Electrical Dysfunction, Externalized Conductors, Visual Lead Anomalies
Conditions
Keywords
Externalized Conductors, visual lead an
Brief summary
The Cardiac Lead Assessment Study (CLAS) is designed as a prospective, non-randomized, multi-center, international postmarket surveillance study under 522 Order PS120111 (August 16, 2012).
Detailed description
The objective of this study was to determine the prevalence and incidence of externalized conductors and other visual lead anomalies (as evidenced by imaging) and the risk of progression to electrical dysfunction in subjects with one or more of the following St. Jude Medical (SJM) cardiac leads: Riata leads, Riata ST leads, Durata and/or QuickSite/QuickFlex Cardiac Resynchronization Therapy (CRT) leads.
Interventions
Riata leads implanted between 2002-2009 Riata ST leads implanted between 2006-2009 Quicksite or QuickFlex leads implanted between 2006-2010 Durata leads implanted between 2008-2010 Subjects were followed for three years and cinefluoroscopy imaging was done annually to determine prevalence of externalized conductors or other visual lead anomalies
Sponsors
Study design
Eligibility
Inclusion criteria
1. Patient has a market released SJM Implantable Cardioverter Defibrillator (ICD), CRT-D or CRT-P already implanted. 2. Patient has at least one market released Riata/ Riata ST/QuickSite/QuickFlex/Durata already implanted in which the lead model number and lead implant date is indicated in Table 1 (see protocol attached.) 3. Have the ability to provide informed consent for study participation and be willing to comply with the prescribed evaluations as detailed in this study plan. 4. Are 18 years or above
Exclusion criteria
1. Patient is currently pregnant. 2. Enrolled or intend to participate in a clinical drug and/or device study, which could confound the results of this trial as determined by SJM, during the course of this clinical study
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Number of Leads With Electrical Dysfunction | This outcome is assessed at enrollment and over the course of 3 years of follow-up | The prevalence of electrical dysfunction was calculated as the number of leads that have electrical dysfunction through three years of follow-up for the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups. |
| Number of Leads With Externalized Conductors | This outcome is assessed at enrollment | The prevalence of externalized conductors was calculated as the number of leads that have been shown to have externalized conductors by imaging at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups |
| Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities) | This outcome is assessed at enrollment | The prevalence of other visual lead anomalies (by each subcategory) was calculated as the number of leads that have been shown to have other visual anomaly (by each subcategory) at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups. |
| The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | This outcome is assessed from enrollment through the course of 3 years of follow-up | Annual hazard rate is the proportion of number of leads with electrical dysfunction to the total number of lead follow-up years. |
| The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | year 1 follow-up | Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from enrollment to year 1 |
| The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | year 1 to year 2 follow-up | Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 1 to year 2 |
| The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | year 2 to year 3 follow-up | Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 2 to year 3 |
| The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | This outcome is assessed from enrollment to year 1 | Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from enrollment to year 1 |
| The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | This outcome is assessed from year 1 to year 2 of follow-up | Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 1 to year 2 |
| The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | This outcome is assessed from year 2 to year 3 follow-up | Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 2 to year 3 |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | This outcome is assessed from time of detecting EC till the course of 3 years of follow-up | Kaplan-Meier analysis of time from externalized conductors (EC) to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting EC |
| Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | This outcome is assessed from time of detecting other visual lead anomaly till the course of 3 years of follow-up | Kaplan-Meier analysis of time from other visual lead anomaly to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting other visual lead anomaly |
| Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | within 30 days associated with study lead related interventions | Number of participants with adverse events within 30 days associated with study lead related interventions for each lead group |
| Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | This outcome is assessed from time of detection of externalized conductor to 3 years follow-up | time from externalized conductors to clinical intervention for all lead subgroups. |
| Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | This outcome is assessed from time of detection of other visual lead anomaly until 3 years of follow-up | Kaplan-Meier analysis of time from visual lead anomalies for all lead and lead age subcategories to clinical intervention |
| Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC) | This outcome is assessed from enrollment till the course of 3 years of follow-up | Comparison of patients with EC to those without EC: lead size effect on number of patients with or without EC (7F and 8F are lead sizes, F = French). Riata leads have 8F diameter and Riata ST leads have 7F diameter. The Riata/Riata ST families of leads were used for this comparison only to determine whether the difference in lead size (with the silicon material and other lead design factors controlled for) had an effect on externalization of conductors. |
| Comparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED). | This outcome is assessed from enrollment till the course of 3 years of follow-up | Comparison of patients with ED to those without ED: subject age effect on ED |
Countries
Canada, Japan, United States
Participant flow
Recruitment details
The study enrolled a total of 2,216 subjects at 45 sites to achieve an evaluable sample size of at least 1,401 subjects. The first subject enrolled in this study on March 20, 2013 and enrollment was completed in June, 2016 with the last follow-up completed in June 13 2019.
