Cryoablation of Atrial Fibrillation Using a Novel Cryoablation System

Cryoablation of Atrial Fibrillation Using a Novel Cryoablation System

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01687036
Enrollment
10
Registered
2012-09-18
Start date
2012-09-30
Completion date
2013-06-30
Last updated
2018-12-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Atrial Fibrillation

Keywords

Atrial fibrillation, Cryoablation

Brief summary

Primary Endpoint Safety and tolerability of ablation using the AFreeze Cryoablation System consisting of the ablation CoolLoop® catheter, its steerable sheath and the cryoconsole Cryo-Caddy will be assessed and expressed in number of participants with Adverse Events (AEs).

Detailed description

Therefore, the primary endpoint will be a measurement of the following parameters and expressed in number of participants with AE: * deformation of the catheter loop resulting in entrapment of the catheter in the heart and difficulties removing the catheter during visit 3 (treatment) * phrenic nerve palsy during visit 3 (treatment)- onset and time course between visit 3 (treatment) and visit 5 (discharge) of death, pericardial tamponade, valve damage requiring surgery and hemorrhage requiring transfusion. * onset and time course between visit 3 (treatment) and visit 9 (final visit) of atrioesophageal fistula, sepsis, abscesses, endocarditis, stroke, transient ischemic attack, and pulmonary vein (PV) stenosis requiring intervention. Secondary Endpoint The following parameters will be assessed and expressed in number of participants with AE/duration time: * feasibility, defined by the ability to position the catheter in the proximal part in each of the PVs using an over-the-wire technique, forming the cryo-applicator of the catheter to a loop by operating the catheter handle, position-ing the loop at the wall of the PV antrum and delivering cryothermia. * acute efficacy of catheter ablation is defined as the percentage of electrodes within the PV without measureable signals. * clinical efficacy of catheter ablation defined by freedom of Atrial Fibrillation (AF) four, eight and twelve weeks after ablation. * procedure time defined from introduction of the CoolLoop® catheter into the left atrium until removal of the CoolLoop® catheter from the left atrium after termination of the last cryoapplication. * fluoroscopy time defined from introduction of the CoolLoop® catheter into the left atrium until removal of the CoolLoop® catheter from the left atrium after termination of the last cryoapplication. * cumulative cryoablation time. * onset and time course between visit 3 (treatment) and visit 5 (discharge) of pneumothorax, hemothorax, femoral pseudoaneurysm, arteriovenous fistulae, hemorrhage not requiring transfusion.

Interventions

DEVICECryoablation

Cryoablation of Atrial Fibrillation Using a Novel Cryoablation System

Sponsors

afreeze GmbH
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 70 Years
Healthy volunteers
No

Inclusion criteria

* age 18 - 70 years * symptomatic paroxysmal AF, according to the actual guidelines of the Eu-ropean Society of Cardiology, despite treatment with at least one antiarr- hythmic drug * at least one episode of AF within the last 3 months documented by ECG * signed and dated informed consent documented by the patient, indicating that the patient has been informed of all the pertinent aspects of the trial prior to study enrollment * female subjects of childbearing potential can be included, if they: agree to use, and be able to comply with, effective contraception without in-terruption, from study enrollment, throughout hospitalization and until 12 weeks after catheter ablation (day 0, visit 3). This applies unless the subject commits to absolute and continued abstinence. The following are effective methods of contraception: * Implant, * Levonorgestrel-releasing intrauterine system (IUS), * Medroxyprogesterone acetate depot, * Tubal sterilization, * Sexual intercourse with a vasectomized male partner only; vasectomy must be confirmed by two negative semen ana-lyses, * Ovulation inhibitory progesterone-only pills (i.e., desoge-strel). (Combined oral contraceptive pills are not recommended. If a subject was using combined oral contraception, she must switch to one of the methods above. The increased risk of venous thromboembolism (VTE) continues for 4 to 6 weeks after stopping combined oral contraception.

Exclusion criteria

* left atrial diameter \> 50 mm as assessed by transthoracic echocardiogra-phy, * electrical cardioversion performed later than seven days after onset of AF, * advanced structural heart disease including * moderate-to-severe valvular stenosis or regurgitation, * previous myocardial infarction with impaired left ventricular systolic function, * congenital heart disease, * left ventricular ejection fraction \< 45% during sinus rhythm, * coronary artery bypass graft surgery within the last 3 months. * chronic obstructive pulmonary disease treated with beta-sympathomimetic drugs, * severe respiratory insufficiency, * known bleeding diathesis or intolerance of heparin or oral anticoagulation, * previous AF ablation, * left atrial thrombus, * severe comorbidity, * hyperthyreosis, * any condition, including the presence of laboratory abnormalities, which would place the subject at unacceptable risk if he / she were to participate in the study or confound the ability to interpret data from the study, * any serious medical condition, laboratory abnormality, or psychiatric illness, that would prevent the subject from signing the Informed Consent Form, * pregnant or lactating females, * other severe conditions, which makes the patient unsuitable to participate in the study, as judged by the Investigator, * the patient is active in another clinical trial.

