Surgical Treatment of Aortic Stenosis With a Next Generation, Rapid Deployment Surgical Aortic Valve

Multi-CenTer Experience With the Rapid Deployment EDWARDS INTUITY Valve System FOR Aortic Valve ReplaceMent

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01700439
Acronym
TRANSFORM™
Enrollment
934
Registered
2012-10-04
Start date
2012-09-26
Completion date
2021-05-26
Last updated
2023-05-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Aortic Valve Incompetence, Aortic Valve Stenosis, Aortic Valve Stenosis With Insufficiency, Regurgitation, Aortic Valve

Keywords

Aortic valve, Heart valve, Tissue valve, Bioprosthesis, Valve disorder, Valve disease, Cardiac surgery, Less invasive cardiac surgery, Aortic valve stenosis, Aortic valve regurgitation, Bovine pericardium

Brief summary

The purpose of the clinical study is to prove that the heart valve device is safe, effective, and performs as intended.

Detailed description

This is a prospective, non-randomized, multi-center trial. Up to 950 subjects will be enrolled at up to 35 centers in the US. After replacement of their aortic heart valve with the EDWARDS INTUITY valve system, each patient will have routine follow-up tests at the following intervals: discharge, 3 months, 1 year, and annually thereafter for a minimum of five years.

Interventions

DEVICEEDWARDS INTUITY valve

Surgical replacement of the aortic valve with the EDWARDS INTUITY valve.

Sponsors

Edwards Lifesciences
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Inclusion Criteria - Subjects will be required to meet all inclusion criteria: 1. Male or female, age 18 years or older 2. Has aortic stenosis or stenosis-insufficiency of an aortic valve requiring a planned replacement as indicated in the preoperative evaluation 3. Is scheduled to undergo planned aortic valve replacement with or without concomitant coronary bypass surgery 4. Provide written informed consent 5. Geographically stable and agrees to attend follow-up assessments until all subjects have completed 5 years of follow up

Exclusion criteria

\-

Design outcomes

Primary

MeasureTime frameDescription
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortEvents occurring ≥ 31 days and up through 8 years post-implantLate patient years are calculated from 31 days post-implant to the date of the last follow-up visits (or contact) or adverse events. Late Patient year calculation:\[(Number of late adverse events/sum of late patient years) x 100\]

Secondary

MeasureTime frameDescription
Average Number of Days Subjects Were in the Intensive Care Unit (ICU)Day of procedure through discharge from the hospitalLength of time surgical subjects were in the intensive care unit (ICU) after their heart valve replacement procedure.
Subject's Effective Orifice Area (EOA) Measurement Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upEffective orifice area represents the cross-sectional area of the blood flow downstream of the aortic valve. Effective orifice area is evaluated by echocardiography over time. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size and type of valve.
Subject's Effective Orifice Area Index (EOAI) Measurement Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upEffective orifice area index represents the minimal cross-sectional area of the blood flow downstream of the mitral valve divided by the person's body surface area. Effective orifice area index is evaluated by echocardiography over time. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient and the size and type of valve.
Subject's Performance Index Measurement Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upPerformance index is defined as the subject's effective orifice area (the cross-sectional area of the blood flow downstream of the aortic valve) divided by the subject's native orifice area. Performance Index is evaluated by echocardiography over time and are dependent on the size and type of the valve. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient and the size and type of valve.
Subject's Cardiac Output Measurement at Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upCardiac output is the amount of blood the heart pumps through the circulatory system in a minute. Cardiac output is evaluated by echocardiography over time. Normal cardiac output in healthy volunteers over 60 years of age was reported between 3.1 and 6.4 liters per minute. In general, a higher value is considered better, and a lower value is considered worse.
Subject's Cardiac Index Measurement Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upHemodynamic performance - Cardiac index (CI) evaluated by echocardiography. Cardiac index is an assessment of the cardiac output value based on the patient's size. To obtain the cardiac index, divide the cardiac output by the person's body surface area (BSA). Cardiac Index is a useful marker of how well the heart is functioning as a pump by directly correlating the volume of blood pumped by the heart with an individual's body surface area. Normal cardiac output index in healthy volunteers over 60 years of age was reported between 1.1 and 3.2 L/min/m\^2. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient.
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upValvular regurgitation occurs when the valve in the heart does not close tightly allowing some of the blood that was pumped out of the heart to leak back into it. Valvular regurgitation is evaluated by echocardiography over time. It is assessed on a scale from 0 to 4, where 0 represents no regurgitation and 4 represents severe regurgitation. Higher numbers on the scale show a worsening outcome.
Percentage of Subjects With Edwards INTUITY Surgical Aortic Heart Valve Device Technical SuccessDay of procedureDevice technical success is defined as the successful delivery and deployment of the aortic trial heart valve with maximum of two attempts and subject leaving the operating room (OR) with valve in place.
Percentage of Subjects With Edwards INTUITY Surgical Heart Valve Procedural SuccessDay of procedure through discharge or 10 days post index procedure, whichever comes first.Procedural success is defined as device technical success followed by the absence of adverse events resulting in device reoperation implant of permanent pacemaker (with baseline sinus rhythm and no other pre-existing conduction issues), or valve-related death within discharge or 10 days post index procedure, whichever comes first.
Average Amount of Time Subject Spent on Cardiopulmonary BypassDay of procedureSurgical and hospitalization factors - Cardiopulmonary bypass time
Average Subject Time on Cardiopulmonary Cross ClampDay of procedureSurgical and hospitalization factors - Cardiopulmonary cross clamp time
Subject's New York Heart Association (NYHA) Classification Over TimeBaseline, 3-months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8 YearsThe New York Heart Association (NYHA) functional classification system relates symptoms to everyday activities and the patient's quality of life. Class I. Patients with cardiac disease but without resulting limitation of physical activity. Class II. Patients with cardiac disease resulting in slight limitation of physical activity. They are comfortable at rest. Class III. Patients with cardiac disease resulting in marked limitation of physical activity. They are comfortable at rest. Class IV. Patients with cardiac disease resulting in inability to carry on any physical activity without discomfort. Symptoms of heart failure or the anginal syndrome may be present even at rest.
Subject's Average Mean Gradient Measurements Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8-Years follow-upMean gradient is the average flow of blood through the aortic valve measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. In general, a higher value is considered worse, and a lower value is considered better but the value is dependent on the size and type of valve.
Subject's Average Peak Gradient Measurements Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8-Years follow-upPeak gradient is the maximum value measured of flow of blood through the aortic valve as measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. In general, a higher valve is considered worse, and a lower value is considered better, but the value is dependent on the size and type of valve.

