Aortic Valve Stenosis
Conditions
Keywords
Valvular Heart Disease, Severe Aortic Stenosis, Aortic Valve Replacement
Brief summary
The primary objective of the present trial is to demonstrate the safety and effectiveness of the MDT-2111 in the treatment of symptomatic severe aortic stenosis in subjects with small aortic annuli and deemed difficult for surgical operation.
Detailed description
The purpose of this trial is to evaluate the effectiveness and safety of the MDT-2111 TAV system in subjects with small annuli and symptomatic severe AS deemed difficult for surgical intervention. The primary endpoint is a composite of functional effectiveness as measured by improvement of at least 1 New York Heart Association (NYHA) class from baseline to 6 months and anatomical effectiveness as measured by Effective Orifice Area (EOA) ≥1.0 cm² at 6 months.
Interventions
Device: 23 mm MDT-2111 System for Transcatheter Aortic Valve Implantation (TAVI) Transcatheter Aortic Valve Implantation (TAVI) using the 23 mm MDT-2111 system.
Sponsors
Study design
Eligibility
Inclusion criteria
1. Subject must have co-morbidities such that one cardiologist and one cardiac surgeon agree that medical factors preclude operation, based on a conclusion that the probability of death or serious morbidity exceeds the probability of meaningful improvement. 2. Subject has senile degenerative aortic valve stenosis with: mean gradient \> 40 mmHg or jet velocity greater than 4.0 m/s by either resting or dobutamine stress echocardiogram, or simultaneous pressure recordings at cardiac catheterization (either resting or dobutamine stress), AND an initial aortic valve area of ≤ 0.8 cm² (or aortic valve area index ≤ 0.5 cm²/m²) by resting echocardiogram. 3. Subject is symptomatic from his/her aortic valve stenosis, as demonstrated by NYHA Functional Class Class III or greater. If screening committee agrees the eligibility of a patient with class II , based on medical factors, he/she can be enrolled. 4. Patient has been informed of the nature of the trial and has signed an Informed Consent Form. 5. Patient agrees to comply with specified follow-up evaluations and to return to the same investigational site where the procedure was performed.
Exclusion criteria
1. Evidence of an acute myocardial infarction ≤ 30 days prior to the intended treatment. 2. Any percutaneous coronary or peripheral interventional procedure performed within 30 days prior to the procedure. 3. Blood dyscrasias as defined: * Leukopenia (WBC count \< 1,000 cells/mm³) * Thrombocytopenia (platelet count \<50,000 cells/mm³) * History of bleeding diathesis or coagulopathy * Hypercoagulable states 4. Untreated clinically significant coronary artery disease requiring revascularization. 5. Cardiogenic shock manifested by low cardiac output, vasopressor dependence, or mechanical hemodynamic support. 6. Need for emergency surgery for any reason. 7. Severe ventricular dysfunction with left ventricular ejection fraction (LVEF) \< 20% as measured by resting echocardiogram. 8. Recent (within 6 months) cerebrovascular accident (CVA) or transient ischemic attack(TIA). 9. End stage renal disease requiring chronic dialysis. 10. GI bleeding within the past 3 months. 11. A known hypersensitivity or contraindication to any of the following which cannot be adequately pre-medicated: * Aspirin * Ticlopidine * Heparin * Contrast media * Nitinol (titanium and nickel alloy) 12. Ongoing sepsis, including active endocarditis. 13. Subject refuses a blood transfusion. 14. Life expectancy \< 12 months due to associated non-cardiac co-morbid conditions. 15. Other medical, social, or psychological conditions that in the opinion of an Investigator precludes the subject from appropriate consent. 16. Severe dementia (resulting in either inability to provide informed consent for the trial/procedure, prevents independent lifestyle outside of a chronic care facility, or will fundamentally complicate rehabilitation from the procedure or compliance with follow-up visits). 