Skip to content

St.Jude Medical Cardiac Lead Assessment Study

St.Jude Medical Cardiac Lead Assessment Study

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT01507987
Acronym
CLAS
Enrollment
2216
Registered
2012-01-11
Start date
2013-03-31
Completion date
2019-09-30
Last updated
2021-08-23

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Electrical Dysfunction, Externalized Conductors, Visual Lead Anomalies

Keywords

Externalized Conductors, visual lead an

Brief summary

The Cardiac Lead Assessment Study (CLAS) is designed as a prospective, non-randomized, multi-center, international postmarket surveillance study under 522 Order PS120111 (August 16, 2012).

Detailed description

The objective of this study was to determine the prevalence and incidence of externalized conductors and other visual lead anomalies (as evidenced by imaging) and the risk of progression to electrical dysfunction in subjects with one or more of the following St. Jude Medical (SJM) cardiac leads: Riata leads, Riata ST leads, Durata and/or QuickSite/QuickFlex Cardiac Resynchronization Therapy (CRT) leads.

Interventions

DEVICEImplanted with one or more of the following SJM leads: Riata leads, RiataST leads, Durata leads, and QuickSite/QuickFlex leads

Riata leads implanted between 2002-2009 Riata ST leads implanted between 2006-2009 Quicksite or QuickFlex leads implanted between 2006-2010 Durata leads implanted between 2008-2010 Subjects were followed for three years and cinefluoroscopy imaging was done annually to determine prevalence of externalized conductors or other visual lead anomalies

Sponsors

Abbott Medical Devices
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Patient has a market released SJM Implantable Cardioverter Defibrillator (ICD), CRT-D or CRT-P already implanted. 2. Patient has at least one market released Riata/ Riata ST/QuickSite/QuickFlex/Durata already implanted in which the lead model number and lead implant date is indicated in Table 1 (see protocol attached.) 3. Have the ability to provide informed consent for study participation and be willing to comply with the prescribed evaluations as detailed in this study plan. 4. Are 18 years or above

Exclusion criteria

1. Patient is currently pregnant. 2. Enrolled or intend to participate in a clinical drug and/or device study, which could confound the results of this trial as determined by SJM, during the course of this clinical study

Design outcomes

Primary

MeasureTime frameDescription
Number of Leads With Electrical DysfunctionThis outcome is assessed at enrollment and over the course of 3 years of follow-upThe prevalence of electrical dysfunction was calculated as the number of leads that have electrical dysfunction through three years of follow-up for the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.
Number of Leads With Externalized ConductorsThis outcome is assessed at enrollmentThe prevalence of externalized conductors was calculated as the number of leads that have been shown to have externalized conductors by imaging at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups
Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)This outcome is assessed at enrollmentThe prevalence of other visual lead anomalies (by each subcategory) was calculated as the number of leads that have been shown to have other visual anomaly (by each subcategory) at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.
The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.This outcome is assessed from enrollment through the course of 3 years of follow-upAnnual hazard rate is the proportion of number of leads with electrical dysfunction to the total number of lead follow-up years.
The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.year 1 follow-upAnnual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from enrollment to year 1
The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.year 1 to year 2 follow-upAnnual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 1 to year 2
The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.year 2 to year 3 follow-upAnnual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 2 to year 3
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata LeadsThis outcome is assessed from enrollment to year 1Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from enrollment to year 1
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata LeadsThis outcome is assessed from year 1 to year 2 of follow-upAnnual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 1 to year 2
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata LeadsThis outcome is assessed from year 2 to year 3 follow-upAnnual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 2 to year 3

Secondary

MeasureTime frameDescription
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial ImplantThis outcome is assessed from time of detecting EC till the course of 3 years of follow-upKaplan-Meier analysis of time from externalized conductors (EC) to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting EC
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial ImplantThis outcome is assessed from time of detecting other visual lead anomaly till the course of 3 years of follow-upKaplan-Meier analysis of time from other visual lead anomaly to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting other visual lead anomaly
Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).within 30 days associated with study lead related interventionsNumber of participants with adverse events within 30 days associated with study lead related interventions for each lead group
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial ImplantThis outcome is assessed from time of detection of externalized conductor to 3 years follow-uptime from externalized conductors to clinical intervention for all lead subgroups.
Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial ImplantThis outcome is assessed from time of detection of other visual lead anomaly until 3 years of follow-upKaplan-Meier analysis of time from visual lead anomalies for all lead and lead age subcategories to clinical intervention
Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)This outcome is assessed from enrollment till the course of 3 years of follow-upComparison of patients with EC to those without EC: lead size effect on number of patients with or without EC (7F and 8F are lead sizes, F = French). Riata leads have 8F diameter and Riata ST leads have 7F diameter. The Riata/Riata ST families of leads were used for this comparison only to determine whether the difference in lead size (with the silicon material and other lead design factors controlled for) had an effect on externalization of conductors.
Comparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED).This outcome is assessed from enrollment till the course of 3 years of follow-upComparison of patients with ED to those without ED: subject age effect on ED