Other

MeasureTime frameDescription
Subject's Average Red Blood Cell Count Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upThe red blood cell (RBC) count is a test that measures the number of oxygen-carrying blood cells in your blood. The red blood cell count normal range for women is 4.2 to 5.4 10\^6 cells/microliters and for men is 4.7 to 6.1 10\^6 cells/microliters. A high red blood cell count may indicate that you have a condition that's preventing you from getting enough oxygen. A low RBC may be caused by an infection, or a medical condition related to anemia. In general, results either lower or higher than the normal range can indicate something is wrong.
Subject's Average Hematocrit Percentage Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upLaboratory Analysis of Hematocrit Percentage on blood drawn from subjects. A hematocrit test measures the proportion of red blood cells in your blood. Red blood cells carry oxygen throughout your body. Having too few or too many red blood cells can be a sign of certain diseases. Normal hematocrit ranges for adult females are 36% to 44% and for adult males are 41% to 50%. In general, results either lower or higher than the normal range can indicate something is wrong.
Subject's Average Hemoglobin Percentage Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upLaboratory Analysis of Hemoglobin Count on blood drawn from subjects. Hemoglobin is an oxygen-carrying protein in red blood cells. In males, a healthy hemoglobin level is between 13.2 and 16.6 g/dL. In females, a hemoglobin count of between 11.6 and 15 g/dL is considered normal and healthy. In general, results either lower or higher than the normal range can indicate something is wrong.
Subject's Average Plasma Free Hemoglobin Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upLaboratory Analysis of Plasma Free Hemoglobin (Hgb) of blood drawn from subject. This blood test measures the level of free hemoglobin in the plasma (liquid portion of the blood). The reference range for plasma free hemoglobin is 0.0 to 15.2 mg/dL. In general, values higher than this can indicate something is wrong.
Subject's Average Score at Baseline and 1 Year on the Quality of Life SurveyBaseline and one year follow-upThe Medical Outcomes Study Short-Form 12 (SF-12) - physical and metal states. The SF-12 questionnaire scale ranges from 100, which reflects the best health status to 0, which reflects the worst health status.
Subject's Average White Blood Cell Count Over TimeDischarge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-upLaboratory analysis of White Blood Cell (WBC) count on blood drawn from subject; WBC fight infection. The normal reference range for WBC count is 4.5 to 10 10\^3 cells/microliters. In general, results either lower or higher than the normal range can indicate something is wrong.

Countries

United States

Participant flow

Participants by arm

ArmCount
Edwards TRANSFORM Trial Cohort
Subjects who were enrolled in the TRANSFORM trial and were eligible to receive an Edwards INTUITY surgical aortic heart valve.
934
Total934

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyDeath186
Overall StudyLost to Follow-up47
Overall StudyWithdrawal by Subject183

Baseline characteristics

CharacteristicEdwards TRANSFORM Trial Cohort
Age, Continuous73.4 Years
STANDARD_DEVIATION 8.3
Ethnicity (NIH/OMB)
Hispanic or Latino
4 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
865 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
65 Participants
Region of Enrollment
United States
934 participants
Sex: Female, Male
Female
336 Participants
Sex: Female, Male
Male
598 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
178 / 8858 / 49
other
Total, other adverse events
868 / 88547 / 49
serious
Total, serious adverse events
756 / 88537 / 49

Outcome results

Primary

Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort

Late patient years are calculated from 31 days post-implant to the date of the last follow-up visits (or contact) or adverse events. Late Patient year calculation:\[(Number of late adverse events/sum of late patient years) x 100\]

Time frame: Events occurring ≥ 31 days and up through 8 years post-implant

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve.

