Artificial Pancreas Control System in an Inpatient Setting

Sensor-controlled Insulin- and Glucagon Delivery in Subjects With Type 1 Diabetes: Testing of an Automated System in a Supervised Inpatient Setting

Status
Completed
Phases
Phase 2
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01552603
Enrollment
14
Registered
2012-03-13
Start date
2012-03-31
Completion date
2013-09-30
Last updated
2016-06-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Type I Diabetes Mellitus

Keywords

glucose sensors, artificial pancreas, Diabetes Mellitus

Brief summary

The purpose of this study is to verify an automated system of blood glucose control in Type I Diabetics. The automated system consists of the investigational Artificial Pancreas Control software (APC), two blood glucose sensors, and two hormone pumps, one for delivering insulin to lower blood sugar, and the second for delivering glucagon to raise blood sugar. The blood glucose sensors relays information to the Artificial Pancreas software, which uses the Adaptive Proportional Device algorithm to determine the rate of insulin and glucagon infusion by the hormone pumps. In prior studies, the Adaptive Proportional Device algorithm has been verified, but required manual input into the computer and hormone pumps. This study differs in that it uses a fully automated system under the control of the Artificial Pancreas Control software. The importance of this change is that it is the next step to enable outpatient use of automated, closed loop blood glucose control.

Detailed description

The objective of the current human study is to verify the components of the Artificial Pancreas Control system during an inpatient study. This master controller software is designed to be used in conjunction with two subcutaneous continuous glucose monitoring systems to regulate blood glucose levels as well as two Omnipod pumps, one for administering insulin and one for administering glucagon. The sensors communicate wirelessly with two sensor receivers which will be interfaced with the APC by wireless USB connection. The insulin and glucagon pumps will be controlled by the APC through a wireless USB connection. The algorithm included in the APC is an automated version of the Adaptive Proportional Derivative (APD) insulin and glucagon control algorithm, which was previously studied as an investigational device. The APD has been studied in vivo (in 28 experiments, each 33 hr in length, with manual adjustment of pumps) and no serious adverse effects were noted. Manual input of the glucose sensor data and insulin/glucagon infusion rates will no longer be necessary. The APC will be tested in vivo during 28 hour experiments in an inpatient setting in preparation for outpatient testing.

Interventions

This master controller software is used in conjunction with two subcutaneous continuous glucose monitoring systems and two Omnipod pumps, one for administering aspart insulin (NovoLog) and one for administering glucagon (GlucaGen), to control blood glucose levels. The insulin and glucagon infusion rates are determined by an automated version of the Adaptive Proportional Derivative (APD) insulin and glucagon control algorithm.

Sponsors

Juvenile Diabetes Research Foundation
CollaboratorOTHER
Legacy Health System
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
21 Years to 65 Years
Healthy volunteers
No

Inclusion criteria

* Diagnosis of type 1 diabetes mellitus for at least 1 year. * Male or female subjects 21 to 65 years of age. * Current use of an insulin pump. * Willingness to follow all study procedures, including attending all clinic visits. * Willingness to sign informed consent and HIPAA documents.

