Safety of Body Composition Analysis in Heart Failure Patients With Implantable Cardioverter Defibrillators (ICDs)

Safety of Body Composition Analysis Using Bioimpedance in Heart Failure Patients With Implantable Cardioverter Defibrillators (ICDs)

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01568606
Enrollment
20
Registered
2012-04-02
Start date
2011-01-31
Completion date
2011-03-31
Last updated
2020-03-10

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Heart Failure, Obesity

Keywords

Heart Failure, Obesity, Implantable Cardioverter Defibrillator

Brief summary

The purpose of this study is to define the safety of using bioimpedance for analysis of body composition in heart failure patients with ICD devices. The Biospace America InBody 520 scale, using direct segmental multifrequency bioimpedance, will be utilized to assess patients' fat mass, lean muscle mass, and edema status. Although the Biospace America InBody 520 scale is routinely used to analyze body composition in various settings including the Ahmanson-UCLA Cardiomyopathy clinic, due to theoretical concerns of safety, bioimpedance has not been routinely used in patients with ICDs. The investigators hope that this study will allow us to routinely analyze body composition in heart failure patients with ICDs, information which can be used to help guide dietary, exercise, and medical prescriptions for the investigators heart failure patients.

Detailed description

Up to one half of heart failure (HF) patients are overweight and obese, as measured by body mass index (BMI). Comprehensive body composition assessment utilizing bioimpedance scales wuch as Biospace America InBody 520 scale, can give accurate measurements of fat mass, lean muscle mass, and edematous water weight. Body composition analysis is routinely used in many clinical settings, including the Ahmanson-UCLA Cardiomyopathy Center. However, bioimpedance scales are generally not used in patients with implantable cardioverter-defibrillators (ICDs), a group that represents a majority of our HF patient population. The safety of InBody 520 technology - direct segmental multifrequency (5, 100, 500 KHz) bioimpedance analysis - in patients with ICDs has not been formally evaluated. There is a theoretic concern that the ICD may misinterpret the applied currents for a cardiac arrhythmia and possibly cause a discharge from the ICD. However, there has been no documentation of any adverse events in patients with ICDs using the InBody 520 or other bioimpedance scales. Furthermore, thoracic bioimpedance analysis (BioZ ICG, 70 Khz) has been documented to be successfully utilized without adverse effects in patients with HF and ICDs. This study will define the safety of using the InBody 520 scale for analysis of body composition in patients with ICD devices. Subjects with HF and and ICD will have their ICD interrogated and continually monitored by an electrophysiologist before, during, and after the 30-50 seconds of bioimpedance analysis, which involves standing on the InBody 520 scale. If there is any suggestion of the ICD interpreting applied frequencies as a cardiac arrhythmia, the patient will be removed from the scale and the electrophysiologist will simultaneously temporarily deactivate their ICD device to prevent discharge. Twenty subjects (from the following four ICD companies: Medtronic, St. Jude, Guidant, and Biotronik) will participate in our study.

Interventions

DEVICEBody Composition Analysis InBody Scale

Patients will have their ICD interrogated using a pacemaker Programmer specifically designed from one of the 4 pacemaker manufacturers: St. Jude, Medtronic, Guidant, and Biotronik. Depending on the type of pacemaker, interrogation can be done either wirelessly or by placing a wand over the pacemaker. The ICD will be turned to a monitor-only mode to disable risk of ICD shock by the supervising electrophysiologist. Patient will be asked to step on the InBody 520 scale to have their body composition analyzed. Patient will continue to have their ICD actively monitored. Analysis will take approximately 30 - 50 seconds. At the end of analysis, ICD device will be returned to default settings by supervising electrophysiologist.

Sponsors

University of California, Los Angeles
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
SCREENING
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* HF patients of any etiology who currently have an implanted ICD

Exclusion criteria

* patients who are non-ambulatory * have a physical disability making them unable to stand on the InBody * those who are above the height and weight maximums for the device \[Height \> 220cm (7'2.6''), Weight \> 250kg (551lb)\] * those who are pregnant

Design outcomes

Primary

MeasureTime frameDescription
Number of Participants With Arrhythmia as Detected by ICD1 day, on day of study ICD interrogated and continually monitored before, during, and after 30-50 seconds of bioimpedance analysisSubjects with HF and ICD had their ICD interrogated and continually monitored by an electrophysiologist before, during, and after 30-50 seconds of bioimpedance analysis, which involves standing on the InBody 520 scale. Outcome measure was assessed during a one day visit. There is no further follow-up.

Countries

United States

Participant flow

Participants by arm

ArmCount
Body Composition Analysis InBody Scale
Subjects with HF and ICD had their ICD interrogated and continually monitored by an electrophysiologist before, during, and after the 30-50 seconds of bioimpedance analysis, which involves standing on the InBody 520 scale. Outcome measure was assessed during a one day visit. There is no further follow-up.
20
Total20

Baseline characteristics

CharacteristicBody Composition Analysis InBody Scale
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
20 Participants
Region of Enrollment
United States
20 participants
Sex: Female, Male
Female
2 Participants
Sex: Female, Male
Male
18 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 20
other
Total, other adverse events
0 / 20
serious
Total, serious adverse events
0 / 20

Outcome results

Primary

Number of Participants With Arrhythmia as Detected by ICD

Subjects with HF and ICD had their ICD interrogated and continually monitored by an electrophysiologist before, during, and after 30-50 seconds of bioimpedance analysis, which involves standing on the InBody 520 scale. Outcome measure was assessed during a one day visit. There is no further follow-up.

Time frame: 1 day, on day of study ICD interrogated and continually monitored before, during, and after 30-50 seconds of bioimpedance analysis

ArmMeasureValue (NUMBER)
Body Composition Analysis InBody ScaleNumber of Participants With Arrhythmia as Detected by ICD0 participants