ArmMeasureGroupValue (NUMBER)
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortConduction defects with permanent pacemaker0.1 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortConduction defects without permanent pacemaker0.3 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortAll cause mortality3.9 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortValve related mortality0.9 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortStudy valve reoperation0.6 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortStudy valve explant0.3 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortThromboembolism1.8 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortValve thrombosis0.0 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortEndocarditis0.1 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortAll bleeding4.7 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortMajor bleeding2.4 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortAll paravalvular leak0.5 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortMajor paravalvular leak (PVL)0.3 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortHemolysis0.2 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortNon-structural valve dysfunction other than PVL0.1 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortValve migration/embolization0.0 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortValve malposition0.0 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortValve instability0.0 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortValve dislodgement0.0 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortValve stenosis0.1 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortStructural valve deterioration0.4 Percentage of events/late patient years
Edwards INTUITY Surgical Aortic Heart ValveNumber of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted CohortConduction defects0.4 Percentage of events/late patient years
Secondary

Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size

Valvular regurgitation occurs when the valve in the heart does not close tightly allowing some of the blood that was pumped out of the heart to leak back into it. Valvular regurgitation is evaluated by echocardiography over time. It is assessed on a scale from 0 to 4, where 0 represents no regurgitation and 4 represents severe regurgitation. Higher numbers on the scale show a worsening outcome.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsNone (0)/Trival (+1)7 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsMild (+2)5 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsMild (+2)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsNone (0)/Trival (+1)1 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsMild (+2)4 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsNone (0)/Trival (+1)20 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsNone (0)/Trival (+1)31 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsNone (0)/Trival (+1)43 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearNone (0)/Trival (+1)40 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearMild (+2)5 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsNone (0)/Trival (+1)38 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeNone (0)/Trival (+1)46 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsNone (0)/Trival (+1)26 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsNone (0)/Trival (+1)33 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsMild (+2)5 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart ValveAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsMild (+2)12 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsMild (+2)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearModerate (+3)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsNone (0)/Trival (+1)25 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsMild (+2)8 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsNone (0)/Trival (+1)6 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsNone (0)/Trival (+1)138 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsMild (+2)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsMild (+2)7 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsNone (0)/Trival (+1)87 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsNone (0)/Trival (+1)156 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsMild (+2)7 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeNone (0)/Trival (+1)171 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsNone (0)/Trival (+1)104 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsNone (0)/Trival (+1)53 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsNone (0)/Trival (+1)120 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearNone (0)/Trival (+1)156 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsMild (+2)9 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeModerate (+3)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearMild (+2)9 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeMild (+2)3 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsMild (+2)9 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsModerate (+3)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsNone (0)/Trival (+1)229 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearNone (0)/Trival (+1)245 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsMild (+2)17 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearMild (+2)18 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsModerate (+3)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsNone (0)/Trival (+1)87 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsNone (0)/Trival (+1)146 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsMild (+2)15 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsModerate (+3)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsNone (0)/Trival (+1)30 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsMild (+2)5 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsNone (0)/Trival (+1)10 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsMild (+2)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearModerate (+3)5 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsMild (+2)19 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsModerate (+3)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsNone (0)/Trival (+1)211 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsMild (+2)18 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsModerate (+3)3 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeNone (0)/Trival (+1)266 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeMild (+2)12 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeModerate (+3)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeSevere (+4)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsNone (0)/Trival (+1)191 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsNone (0)/Trival (+1)245 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsMild (+2)16 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIIAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsSevere (+4)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsNone (0)/Trival (+1)78 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsSevere (+4)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearMild (+2)21 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsNone (0)/Trival (+1)193 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeMild (+2)13 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsNone (0)/Trival (+1)179 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsMild (+2)19 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearNone (0)/Trival (+1)200 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsNone (0)/Trival (+1)219 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsMild (+2)18 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsSevere (+4)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsMild (+2)18 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsMild (+2)10 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeSevere (+4)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearSevere (+4)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsModerate (+3)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsModerate (+3)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsMild (+2)6 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsMild (+2)18 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearModerate (+3)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsNone (0)/Trival (+1)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsNone (0)/Trival (+1)22 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsModerate (+3)6 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsMild (+2)23 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsSevere (+4)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsNone (0)/Trival (+1)160 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeNone (0)/Trival (+1)235 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsNone (0)/Trival (+1)129 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearModerate (+3)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsMild (+2)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsMild (+2)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearSevere (+4)1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsNone (0)/Trival (+1)86 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsNone (0)/Trival (+1)97 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsMild (+2)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsMild (+2)5 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsModerate (+3)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsNone (0)/Trival (+1)61 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size2 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsNone (0)/Trival (+1)31 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsNone (0)/Trival (+1)79 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsMild (+2)5 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsNone (0)/Trival (+1)77 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsModerate (+3)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size4 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsMild (+2)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeNone (0)/Trival (+1)97 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearMild (+2)5 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeMild (+2)3 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size1 YearNone (0)/Trival (+1)89 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeModerate (+3)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsMild (+2)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 monthsModerate (+3)4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve SizeDischargeSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsNone (0)/Trival (+1)2 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size3 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsSevere (+4)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsMild (+2)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size7 YearsNone (0)/Trival (+1)6 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size6 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size8 YearsModerate (+3)0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - 27mmAmount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size5 YearsModerate (+3)1 Participants
Secondary

Average Amount of Time Subject Spent on Cardiopulmonary Bypass

Surgical and hospitalization factors - Cardiopulmonary bypass time

Time frame: Day of procedure

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve.

ArmMeasureValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveAverage Amount of Time Subject Spent on Cardiopulmonary Bypass86.6 MinutesStandard Deviation 37.4
Secondary

Average Number of Days Subjects Were in the Intensive Care Unit (ICU)

Length of time surgical subjects were in the intensive care unit (ICU) after their heart valve replacement procedure.

Time frame: Day of procedure through discharge from the hospital

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve. Data is not available for 4 subjects.

ArmMeasureValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveAverage Number of Days Subjects Were in the Intensive Care Unit (ICU)3.1 DaysStandard Deviation 3.7
Secondary

Average Subject Time on Cardiopulmonary Cross Clamp

Surgical and hospitalization factors - Cardiopulmonary cross clamp time

Time frame: Day of procedure

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve. Data is not available for one subject.

ArmMeasureValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveAverage Subject Time on Cardiopulmonary Cross Clamp65.4 MinutesStandard Deviation 29.7
Secondary

Percentage of Subjects With Edwards INTUITY Surgical Aortic Heart Valve Device Technical Success

Device technical success is defined as the successful delivery and deployment of the aortic trial heart valve with maximum of two attempts and subject leaving the operating room (OR) with valve in place.

Time frame: Day of procedure

Population: Patients who were eligible to receive an Edwards INTUITY surgical aortic heart valve.

ArmMeasureValue (NUMBER)
Edwards INTUITY Surgical Aortic Heart ValvePercentage of Subjects With Edwards INTUITY Surgical Aortic Heart Valve Device Technical Success94.8 Percentage of subjects
Secondary

Percentage of Subjects With Edwards INTUITY Surgical Heart Valve Procedural Success

Procedural success is defined as device technical success followed by the absence of adverse events resulting in device reoperation implant of permanent pacemaker (with baseline sinus rhythm and no other pre-existing conduction issues), or valve-related death within discharge or 10 days post index procedure, whichever comes first.

Time frame: Day of procedure through discharge or 10 days post index procedure, whichever comes first.

Population: Patients who were eligible to receive an Edwards INTUITY surgical aortic heart valve.

ArmMeasureValue (NUMBER)
Edwards INTUITY Surgical Aortic Heart ValvePercentage of Subjects With Edwards INTUITY Surgical Heart Valve Procedural Success94.4 Percentage of subjects
Secondary

Subject's Average Mean Gradient Measurements Over Time

Mean gradient is the average flow of blood through the aortic valve measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. In general, a higher value is considered worse, and a lower value is considered better but the value is dependent on the size and type of valve.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8-Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time4 Years12.8 mmHgStandard Deviation 3.8
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time5 Years12.4 mmHgStandard Deviation 3.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time3 Years13.7 mmHgStandard Deviation 4.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time6 Years13.1 mmHgStandard Deviation 3.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over TimeDischarge17.7 mmHgStandard Deviation 6.2
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time3 months14.4 mmHgStandard Deviation 4.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time7 Years12.1 mmHgStandard Deviation 4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time1 Year13.9 mmHgStandard Deviation 4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time2 Years13.1 mmHgStandard Deviation 4.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Mean Gradient Measurements Over Time8 Years11.2 mmHgStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time7 Years10.8 mmHgStandard Deviation 3.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time3 Years11.0 mmHgStandard Deviation 4.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time3 months11.1 mmHgStandard Deviation 3.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time6 Years10.5 mmHgStandard Deviation 3.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time1 Year11.3 mmHgStandard Deviation 3.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time5 Years10.7 mmHgStandard Deviation 3.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time8 Years13.2 mmHgStandard Deviation 5
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time2 Years11.1 mmHgStandard Deviation 3.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over Time4 Years10.8 mmHgStandard Deviation 3.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Mean Gradient Measurements Over TimeDischarge13.0 mmHgStandard Deviation 4.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time6 Years11.1 mmHgStandard Deviation 6.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over TimeDischarge12.3 mmHgStandard Deviation 4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time7 Years12.8 mmHgStandard Deviation 7.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time3 Years10.0 mmHgStandard Deviation 3.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time3 months10.3 mmHgStandard Deviation 3.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time2 Years10.0 mmHgStandard Deviation 3.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time4 Years10.3 mmHgStandard Deviation 6.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time8 Years12.3 mmHgStandard Deviation 8.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time5 Years10.0 mmHgStandard Deviation 5.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Mean Gradient Measurements Over Time1 Year10.0 mmHgStandard Deviation 3.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time3 months9.7 mmHgStandard Deviation 3.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time7 Years9.8 mmHgStandard Deviation 4
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time8 Years12.9 mmHgStandard Deviation 8.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over TimeDischarge11.2 mmHgStandard Deviation 3.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time1 Year9.3 mmHgStandard Deviation 3
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time2 Years9.3 mmHgStandard Deviation 2.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time3 Years9.5 mmHgStandard Deviation 3.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time4 Years9.5 mmHgStandard Deviation 4
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time5 Years9.8 mmHgStandard Deviation 6.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Mean Gradient Measurements Over Time6 Years9.3 mmHgStandard Deviation 4
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time4 Years8.0 mmHgStandard Deviation 3.7
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time1 Year8.1 mmHgStandard Deviation 3.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time8 Years9.4 mmHgStandard Deviation 2.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time6 Years9.2 mmHgStandard Deviation 6.5
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time3 months8.3 mmHgStandard Deviation 2.5
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time2 Years8.1 mmHgStandard Deviation 3.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over TimeDischarge9.7 mmHgStandard Deviation 3.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time7 Years7.7 mmHgStandard Deviation 1.8
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time3 Years7.9 mmHgStandard Deviation 2.9
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Mean Gradient Measurements Over Time5 Years7.9 mmHgStandard Deviation 4.4
Secondary

