Abdominal Surgery
Conditions
Keywords
impedance cardiography, Flotrac/Vigileo, cardiac output
Brief summary
The investigators validated ICG's(impedance cardiography) usefulness during the abdominal surgery, by comparing it with FloTrac/Vigileo™ (Edwards Lifesciences,Irvine, CA, USA)
Detailed description
Intrathoracic structures have relatively constant resistance except the influence of changing diameter of aorta. Great vessels in thorax has the lowest resistance, and their diameter changes by the beats of the heart. As their diameter increases, their resistance decreases. ICG measures the difference between the resistances caused by heartbeats, and calculates cardiac output. But during the surgery, there are many interfering electrical currents made by electro cautery, patient positioning, and internal organ handling. Therefore, it has been thought that ICG is not a valid tool for measuring cardiac output during the surgery. The investigators validated ICG's usefulness during the surgery, by comparing it with Floctac/Vigileo.
Interventions
Sponsors
Study design
Eligibility
Inclusion criteria
* Patients undergoing open gastrectomy with arterial line * ASA 1 & 2
Exclusion criteria
* severe valve disease * severe lung disease
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Cardiac output | — | Cardiac output will be measured using an ICG and compared with cardiac output measured using Flotrac/vigileo |
Countries
South Korea