Performance Evaluation of an Investigational Blood Glucose Monitoring System

Performance of the CONTOUR® PLUS Blood Glucose Monitoring System With CONTOUR® PLUS Strip

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01598610
Enrollment
220
Registered
2012-05-15
Start date
2012-05-31
Completion date
2012-05-31
Last updated
2016-02-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study is to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.

Interventions

DEVICECONTOUR® PLUS Investigational BG Monitoring System

Untrained subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the CONTOUR® PLUS Investigational BG Monitoring System. Study staff test subject venous blood. All Blood Glucose (BG) results are compared to a reference laboratory glucose method.

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Males and females, 18 years of age and older * Type 1 or type 2 diabetes * Able to speak, read, and understand English * Willing to complete all study procedures

Exclusion criteria

* Pregnancy * Hemophilia or any other bleeding disorder * Previously participated in a study using the CONTOUR PLUS system * Working for a medical laboratory, hospital or other clinical setting that involves training on and clinical use of blood glucose monitors. * Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company. * A condition, which in the opinion of the investigator, would put the person or study conduct at risk (reason for exclusion will be clearly documented by investigator or designee).

Design outcomes

Primary

MeasureTime frameDescription
Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma) of the reference method results.

Secondary

MeasureTime frameDescription
Percent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-test Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma) of the reference method results.
Percent of Venous Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method1 hourStudy staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. Venous plasma BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI venous plasma) or +/- 20% (\>=75mg/dL YSI venous plasma) of the reference method results.
Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of BGMS results within +/- 5to15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 5to15% (\>=100mg/dL YSI capillary plasma) of the reference method results.
Percent of Subject Fingerstick BG Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method When Tested by Study Staff1 hourStudy staff test subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma) of the reference method results.
Number of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire Statements1 hourSubjects respond to statements read by study staff to provide feedback on the labeling materials and system ease of use. Subjects may respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree'.

Countries

United States

Participant flow

Participants by arm

ArmCount
Intended Users of the Monitoring System
Untrained subjects with diabetes use CONTOUR® PLUS Investigational BG Monitoring System.
220
Total220

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyAdverse Event1

Baseline characteristics

CharacteristicIntended Users of the Monitoring System
Age, Continuous60 years
Region of Enrollment
United States
220 participants
Sex: Female, Male
Female
118 Participants
Sex: Female, Male
Male
102 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
2 / 220
serious
Total, serious adverse events
0 / 220

Outcome results

Primary

Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma) of the reference method results.

Time frame: 1 hour

Population: 216 (220-4) blood test results were analyzed. Blood data for 2 subjects were not evaluable because time, defined in protocol, was exceeded between meter test and blood sample preparation for reference method. No fingerstick results were obtained for 2 subjects.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method100 percentage of BG Test Results
Secondary

Number of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire Statements

Subjects respond to statements read by study staff to provide feedback on the labeling materials and system ease of use. Subjects may respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree'.

Time frame: 1 hour

Population: Per protocol

ArmMeasureGroupValue (NUMBER)
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsEasy to handle the individual test strips.218 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsInstructions (User Guide) easy to understand.218 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsInstructions clearly explain how to run a test.220 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsInstructions clearly explain meter error messages.218 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsI found it easy to do a blood test with this meter215 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsMeter display(test result) is easy to see and read220 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsIt was easy to understand my test results.218 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsI like the overall meter design.218 participants
Intended Users of the Monitoring SystemNumber of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire StatementsI find the meter easy to handle.218 participants
Secondary

Percent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-test Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma) of the reference method results.

Time frame: 1 hour

Population: 213 (220-7) blood test results were analyzed. Four(4) subjects had low blood sugar and AST results were not evaluable per protocol. One(1) subject with low blood sugar (AE) did not attempt AST testing per protocol. No AST palm results were obtained for 2 subjects.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method98.1 percentage of BG Test Results
Secondary

Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of BGMS results within +/- 5to15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 5to15% (\>=100mg/dL YSI capillary plasma) of the reference method results.

Time frame: 1 hour

Population: 108 (110-2) blood test results from one test strip lot were analyzed. Blood data for 1 subject was not evaluable because time, defined in protocol, was exceeded between meter test and blood sample preparation for reference method. No fingerstick result was obtained for 1 subject.

ArmMeasureGroupValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Methodw/in +/-10mg/dL(<100mg/dL) or +/- 10% (>=100mg/dL)86.1 percentage of BG Test Results
Intended Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Methodw/in +/-5mg/dL(<100mg/dL) or +/- 5% (>=100mg/dL)53.7 percentage of BG Test Results
Intended Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Method% within +/-15mg/dL(<100mg/dL) or 15% (>=100mg/dL)98.2 percentage of BG Test Results
Intended Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Methodw/in +/-12.5mg/dL(<100mg/dL) or 12.5% (>=100mg/dL)93.5 percentage of BG Test Results
Secondary

Percent of Subject Fingerstick BG Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method When Tested by Study Staff

Study staff test subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma) of the reference method results.

Time frame: 1 hour

Population: 217 (220-3) blood test results were analyzed. Blood data for 2 subjects were not evaluable because time, defined in protocol, was exceeded between meter test and blood sample preparation for reference method. No fingerstick results were obtained for 1 subject.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Subject Fingerstick BG Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method When Tested by Study Staff99.5 percentage of BG Test Results
Secondary

Percent of Venous Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. Venous plasma BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI venous plasma) or +/- 20% (\>=75mg/dL YSI venous plasma) of the reference method results.

Time frame: 1 hour

Population: 213 (220-7) blood test results were analyzed. Venipuncture was unsuccessful for 6 subjects. Blood data for 1 subject was not evaluable because time, defined in protocol, was exceeded between meter test and blood sample preparation for reference method.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Venous Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method99.5 percentage of BG Test Results