17. Currently participating in an investigational drug or another device trial. Note: Trials requiring extended follow-up for products that were investigational, but have since become commercially available, are not considered investigational trials. 18. Symptomatic carotid or vertebral artery disease. 19. Native aortic annulus size \< 18 mm or \> 20 mm per the screening diagnostic imaging. 20. Pre-existing prosthetic heart valve in any position. 21. Mixed aortic valve disease (aortic stenosis and aortic regurgitation with predominant aortic regurgitation). 22. Mitral regurgitation (moderate to severe) or severe tricuspid regurgitation. 23. Moderate to severe mitral stenosis. 24. Hypertrophic obstructive cardiomyopathy. 25. Echocardiographic evidence of intracardiac mass, thrombus or vegetation. 26. Severe basal septal hypertrophy with an outflow gradient. 27. Ascending aorta diameter \> 34 mm 28. Congenital bicuspid or unicuspid valve verified by echocardiography. 29. For patients with native coronary artery dependent circulation: * Sinus of valsalva width \< 25 mm OR * Height of the left or right coronary sinus of valsalva (to the tubular aorta) \< 15mm. 30. Femoral or iliac artery of the first choice corresponding to any one of the following: * Angle at aortic root (the angle between aortic valve annulus plane and horizontal plane/vertebra) exceeds 70°. * Vessel diameter of femoral or iliac artery is less than 6 mm. * Aorta has severe calcification, excess tortuosity or severe atherosclerosis. * Transarterial access not able to accommodate an 18Fr sheath. 31. Subclavian artery of the second choice corresponding to any one of the following: * Angle at aortic root (the angle between aortic valve annulus plane and horizontal plane/vertebra) exceeds 70° (in the case of left subclavian artery) and 30° (in the case of right subclavian artery). * Vessel diameter of subclavian artery is less than 6 mm. * Transarterial access not able to accommodate an 18Fr sheath. 32. Direct Aortic Artery as third line choice of access. Patients are excluded from Direct Aortic access if: * Access site is less than 6 cm from the aortic valve basal plane * Access site has calcification or porcelain aorta * Access site and delivery trajectory contain RIMA or patent RIMA graft
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Composite Score of Change in New York Heart Association (NYHA) Class and Effective Orifice Area (EOA). | baseline and 6 months | The primary endpoint is a composite of functional effectiveness as measured by improvement of at least 1 NYHA class from baseline to 6 months and anatomical effectiveness as measured by Effective Orifice Area (EOA) ≥1.0 cm² at 6 months. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Major Adverse Cardiovascular and Cerebrovascular Event (MACCE) | 0 day to 30 days | MACCE is defined as a composite of: * all-cause death * myocardial infarction (MI) * all stroke, and * reintervention (defined as any cardiac surgery or percutaneous reintervention catheter procedure that repairs, otherwise alters or adjusts, or replaces a previously implanted valve) |
| Device Success as Defined in the Description. | after procedure or discharge | The following components must be satisfied for device success: 1. successful vascular access, delivery and deployment of device and successful retrieval of delivery system 2. correct position of device in the proper anatomical location 3. EOA≥1.0 cm² AND mean gradient \<20 mmHg or peak velocity \<3 m/s, without moderate or severe AR 4. only one valve implanted. |
| Procedural Success, Defined as Device Success and Absence of In-hospital MACCE | from admission for procedure to discharge | Procedural success is defined as device success and absence of in-hospital MACCE. |
| Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient | 30 days | — |
| Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA) | 30 days | — |
| New York Heart Classification (NYHA) Over Time | 30 days | NEW YORK HEART ASSOCIATION CLASSIFICATION (NYHA) Class I: Subject with cardiac disease but without resulting limitations of physical activity. Class II: Subjects with cardiac disease resulting in slight limitation of physical activity. Class III: Subjects with cardiac disease resulting in marked limitation of physical activity. Class IV: Subjects with cardiac disease resulting in inability to carry on any physical activity without discomfort. |
| Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | 30 days | — |
| Repeat Hospitalization | 0 day to 30 days | — |
| Valve-related Deaths | 0 day to 30 days | — |
| Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline) | Baseline to 30 days | The SF-36 assessment was used to evaluate subject Quality of life (QoL) by assessing change in physical function and general health status. The SF-36 v2 ͭ ͫ Scoring Program was used to convert raw scores ranging from 0 to 100 into norm-based scores, allowing direct comparison to the reference values for the Japanese population. The Z-score of each subdomain is calculated as the numbers of standard deviation away from the Japanese population mean of the corresponding raw score. Norm-based score is then derived as fifty plus 10 times of the z-score. A norm-based score of less than 50 was interpreted as below average when compared to the Japanese population whereas norm-based scores greater than 50 were interpreted as above average. |
| Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF) | 30 days | — |
Countries
Japan
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| MDT-2111 TAVI 23 mm Transcatheter Aortic Valve Implantation (TAVI) using the 23 mm MDT-2111 system. | 20 |
| Total | 20 |
Baseline characteristics
| Characteristic | MDT-2111 TAVI 23 mm |
|---|---|
| Age, Continuous | 81.0 years STANDARD_DEVIATION 6.6 |
| Body Surface Area | 1.4 m² STANDARD_DEVIATION 0.1 |
| Logistic EuroScore | 22.3 percent STANDARD_DEVIATION 10.1 |
| NYHA Class NYHA Class II | 0 participants |
| NYHA Class NYHA Class III | 19 participants |
| NYHA Class NYHA IV | 1 participants |
| Region of Enrollment Japan | 20 participants |
| Sex: Female, Male Female | 19 Participants |
| Sex: Female, Male Male | 1 Participants |
| STS Score | 7.0 percent STANDARD_DEVIATION 3.3 |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk | EG002 affected / at risk |
|---|---|---|---|
| deaths Total, all-cause mortality | — / — | — / — | — / — |
| other Total, other adverse events | 16 / 16 | 4 / 4 | 20 / 20 |
| serious Total, serious adverse events | 13 / 16 | 3 / 4 | 16 / 20 |
Outcome results
Composite Score of Change in New York Heart Association (NYHA) Class and Effective Orifice Area (EOA).
The primary endpoint is a composite of functional effectiveness as measured by improvement of at least 1 NYHA class from baseline to 6 months and anatomical effectiveness as measured by Effective Orifice Area (EOA) ≥1.0 cm² at 6 months.
Time frame: baseline and 6 months
Population: Implanted subjects
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Composite Score of Change in New York Heart Association (NYHA) Class and Effective Orifice Area (EOA). | 75.0 percentage of participants analyzed |
| Iliofemoral Implanted Subjects | Composite Score of Change in New York Heart Association (NYHA) Class and Effective Orifice Area (EOA). | 87.5 percentage of participants analyzed |
Device Success as Defined in the Description.
The following components must be satisfied for device success: 1. successful vascular access, delivery and deployment of device and successful retrieval of delivery system 2. correct position of device in the proper anatomical location 3. EOA≥1.0 cm² AND mean gradient \<20 mmHg or peak velocity \<3 m/s, without moderate or severe AR 4. only one valve implanted.
