Skip to content

Three-day, In-clinic Evaluation of the BD 2nd Generation Continuous Glucose Sensor Device in Type 1 Diabetics

Three-day, In-clinic, Clamp Evaluation of the BD 2nd Generation Continuous Glucose Sensor in Subjects With Type 1 Diabetes

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01645696
Enrollment
30
Registered
2012-07-20
Start date
2012-06-30
Completion date
2012-09-30
Last updated
2021-12-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Keywords

Diabetes, Type 1 Diabetes, Continuous glucose monitoring, Blood glucose, Blood glucose sensor, Glucose Binding Protein (GBP)

Brief summary

The purpose of this study is to investigate the accuracy and performance of a new subcutaneous continuous glucose monitor (BD-CGM, Becton Dickinson) in hyperglycemic (high blood sugar) and hypoglycemic (low blood sugar) clamp conditions and during meal excursions over the course of 72 hours as compared to a commercially available monitor.

Detailed description

This is a single site, non-randomized study. The study consists of a Screening Visit (Visit 1) during which the subject will be consented and the inclusion exclusion criteria confirmed. An interventional visit (Visit 2)which consists of a 72 hour in-clinic stay and a Follow-up Visit (Visit 3). Subjects eligible for the study will be admitted to the clinic for the Study Visit 2 in the afternoon on the day before the first Clamp is performed. An IV line for blood sampling will be established. One blood sample will be obtained for glucose determination and a second blood sample will be collected for immunoassay development before any sensors are inserted. Two BD-Glucose Binding Protein-Continuous Glucose Monitor Sensors(BD-GBP-CGM), with and without outer layer, and one commercial CGM sensor will be inserted in the subcutaneous tissue in the abdomen shortly thereafter. Blood sampling intervals will be adjusted over the 3 study days as determined by the study event (i.e. clamp period, meal excursion, nighttime). During the hyper- /hypo-glycemic clamps periods on Day 1 and Day 3 blood samples will be taken more frequently, every 5-10 minutes. During the breakfast meal on Day 2 sampling will occur at 10-15 minute intervals for 4 hours to capture the meal excursion. Sampling will be less frequent during the evening meal and at night during sleep hours.

Interventions

DEVICEBD CGM with Outer Layer

continuous (every 2 minutes) subcutaneous glucose monitoring for 72 hours.

DEVICEBD CGM without Outer Layer

continuous (every 2 minutes) subcutaneous glucose monitoring over 72 hours.

DEVICEMedtronic iPro2 Professional CGM

continuous subcutaneous glucose monitoring for 72 hours

Sponsors

Becton, Dickinson and Company
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
FACTORIAL
Primary purpose
DEVICE_FEASIBILITY
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
No

Inclusion criteria

1. Clinical diagnosis of type 1 diabetes mellitus for ≥1 year. For an individual to be enrolled at least one criterion from each list must be met. 2. Criteria for documented hyperglycemia (at least 1 must be met): 1. Fasting glucose ≥ 7 mmol/L \[126 mg/dL\] - confirmed 2. Two-hour OGTT (oral glucose tolerance test) glucose ≥ 11.1 mmol/L \[200 mg/dL\] - confirmed 3. HbA1c ≥6.5% documented - confirmed 4. Random glucose ≥ 11.1 mmol/L \[200 mg/dL\] with symptoms 5. No data at diagnosis is available but the participant has a convincing history of hyperglycemia consistent with diabetes 3. Criteria for requiring insulin at diagnosis (1 must be met): 1. Participant required insulin at diagnosis and continually thereafter 2. Participant did not start insulin at diagnosis but upon investigator review likely needed insulin (significant hyperglycemia that did not respond to oral agents) and did require insulin eventually and used continually 3. Participant did not start insulin at diagnosis but continued to be hyperglycemic, had positive islet cell antibodies - consistent with latent autoimmune diabetes in adults (LADA) and did require insulin eventually and used continually 4. Signed informed consent 5. Age ≥18 and ≤65 years old 6. Body mass index between 19 and 35 kg/m2, inclusive 7. HbA1c ≤ 10.0%

Exclusion criteria

1. Uncontrolled arterial hypertension (diastolic blood pressure \> 90 mm Hg and/or systolic blood pressure \> 160 mm Hg) 2. Impaired hepatic function measured as alanine aminotransferase or aspartate aminotransferase ≥ three times the upper reference limit 3. Impaired renal function measured as creatinine \> 1.2 times above the upper limit of normal 4. Diabetic ketoacidosis in the past 6 months 5. Severe hypoglycemia resulting in a seizure or loss of consciousness in the 6 months prior to enrollment 6. Conditions which may increase the risk of hypoglycemia or conditions of known microvascular (diabetic) complications will be assessed on an individual basis with exclusion based on the discretion of the principal investigator. 7. Current use of medications containing \> 4000 mg acetaminophen per day. 8. Current use of MAO (monoamine oxidase) inhibitors. 9. Known allergy to eggs 10. Pregnancy, breast-feeding or intention of becoming pregnant 11. Current or recent alcohol or drug abuse by subject history. 12. Blood donation of more than 473 ml within the last 56 days 13. Mental incapacity, unwillingness or language barriers precluding adequate understanding or cooperation 14. Any skin condition that prevents sensor placement on the abdomen (e.g., bad sunburn, pre-existing dermatitis, intertrigo, psoriasis, extensive scarring, cellulitis) 15. Known allergy to medical adhesives, e.g. Tegaderm 16. Hematocrit \< 38% (males) and \< 36% (females) 17. Potassium \< 3.4 mmol/L 18. Active enrollment in another clinical trial

Design outcomes

Primary

MeasureTime frameDescription
Blood Glucose72 hoursBlood glucose will be measured by the BD-Continuous Glucose Monitor, with and without outer layer, the commercially available Medtronic iPro2 and the YSI (Yellow Springs Instrument) Glucose analyzer (control) for 72 hours. Blood glucose will be used to determine performance of the device to include warm up behavior, lag time and accuracy of the blood glucose monitor over 72 hours.
Number of participants with adverse eventsup to 89 days or until the subject is dischargedAt each study contact, subjects will be questioned about any adverse events that may have occurred and are potentially related to the device.

Secondary

MeasureTime frameDescription
antibodies against the glucose binding protein36 daysA blood sample will be taken at the beginning of Visit 2 and at Visit 3 to test for antibody production following exposure to the sensor's glucose binding protein.
Skin Effects-Draize Scoring for Skin IrritationUp to 36 daysLocal reaction at insertion sites will be scored for redness and swelling at the following timepoints: Visit 2-pre-insertion of devices, immediately post insertion of the devices, each morning of Day 1, 2 and 3, immediately after removal of the devices and at Visit 3.
Skin thickness using ultrasoundUpon removal of the devicesSkin thickness will be measured at the sensor sites and a control site (on the abdomen) immediately after removal of the device.
Insulin levels72 hoursBlood samples will be taken at pre-determined times following insulin dosings to test for insulin levels:

Countries

Canada

Outcome results