Pre-assignment details
A subject that has at least one market released lead implanted was eligible for enrollment in the study. Subjects could have more than one study lead implanted. Each lead group was considered separately for independent analyses.
Participants by arm
| Arm | Count |
|---|---|
| All Subjects Enrolled in the Study Demographics for all enrolled patients (N=2216) | 2,216 |
| Total | 2,216 |
Baseline characteristics
| Characteristic | All Subjects Enrolled in the Study |
|---|---|
| Age, Continuous | 67 years STANDARD_DEVIATION 12 |
| Ethnicity (NIH/OMB) Hispanic or Latino | 56 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 2158 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 2 Participants |
| Race (NIH/OMB) American Indian or Alaska Native | 8 Participants |
| Race (NIH/OMB) Asian | 91 Participants |
| Race (NIH/OMB) Black or African American | 335 Participants |
| Race (NIH/OMB) More than one race | 0 Participants |
| Race (NIH/OMB) Native Hawaiian or Other Pacific Islander | 3 Participants |
| Race (NIH/OMB) Unknown or Not Reported | 9 Participants |
| Race (NIH/OMB) White | 1770 Participants |
| Region of Enrollment Canada | 48 participants |
| Region of Enrollment Japan | 83 participants |
| Region of Enrollment United States | 2085 participants |
| Sex: Female, Male Female | 600 Participants |
| Sex: Female, Male Male | 1616 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk | EG002 affected / at risk | EG003 affected / at risk |
|---|---|---|---|---|
| deaths Total, all-cause mortality | 130 / 982 | 111 / 670 | 64 / 458 | 135 / 737 |
| other Total, other adverse events | 20 / 982 | 14 / 670 | 9 / 458 | 11 / 737 |
| serious Total, serious adverse events | 26 / 982 | 28 / 670 | 22 / 458 | 7 / 737 |
Outcome results
Number of Leads With Electrical Dysfunction
The prevalence of electrical dysfunction was calculated as the number of leads that have electrical dysfunction through three years of follow-up for the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.
Time frame: This outcome is assessed at enrollment and over the course of 3 years of follow-up
Population: All SJM leads enrolled in the study with Durata, Riata, Riata ST, or QuickSite/QuickFlex leads
| Arm | Measure | Value (COUNT_OF_UNITS) |
|---|---|---|
| Durata Leads | Number of Leads With Electrical Dysfunction | 24 implanted leads |
| Riata Leads | Number of Leads With Electrical Dysfunction | 27 implanted leads |
| Riata ST | Number of Leads With Electrical Dysfunction | 15 implanted leads |
| QuickSite/QuickFlex | Number of Leads With Electrical Dysfunction | 2 implanted leads |
Number of Leads With Externalized Conductors
The prevalence of externalized conductors was calculated as the number of leads that have been shown to have externalized conductors by imaging at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups
Time frame: This outcome is assessed at enrollment
Population: All SJM leads enrolled in the study with Durata, Riata, Riata ST, or QuickSite/QuickFlex leads
| Arm | Measure | Value (COUNT_OF_UNITS) |
|---|---|---|
| Durata Leads | Number of Leads With Externalized Conductors | 0 implanted leads |
| Riata Leads | Number of Leads With Externalized Conductors | 144 implanted leads |
| Riata ST | Number of Leads With Externalized Conductors | 44 implanted leads |
| QuickSite/QuickFlex | Number of Leads With Externalized Conductors | 10 implanted leads |
Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)
The prevalence of other visual lead anomalies (by each subcategory) was calculated as the number of leads that have been shown to have other visual anomaly (by each subcategory) at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.