Countries

Canada, Japan, United States

Participant flow

Recruitment details

The study enrolled a total of 2,216 subjects at 45 sites to achieve an evaluable sample size of at least 1,401 subjects. The first subject enrolled in this study on March 20, 2013 and enrollment was completed in June, 2016 with the last follow-up completed in June 13 2019.

Pre-assignment details

A subject that has at least one market released lead implanted was eligible for enrollment in the study. Subjects could have more than one study lead implanted. Each lead group was considered separately for independent analyses.

Participants by arm

ArmCount
All Subjects Enrolled in the Study
Demographics for all enrolled patients (N=2216)
2,216
Total2,216

Baseline characteristics

CharacteristicAll Subjects Enrolled in the Study
Age, Continuous67 years
STANDARD_DEVIATION 12
Ethnicity (NIH/OMB)
Hispanic or Latino
56 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
2158 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
2 Participants
Race (NIH/OMB)
American Indian or Alaska Native
8 Participants
Race (NIH/OMB)
Asian
91 Participants
Race (NIH/OMB)
Black or African American
335 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
3 Participants
Race (NIH/OMB)
Unknown or Not Reported
9 Participants
Race (NIH/OMB)
White
1770 Participants
Region of Enrollment
Canada
48 participants
Region of Enrollment
Japan
83 participants
Region of Enrollment
United States
2085 participants
Sex: Female, Male
Female
600 Participants
Sex: Female, Male
Male
1616 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
deaths
Total, all-cause mortality
130 / 982111 / 67064 / 458135 / 737
other
Total, other adverse events
20 / 98214 / 6709 / 45811 / 737
serious
Total, serious adverse events
26 / 98228 / 67022 / 4587 / 737

Outcome results

Primary

Number of Leads With Electrical Dysfunction

The prevalence of electrical dysfunction was calculated as the number of leads that have electrical dysfunction through three years of follow-up for the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.

Time frame: This outcome is assessed at enrollment and over the course of 3 years of follow-up

Population: All SJM leads enrolled in the study with Durata, Riata, Riata ST, or QuickSite/QuickFlex leads

ArmMeasureValue (COUNT_OF_UNITS)
Durata LeadsNumber of Leads With Electrical Dysfunction24 implanted leads
Riata LeadsNumber of Leads With Electrical Dysfunction27 implanted leads
Riata STNumber of Leads With Electrical Dysfunction15 implanted leads
QuickSite/QuickFlexNumber of Leads With Electrical Dysfunction2 implanted leads
Primary

Number of Leads With Externalized Conductors

The prevalence of externalized conductors was calculated as the number of leads that have been shown to have externalized conductors by imaging at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups

Time frame: This outcome is assessed at enrollment

Population: All SJM leads enrolled in the study with Durata, Riata, Riata ST, or QuickSite/QuickFlex leads

ArmMeasureValue (COUNT_OF_UNITS)
Durata LeadsNumber of Leads With Externalized Conductors0 implanted leads
Riata LeadsNumber of Leads With Externalized Conductors144 implanted leads
Riata STNumber of Leads With Externalized Conductors44 implanted leads
QuickSite/QuickFlexNumber of Leads With Externalized Conductors10 implanted leads
Primary

Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)

The prevalence of other visual lead anomalies (by each subcategory) was calculated as the number of leads that have been shown to have other visual anomaly (by each subcategory) at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.

Time frame: This outcome is assessed at enrollment

ArmMeasureValue (COUNT_OF_UNITS)
Durata LeadsNumber of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)7 implanted leads
Riata LeadsNumber of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)19 implanted leads
Riata STNumber of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)7 implanted leads
QuickSite/QuickFlexNumber of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)4 implanted leads
Primary

The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.