Time frame: after procedure or discharge
Population: This includes subjects with an index procedure. Index procedure is defined as the first procedure that Medtronic MDT-2111 TAV system delivery catheter is introduced.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Device Success as Defined in the Description. | 65.0 percentage of participants analyzed |
Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA)
Time frame: 12 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA) | 1.2 cm² | Standard Deviation 0.3 |
Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA)
Time frame: 30 days
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA) | 1.2 cm² | Standard Deviation 0.3 |
Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA)
Time frame: 6 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA) | 1.2 cm² | Standard Deviation 0.3 |
Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA)
Time frame: 24 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Effective Orifice Area (EOA) | 1.3 cm² | Standard Deviation 0.3 |
Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF)
Time frame: 24 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF) | 67.9 percent | Standard Deviation 4.9 |
Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF)
Time frame: 12 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF) | 67.5 percent | Standard Deviation 7.9 |
Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF)
Time frame: 6 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF) | 66.8 percent | Standard Deviation 5.4 |
Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF)
Time frame: 30 days
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Left Ventricular Ejection Fraction (LVEF) | 64.5 percent | Standard Deviation 6.5 |
Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient
Time frame: 6 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient | 17.1 mmHg | Standard Deviation 7.7 |
Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient
Time frame: 30 days
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient | 16.8 mmHg | Standard Deviation 4.8 |
Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient
Time frame: 12 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient | 15.2 mmHg | Standard Deviation 5.6 |
Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient
Time frame: 24 months
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Mean Gradient | 13.7 mmHg | Standard Deviation 5.6 |
Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR)
Time frame: 30 days
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | None | 10.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Trace | 0.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Mild | 75.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Moderate | 15.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Severe | 0.0 percentage of participants analyzed |
Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR)
Time frame: 24 months
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | None | 11.1 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Trace | 38.9 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Mild | 44.4 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Moderate | 5.6 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Severe | 0.0 percentage of participants analyzed |
Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR)
Time frame: 12 months
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | None | 5.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Trace | 15.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Mild | 70.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Moderate | 10.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Severe | 0.0 percentage of participants analyzed |
Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR)
Time frame: 6 months
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | None | 0.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Trace | 30.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Mild | 55.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Moderate | 15.0 percentage of participants analyzed |
| All Implanted Subjects | Echocardiographic Assessment of Prosthetic Valve Performance - Total Aortic Regurgitation (Transvalvular & Paravalvular) (Total AR) | Severe | 0.0 percentage of participants analyzed |
Major Adverse Cardiovascular and Cerebrovascular Event (MACCE)
MACCE is defined as a composite of: * all-cause death * myocardial infarction (MI) * all stroke, and * reintervention (defined as any cardiac surgery or percutaneous reintervention catheter procedure that repairs, otherwise alters or adjusts, or replaces a previously implanted valve)
Time frame: 0 day to 30 days
Population: The Kaplan-Meier Method was used to calculate the number.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Major Adverse Cardiovascular and Cerebrovascular Event (MACCE) | 100 prob of freedom from event @ 30 days |
Major Adverse Cardiovascular and Cerebrovascular Event (MACCE)
MACCE is defined as a composite of: * all-cause death * myocardial infarction (MI) * all stroke, and * reintervention (defined as any cardiac surgery or percutaneous reintervention catheter procedure that repairs, otherwise alters or adjusts, or replaces a previously implanted valve)
Time frame: 0 day to 6 months
Population: The Kaplan-Meier Method was used to calculate the number.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Major Adverse Cardiovascular and Cerebrovascular Event (MACCE) | 95.0 prob of freedom from event @ 183 days |
Major Adverse Cardiovascular and Cerebrovascular Event (MACCE)
MACCE is defined as a composite of: * all-cause death * myocardial infarction (MI) * all stroke, and * reintervention (defined as any cardiac surgery or percutaneous reintervention catheter procedure that repairs, otherwise alters or adjusts, or replaces a previously implanted valve)
Time frame: 0 day to 12 months
Population: The Kaplan-Meier Method was used to calculate the number.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Major Adverse Cardiovascular and Cerebrovascular Event (MACCE) | 95.0 prob of freedom from event @ 365 days |
Major Adverse Cardiovascular and Cerebrovascular Event (MACCE)
MACCE is defined as a composite of: * all-cause death * myocardial infarction (MI) * all stroke, and * reintervention (defined as any cardiac surgery or percutaneous reintervention catheter procedure that repairs, otherwise alters or adjusts, or replaces a previously implanted valve)
Time frame: 0 day to 24 months
Population: The Kaplan-Meier Method was used to calculate the number.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Major Adverse Cardiovascular and Cerebrovascular Event (MACCE) | 90.0 prob of freedom from event @ 730 days |
New York Heart Classification (NYHA) Over Time
NEW YORK HEART ASSOCIATION CLASSIFICATION (NYHA) Class I: Subject with cardiac disease but without resulting limitations of physical activity. Class II: Subjects with cardiac disease resulting in slight limitation of physical activity. Class III: Subjects with cardiac disease resulting in marked limitation of physical activity. Class IV: Subjects with cardiac disease resulting in inability to carry on any physical activity without discomfort.