Subject's Average Peak Gradient Measurements Over Time

Peak gradient is the maximum value measured of flow of blood through the aortic valve as measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. In general, a higher valve is considered worse, and a lower value is considered better, but the value is dependent on the size and type of valve.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8-Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over TimeDischarge32.8 mmHgStandard Deviation 11.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time8 Years21.3 mmHgStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time3 Years24.7 mmHgStandard Deviation 8.2
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time7 Years22.1 mmHgStandard Deviation 7.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time6 Years23.6 mmHgStandard Deviation 6.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time5 Years22.7 mmHgStandard Deviation 6.3
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time4 Years23.0 mmHgStandard Deviation 6.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time2 Years23.4 mmHgStandard Deviation 7.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time1 Year25.0 mmHgStandard Deviation 7.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Peak Gradient Measurements Over Time3 months26.0 mmHgStandard Deviation 8
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time7 Years19.6 mmHgStandard Deviation 5.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time1 Year20.5 mmHgStandard Deviation 6.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over TimeDischarge24.1 mmHgStandard Deviation 7.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time3 months19.9 mmHgStandard Deviation 6.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time2 Years20.2 mmHgStandard Deviation 6.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time3 Years20.0 mmHgStandard Deviation 7.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time4 Years19.8 mmHgStandard Deviation 6.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time5 Years19.5 mmHgStandard Deviation 7
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time6 Years19.1 mmHgStandard Deviation 6.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Average Peak Gradient Measurements Over Time8 Years24.4 mmHgStandard Deviation 8.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time3 months19.0 mmHgStandard Deviation 6.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time5 Years18.4 mmHgStandard Deviation 10
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time1 Year18.2 mmHgStandard Deviation 6
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time7 Years23.5 mmHgStandard Deviation 13.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time2 Years18.4 mmHgStandard Deviation 6.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time4 Years18.9 mmHgStandard Deviation 10.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time8 Years21.0 mmHgStandard Deviation 12.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time6 Years19.8 mmHgStandard Deviation 9.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over Time3 Years18.5 mmHgStandard Deviation 6.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Average Peak Gradient Measurements Over TimeDischarge22.9 mmHgStandard Deviation 7.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time6 Years17.1 mmHgStandard Deviation 7.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over TimeDischarge20.8 mmHgStandard Deviation 6.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time7 Years18.5 mmHgStandard Deviation 7.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time1 Year17.0 mmHgStandard Deviation 5.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time4 Years17.6 mmHgStandard Deviation 7.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time3 Years17.4 mmHgStandard Deviation 6.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time8 Years22.7 mmHgStandard Deviation 14.7
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time3 months17.6 mmHgStandard Deviation 5.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time2 Years17.0 mmHgStandard Deviation 5.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Average Peak Gradient Measurements Over Time5 Years17.9 mmHgStandard Deviation 10.7
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time2 Years14.6 mmHgStandard Deviation 6.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time8 Years17.2 mmHgStandard Deviation 3.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time3 Years14.5 mmHgStandard Deviation 5.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over TimeDischarge17.9 mmHgStandard Deviation 5.8
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time5 Years14.5 mmHgStandard Deviation 7.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time6 Years16.9 mmHgStandard Deviation 10.9
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time1 Year14.6 mmHgStandard Deviation 5.8
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time4 Years14.6 mmHgStandard Deviation 6.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time3 months15.0 mmHgStandard Deviation 4.4
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Average Peak Gradient Measurements Over Time7 Years14.6 mmHgStandard Deviation 3.7
Secondary

Subject's Cardiac Index Measurement Over Time

Hemodynamic performance - Cardiac index (CI) evaluated by echocardiography. Cardiac index is an assessment of the cardiac output value based on the patient's size. To obtain the cardiac index, divide the cardiac output by the person's body surface area (BSA). Cardiac Index is a useful marker of how well the heart is functioning as a pump by directly correlating the volume of blood pumped by the heart with an individual's body surface area. Normal cardiac output index in healthy volunteers over 60 years of age was reported between 1.1 and 3.2 L/min/m\^2. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time4 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time3 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over TimeDischarge2.5 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time6 Years2.5 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time8 Years1.5 Liters/minutes/meters squaredStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time7 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time2 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time1 Year2.5 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time5 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Index Measurement Over Time3 months2.6 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time5 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time1 Year2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time2 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over TimeDischarge2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time3 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time4 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time3 months2.4 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time7 Years2.2 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time8 Years2.6 Liters/minutes/meters squaredStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Index Measurement Over Time6 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time2 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time4 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time5 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over TimeDischarge2.7 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time1 Year2.4 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time3 months2.5 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time6 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time7 Years2.5 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time8 Years2.6 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Index Measurement Over Time3 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over TimeDischarge2.6 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time8 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time5 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time3 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time4 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time2 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time6 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time3 months2.5 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time7 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Index Measurement Over Time1 Year2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time2 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.4
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over TimeDischarge2.7 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time3 months2.5 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time4 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time6 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time1 Year2.5 Liters/minutes/meters squaredStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time8 Years2.8 Liters/minutes/meters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time7 Years2.3 Liters/minutes/meters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time3 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Index Measurement Over Time5 Years2.4 Liters/minutes/meters squaredStandard Deviation 0.4
Secondary

