Comparison of CardioQ and Thermodilution Derived Cardiac Output Measurements
Conditions
Keywords
cardiac output, thermodilution, esophageal doppler, trending ability
Brief summary
Minimal invasive monitoring systems of central hemodynamics are gaining increasing popularity. The present study investigates the precision of the esophageal doppler (CardioQ) derived cardiac output and its agreement with pulmonary artery catheter thermodilution (PAC TD) for measuring CO during steady state and with induced hemodynamic changes in patients scheduled for elective cardiac surgery.
Detailed description
Twenty-five patients are planned to be enrolled. After induction of anesthesia, insertion of the PAC and the esophageal-doppler probe, the patient are placed in the following successive positions: a) supine, b) head-down tilt, c) head-up tilt, d) supine, e) supine with phenylephrine administration f) pace heart rate 80 bpm, g) pace heart rate 110 bpm, and CO are measured simultaneously using the CardioQ and PAC TD.
Interventions
Twenty-five patients are planned enrolled. After induction of anesthesia, insertion of the PAC and the CardioQ probe, the patient are placed in the following successive positions: a) supine, b) head-down tilt, c) head-up tilt, d) supine, e) supine with phenylephrine administration f) pace heart rate 80 bpm, g) pace heart rate 110 bpm. CO are measured simultaneously using the TEE and thermodilution technique.
Sponsors
Study design
Eligibility
Inclusion criteria
* patients scheduled for elective coronary artery bypass grafting (CABG), who are in stable sinus rhythm, who have a left ventricular ejection fraction \> 0.40 measured by transthoracic echocardiography, and who have no significant valve pathology
Exclusion criteria
\-
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Agreement of cardiac output determined with CardioQ, compared with thermodilution | Data are collected in the perioperative period (30min.) | Data are collected for each patient in the perioperative period. Patients are followed in the first 24 hours. Data will after inclusion of the last data be analysed and appropiate statistics aplied. |
Countries
Denmark