Time frame: This outcome is assessed at enrollment
| Arm | Measure | Value (COUNT_OF_UNITS) |
|---|---|---|
| Durata Leads | Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities) | 7 implanted leads |
| Riata Leads | Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities) | 19 implanted leads |
| Riata ST | Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities) | 7 implanted leads |
| QuickSite/QuickFlex | Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities) | 4 implanted leads |
The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
Annual hazard rate is the proportion of number of leads with electrical dysfunction to the total number of lead follow-up years.
Time frame: This outcome is assessed from enrollment through the course of 3 years of follow-up
Population: The annual hazard rate was assessed separately in each of the 4 lead groups.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Durata Leads | The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.010 leads with electrical dysfunction/years |
| Riata Leads | The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.019 leads with electrical dysfunction/years |
| Riata ST | The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.014 leads with electrical dysfunction/years |
| QuickSite/QuickFlex | The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.001 leads with electrical dysfunction/years |
The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from enrollment to year 1
Time frame: year 1 follow-up
Population: The annual hazard rate was assessed separately in each of the 4 lead groups.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Durata Leads | The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0025 leads with externalized conductors/years |
| Riata Leads | The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0480 leads with externalized conductors/years |
| Riata ST | The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0158 leads with externalized conductors/years |
| QuickSite/QuickFlex | The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0288 leads with externalized conductors/years |
The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 1 to year 2
Time frame: year 1 to year 2 follow-up
Population: The annual hazard rate was assessed separately in each of the 4 lead groups.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Durata Leads | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0015 leads with externalized conductors/years |
| Riata Leads | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0897 leads with externalized conductors/years |
| Riata ST | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0234 leads with externalized conductors/years |
| QuickSite/QuickFlex | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0087 leads with externalized conductors/years |
The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 2 to year 3
Time frame: year 2 to year 3 follow-up
Population: The annual hazard rate was assessed separately in each of the 4 lead groups.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Durata Leads | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0035 leads with externalized conductors/years |
| Riata Leads | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0658 leads with externalized conductors/years |
| Riata ST | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0097 leads with externalized conductors/years |
| QuickSite/QuickFlex | The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads. | 0.0110 leads with externalized conductors/years |
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads
Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from enrollment to year 1
Time frame: This outcome is assessed from enrollment to year 1
Population: The annual hazard rate was assessed separately in each of the 4 lead groups.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Durata Leads | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0025 leads w other visual lead anomaly/years |
| Riata Leads | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0108 leads w other visual lead anomaly/years |
| Riata ST | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0117 leads w other visual lead anomaly/years |
| QuickSite/QuickFlex | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0053 leads w other visual lead anomaly/years |
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads
Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 1 to year 2
Time frame: This outcome is assessed from year 1 to year 2 of follow-up
Population: The annual hazard rate was assessed separately in each of the 4 lead groups.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Durata Leads | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0 leads w other visual lead anomaly/years |
| Riata Leads | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0027 leads w other visual lead anomaly/years |
| Riata ST | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0070 leads w other visual lead anomaly/years |
| QuickSite/QuickFlex | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0021 leads w other visual lead anomaly/years |
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads
Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 2 to year 3
Time frame: This outcome is assessed from year 2 to year 3 follow-up
Population: The annual hazard rate was assessed separately in each of the 4 lead groups.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Durata Leads | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0 leads w other visual lead anomaly/years |
| Riata Leads | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0.0066 leads w other visual lead anomaly/years |
| Riata ST | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0 leads w other visual lead anomaly/years |
| QuickSite/QuickFlex | The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads | 0 leads w other visual lead anomaly/years |
Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)
Comparison of patients with EC to those without EC: lead size effect on number of patients with or without EC (7F and 8F are lead sizes, F = French). Riata leads have 8F diameter and Riata ST leads have 7F diameter. The Riata/Riata ST families of leads were used for this comparison only to determine whether the difference in lead size (with the silicon material and other lead design factors controlled for) had an effect on externalization of conductors.
Time frame: This outcome is assessed from enrollment till the course of 3 years of follow-up
Population: Patients with Riata or Riata ST leads.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Durata Leads | Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC) | 7F lead size (Riata ST) | 57 Participants |
| Durata Leads | Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC) | 8F lead size (Riata) | 204 Participants |
| Riata Leads | Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC) | 7F lead size (Riata ST) | 395 Participants |
| Riata Leads | Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC) | 8F lead size (Riata) | 456 Participants |
Comparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED).