Time frame: 24 months
Population: All Implanted Subjects
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA I | 47.4 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA II | 47.4 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA III | 0.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA IV | 0.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | Death | 5.3 percentage of participants analyzed |
New York Heart Classification (NYHA) Over Time
NEW YORK HEART ASSOCIATION CLASSIFICATION (NYHA) Class I: Subject with cardiac disease but without resulting limitations of physical activity. Class II: Subjects with cardiac disease resulting in slight limitation of physical activity. Class III: Subjects with cardiac disease resulting in marked limitation of physical activity. Class IV: Subjects with cardiac disease resulting in inability to carry on any physical activity without discomfort.
Time frame: 12 months
Population: All Implanted Subjects
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA I | 40.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA II | 60.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA III | 0.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA IV | 0.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | Death | 0.0 percentage of participants analyzed |
New York Heart Classification (NYHA) Over Time
NEW YORK HEART ASSOCIATION CLASSIFICATION (NYHA) Class I: Subject with cardiac disease but without resulting limitations of physical activity. Class II: Subjects with cardiac disease resulting in slight limitation of physical activity. Class III: Subjects with cardiac disease resulting in marked limitation of physical activity. Class IV: Subjects with cardiac disease resulting in inability to carry on any physical activity without discomfort.
Time frame: 6 months
Population: All Implanted Subjects
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA I | 55.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA II | 45.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA III | 0.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA IV | 0.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | Death | 0.0 percentage of participants analyzed |
New York Heart Classification (NYHA) Over Time
NEW YORK HEART ASSOCIATION CLASSIFICATION (NYHA) Class I: Subject with cardiac disease but without resulting limitations of physical activity. Class II: Subjects with cardiac disease resulting in slight limitation of physical activity. Class III: Subjects with cardiac disease resulting in marked limitation of physical activity. Class IV: Subjects with cardiac disease resulting in inability to carry on any physical activity without discomfort.
Time frame: 30 days
Population: All Implanted Subjects
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA I | 45.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA II | 50.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA III | 5.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | NYHA IV | 0.0 percentage of participants analyzed |
| All Implanted Subjects | New York Heart Classification (NYHA) Over Time | Death | 0.0 percentage of participants analyzed |
Procedural Success, Defined as Device Success and Absence of In-hospital MACCE
Procedural success is defined as device success and absence of in-hospital MACCE.
Time frame: from admission for procedure to discharge
Population: This includes subjects with an index procedure. Index procedure is defined as the first procedure that Medtronic MDT-2111 TAV system delivery catheter is introduced.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Procedural Success, Defined as Device Success and Absence of In-hospital MACCE | 65.0 percent |
Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline)
The SF-36 assessment was used to evaluate subject Quality of life (QoL) by assessing change in physical function and general health status. The SF-36 v2 ͭ ͫ Scoring Program was used to convert raw scores ranging from 0 to 100 into norm-based scores, allowing direct comparison to the reference values for the Japanese population. The Z-score of each subdomain is calculated as the numbers of standard deviation away from the Japanese population mean of the corresponding raw score. Norm-based score is then derived as fifty plus 10 times of the z-score. A norm-based score of less than 50 was interpreted as below average when compared to the Japanese population whereas norm-based scores greater than 50 were interpreted as above average.