Annual hazard rate is the proportion of number of leads with electrical dysfunction to the total number of lead follow-up years.

Time frame: This outcome is assessed from enrollment through the course of 3 years of follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

ArmMeasureValue (NUMBER)
Durata LeadsThe Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.010 leads with electrical dysfunction/years
Riata LeadsThe Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.019 leads with electrical dysfunction/years
Riata STThe Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.014 leads with electrical dysfunction/years
QuickSite/QuickFlexThe Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.001 leads with electrical dysfunction/years
Primary

The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.

Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from enrollment to year 1

Time frame: year 1 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

ArmMeasureValue (NUMBER)
Durata LeadsThe Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0025 leads with externalized conductors/years
Riata LeadsThe Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0480 leads with externalized conductors/years
Riata STThe Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0158 leads with externalized conductors/years
QuickSite/QuickFlexThe Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0288 leads with externalized conductors/years
Primary

The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.

Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 1 to year 2

Time frame: year 1 to year 2 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

ArmMeasureValue (NUMBER)
Durata LeadsThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0015 leads with externalized conductors/years
Riata LeadsThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0897 leads with externalized conductors/years
Riata STThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0234 leads with externalized conductors/years
QuickSite/QuickFlexThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0087 leads with externalized conductors/years
Primary

The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.

Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 2 to year 3

Time frame: year 2 to year 3 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

ArmMeasureValue (NUMBER)
Durata LeadsThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0035 leads with externalized conductors/years
Riata LeadsThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0658 leads with externalized conductors/years
Riata STThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0097 leads with externalized conductors/years
QuickSite/QuickFlexThe Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.0.0110 leads with externalized conductors/years
Primary

The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads

Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from enrollment to year 1

Time frame: This outcome is assessed from enrollment to year 1

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

ArmMeasureValue (NUMBER)
Durata LeadsThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0025 leads w other visual lead anomaly/years
Riata LeadsThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0108 leads w other visual lead anomaly/years
Riata STThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0117 leads w other visual lead anomaly/years
QuickSite/QuickFlexThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0053 leads w other visual lead anomaly/years
Primary

The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads

Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 1 to year 2

Time frame: This outcome is assessed from year 1 to year 2 of follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

ArmMeasureValue (NUMBER)
Durata LeadsThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0 leads w other visual lead anomaly/years
Riata LeadsThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0027 leads w other visual lead anomaly/years
Riata STThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0070 leads w other visual lead anomaly/years
QuickSite/QuickFlexThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0021 leads w other visual lead anomaly/years
Primary

The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads

Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 2 to year 3

Time frame: This outcome is assessed from year 2 to year 3 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

ArmMeasureValue (NUMBER)
Durata LeadsThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0 leads w other visual lead anomaly/years
Riata LeadsThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0.0066 leads w other visual lead anomaly/years
Riata STThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0 leads w other visual lead anomaly/years
QuickSite/QuickFlexThe Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads0 leads w other visual lead anomaly/years
Secondary

Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)

Comparison of patients with EC to those without EC: lead size effect on number of patients with or without EC (7F and 8F are lead sizes, F = French). Riata leads have 8F diameter and Riata ST leads have 7F diameter. The Riata/Riata ST families of leads were used for this comparison only to determine whether the difference in lead size (with the silicon material and other lead design factors controlled for) had an effect on externalization of conductors.

Time frame: This outcome is assessed from enrollment till the course of 3 years of follow-up

Population: Patients with Riata or Riata ST leads.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Durata LeadsComparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)7F lead size (Riata ST)57 Participants
Durata LeadsComparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)8F lead size (Riata)204 Participants
Riata LeadsComparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)7F lead size (Riata ST)395 Participants
Riata LeadsComparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)8F lead size (Riata)456 Participants
Secondary

Comparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED).

Comparison of patients with ED to those without ED: subject age effect on ED

Time frame: This outcome is assessed from enrollment till the course of 3 years of follow-up

Population: Patients implanted with SJM Riata, Riata ST, Durata, or QuickSite/QuickFlex leads. Leads were combined for this analysis to compare patient groups with and without electrical dysfunction based on age (irrespective of lead implant type)

ArmMeasureValue (MEAN)Dispersion
Durata LeadsComparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED).61.8 yearsStandard Deviation 13.4
Riata LeadsComparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED).67.2 yearsStandard Deviation 11.9
Secondary

Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant

time from externalized conductors to clinical intervention for all lead subgroups.