Subject's Cardiac Output Measurement at Over Time

Cardiac output is the amount of blood the heart pumps through the circulatory system in a minute. Cardiac output is evaluated by echocardiography over time. Normal cardiac output in healthy volunteers over 60 years of age was reported between 3.1 and 6.4 liters per minute. In general, a higher value is considered better, and a lower value is considered worse.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time5 Years4.2 Liters/minutesStandard Deviation 0.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time7 Years3.9 Liters/minutesStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time6 Years4.3 Liters/minutesStandard Deviation 0.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time3 months4.3 Liters/minutesStandard Deviation 0.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over TimeDischarge4.3 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time2 Years4.1 Liters/minutesStandard Deviation 0.8
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time8 Years2.9 Liters/minutesStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time3 Years4.0 Liters/minutesStandard Deviation 0.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time4 Years4.1 Liters/minutesStandard Deviation 0.8
Edwards INTUITY Surgical Aortic Heart ValveSubject's Cardiac Output Measurement at Over Time1 Year4.3 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time8 Years4.9 Liters/minutesStandard Deviation 1.6
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time5 Years4.3 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time7 Years4.2 Liters/minutesStandard Deviation 0.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time2 Years4.3 Liters/minutesStandard Deviation 0.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time3 Years4.3 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time1 Year4.4 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time6 Years4.2 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over TimeDischarge4.6 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time4 Years4.1 Liters/minutesStandard Deviation 0.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Cardiac Output Measurement at Over Time3 months4.4 Liters/minutesStandard Deviation 0.8
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time4 Years4.9 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time1 Year4.9 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time7 Years5.0 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time2 Years4.9 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time3 months5.0 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time5 Years4.8 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time8 Years5.4 Liters/minutesStandard Deviation 1.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time3 Years4.9 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over Time6 Years4.9 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Cardiac Output Measurement at Over TimeDischarge5.4 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time7 Years5.1 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over TimeDischarge5.6 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time3 months5.2 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time1 Year5.1 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time2 Years5.0 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time3 Years5.1 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time4 Years5.1 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time5 Years5.0 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time6 Years5.0 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Cardiac Output Measurement at Over Time8 Years4.9 Liters/minutesStandard Deviation 1.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time5 Years5.2 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time4 Years5.2 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time3 Years5.2 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time3 months5.4 Liters/minutesStandard Deviation 1.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time7 Years5.0 Liters/minutesStandard Deviation 0.9
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time2 Years5.3 Liters/minutesStandard Deviation 1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time1 Year5.5 Liters/minutesStandard Deviation 1.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over TimeDischarge5.9 Liters/minutesStandard Deviation 1.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time8 Years5.5 Liters/minutesStandard Deviation 1.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Cardiac Output Measurement at Over Time6 Years5.3 Liters/minutesStandard Deviation 1.2
Secondary

Subject's Effective Orifice Area (EOA) Measurement Over Time

Effective orifice area represents the cross-sectional area of the blood flow downstream of the aortic valve. Effective orifice area is evaluated by echocardiography over time. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size and type of valve.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time1 Year1.1 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over TimeDischarge1.1 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time4 Years1.2 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time3 months1.1 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time6 Years1.2 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time5 Years1.2 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time3 Years1.1 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time7 Years1.2 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time2 Years1.1 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area (EOA) Measurement Over Time8 Years1.1 Centimeters squaredStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time1 Year1.4 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time3 months1.3 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time2 Years1.4 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over TimeDischarge1.3 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time4 Years1.4 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time7 Years1.4 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time3 Years1.4 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time5 Years1.4 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time6 Years1.4 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area (EOA) Measurement Over Time8 Years1.4 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time2 Years1.7 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time5 Years1.7 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time6 Years1.6 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time7 Years1.6 Centimeters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time8 Years1.6 Centimeters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over TimeDischarge1.7 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time3 months1.7 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time1 Year1.7 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time3 Years1.7 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area (EOA) Measurement Over Time4 Years1.7 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time3 months1.9 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time8 Years1.7 Centimeters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time1 Year1.9 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time7 Years1.8 Centimeters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time2 Years1.9 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time6 Years1.9 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time4 Years1.9 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time3 Years1.9 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over TimeDischarge1.9 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area (EOA) Measurement Over Time5 Years1.8 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time2 Years2.2 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time3 months2.2 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time3 Years2.1 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time7 Years2.1 Centimeters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time4 Years2.1 Centimeters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time5 Years2.1 Centimeters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time1 Year2.2 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over TimeDischarge2.2 Centimeters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time6 Years2.0 Centimeters squaredStandard Deviation 0.3
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area (EOA) Measurement Over Time8 Years2.1 Centimeters squaredStandard Deviation 0.1
Secondary

Subject's Effective Orifice Area Index (EOAI) Measurement Over Time

Effective orifice area index represents the minimal cross-sectional area of the blood flow downstream of the mitral valve divided by the person's body surface area. Effective orifice area index is evaluated by echocardiography over time. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient and the size and type of valve.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time4 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time8 Years0.6 centimeters squared/meters squaredStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 months0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over TimeDischarge0.6 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time7 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time1 Year0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time6 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time2 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Effective Orifice Area Index (EOAI) Measurement Over Time5 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time7 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time2 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time4 Years0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over TimeDischarge0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time8 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time5 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time1 Year0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 months0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Effective Orifice Area Index (EOAI) Measurement Over Time6 Years0.7 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time1 Year0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over TimeDischarge0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 months0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time2 Years0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 Years0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time4 Years0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time5 Years0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time6 Years0.8 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time7 Years0.8 centimeters squared/meters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Effective Orifice Area Index (EOAI) Measurement Over Time8 Years0.8 centimeters squared/meters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over TimeDischarge0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 months0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time8 Years0.8 centimeters squared/meters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time7 Years0.9 centimeters squared/meters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time1 Year0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time5 Years0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 Years0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time6 Years0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time4 Years0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Effective Orifice Area Index (EOAI) Measurement Over Time2 Years0.9 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time4 Years1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time5 Years1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time1 Year1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time8 Years1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time6 Years0.9 centimeters squared/meters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 months1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time7 Years1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time3 Years1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over TimeDischarge1.0 centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Effective Orifice Area Index (EOAI) Measurement Over Time2 Years1.0 centimeters squared/meters squaredStandard Deviation 0.1
Secondary