Time frame: Baseline to 24 months
| Arm | Measure | Value (MEDIAN) |
|---|---|---|
| All Implanted Subjects | Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline) | 5.3 points |
Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline)
The SF-36 assessment was used to evaluate subject Quality of life (QoL) by assessing change in physical function and general health status. The SF-36 v2 ͭ ͫ Scoring Program was used to convert raw scores ranging from 0 to 100 into norm-based scores, allowing direct comparison to the reference values for the Japanese population. The Z-score of each subdomain is calculated as the numbers of standard deviation away from the Japanese population mean of the corresponding raw score. Norm-based score is then derived as fifty plus 10 times of the z-score. A norm-based score of less than 50 was interpreted as below average when compared to the Japanese population whereas norm-based scores greater than 50 were interpreted as above average.
Time frame: Baseline to 30 days
| Arm | Measure | Value (MEDIAN) |
|---|---|---|
| All Implanted Subjects | Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline) | 5.1 points |
Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline)
The SF-36 assessment was used to evaluate subject Quality of life (QoL) by assessing change in physical function and general health status. The SF-36 v2 ͭ ͫ Scoring Program was used to convert raw scores ranging from 0 to 100 into norm-based scores, allowing direct comparison to the reference values for the Japanese population. The Z-score of each subdomain is calculated as the numbers of standard deviation away from the Japanese population mean of the corresponding raw score. Norm-based score is then derived as fifty plus 10 times of the z-score. A norm-based score of less than 50 was interpreted as below average when compared to the Japanese population whereas norm-based scores greater than 50 were interpreted as above average.
Time frame: Baseline to 6 months
| Arm | Measure | Value (MEDIAN) |
|---|---|---|
| All Implanted Subjects | Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline) | 1.7 points |
Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline)
The SF-36 assessment was used to evaluate subject Quality of life (QoL) by assessing change in physical function and general health status. The SF-36 v2 ͭ ͫ Scoring Program was used to convert raw scores ranging from 0 to 100 into norm-based scores, allowing direct comparison to the reference values for the Japanese population. The Z-score of each subdomain is calculated as the numbers of standard deviation away from the Japanese population mean of the corresponding raw score. Norm-based score is then derived as fifty plus 10 times of the z-score. A norm-based score of less than 50 was interpreted as below average when compared to the Japanese population whereas norm-based scores greater than 50 were interpreted as above average.
Time frame: Baseline to 12 months
| Arm | Measure | Value (MEDIAN) |
|---|---|---|
| All Implanted Subjects | Quality of Life Assessment Using SF-36 Questionnaire - Physical Component Summary (Paired Change From Baseline) | 4.6 points |
Repeat Hospitalization
Time frame: 0 day to 24 months
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Repeat Hospitalization | 30.0 prob of freedom from event @ 730 days |
Repeat Hospitalization
Time frame: 0 day to 30 days
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Repeat Hospitalization | 80.0 prob of freedom from event @ 30 days |
Repeat Hospitalization
Time frame: 0 day to 6 months
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Repeat Hospitalization | 55.0 prob of freedom from event @ 183 days |
Repeat Hospitalization
Time frame: 0 day to 12 months
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Repeat Hospitalization | 35.0 prob of freedom from event @ 365 days |
Valve-related Deaths
Time frame: 0 day to 30 days
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Valve-related Deaths | 100.0 prob of freedom from event @ 30 days |
Valve-related Deaths
Time frame: 0 day to 6 months
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Valve-related Deaths | 100.0 prob of freedom from event @ 183 days |
Valve-related Deaths
Time frame: 0 day to 12 months
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Valve-related Deaths | 100.0 prob of freedom from event @ 365 days |
Valve-related Deaths
Time frame: 0 day to 24 months
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Implanted Subjects | Valve-related Deaths | 100 prob of freedom from event @ 730 days |