An Outpatient Pump Shutoff Pilot Feasibility and Safety Study

Reduction of Nocturnal Hypoglycemia by Using Predictive Algorithms and Pump Suspension: An Outpatient Pilot Feasibility and Safety Study

Status
Completed
Phases
Phase 2
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01736930
Acronym
PSO1
Enrollment
20
Registered
2012-11-29
Start date
2012-01-31
Completion date
2012-06-30
Last updated
2016-08-30

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Type 1 Diabetes

Keywords

Type 1 Diabetes, Hypoglycemia, Continuous Glucose Monitoring, Pump Suspension

Brief summary

The purpose of this study is to evaluate an overnight system that will turn off the insulin pump automatically if the system predicts that a low blood sugar is likely. The study system includes a combination continuous glucose monitor (CGM)/ insulin pump made by Medtronic MiniMed, Inc and a regular laptop computer that runs a computer program that predicts low blood sugar. It works by (1) measuring the glucose levels under the skin with a continuous glucose monitor, (2) using a computer program on a laptop to predict what will happen to the glucose level over the next 35-55 minutes, and (3) turning off the insulin pump when the computer program predicts that low blood sugar will occur. We have tested this system overnight in the hospital and are ready to test the system in the home environment to learn more about how well it will work and to make sure that the blood sugar does not go too high when the pump shuts off. This study has several phases and will take about a month or a little more for a patient to complete. Patients will use the study system for about 5 days at home to show that the patient is able to use it correctly. After that, the patient will be asked to use the study system each night for an additional 3-4 weeks. During this time, the system will be active for two-thirds of the nights and not active for one-third of the nights. When the system is active and predicts that your blood sugar will become low, the insulin pump will shut off for up to 2 hours. The study will include 2 clinical centers in the United States.

Detailed description

Patients who meet all eligibility criteria will use the study system overnight at home for about 5 nights to demonstrate their ability to use the system and submit study data to the Coordinating Center. Patients who successfully demonstrate their ability to use the system as described above will be eligible for the randomized trial phase. This phase consists of use of the full system as an outpatient for approximately 21 nights: * Each night the blood glucose level will be checked with the study blood glucose (BG) meter and used to perform a calibration of the CGM. The calibration must occur no more than 90 minutes prior to activation of the system. NOTE: Patients will be instructed to calibrate the CGM per manufacturer guidelines. * Then the system will be activated, linking the CGM and insulin pump to the computer at the bedside. * A randomization schedule on the laptop will be used to determine whether the 'pump shut off' application will be active that night or not. * Patients will be blinded as to whether the pump shut off is active when a session is initiated each night. * There will not be an alarm if the pump shuts off. The CGM alarm will be set to 60 mg/dL. When a CGM alarm occurs, the patient will be asked to measure the blood glucose with a BG meter, if he/she is aware of the alarm. * The time period for outcome assessment each night will be from the time the system is activated until it is turned off in the morning. * Pump shut off, when it occurs, will be for up to 2 hours. Multiple instances of pump suspension can occur if there are recurrent predictions of hypoglycemia during the night. * Patients will be asked to check blood glucose with the study BG meter, blood ketones with the study ketone meter, and urine ketones with a ketone strip each morning prior to breakfast and enter the results on the study laptop. The patient will be instructed to contact the study physician if the blood glucose or ketone readings are out of an expected range. Patients will be contacted if these morning safety values are not reported as required or are out of range. * Patients will be asked to record all overnight carbohydrate intake on the study laptop. * Patients will be asked to perform periodic data uploads using the study laptop. Monitoring processes will ensure that the patient is contacted if these uploads do not occur as required, or if review of an upload reveals any extreme, prolonged episodes of low or high blood glucose readings, or elevated morning blood glucose, blood ketone, or urine ketone values. Upon completion of the study, patients as well as study clinicians will be asked to complete a questionnaire regarding use of the study system. There will a follow-up visit after completion of 21 successful nights of study system use. A successful night of study system use is defined as use of the system for at least four hours. Phone contacts with the patients will be made once a week.

