelNo Response1.23 percentage of participants
Attain Performa LV Lead Model 4598Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelAcceptable98.67 percentage of participants
Attain Performa LV Lead Model 4598Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelNo Response1.00 percentage of participants
Attain Performa LV Lead Model 4598Rate of Overall Acceptable Lead Handling Per Attain Performa Lead ModelUnacceptable0.33 percentage of participants