Skip to content

User Performance Evaluation of an Investigational Blood Glucose Monitoring System

Ninja 2 User Performance Evaluation

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01410773
Enrollment
110
Registered
2011-08-05
Start date
2011-07-31
Completion date
2011-08-31
Last updated
2016-02-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study is to demonstrate that untrained subjects with diabetes can use an investigational Blood Glucose Monitoring System (BGMS) with capillary blood obtained from fingerstick and from Alternative Site (AST) Palm .

Interventions

The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Males and females, 18 years of age and above * Diagnosed as having type 1 or type 2 diabetes * Currently self-testing blood glucose at home * Able to speak, read, and understand English * Willing to complete all study procedures

Exclusion criteria

* Blood Borne infections like hepatitis or HIV or infections such as tuberculosis * Hemophilia or any other bleeding disorder * Employee of competitive medical device company,immediate family member of such, or living within the household of such * Pregnancy

Design outcomes

Primary

MeasureTime frameDescription
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM). BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or within +/- 20%(for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).

Secondary

MeasureTime frameDescription
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-test subject Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGM meter results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the YSI capillary plasma reference method results.

Countries

United States

Participant flow

Participants by arm

ArmCount
Intended Users of the System
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm. Ninja 2 Investigational Blood Glucose Monitoring System : The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
110
Total110

Baseline characteristics

CharacteristicIntended Users of the System
Age, Customized
Mean (Full range of ages)
53.1 years
Region of Enrollment
United States
110 participants
Sex: Female, Male
Female
58 Participants
Sex: Female, Male
Male
52 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 110
serious
Total, serious adverse events
0 / 110

Outcome results

Primary

Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM). BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or within +/- 20%(for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).

Time frame: 1 hour

Population: 110 subjects tested one of 3 test strip lots on the BGM system. 110 (1x110) test results are available.

ArmMeasureGroupValue (NUMBER)
Intended Users of the SystemNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method# within +/-15mg/dL(<75mg/dL)or 20% (>=75mg/dL)110 participants
Intended Users of the SystemNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method# within +/-15mg/dL(<75mg/dL)or 15% (>=75mg/dL)109 participants
Intended Users of the SystemNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method# within +/-10mg/dL(<75mg/dL)or 10% (>=75mg/dL)109 participants
Intended Users of the SystemNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method# within +/-5mg/dL(<75mg/dL) or 5% (>=75mg/dL)90 participants
Secondary

Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-test subject Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGM meter results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the YSI capillary plasma reference method results.

Time frame: 1 hour

Population: One subject had low blood sugar. The protocol (and User Guide) do not allow alternative site testing when blood sugar is low. The remaining 109 subjects tested one strip lot on the BGM system. 109 test results were available.

ArmMeasureGroupValue (NUMBER)
Intended Users of the SystemNumber of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method# within +/- 15mg/dL(<75mg/dL) or 20% (>=75mg/dL)106 participants
Intended Users of the SystemNumber of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method# within +/- 15mg/dL(<75mg/dL) or 15% (>=75mg/dL)105 participants
Intended Users of the SystemNumber of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method# within +/- 10mg/dL(<75mg/dL) or 10% (>=75mg/dL)97 participants
Intended Users of the SystemNumber of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method# within +/- 5mg/dL(<75mg/dL) or 5% (>=75mg/dL)67 participants

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026