Study of an Investigational Glucose Meter System

Clinical Study of the Apollo Evolution BGMS With Tatsu Blood Glucose Test Strips

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01466075
Enrollment
207
Registered
2011-11-07
Start date
2011-11-30
Completion date
2011-12-31
Last updated
2016-02-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study is to demonstrate that untrained subjects who have diabetes or pre-diabetes can operate the investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.

Interventions

DEVICEApollo Evolution Investigational BG Monitoring System

Untrained subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick blood and Alternative Site Testing (AST) of the palm using the Apollo Evolution meter and an investigational sensor. Study staff test subject venous blood and all BG results are compared to a reference laboratory glucose method. Untrained subjects complete basic tasks using the User Guide and provide feedback.

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Have a diagnosis of diabetes or pre-diabetes. * Be 18 years of age or older * Be able to speak, read and understand English and understand the Informed Consent document. * Be willing to complete study procedures

Exclusion criteria

* Pregnancy * Infections or skin disorders at the site of the venipuncture (at the discretion of the phlebotomist). * Hemophilia, bleeding disorder, or clotting problems. Persons taking aspirin (81 mg or 325 mg) daily or Plavix® are not reason for exclusion. * Physical (dexterity), visual, or neurological impairments that would make the person unable to perform testing with the BGM (at the discretion of the site professional staff). * A condition, in the opinion of the Investigator, would put the subject at risk or influence the conduct of the study or interpretation of the results. The reason for exclusion will be documented by the professional staff. * Previously participated in a blood glucose monitoring study using a similar device or use a similar device for personal use when monitoring blood glucose. * Working for a competitive medical device company or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company * Working for a medical laboratory, hospital or other clinical setting that involves training on and clinical use of blood glucose monitors

Design outcomes

Primary

MeasureTime frameDescription
Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-test fingerstick blood using the Apollo Evolution Investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma). Site staff tested in parallel after subjects.

Secondary

MeasureTime frameDescription
Percent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-test Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma).
Percent of Venous Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method1 hourStudy staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI venous plasma results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI venous plasma) or +/- 20% (\>=75mg/dL YSI venous plasma).
Number of Subjects Able to Perform Given Tasks Using Product Labeling for Instruction1 hourAfter reading the instructions for use, and without assistance from the study staff, subjects use the BGMS to perform basic tasks considered to be essential for the operation of the system.

Countries

United States

Participant flow

Participants by arm

ArmCount
Intended Users of the Monitoring System
Untrained subjects with diabetes use the Apollo Evolution Investigational Blood Glucose Monitoring System. Apollo Evolution Investigational BG Monitoring System : Untrained subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick blood and AST of the palm using the Apollo Evolution meter and an investigational sensor. Study staff test subject venous blood. All BG results are compared to a reference laboratory glucose method. Untrained subjects complete basic tasks using the User Guide and provide feedback.
207
Total207

Baseline characteristics

CharacteristicIntended Users of the Monitoring System
Age, Continuous55.7 years
Region of Enrollment
United States
207 participants
Sex: Female, Male
Female
117 Participants
Sex: Female, Male
Male
90 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
1 / 207
serious
Total, serious adverse events
0 / 207

Outcome results

Primary

Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-test fingerstick blood using the Apollo Evolution Investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma). Site staff tested in parallel after subjects.

Time frame: 1 hour

Population: Blood data for three subjects were not evaluable because the difference between the replicates of the reference YSI analyzer measurements exceeded the protocol defined criteria. Remaining 204 subjects each tested one of three test strip lots on the system. 204 (207-3) test results were available.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method99.51 percentage of Blood Glucose Test Results
Secondary

Number of Subjects Able to Perform Given Tasks Using Product Labeling for Instruction

After reading the instructions for use, and without assistance from the study staff, subjects use the BGMS to perform basic tasks considered to be essential for the operation of the system.

Time frame: 1 hour

Population: Blood data for three subjects were not considered evaluable because the difference between the replicates of the reference YSI analyzer measurements exceeded the protocol defined criteria. Remaining 204 subjects each tested one of three test strip lots on the system. 204 (207-3) test results were available.

ArmMeasureGroupValue (NUMBER)
Intended Users of the Monitoring SystemNumber of Subjects Able to Perform Given Tasks Using Product Labeling for InstructionSuccessfully performed fingerstick BG test203 participants
Intended Users of the Monitoring SystemNumber of Subjects Able to Perform Given Tasks Using Product Labeling for InstructionSuccessfully performed palm AST BG test203 participants
Secondary

Percent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-test Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: Blood data for three subjects were not evaluable because the difference between the replicates of the reference YSI analyzer measurements exceeded the protocol defined criteria. The remaining 204 subjects each tested one of three test strip lots on the system. 204 (207-3) test results were available.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method99.51 percentage of Blood Glucose Test Results
Secondary

Percent of Venous Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI venous plasma results are used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI venous plasma) or +/- 20% (\>=75mg/dL YSI venous plasma).

Time frame: 1 hour

Population: 609 venous BG results(203 subjects x 3 test strip lots) were available. Three subjects did not have successful venipunctures so no blood obtained. BGMS testing was not performed with one subject's blood sample. Study staff tested each subject venous blood sample using 3 test strip lots.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemPercent of Venous Blood Glucose Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method100.0 percentage of Blood Glucose Test Results