Exclusion criteria

* Pregnancy or Lactation: For women of childbearing potential: there is a requirement for a negative urine pregnancy test and for agreement to use contraception during the study and for at least 1 month after participating in the study. Acceptable contraception includes birth control pill / patch / vaginal ring, Depo-Provera, Norplant, an IUD, the double barrier method (the woman uses a diaphragm and spermicide and the man uses a condom), or abstinence. * Renal insufficiency (serum creatinine of 2.0 mg/dL or greater). * Serum ALT or AST equal to or greater than 3 times the upper limit of normal; hepatic synthetic insufficiency as defined as a serum albumin of less than 3.3 g/dL; or serum bilirubin of over 2. * Adrenal insufficiency * Hematocrit of less than or equal to 34%. * A history of cerebrovascular disease or coronary artery disease regardless of the time since occurrence. * Congestive heart failure, NYHA class III or IV. * Diagnosis of 2nd or 3rd degree heart block or any arrhythmia judged by the investigator to be exclusionary. * Any active infection. * Visual impairment preventing reading of glucose meter values or continuous glucose monitoring device. * Physical impairment impeding the ability to use a glucose meter or glucose monitoring device. * Active foot ulceration. * Severe peripheral arterial disease characterized by ischemic rest pain or severe claudication. * Active alcohol abuse, substance abuse, or severe mental illness (as judged by the principal investigator). * Active malignancy, except basal cell or squamous cell skin cancers. * Major surgical operation within 30 days prior to screening. * Seizure disorder. * Chronic usage of any immunosuppressive medication (such as cyclosporine, azathioprine, sirolimus, or tacrolimus). * Current administration of oral or parenteral corticosteroids. * Use of an investigational drug within 30 days prior to screening. * Bleeding disorder, treatment with warfarin, or platelet count below 50,000. * Allergy to aspart insulin. * Allergy to glucagon. * Past history of pheochromocytoma or a family history of MEN 2, neurofibromatosis, or von Hippel-Lindau disease. * Insulin resistance requiring more than 200 units per day. * Need for uninterrupted treatment of acetaminophen. * Intolerance of mild hypoglycemia (glucose 60-70 mg/dl). * History of hypoglycemic unawareness. * Insulin antibody level of ≥ 100 µUnits/ml. * C peptide level of ≥0.5 ng/ml. * Any concurrent illness, other than diabetes, that is not controlled by a stable therapeutic regimen. * Any reason the principal investigator deems exclusionary

Design outcomes

Primary

MeasureTime frameDescription
Mean Percent of Time in Target Blood Glucose Rangeall 28 hour studiesMean percent of time venous blood glucose was sampled between 70-180 mg/dl

Secondary

MeasureTime frameDescription
Mean Deviation From Target Blood Glucoseall 28 hour studiesMean difference of venous blood glucose from glucose target. Daytime venous blood glucose values (7am-11pm) subtracted from daytime target of 115 mg/dl. Nighttime venous blood glucose values (11pm-7am) subtracted from nighttime target of 140 mg/dl. This is a metric of how successful the closed loop algorithm was at controlling glucose.

Countries

United States

Participant flow

Participants by arm

ArmCount
Artificial Pancreas Control
Type 1 Diabetes Mellitus subjects who fit the inclusion/exclusion criteria will undergo artificial pancreas closed-loop study for 28 hours. For the entire study, the adaptive component of Artificial Pancreas Control Software will be used to control the subject's blood glucose. Artificial Pancreas Control Software: This master controller software is used in conjunction with two subcutaneous continuous glucose monitoring systems and two Omnipod pumps, one for administering aspart insulin (NovoLog) and one for administering glucagon (GlucaGen), to control blood glucose levels. The insulin and glucagon infusion rates are determined by an automated version of the Adaptive Proportional Derivative (APD) insulin and glucagon control algorithm.
14
Total14

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyPhysician Decision1

Baseline characteristics

CharacteristicArtificial Pancreas Control
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
14 Participants
Sex: Female, Male
Female
9 Participants
Sex: Female, Male
Male
5 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
1 / 14
serious
Total, serious adverse events
0 / 14

Outcome results

Primary

Mean Percent of Time in Target Blood Glucose Range

Mean percent of time venous blood glucose was sampled between 70-180 mg/dl

Time frame: all 28 hour studies

ArmMeasureValue (MEAN)Dispersion
Artificial Pancreas ControlMean Percent of Time in Target Blood Glucose Range62.4 percentage of timeStandard Deviation 7
Secondary

Mean Deviation From Target Blood Glucose

Mean difference of venous blood glucose from glucose target. Daytime venous blood glucose values (7am-11pm) subtracted from daytime target of 115 mg/dl. Nighttime venous blood glucose values (11pm-7am) subtracted from nighttime target of 140 mg/dl. This is a metric of how successful the closed loop algorithm was at controlling glucose.

Time frame: all 28 hour studies

ArmMeasureValue (MEAN)Dispersion
Artificial Pancreas ControlMean Deviation From Target Blood Glucose59.9 mg/dlStandard Deviation 51.6