Subject's New York Heart Association (NYHA) Classification Over Time

The New York Heart Association (NYHA) functional classification system relates symptoms to everyday activities and the patient's quality of life. Class I. Patients with cardiac disease but without resulting limitation of physical activity. Class II. Patients with cardiac disease resulting in slight limitation of physical activity. They are comfortable at rest. Class III. Patients with cardiac disease resulting in marked limitation of physical activity. They are comfortable at rest. Class IV. Patients with cardiac disease resulting in inability to carry on any physical activity without discomfort. Symptoms of heart failure or the anginal syndrome may be present even at rest.

Time frame: Baseline, 3-months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8 Years

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over TimeBaseline135 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time8 Years28 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time1 Year648 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time3 months671 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time6 Years251 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time2 Years606 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time5 Years420 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time3 Years551 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time7 Years109 Participants
Edwards INTUITY Surgical Aortic Heart ValveSubject's New York Heart Association (NYHA) Classification Over Time4 Years487 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time3 Years122 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time5 Years109 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time7 Years31 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time6 Years63 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time8 Years5 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time2 Years133 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time3 months148 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time1 Year145 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over Time4 Years123 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's New York Heart Association (NYHA) Classification Over TimeBaseline468 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over TimeBaseline264 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time3 months20 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time1 Year18 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time2 Years22 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time3 Years19 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time7 Years4 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time8 Years1 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time4 Years21 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time5 Years14 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's New York Heart Association (NYHA) Classification Over Time6 Years14 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time6 Years0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time2 Years3 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time7 Years0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time5 Years0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time1 Year3 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time3 months0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over TimeBaseline15 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time8 Years0 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time4 Years3 Participants
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's New York Heart Association (NYHA) Classification Over Time3 Years1 Participants
Secondary

Subject's Performance Index Measurement Over Time

Performance index is defined as the subject's effective orifice area (the cross-sectional area of the blood flow downstream of the aortic valve) divided by the subject's native orifice area. Performance Index is evaluated by echocardiography over time and are dependent on the size and type of the valve. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient and the size and type of valve.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time1 Year0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time3 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time7 Years1.0 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time3 months0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time6 Years1.0 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time5 Years1.0 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time8 Years0.9 Centimeters squared/meters squaredStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time2 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over TimeDischarge0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Performance Index Measurement Over Time4 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over TimeDischarge0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time1 Year0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time3 months0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time4 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time3 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time7 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time5 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time8 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time2 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IISubject's Performance Index Measurement Over Time6 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time4 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time3 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time5 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time6 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over TimeDischarge0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time3 months0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time1 Year0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time2 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time7 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IIISubject's Performance Index Measurement Over Time8 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.2
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time3 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time1 Year0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time6 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time5 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time2 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time4 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time8 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time7 Years0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over Time3 months0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - Class IVSubject's Performance Index Measurement Over TimeDischarge0.9 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time4 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time3 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time1 Year0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over TimeDischarge0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time5 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time6 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time3 months0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time8 Years0.8 Centimeters squared/meters squaredStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time7 Years0.8 Centimeters squared/meters squaredStandard Deviation 0
Edwards INTUITY Surgical Aortic Heart Valve - 27mmSubject's Performance Index Measurement Over Time2 Years0.8 Centimeters squared/meters squaredStandard Deviation 0.1
Other Pre-specified

Subject's Average Hematocrit Percentage Over Time

Laboratory Analysis of Hematocrit Percentage on blood drawn from subjects. A hematocrit test measures the proportion of red blood cells in your blood. Red blood cells carry oxygen throughout your body. Having too few or too many red blood cells can be a sign of certain diseases. Normal hematocrit ranges for adult females are 36% to 44% and for adult males are 41% to 50%. In general, results either lower or higher than the normal range can indicate something is wrong.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time3 Years40.42 Percentage of RBCStandard Deviation 4.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time4 Years40.40 Percentage of RBCStandard Deviation 4.2
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over TimeDischarge29.94 Percentage of RBCStandard Deviation 4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time3 months39.28 Percentage of RBCStandard Deviation 4.2
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time1 Year40.25 Percentage of RBCStandard Deviation 4.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time2 Years40.30 Percentage of RBCStandard Deviation 4.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time5 Years40.66 Percentage of RBCStandard Deviation 4.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time6 Years40.41 Percentage of RBCStandard Deviation 4.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time7 Years40.00 Percentage of RBCStandard Deviation 4.8
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hematocrit Percentage Over Time8 Years39.31 Percentage of RBCStandard Deviation 5.4
Other Pre-specified