Interventions

The study laptop will communicate to the pump causing suspension based on output from the algorithm which predicts hypoglycemia based on the continuous glucose sensor trend.

Sponsors

National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK)
CollaboratorNIH
Jaeb Center for Health Research
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
PREVENTION
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Clinical diagnosis of type 1 diabetes and using daily insulin therapy for at least one year and an insulin infusion pump for at least 6 months * Age \>=18.0 years * HbA1c \<=8.0% * Availability of internet access for periodic upload of study data * Experience using the Medtronic pump and CGM and investigator is confident that subject will use CGM on a near-daily basis during the study * Nocturnal hypoglycemia meeting the following criterion based on CGM download: during the most recent 15 nights with CGM glucose data (must be within the past 42 days), one or more nights with a sensor glucose value \<=70 mg/dL.

Exclusion criteria

* Diabetic ketoacidosis in the past 3 months * Hypoglycemic seizure or loss of consciousness in the past 6 months * History of seizure disorder (except for hypoglycemic seizure) * Coronary artery disease or heart failure * Cystic fibrosis * Current use of oral/inhaled glucocorticoids, beta-blockers or other medications, which in the judgment of the investigator would be a contraindication to participation in the study. * History of ongoing renal disease (other than microalbuminuria). Creatinine level to have been obtained within the last year if subject has diabetes of \>10 years duration or is over 50 years of age. If creatinine is \> 1.5 mg/dL, the subject is excluded. * History of liver disease * Medical or psychiatric condition that in the judgment of the investigator might interfere with the completion of the protocol such as: inpatient psychiatric treatment in the past 6 months, uncontrolled adrenal disorder, and/or abuse of alcohol * Pregnancy: A negative urine pregnancy test will be required for all premenopausal women who are not surgically sterile. Subjects who become pregnant will be discontinued from the study.

Design outcomes

Primary

MeasureTime frameDescription
Mornings With Urine Ketones Characterized as Moderate or Large - Algorithm 321 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with urine ketones characterized as moderate or large based on measurement results of a urine dipstick test taken using Ketostix. Moderate is considered approximately 30 - 40 mg/dL and Large \>80 mg/dL.
Percent Morning Blood Glucose >250 mg/dL - Algorithm 321 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as overall percentage of mornings glucose measured with home glucose meter \>250 mg/dL.
Mornings With Blood Ketones >0.6 mmol/L - Algorithm 121 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with blood ketones \>0.6 mmol/L.
Mornings With Blood Ketones >0.6 mmol/L - Algorithm 221 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with blood ketones \>0.6 mmol/L.
Mornings With Blood Ketones >0.6 mmol/L - Algorithm 321 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with blood ketones \>0.6 mmol/L.
Mornings With Urine Ketones Characterized as Moderate or Large - Algorithm 121 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with urine ketones characterized as moderate or large based on measurement results of a urine dipstick test taken using Ketostix. Moderate is considered approximately 30 - 40 mg/dL and Large \>80 mg/dL.
Mornings With Urine Ketones Characterized as Moderate or Large - Algorithm 221 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with urine ketones characterized as moderate or large based on measurement results of a urine dipstick test taken using Ketostix. Moderate is considered approximately 30 - 40 mg/dL and Large \>80 mg/dL.
Mean Morning Blood Glucose (mg/dL)- Algorithm 321 study nightsThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as mean morning blood glucose (mg/dL). The hypoglycemia prediction horizon was reduced further in algorithm 3 to 30 minutes. A total of 114 study nights (37 Control nights and 77 Intervention nights) using algorithm 3.
Mean Morning Blood Glucose (mg/dL)- Algorithm 121 study nightsThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as mean morning blood glucose (mg/dL). An objective was to evaluate and refine the control algorithm. The data were reviewed periodically during the study with the pre-stated goal of determining whether any changes should be made in the control algorithm. Algorithm 1 was used for the first 105 nights of the study (38 Control nights and 67 Intervention nights). The horizon prediction time of algorithm 1 was set at 70 minutes.
Mean Morning Blood Glucose (mg/dL)- Algorithm 221 study nightsThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as mean morning blood glucose (mg/dL). Algorithm 1 was modified to reduce the hypoglycemia prediction horizon from 70 minutes to 50 minutes, to suspend the pump only when the continuous glucose monitor sensor glucose value was ≤ 230 mg/dl, not suspend if there was a drop of \>40 mg/dl in consecutive sensor glucose readings, and to resume insulin delivery at the first rise in sensor glucose following a suspension. There was 156 nights of study data collected (48 Control nights and 108 Intervention nights) using algorithm 2.
Percent Morning Blood Glucose >250 mg/dL - Algorithm 121 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as the overall percentage of mornings glucose measured with home glucose meter \>250 mg/dL.
Percent Morning Blood Glucose >250 mg/dL - Algorithm 221 daysThe primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as overall percentage of mornings glucose measured with home glucose meter \>250 mg/dL.