Time frame: This outcome is assessed from time of detection of externalized conductor to 3 years follow-up

Population: N at risk from time of externalized conductors at each follow-up period

ArmMeasureGroupValue (NUMBER)
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting ECNA lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting ECNA lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC1 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC1 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC1 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC1 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC1 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC1 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC3 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC2 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC4 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC7 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC5 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC8 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC15 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC11 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC18 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC1 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC3 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC6 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC6 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC4 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC4 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC6 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC1 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC1 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant1 year from detecting EC0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant0 years from detecting EC0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant2 years from detecting EC0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant3 years from detecting EC0 lead anomaly events
Secondary

Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant

Kaplan-Meier analysis of time from externalized conductors (EC) to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting EC

Time frame: This outcome is assessed from time of detecting EC till the course of 3 years of follow-up

Population: N at risk from detecting externalized conductors at each follow-up time point

ArmMeasureGroupValue (NUMBER)
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor0 lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductorNA lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductorNA lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductorNA lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductorNA lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductorNA lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductorNA lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductorNA lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductorNA lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor2 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor1 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor2 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductor2 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor1 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor1 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor2 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductor3 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor3 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor5 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductor6 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor6 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductor1 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor1 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductor4 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor2 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor2 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor1 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductorNA lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 2 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 1 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 0 from detecting externalized conductor0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implantyear 3 from detecting externalized conductor0 lead anomaly events
Secondary

Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant

Kaplan-Meier analysis of time from other visual lead anomaly to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting other visual lead anomaly

Time frame: This outcome is assessed from time of detecting other visual lead anomaly till the course of 3 years of follow-up

Population: N at risk from detecting other visual lead anomaly

ArmMeasureGroupValue (NUMBER)
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
Durata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata STEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly1 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly1 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly1 lead anomaly events
QuickSite/QuickFlexEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly1 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly1 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly1 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant3 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant2 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant1 years follow up from detecting other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant0 years follow up from detecting other visual lead anomaly0 lead anomaly events
Secondary

Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant

Kaplan-Meier analysis of time from visual lead anomalies for all lead and lead age subcategories to clinical intervention

Time frame: This outcome is assessed from time of detection of other visual lead anomaly until 3 years of follow-up

Population: N at risk from time of detecting visual lead anomaly at each follow-up period

ArmMeasureGroupValue (NUMBER)
Durata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
Durata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
Durata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
Durata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
Riata LeadsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
Riata STEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
Riata STEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
Riata STEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
Riata STEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlexEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly1 lead anomaly events
QuickSite/QuickFlexEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlexEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly1 lead anomaly events
QuickSite/QuickFlexEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly1 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 6-less Than 7 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly2 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
Riata Leads Lead Age: 7 Years or GreaterEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly2 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
Riata ST Lead Age: Less Than 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly1 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly2 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly2 lead anomaly events
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Lead Age: Less Than 5 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant0 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant3 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant1 years from detection other visual lead anomaly0 lead anomaly events
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 YearsEstimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant2 years from detection other visual lead anomaly0 lead anomaly events
Secondary

Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).

Number of participants with adverse events within 30 days associated with study lead related interventions for each lead group

Time frame: within 30 days associated with study lead related interventions

Population: Subject has at least one market released Riata, Riata ST, QuickSite/QuickFlex, or Durata lead implanted

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Durata LeadsNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).lead insulation damage0 Participants
Durata LeadsNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).phrenic nerve/diaphragmatic stimulation0 Participants
Durata LeadsNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).Lead dislodgement or migration1 Participants
Riata LeadsNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).Lead dislodgement or migration0 Participants
Riata LeadsNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).lead insulation damage0 Participants
Riata LeadsNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).phrenic nerve/diaphragmatic stimulation0 Participants
Riata STNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).lead insulation damage0 Participants
Riata STNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).Lead dislodgement or migration0 Participants
Riata STNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).phrenic nerve/diaphragmatic stimulation0 Participants
QuickSite/QuickFlexNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).Lead dislodgement or migration0 Participants
QuickSite/QuickFlexNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).phrenic nerve/diaphragmatic stimulation1 Participants
QuickSite/QuickFlexNumber of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).lead insulation damage1 Participants

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026