Effect of Different Plasma Insulin Levels on the Accuracy of Continuous Subcutaneous Glucose Monitoring

New Strategies for Automated Glycaemic Control: the Issue of Continuous Glucose Monitoring Accuracy Under Hypoglycaemic Conditions

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01714895
Enrollment
14
Registered
2012-10-26
Start date
2011-10-31
Completion date
2012-10-31
Last updated
2012-10-26

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Continuous Glucose Monitoring, Hypoglycemia, Type 1 Diabetes

Keywords

Interstitial glucose dynamics, Accuracy of continuous glucose monitoring under hypoglycemia, Glucose clamp

Brief summary

Achieving near-normoglycaemia has been established as the main objective for most patients with diabetes. However, it is well known that intensification of treatment is associated with an increase in the frequency of hypoglycemia, especially in the context of insulin therapy. The burden of hypoglycemia in terms of psychological implications, morbidity and even mortality, explains why it has been defined as the main limiting factor to achievement of good metabolic control. Continuous subcutaneous glucose monitoring (CGM) devices have been claimed to be useful in hypoglycemia detection/prevention, allowing theoretically for safer intensification of therapy in diabetic patients. However, accuracy of CGM devices, especially in the hypoglycemic range, raises some concerns. In fact, commercially available CGM devices estimate plasma glucose from measurements in the interstitial fluid and not in plasma. However, the relationship between plasma and interstitial glucose is not fully understood, especially under dynamic conditions, and this may explain the poor CGM performance during rapid changes in blood glucose and hypoglycemia. In this project, the relationship between plasma and interstitial glucose will be evaluated under conditions of normal glucose concentrations and hypoglycemia. Experiments will be performed to assess the role, if any, of different plasma insulin concentrations on the accuracy of CGM. All the information obtained may be relevant to the improvement of the ability of CGM devices to detect hypoglycemia and hypoglycemic risk.

Interventions

OTHERHigh insulin eu-hypoglycemic clamp

Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.

OTHERLow insulin eu-hypoglycemic clamp

Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.

Sponsors

Universitat Politècnica de València
CollaboratorOTHER
Fundación para la Investigación del Hospital Clínico de Valencia
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
BASIC_SCIENCE
Masking
DOUBLE (Subject, Outcomes Assessor)

Eligibility

Sex/Gender
ALL
Age
18 Years to 60 Years
Healthy volunteers
No

Inclusion criteria

* Subjects with type 1 diabetes mellitus meeting all of the following criteria will be considered for admission to the study: * Aged between 18 and 60 years * Under CSII or MDI treatment for at least six months before Visit 1 * Body mass index of between 18 and 30 kg/m2 * HbA1c 6.0-8.5% at Visit 1 * Normal laboratory values, ECG, and vital signs unless the investigator considered an abnormality to be clinically irrelevant * Women postmenopausal or using contraception judged by the investigator to be adequate (e.g., oral contraceptives, intra-uterine device or surgical treatment)

Exclusion criteria

* Subjects meeting any of the following criteria will not be included in the study: * Pregnancy and lactation * History of hypersensitivity to the study medications or to drugs with similar chemical structures * Confirmed hypoglycaemia unawareness * Progressive fatal diseases * History of drug or alcohol abuse * History of positive HIV or hepatitis B or C test * Impaired hepatic function, as shown by, but not limited to, SGPT or SGOT of more than twice the upper limit of the normal range at visit 1 * Impaired renal function, as shown by, but not limited to, serum creatinine \> 1.5 mg/dL at visit 1 * Clinically relevant microvascular, cardiovascular, hepatic, neurologic, endocrine or other major systemic diseases other than T1DM which could hinder implementation of the clinical study protocol or interpretation of the study results * Pre-planned surgery during the study * Blood donation of more than 500 ml during the past three months for men, or during the past six months for women * Mental condition rendering the subject unable to understand the nature, scope and possible consequences of the study * Subject unlikely to comply with clinical study protocol, e.g., uncooperative attitude, inability to return for follow-up visits, or poor likelihood of completing the study * Receipt of an experimental drug or use of an experimental device during the past 30 days.

Design outcomes

Primary

MeasureTime frameDescription
Normalized current signal from the CGM devices330 minutesCGM devices are electrochemical sensors. Currently available sensors are placed in the subcutaneous tissue where they react with the glucose of the interstitial fluid giving a current signal (in nano Amperes), which is proportional to glucose concentrations. The current signal is then filtered and transformed into plasma glucose estimations through a calibration algorithm built into the CGM device. In this study we will analyze the current signal from CGM obtained under two different experimental conditions: 1. Eu-hypoglycemic clamp with High plasma insulin concentrations 2. Eu-hypoglycemic clamp with relatively Low plasma insulin concentrations This will allow for establishing the role, if any, of different plasma insulin concentrations on the accuracy of CGM to identify plasma glucose falls resulting in hypoglycaemia. Since CGM sensors can exhibit different sensitivities to glucose concentrations, the current signal will be normalized before statistical analysis

Countries

Spain

Outcome results

None listed