Secondary

MeasureTime frame
Mean Sensor Glucose Overnight - Algorithm 2Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use
Mean Sensor Glucose Overnight - Algorithm 3Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use
Percentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 1Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use
Percentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 2Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use
Percentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 3Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use
Mean Sensor Glucose Overnight - Algorithm 1Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use

Countries

United States

Participant flow

Recruitment details

The study was conducted at Stanford University and the Barbara Davis Center. A total of 20 subjects were enrolled between January 1, 2012 and June 30, 2012. Nineteen participants 18-56 years old with type 1 diabetes (HbA1c 6.0%-7.7%) completed the trial.

Pre-assignment details

After the run-in phase, a 21-night randomized trial was conducted in which each night was randomly assigned 2:1 to have either the predictive low glucose suspend system active (intervention) or inactive (control). Participants were current users of the MiniMed Paradigm® REAL-Time Revel™ System and Sof-sensor® glucose sensor at time of enrollment.

Participants by arm

ArmCount
Randomized Nights- Treatment or Control
Each study night will be randomized to have either Predictive Low Glucose Suspend or to be inactive (control). On nights randomized to the intervention treatment, the study laptop will be running actively during the night and suspending the patient's pump if the algorithm predicts hypoglycemia based on the patient's continuous glucose sensor trend. (Pump suspension : The study laptop will communicate to the pump causing suspension based on output from the algorithm which predicts hypoglycemia based on the continuous glucose sensor trend.) On control nights, the algorithm will run passively and not recommend suspensions or resumption to the patient's pump.
20
Total20

Baseline characteristics

CharacteristicRandomized Nights- Treatment or Control
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
20 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
0 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
1 Participants
Race (NIH/OMB)
White
19 Participants
Sex: Female, Male
Female
11 Participants
Sex: Female, Male
Male
9 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
— / —— / —
other
Total, other adverse events
0 / 200 / 20
serious
Total, serious adverse events
0 / 200 / 20

Outcome results

Primary

Mean Morning Blood Glucose (mg/dL)- Algorithm 1

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as mean morning blood glucose (mg/dL). An objective was to evaluate and refine the control algorithm. The data were reviewed periodically during the study with the pre-stated goal of determining whether any changes should be made in the control algorithm. Algorithm 1 was used for the first 105 nights of the study (38 Control nights and 67 Intervention nights). The horizon prediction time of algorithm 1 was set at 70 minutes.