Design outcomes

Primary

MeasureTime frameDescription
Safety of Ablation Using the AFreeze Cryoablation System Consisting of the CoolLoop® Ablation Catheter, Its Steerable Sheath and the Cryoconsole Cryo-Caddy Assessed by Recording All Serious Adverse Events (SAEs)3 monthsThe primary study objective is assessed by recording all Serious Adverse Events (SAEs). Primary endpoint: measurement of the following parameters: * deformation of the catheter loop resulting in entrapment of the catheter in the heart and difficulties removing the catheter during visit 3 (treatment). * phrenic nerve palsy during visit 3 (treatment). * onset and time course between visit 3 (treatment) and visit 5 (discharge) of death, pericardial tamponade, valve damage requiring surgery and hemorrhage requiring transfusion. * onset and time course between visit 3 (treatment) and visit 9 (final visit) of atrioesophageal fistula, sepsis, abscesses, endocarditis, stroke, transient ischemic attack, and PV stenosis requiring intervention.
Tolerability of Ablation Using the AFreeze Cryoablation SystemTreatment duration, up to 6 hoursThe primary study objective is assessed by recording all Adverse Events (AEs). Primary endpoint: measurement of the following parameters: \- deformation of the catheter loop resulting in entrapment of the catheter in the heart and difficulties removing the catheter during visit 3 (treatment).

Secondary

MeasureTime frameDescription
Clinical Efficacy of Catheter AblationFirst patient - first visit: September 11, 2012; Last patient - last visit: March 26, 2013During follow-up recurrence of AF was reported by the patient and documented by ECG in 8 patients (80%). 2 patients remained free of recurrences of AF.
Procedure TimeAverage procedure time: 251 min. 06 sec. (range 126 - 320 min.)Procedure time was defined from introduction of the CoolLoop® catheter into the left atrium until removal of the CoolLoop® catheter from the left atrium after termination of the last cryo-application.
Feasibility of Catheter Ablation Measured by Number of Participants Treated With the AFreeze Cryoablation System.Treatment DurationFeasibility, defined by the ability to position the catheter in the proximal part in each of the pulmonary veins (PVs) using an over-the-wire technique, forming the cryo-applicator of the catheter to a loop by operating the catheter handle, positioning the loop at the wall of the PV antrum and delivering cryothermia.
Cumulative Cryoablation TimeFirst patient - first visit: September 11, 2012; Last patient - last visit: March 26, 2013Cryoablation time was defined as the cumulative duration of all cryoapplications in each single study patient. Average cryoablation time was 114 min. 33 sec. (range 80 - 130 min.).
AEFirst patient - first visit: September 11, 2012; Last patient - last visit: March 26, 201318 AEs were recorded in 7 out of 10 patients. 3 AEs were definitely related to treatment and one AE was definitely device related. All patients recovered completely.
Fluoroscopy TimeTreatment DurationFluoroscopy time was defined from introduction of the CoolLoop® catheter into the left atrium until removal of the CoolLoop® catheter from the left atrium after termination of the last cryo - application. Accumulated time using fluoroscopy during procedure time. Average fluoroscopy time was 44 min. 01 sec. (range 17 min. 03 sec. - 59 min.).
Acute Efficacy of Catheter AblationTreatment DurationAbsolute percentage of PVs isolated with the CoolLoop® catheter.

Countries

Austria

Participant flow

Recruitment details

Recruitment period: September 3,2012 - December 14, 2012 Locations: Medical University Innsbruck, General Hospital Linz

Participants by arm

ArmCount
Cryoablation
Cryoablation Cryoablation : Cryoablation of Atrial Fibrillation Using a Novel Cryoablation System
10
Total10

Baseline characteristics

CharacteristicCryoablation
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
2 Participants
Age, Categorical
Between 18 and 65 years
8 Participants
Age, Continuous57.6 years
STANDARD_DEVIATION 7.6
Region of Enrollment
Austria
10 participants
Sex: Female, Male
Female
4 Participants
Sex: Female, Male
Male
6 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
7 / 10
serious
Total, serious adverse events
6 / 10

Outcome results

Primary

Safety of Ablation Using the AFreeze Cryoablation System Consisting of the CoolLoop® Ablation Catheter, Its Steerable Sheath and the Cryoconsole Cryo-Caddy Assessed by Recording All Serious Adverse Events (SAEs)

The primary study objective is assessed by recording all Serious Adverse Events (SAEs). Primary endpoint: measurement of the following parameters: * deformation of the catheter loop resulting in entrapment of the catheter in the heart and difficulties removing the catheter during visit 3 (treatment). * phrenic nerve palsy during visit 3 (treatment). * onset and time course between visit 3 (treatment) and visit 5 (discharge) of death, pericardial tamponade, valve damage requiring surgery and hemorrhage requiring transfusion. * onset and time course between visit 3 (treatment) and visit 9 (final visit) of atrioesophageal fistula, sepsis, abscesses, endocarditis, stroke, transient ischemic attack, and PV stenosis requiring intervention.

Time frame: 3 months

ArmMeasureValue (NUMBER)
CryoablationSafety of Ablation Using the AFreeze Cryoablation System Consisting of the CoolLoop® Ablation Catheter, Its Steerable Sheath and the Cryoconsole Cryo-Caddy Assessed by Recording All Serious Adverse Events (SAEs)6 participants
Primary

Tolerability of Ablation Using the AFreeze Cryoablation System

The primary study objective is assessed by recording all Adverse Events (AEs). Primary endpoint: measurement of the following parameters: \- deformation of the catheter loop resulting in entrapment of the catheter in the heart and difficulties removing the catheter during visit 3 (treatment).

Time frame: Treatment duration, up to 6 hours

ArmMeasureValue (NUMBER)
CryoablationTolerability of Ablation Using the AFreeze Cryoablation System1 AE (device related)
Secondary

Acute Efficacy of Catheter Ablation

Absolute percentage of PVs isolated with the CoolLoop® catheter.

Time frame: Treatment Duration

ArmMeasureValue (NUMBER)
CryoablationAcute Efficacy of Catheter Ablation57.5 percentage of isolated PVs
Secondary

AE

18 AEs were recorded in 7 out of 10 patients. 3 AEs were definitely related to treatment and one AE was definitely device related. All patients recovered completely.

Time frame: First patient - first visit: September 11, 2012; Last patient - last visit: March 26, 2013

ArmMeasureValue (NUMBER)
CryoablationAE7 participants
Secondary

Clinical Efficacy of Catheter Ablation

During follow-up recurrence of AF was reported by the patient and documented by ECG in 8 patients (80%). 2 patients remained free of recurrences of AF.

Time frame: First patient - first visit: September 11, 2012; Last patient - last visit: March 26, 2013

ArmMeasureValue (NUMBER)
CryoablationClinical Efficacy of Catheter Ablation2 participants
Secondary

Cumulative Cryoablation Time

Cryoablation time was defined as the cumulative duration of all cryoapplications in each single study patient. Average cryoablation time was 114 min. 33 sec. (range 80 - 130 min.).

Time frame: First patient - first visit: September 11, 2012; Last patient - last visit: March 26, 2013

ArmMeasureValue (MEAN)Dispersion
CryoablationCumulative Cryoablation Time113.9 minutesStandard Deviation 18.1
Secondary

Feasibility of Catheter Ablation Measured by Number of Participants Treated With the AFreeze Cryoablation System.

Feasibility, defined by the ability to position the catheter in the proximal part in each of the pulmonary veins (PVs) using an over-the-wire technique, forming the cryo-applicator of the catheter to a loop by operating the catheter handle, positioning the loop at the wall of the PV antrum and delivering cryothermia.

Time frame: Treatment Duration

ArmMeasureValue (NUMBER)
CryoablationFeasibility of Catheter Ablation Measured by Number of Participants Treated With the AFreeze Cryoablation System.10 participants
Secondary

Fluoroscopy Time

Fluoroscopy time was defined from introduction of the CoolLoop® catheter into the left atrium until removal of the CoolLoop® catheter from the left atrium after termination of the last cryo - application. Accumulated time using fluoroscopy during procedure time. Average fluoroscopy time was 44 min. 01 sec. (range 17 min. 03 sec. - 59 min.).

Time frame: Treatment Duration

ArmMeasureValue (MEAN)Dispersion
CryoablationFluoroscopy Time44 minutesStandard Deviation 13.2
Secondary

Procedure Time

Procedure time was defined from introduction of the CoolLoop® catheter into the left atrium until removal of the CoolLoop® catheter from the left atrium after termination of the last cryo-application.

Time frame: Average procedure time: 251 min. 06 sec. (range 126 - 320 min.)

ArmMeasureValue (MEAN)Dispersion
CryoablationProcedure Time251.1 minutesStandard Deviation 59.7