Skip to content

Diaphragmatic Pacemaker in Tetraplegic Patients With Spinal Cord Injuries

Evaluation of the NeuRx RA/4 Diaphragmatic Pacemaker in Tetraplegic Patients With Spinal Cord Injuries in Ventilator Support

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01385384
Enrollment
5
Registered
2011-06-30
Start date
2011-06-30
Completion date
2012-12-31
Last updated
2013-03-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diaphragmatic Paralysis, Respiratory Paralysis, Spinal Cord Injury

Brief summary

Permanent dependency of breathing apparatus due to spinal cord injury is traditionally treated with different types of mechanical ventilation. However, the electric ventilation became a possibility through their most current versions, such as diaphragmatic pacemakers. Diaphragmatic pacemakers rhythmically stimulates the diaphragm to replace the functions of the respiratory center that doesn't works well or is inaccessible. However, this modality has the prerequisite that the phrenic nerve and diaphragm muscle are normal. The reason for the development of diaphragmatic pacemaker freeing the patient from the ventilator. By using the mechanical energy of the diaphragm of the patient, the patient may come not need the ventilator tubing, tracheostomy, and with the help of their caregivers, the inconvenient mechanical ventilators.

Interventions

DEVICENeuRx RA/4 diaphragmatic pacemaker

Patients will be implanted with 4 intramuscular electrodes, 2 in each hemi-diaphragm, using laparoscopic techniques. The abdominal cavity is inflated with CO2 and 4 ports are inserted: one for optics, 2 for a probe containing a temporary electrode for mapping the diaphragm and the electrode insertion tool, and a smaller one for the output wires of the electrodes of the abdominal cavity. The intra-abdominal pressure variation during the stimulation test (mapping) will be measured externally by one of the ports. Sites that provide the optimal response (greater region and magnitude) are noted. Once the optimal site for electrode placement is identified two intramuscular electrodes are deployed in each hemi-diaphragm. The guide wires will come out of the peritoneum through the port placed in the xiphoid region. An additional electrode is placed subcutaneously in the upper abdomen. The instruments and ports are then removed and the incision sites closed.

Sponsors

Synapse Biomedical
CollaboratorINDUSTRY
University of Sao Paulo
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* 18 years old or more * Cervical spinal cord injuries patients under mechanical ventilation * Clinically stable after spinal cord injury * Clinical acceptable bilateral phrenic nerve function demonstrated with electromyography and neural conduction time * Fluoroscopic visible diaphragmatic movements under stimulation * Hemodynamically stable * No co-morbidities that can interfere with pacemaker implantation or function * Pregnancy negative test for women * Patient or legal representative informed consent

Exclusion criteria

* Active pulmonary disease * Active cardiovascular disease * Active cerebral disease * Hemodynamic instability or low oxygen levels in ambient air * Hospitalization for infection in the last 3 months * Significant scoliosis or chest disease * Obesity * Poor compliance to the protocol from the patient or the caregiver

Design outcomes

Primary

MeasureTime frameDescription
Volume comparison of the basal with the tidal volume obtained with the NeuRx RA/4 pacemaker in patients with diaphragmatic paralysisOne yearThe ability of the NeuRx RA / 4 to obtain clinically acceptable tidal volume during a period of 4 continuous hours will be evaluated by spirometry. Clinically acceptable volume is defined as a volume in excess of 15% above the basal metabolic requirements of the particular patient. For a male patient, the flow volume to meet the basal metabolic requirement is set at 7 ml/kg body weight, while for a female patient, it is defined in 6ml/Kg body weight.

Secondary

MeasureTime frameDescription
Functional rehabilitation by measure of total usage time, both day and nightOne yearUsing spreadsheets to patients and caregivers, will collect information related to the use of the device, such as total usage time, the preference to use it both day and night instead of a mechanical ventilator, or to use it partial time to perform any activity or acquire independence that would otherwise be difficult with mechanical ventilation.

Countries

Brazil

Outcome results

None listed