Time frame: 21 study nights

ArmMeasureValue (MEAN)Dispersion
Predictive Pump Suspension AlgorithmMean Morning Blood Glucose (mg/dL)- Algorithm 1158 mg/dlStandard Deviation 52
Standard of CareMean Morning Blood Glucose (mg/dL)- Algorithm 1125 mg/dlStandard Deviation 53
Primary

Mean Morning Blood Glucose (mg/dL)- Algorithm 2

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as mean morning blood glucose (mg/dL). Algorithm 1 was modified to reduce the hypoglycemia prediction horizon from 70 minutes to 50 minutes, to suspend the pump only when the continuous glucose monitor sensor glucose value was ≤ 230 mg/dl, not suspend if there was a drop of \>40 mg/dl in consecutive sensor glucose readings, and to resume insulin delivery at the first rise in sensor glucose following a suspension. There was 156 nights of study data collected (48 Control nights and 108 Intervention nights) using algorithm 2.

Time frame: 21 study nights

ArmMeasureValue (MEAN)Dispersion
Predictive Pump Suspension AlgorithmMean Morning Blood Glucose (mg/dL)- Algorithm 2151 mg/dlStandard Deviation 57
Standard of CareMean Morning Blood Glucose (mg/dL)- Algorithm 2138 mg/dlStandard Deviation 63
Primary

Mean Morning Blood Glucose (mg/dL)- Algorithm 3

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as mean morning blood glucose (mg/dL). The hypoglycemia prediction horizon was reduced further in algorithm 3 to 30 minutes. A total of 114 study nights (37 Control nights and 77 Intervention nights) using algorithm 3.

Time frame: 21 study nights

ArmMeasureValue (MEAN)Dispersion
Predictive Pump Suspension AlgorithmMean Morning Blood Glucose (mg/dL)- Algorithm 3144 mg/dlStandard Deviation 48
Standard of CareMean Morning Blood Glucose (mg/dL)- Algorithm 3133 mg/dlStandard Deviation 57
Primary

Mornings With Blood Ketones >0.6 mmol/L - Algorithm 1

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with blood ketones \>0.6 mmol/L.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmMornings With Blood Ketones >0.6 mmol/L - Algorithm 10 number of mornings
Standard of CareMornings With Blood Ketones >0.6 mmol/L - Algorithm 10 number of mornings
Primary

Mornings With Blood Ketones >0.6 mmol/L - Algorithm 2

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with blood ketones \>0.6 mmol/L.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmMornings With Blood Ketones >0.6 mmol/L - Algorithm 23 number of mornings
Standard of CareMornings With Blood Ketones >0.6 mmol/L - Algorithm 21 number of mornings
Primary

Mornings With Blood Ketones >0.6 mmol/L - Algorithm 3

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with blood ketones \>0.6 mmol/L.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmMornings With Blood Ketones >0.6 mmol/L - Algorithm 31 number of mornings
Standard of CareMornings With Blood Ketones >0.6 mmol/L - Algorithm 30 number of mornings
Primary

Mornings With Urine Ketones Characterized as Moderate or Large - Algorithm 1

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with urine ketones characterized as moderate or large based on measurement results of a urine dipstick test taken using Ketostix. Moderate is considered approximately 30 - 40 mg/dL and Large \>80 mg/dL.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmMornings With Urine Ketones Characterized as Moderate or Large - Algorithm 10 number of mornings
Standard of CareMornings With Urine Ketones Characterized as Moderate or Large - Algorithm 11 number of mornings
Primary

Mornings With Urine Ketones Characterized as Moderate or Large - Algorithm 2

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with urine ketones characterized as moderate or large based on measurement results of a urine dipstick test taken using Ketostix. Moderate is considered approximately 30 - 40 mg/dL and Large \>80 mg/dL.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmMornings With Urine Ketones Characterized as Moderate or Large - Algorithm 27 number of mornings
Standard of CareMornings With Urine Ketones Characterized as Moderate or Large - Algorithm 20 number of mornings
Primary