Subject's Average Hemoglobin Percentage Over Time

Laboratory Analysis of Hemoglobin Count on blood drawn from subjects. Hemoglobin is an oxygen-carrying protein in red blood cells. In males, a healthy hemoglobin level is between 13.2 and 16.6 g/dL. In females, a hemoglobin count of between 11.6 and 15 g/dL is considered normal and healthy. In general, results either lower or higher than the normal range can indicate something is wrong.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time2 Years13.37 g/dLStandard Deviation 1.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time8 Years12.87 g/dLStandard Deviation 1.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time3 Years13.42 g/dLStandard Deviation 1.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over TimeDischarge9.95 g/dLStandard Deviation 1.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time3 months12.82 g/dLStandard Deviation 1.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time1 Year13.33 g/dLStandard Deviation 1.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time4 Years13.39 g/dLStandard Deviation 1.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time5 Years13.22 g/dLStandard Deviation 2.2
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time6 Years13.20 g/dLStandard Deviation 1.8
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Hemoglobin Percentage Over Time7 Years13.10 g/dLStandard Deviation 1.8
Other Pre-specified

Subject's Average Plasma Free Hemoglobin Over Time

Laboratory Analysis of Plasma Free Hemoglobin (Hgb) of blood drawn from subject. This blood test measures the level of free hemoglobin in the plasma (liquid portion of the blood). The reference range for plasma free hemoglobin is 0.0 to 15.2 mg/dL. In general, values higher than this can indicate something is wrong.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over TimeDischarge11.89 mg/dLStandard Deviation 14.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time3 months13.12 mg/dLStandard Deviation 13.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time1 Year12.08 mg/dLStandard Deviation 12.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time2 Years12.06 mg/dLStandard Deviation 13.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time3 Years11.90 mg/dLStandard Deviation 15.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time4 Years11.88 mg/dLStandard Deviation 14.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time5 Years12.73 mg/dLStandard Deviation 16.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time6 Years16.32 mg/dLStandard Deviation 14.4
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time7 Years18.43 mg/dLStandard Deviation 14.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Plasma Free Hemoglobin Over Time8 Years22.61 mg/dLStandard Deviation 11.4
Other Pre-specified

Subject's Average Red Blood Cell Count Over Time

The red blood cell (RBC) count is a test that measures the number of oxygen-carrying blood cells in your blood. The red blood cell count normal range for women is 4.2 to 5.4 10\^6 cells/microliters and for men is 4.7 to 6.1 10\^6 cells/microliters. A high red blood cell count may indicate that you have a condition that's preventing you from getting enough oxygen. A low RBC may be caused by an infection, or a medical condition related to anemia. In general, results either lower or higher than the normal range can indicate something is wrong.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over TimeDischarge3.31 10^6 cells/microlitersStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time3 months4.51 10^6 cells/microlitersStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time1 Year4.46 10^6 cells/microlitersStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time2 Years4.45 10^6 cells/microlitersStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time3 Years4.46 10^6 cells/microlitersStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time6 Years4.38 10^6 cells/microlitersStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time7 Years4.33 10^6 cells/microlitersStandard Deviation 0.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time8 Years4.28 10^6 cells/microlitersStandard Deviation 0.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time4 Years4.44 10^6 cells/microlitersStandard Deviation 0.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Red Blood Cell Count Over Time5 Years4.43 10^6 cells/microlitersStandard Deviation 0.6
Other Pre-specified

Subject's Average Score at Baseline and 1 Year on the Quality of Life Survey

The Medical Outcomes Study Short-Form 12 (SF-12) - physical and metal states. The SF-12 questionnaire scale ranges from 100, which reflects the best health status to 0, which reflects the worst health status.

Time frame: Baseline and one year follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve where data is available at both the baseline and 1 year follow-up visits. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyMental Health - Baseline51.3 Units on a scaleStandard Deviation 9.9
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyMental Health - 1 year follow-up54.3 Units on a scaleStandard Deviation 8.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyPhysical Health - Baseline41.8 Units on a scaleStandard Deviation 10.2
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average Score at Baseline and 1 Year on the Quality of Life SurveyPhysical Health - 1 year follow-up47.6 Units on a scaleStandard Deviation 9.7
Other Pre-specified

Subject's Average White Blood Cell Count Over Time

Laboratory analysis of White Blood Cell (WBC) count on blood drawn from subject; WBC fight infection. The normal reference range for WBC count is 4.5 to 10 10\^3 cells/microliters. In general, results either lower or higher than the normal range can indicate something is wrong.

Time frame: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

ArmMeasureGroupValue (MEAN)Dispersion
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time3 months7.27 10^3 cells/microlitersStandard Deviation 2.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time7 Years8.05 10^3 cells/microlitersStandard Deviation 9.8
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over TimeDischarge9.54 10^3 cells/microlitersStandard Deviation 3
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time1 Year7.31 10^3 cells/microlitersStandard Deviation 2.8
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time2 Years7.47 10^3 cells/microlitersStandard Deviation 4.5
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time3 Years7.47 10^3 cells/microlitersStandard Deviation 5.7
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time4 Years7.41 10^3 cells/microlitersStandard Deviation 3
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time5 Years7.37 10^3 cells/microlitersStandard Deviation 3.6
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time6 Years7.47 10^3 cells/microlitersStandard Deviation 6.1
Edwards INTUITY Surgical Aortic Heart ValveSubject's Average White Blood Cell Count Over Time8 Years14.0 10^3 cells/microlitersStandard Deviation 36.8

Source: ClinicalTrials.gov · Data processed: Mar 12, 2026