Effectiveness and Safety Study of the Dexcom G4™ Continuous Glucose Monitoring System

PTL-900590 Effectiveness and Safety Study of the Dexcom G4™ Continuous Glucose Monitoring System

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01514292
Enrollment
72
Registered
2012-01-23
Start date
2011-12-31
Completion date
2012-02-29
Last updated
2015-10-05

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study is to evaluate the effectiveness and safety of the Dexcom G4 System when used as an adjuvant to blood glucose testing over a 7-day period in people \>18 years-old with diabetes mellitus. The primary objective is to characterize the System performance with respect to laboratory reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the G4 System in reference to YSI. Safety of the G4 System will be characterized by Adverse Device Effects and Serious Adverse Device Effects experienced by study participants.

Detailed description

The primary objective is to characterize the System performance with respect to laboratory reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the G4 System in reference to YSI. The trend accuracy of the device performance, i.e., continuous glucose error grid analysis, will be evaluated as well as the temporal G4 System accuracy at different glucose rates of changes and different glucose ranges (hypoglycemic, euglycemic, and hyperglycemic ranges). The primary matched paired (Sensor-YSI) measurements will be collected during multiple in-clinic days.

Interventions

DEVICEContinuous Glucose Monitoring of glucose levels (Dexcom)

Manipulation of glucose levels during multiple clinic days. Achieved through diet, exercise and insulin regimen defined in protocol.Evaluation of CGM performance characteristics in reference to a laboratory standard.

Sponsors

DexCom, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Age 18 years or older; * Diagnosed with diabetes mellitus; * Use one of the following for their diabetes management: * intensive insulin therapy (IIT) or * non-intensive insulin therapy (non-IIT) * Intensive-insulin using (IIT) subjects only - During each in-clinic session, willing to have their blood glucose levels manipulated into high and low glucose levels. Subjects not on IIT will only be observed during each in-clinic session. * For subjects that exercise routinely (at least 3 times per week), willing to exercise on each in-clinic session, if asked; * Willing to take a minimum of 7 fingersticks per day during home use days; * Willing to refrain from the use of acetaminophen during sensor wear period and the day prior to sensor insertion;

Exclusion criteria

* Have extensive skin changes/diseases that preclude wearing the required number of devices on normal skin (e.g., extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis) at the proposed wear sites; * Have a known allergy to medical-grade adhesives; * Are pregnant as demonstrated by a positive pregnancy test within 72 hours of sensor insertion, * Are on dialysis; * Have a hematocrit that is less than 36% or greater than 55% at screening visit; * Currently participating in another investigational study protocol (if a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study); * Have a history of cardiovascular disease (including, but not limited to ischemic heart disease, peripheral vascular disease, cardiomyopathy, cerebrovascular disease, congenital heart disease, or significant arrhythmias), epilepsy, severe migraines in the past 6 months, adrenal disease, syncope, significant hypoglycemia unawareness, or a history of severe hypoglycemia (requiring emergency medical intervention) within the last 6 months; * Have any chronic infectious disease or intercurrent illness that would interfere with their participation in the study or pose an excessive risk to study staff handling venous or capillary blood samples (e.g. HIV/AIDS, Hepatitis B or C); * Have a MRI scan, CT scan, or diathermy scheduled during the week of the study. If any of these procedures are required urgently during the study, subjects will notify the study staff, end their CGM session, and remove their sensor.

Design outcomes

Primary

MeasureTime frameDescription
CGM Relative Differences to Laboratory Reference7 daysThe outcome measure is measured as the relative differences (%) of the CGM glucose value in reference to a laboratory reference, yellow spring instrument( YSI) glucose measurements.

Countries

United States

Participant flow

Participants by arm

ArmCount
Real Time Continuous Glucose Monitoring System
Real Time Continuous Glucose Monitoring(CGM) System Wearing for up to 7 days.
72
Total72

Baseline characteristics

CharacteristicReal Time Continuous Glucose Monitoring System
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
1 Participants
Age, Categorical
Between 18 and 65 years
71 Participants
Age, Continuous42 years
STANDARD_DEVIATION 14
Ethnicity (NIH/OMB)
Hispanic or Latino
8 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
64 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
1 Participants
Race (NIH/OMB)
Black or African American
3 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
68 Participants
Region of Enrollment
United States
72 participants
Sex: Female, Male
Female
28 Participants
Sex: Female, Male
Male
44 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 72
serious
Total, serious adverse events
0 / 72

Outcome results

Primary

CGM Relative Differences to Laboratory Reference

The outcome measure is measured as the relative differences (%) of the CGM glucose value in reference to a laboratory reference, yellow spring instrument( YSI) glucose measurements.

Time frame: 7 days

ArmMeasureValue (MEAN)Dispersion
Real Time Continous Glucose Monitoring SystemCGM Relative Differences to Laboratory Reference13 Percentage of differenceStandard Deviation 6
Comparison: Hence, the hypothesis is established as:~Ho: pi \< 71.5% Ha: pi \>71.5% Thus, the objective is to conclude that the proportion of G4 Sensor-YSI points in the present study meeting the 20 mg/dL/20% criterion is no worse than the existing FDA-approved SEVEN PLUS System. The null hypothesis will be rejected if pi observed in this study is greater than 71.5%, the G4 System performance is no worse than the historical performance of the existing FDA approved CGM system will be concluded.p-value: 0.0001Fisher Exact

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026