Mornings With Urine Ketones Characterized as Moderate or Large - Algorithm 3

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as number of mornings with urine ketones characterized as moderate or large based on measurement results of a urine dipstick test taken using Ketostix. Moderate is considered approximately 30 - 40 mg/dL and Large \>80 mg/dL.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmMornings With Urine Ketones Characterized as Moderate or Large - Algorithm 30 number of mornings
Standard of CareMornings With Urine Ketones Characterized as Moderate or Large - Algorithm 30 number of mornings
Primary

Percent Morning Blood Glucose >250 mg/dL - Algorithm 1

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as the overall percentage of mornings glucose measured with home glucose meter \>250 mg/dL.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmPercent Morning Blood Glucose >250 mg/dL - Algorithm 13 percentage of mornings
Standard of CarePercent Morning Blood Glucose >250 mg/dL - Algorithm 13 percentage of mornings
Primary

Percent Morning Blood Glucose >250 mg/dL - Algorithm 2

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as overall percentage of mornings glucose measured with home glucose meter \>250 mg/dL.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmPercent Morning Blood Glucose >250 mg/dL - Algorithm 26 percentage of mornings
Standard of CarePercent Morning Blood Glucose >250 mg/dL - Algorithm 28 percentage of mornings
Primary

Percent Morning Blood Glucose >250 mg/dL - Algorithm 3

The primary safety outcome will be evaluated by comparing the intervention and control nights and nights with vs. without actual shut-off of the pump for several measures of hyperglycemia such as overall percentage of mornings glucose measured with home glucose meter \>250 mg/dL.

Time frame: 21 days

ArmMeasureValue (NUMBER)
Predictive Pump Suspension AlgorithmPercent Morning Blood Glucose >250 mg/dL - Algorithm 31 percentage of mornings
Standard of CarePercent Morning Blood Glucose >250 mg/dL - Algorithm 33 percentage of mornings
Secondary

Mean Sensor Glucose Overnight - Algorithm 1

Time frame: Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use

ArmMeasureValue (MEAN)Dispersion
Predictive Pump Suspension AlgorithmMean Sensor Glucose Overnight - Algorithm 1158 mg/dlStandard Deviation 34
Standard of CareMean Sensor Glucose Overnight - Algorithm 1145 mg/dlStandard Deviation 42
Secondary

Mean Sensor Glucose Overnight - Algorithm 2

Time frame: Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use

ArmMeasureValue (MEAN)Dispersion
Predictive Pump Suspension AlgorithmMean Sensor Glucose Overnight - Algorithm 2137 mg/dlStandard Deviation 36
Standard of CareMean Sensor Glucose Overnight - Algorithm 2123 mg/dlStandard Deviation 40
Secondary

Mean Sensor Glucose Overnight - Algorithm 3

Time frame: Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use

ArmMeasureValue (MEAN)Dispersion
Predictive Pump Suspension AlgorithmMean Sensor Glucose Overnight - Algorithm 3148 mg/dlStandard Deviation 42
Standard of CareMean Sensor Glucose Overnight - Algorithm 3133 mg/dlStandard Deviation 37
Secondary

Percentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 1

Time frame: Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use

ArmMeasureValue (MEDIAN)
Predictive Pump Suspension AlgorithmPercentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 171 percentage of sensor glucose values
Standard of CarePercentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 176 percentage of sensor glucose values
Secondary

Percentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 2

Time frame: Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use

ArmMeasureValue (MEDIAN)
Predictive Pump Suspension AlgorithmPercentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 290 percentage of sensor glucose values
Standard of CarePercentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 291 percentage of sensor glucose values
Secondary

Percentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 3

Time frame: Overnight from system activation to deactivation in the morning upon awakening for 21 nights of system use

ArmMeasureValue (MEDIAN)
Predictive Pump Suspension AlgorithmPercentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 389 percentage of sensor glucose values
Standard of CarePercentage of Sensor Glucose Values 71 to 180 mg/dL - Algorithm 394 percentage of sensor glucose values

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026