Comparison of patients with ED to those without ED: subject age effect on ED
Time frame: This outcome is assessed from enrollment till the course of 3 years of follow-up
Population: Patients implanted with SJM Riata, Riata ST, Durata, or QuickSite/QuickFlex leads. Leads were combined for this analysis to compare patient groups with and without electrical dysfunction based on age (irrespective of lead implant type)
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Durata Leads | Comparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED). | 61.8 years | Standard Deviation 13.4 |
| Riata Leads | Comparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED). | 67.2 years | Standard Deviation 11.9 |
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant
time from externalized conductors to clinical intervention for all lead subgroups.
Time frame: This outcome is assessed from time of detection of externalized conductor to 3 years follow-up
Population: N at risk from time of externalized conductors at each follow-up period
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | NA lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | NA lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 1 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 1 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 1 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 1 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 1 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 1 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 3 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 2 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 4 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 7 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 5 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 8 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 15 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 11 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 18 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 1 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 3 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 6 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 6 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 4 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 4 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 6 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 1 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 1 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 1 year from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 0 years from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 2 years from detecting EC | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant | 3 years from detecting EC | 0 lead anomaly events |
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant
Kaplan-Meier analysis of time from externalized conductors (EC) to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting EC
Time frame: This outcome is assessed from time of detecting EC till the course of 3 years of follow-up
Population: N at risk from detecting externalized conductors at each follow-up time point
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 0 lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | NA lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | NA lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | NA lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | NA lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | NA lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | NA lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | NA lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | NA lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 2 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 1 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 2 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | 2 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 1 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 1 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 2 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | 3 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 3 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 5 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | 6 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 6 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | 1 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 1 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | 4 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 2 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 2 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 1 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | NA lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 2 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 1 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 0 from detecting externalized conductor | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant | year 3 from detecting externalized conductor | 0 lead anomaly events |
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant
Kaplan-Meier analysis of time from other visual lead anomaly to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting other visual lead anomaly
Time frame: This outcome is assessed from time of detecting other visual lead anomaly till the course of 3 years of follow-up
Population: N at risk from detecting other visual lead anomaly
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Durata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 1 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 1 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 1 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 1 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 1 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 1 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 3 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 2 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 1 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant | 0 years follow up from detecting other visual lead anomaly | 0 lead anomaly events |
Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant
Kaplan-Meier analysis of time from visual lead anomalies for all lead and lead age subcategories to clinical intervention
Time frame: This outcome is assessed from time of detection of other visual lead anomaly until 3 years of follow-up
Population: N at risk from time of detecting visual lead anomaly at each follow-up period
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Durata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
| Durata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| Durata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| Durata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 1 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 1 lead anomaly events |
| QuickSite/QuickFlex | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 1 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 6-less Than 7 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 2 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata Leads Lead Age: 7 Years or Greater | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 2 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST Lead Age: Less Than 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 1 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 2 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 2 lead anomaly events |
| Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Lead Age: Less Than 5 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 0 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 3 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 1 years from detection other visual lead anomaly | 0 lead anomaly events |
| QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years | Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant | 2 years from detection other visual lead anomaly | 0 lead anomaly events |
Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).
Number of participants with adverse events within 30 days associated with study lead related interventions for each lead group
Time frame: within 30 days associated with study lead related interventions
Population: Subject has at least one market released Riata, Riata ST, QuickSite/QuickFlex, or Durata lead implanted
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Durata Leads | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | lead insulation damage | 0 Participants |
| Durata Leads | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | phrenic nerve/diaphragmatic stimulation | 0 Participants |
| Durata Leads | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | Lead dislodgement or migration | 1 Participants |
| Riata Leads | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | Lead dislodgement or migration | 0 Participants |
| Riata Leads | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | lead insulation damage | 0 Participants |
| Riata Leads | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | phrenic nerve/diaphragmatic stimulation | 0 Participants |
| Riata ST | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | lead insulation damage | 0 Participants |
| Riata ST | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | Lead dislodgement or migration | 0 Participants |
| Riata ST | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | phrenic nerve/diaphragmatic stimulation | 0 Participants |
| QuickSite/QuickFlex | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | Lead dislodgement or migration | 0 Participants |
| QuickSite/QuickFlex | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | phrenic nerve/diaphragmatic stimulation | 1 Participants |
| QuickSite/QuickFlex | Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other). | lead insulation damage | 1 Participants |