Closed Loop Glucose Control in Intensive Care Unit

An Open-label, Single-centre, Randomised Controlled Trial to Assess the Safety and Efficacy of Automated Closed-loop Blood Glucose Control in Comparison With Standard Care in Adults in Intensive Care Unit

Status
Completed
Phases
Phase 2
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01440842
Acronym
CLASSIC
Enrollment
24
Registered
2011-09-27
Start date
2012-05-31
Completion date
2012-10-31
Last updated
2013-01-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Stress Hyperglycaemia

Keywords

Stress hyperglycaemia

Brief summary

The purpose of this study is to assess the efficacy, safety and feasibility of a computer-based control algorithm to control glucose levels in adults in intensive care unit, in comparison to standard care.

Detailed description

This will be a single centre, open-label, randomised, parallel design, feasibility study conducted at Neurosciences Critical Care Unit (NCCU), Addenbrooke's hospital, Cambridge, UK. Study will aim for 24 adult subjects (12 participants in each arm of the trial) and study will last for up to 49 hours in each subject. Subjects will have a commercially available Conformité Européenne(CE) marked subcutaneous glucose sensor inserted at the start of the study. Glucose data from the sensor will be transmitted to a small bedside tablet computer, containing the algorithm which will determine insulin infusion rates aimed at maintaining glucose level between 6.0 - 8.0 mmol/L. The advice from the algorithm will be sent to the infusion pump via USB cable and insulin will be delivered intravenously. The system will also deliver intravenous glucose via a second infusion pump at times of low glucose levels. Samples for reference glucose values will be obtained either from an arterial line or central venous cannula and will be analysed using standard blood gas analyser in real time. Subjects randomised to standard care will receive intravenous insulin based on current treatment guidelines at the Neurosciences critical care unit (NCCU), Addenbrooke's Hospital, Cambridge, UK. The primary outcome is time spent in target glucose range between 6.0 to 8.0 mmol/L as recorded by reference glucose. Secondary outcomes are the time spent with glucose levels above and below target, as recorded by reference glucose, and sensor accuracy. Safety includes evaluation of significant hypoglycaemia and hyperglycaemia and other adverse events.

Interventions

Intravenous infusion delivery of Actrapid insulin and dextrose, dose calculated by Model Predictive Control (MPC) algorithm, based on continuous glucose sensor readings.

OTHERStandard IV insulin infusion sliding scale

Standard intravenous insulin infusion sliding scale as per intensive care unit protocol.

Sponsors

European Commission
CollaboratorOTHER
Cambridge University Hospitals NHS Foundation Trust
CollaboratorOTHER
University of Cambridge
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Aged 18 years and older * Admitted to Neurosciences critical care unit; (all patients are eligible regardless of admitting diagnosis except where specified under

Exclusion criteria

) * Stay in intensive care unit expected to be at least 48 hours * At least one of the following conditions applies: * Not on insulin infusion and single confirmed reference blood glucose level \> 10.0 mmol/l * Already on insulin infusion including those subjects with pre-existing diabetes.

Design outcomes

Primary

MeasureTime frameDescription
Percentage of glucose values in target range (6.0 - 8.0 mmol/l) as recorded by reference glucose measurements.Up to 49 hours in each subject.Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer. During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.

Secondary

MeasureTime frameDescription
Percentage of glucose values <4.0 mmol/l and >8.0 mmol/l as recorded by reference glucose measurements.Up to 49 hours in each subject.Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer. During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.
Percentage of glucose values > 10.0 mmol/l as recorded by reference glucose measurements.Up to 49 hours in each subject,Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer. During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.
Mean and standard deviation of reference glucose measured using arterial blood gas analyzerUp to 49 hours in each subject.Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer. This is reported in mmol/L During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation of mean (mmol/L) and standard deviation which will reflect the efficasy of closed loop insulin delivery.
Percentage of glucose values in range (4.0 - 10.0 mmol/l) as recorded by reference glucose measurements.Up to 49 hours in each subjectDefinition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer. During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.
Time to reach target glucoseUp to 49 hours in each subject.
Insulin infusion ratesUp to 49 hours in each subject.
Frequency and magnitude of significant hypoglycaemic (< 3.0 mmol/L), severe hypoglycaemic (<2.0 mmol/L) and significant hyperglycaemic (> 15mmol/l) episodes.Up to 49 hours in each subject.
Mean and median absolute and relative difference between matched pairs of subcutaneous glucose sensor and reference plasma glucose.Up to 49 hours in each subject.For a given subcutaneous glucose sensor value, difference between the sensor and the reference glucose will be calculated. Example - Reference glucose 10 mmol/L and sensor glucose 12 mmol/L, therefore absolute difference will be 2 mmol/L. Mean and median of these deviations will be reported in mmol/L. The term relative implies that data has been converted to a percentage deviation from reference glucose.