Performance Validation of a Non-contact Breathing Frequency Monitor (Model BM07)

Performance Validation of a Non-contact Breathing Frequency Monitor (Model BM07)

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT01319396
Enrollment
24
Registered
2011-03-21
Start date
2010-10-31
Completion date
2011-03-31
Last updated
2021-04-01

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Unspecified Threat to Breathing

Keywords

breathing rate

Brief summary

The purpose of this study is to compare the accuracy of the BiancaMed BM07 breathing frequency indicator device with the respiration rate measured by a SomnoScreen RC Easy device on 20+ human volunteers. This is being carried out in support of a 510(k) submission.

Detailed description

The purpose of this study is to compare the accuracy of the BiancaMed BM07 breathing frequency indicator device with the respiration rate measured by a SomnoScreen RC Easy device on 20+ human volunteers. This is being carried out in support of a 510(k) submission. * The BM07 is a non-contact device, using radio frequency (RF) range measurement to detect the chest wall motion, and thus the respiration rate. The BM07 is a CE Class IIa cleared device. * The SomnoScreen RC Easy is both a CE and FDA (K060708) cleared device, used as an overnight, portable sleep recorder. One of its inputs is a pair of chest respiration effort bands that can be used to record the breathing rate of the subject. * The human volunteers are selected to be a range of age, body mass index (BMI) and gender so as to cover a broad cross section of the population. BMI, gender and age are features that might affect the performance of the RF reflected signal: * High BMI or low BMI could affect the skin RF reflectivity. * High BMI may mask chest wall movement during breathing. * The thinner skin often associated with increasing age may affect RF reflectivity. * Female breast tissue may affect RF reflectivity or mask chest wall movement. The target performance is an agreement between the BM07 and the Somnoscreen of +/- 5 breaths per minute. This target has been selected based upon a literature review, in particular: * Lim et al: Respiratory Rate Measurement in Adults - How Reliable is it? * Droitcour PhD Thesis: Non-Contact Measurement of Heart and Respiration Rates with a Single-Chip Microwave Doppler Radar. There is no risk to the subjects as the RF energy emitted by the BM07 is much less than that of either a mobile phone bluetooth or a domestic WLAN router. The testing is to be carried out on volunteer human subjects, who sign a consent form. The locations of testing are Dublin, Ireland and Belfast, UK. The testing comprises the subjects sitting still for periods of 2 minutes whilst recordings are taken from both the BM07 and Somnoscreen. The test configurations are ranges from the BM07 to subject of 0.5, 1.0 & 1.5m and 1.0m with a thick folded duvet (overblanket/comforter) covering the subject to represent them wearing thick outdoor winter clothing. Each test configuration is carried out twice, making a total of 8 test recordings per subject. The first 2 volunteers will be used to confirm that the maximum effective range of the BM07 is at least 1.5 m and that the target aspect of the subject (ie the sensor being in front, to one side and from behind) does not affect the ability of the BM07 to detect breathing movement. This is an Observational study as the subjects are normal healthy volunteers and no intervention can be made as a result of the study measurements

Interventions

None listed

Sponsors

ResMed
Lead SponsorINDUSTRY

Study design

Observational model
CASE_ONLY
Time perspective
CROSS_SECTIONAL

Eligibility

Sex/Gender
ALL
Age
10 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* volunteer * consent

Exclusion criteria

none

Design outcomes

Primary

MeasureTime frameDescription
Respiration Rate Accuracy - Mean Difference in Breaths Per Minute Between the BM07 5GHz Sensor and SomnoScreen DevicesRecording duration of 2 minutes; measurement made after 45 seconds settling time, for at least 60 secondsThe respiration rate measured contemporaneously by the BM07 5GHz sensor and the reference device, SomnoScreen, are compared on a second-by-second basis, over the recording duration. The difference between the two respiration rate measurements (the delta) for all 24 subjects' recordings, in all positions, are plotted on a normal distribution chart, giving 10,542 data points. The distribution chart reveals the mean difference between the two devices' measurements and the Standard Deviation (SD). The measure of success is if the 95% (ie 2SD) of the difference in the respiration rate indicated by the two devices (the delta) is less than 5 breaths per minute. This measure is derived from the peer review paper by Lim et al, cited in the References section

Countries

Ireland

Participant flow

Participants by arm

ArmCount
5.8 GHz Sensor Spot Respiration Rate
All participants had the difference in indicated breathing rate measured. The 2 devices were the BM07 5GHz non-contact sensor (device under test) and the Somnoscreen (reference device using chest effort bands)
24
Total24

Baseline characteristics

Characteristic5.8 GHz Sensor Spot Respiration Rate
Age, Categorical
<=18 years
2 Participants
Age, Categorical
>=65 years
2 Participants
Age, Categorical
Between 18 and 65 years
20 Participants
Age, Continuous40.6 years
STANDARD_DEVIATION 14.1
Region of Enrollment
Ireland
15 participants
Region of Enrollment
United Kingdom
9 participants
Sex: Female, Male
Female
8 Participants
Sex: Female, Male
Male
16 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 24
serious
Total, serious adverse events
0 / 24

Outcome results

Primary

Respiration Rate Accuracy - Mean Difference in Breaths Per Minute Between the BM07 5GHz Sensor and SomnoScreen Devices

The respiration rate measured contemporaneously by the BM07 5GHz sensor and the reference device, SomnoScreen, are compared on a second-by-second basis, over the recording duration. The difference between the two respiration rate measurements (the delta) for all 24 subjects' recordings, in all positions, are plotted on a normal distribution chart, giving 10,542 data points. The distribution chart reveals the mean difference between the two devices' measurements and the Standard Deviation (SD). The measure of success is if the 95% (ie 2SD) of the difference in the respiration rate indicated by the two devices (the delta) is less than 5 breaths per minute. This measure is derived from the peer review paper by Lim et al, cited in the References section

Time frame: Recording duration of 2 minutes; measurement made after 45 seconds settling time, for at least 60 seconds

ArmMeasureValue (MEAN)Dispersion
5GHz Sensor Spot Respiration RateRespiration Rate Accuracy - Mean Difference in Breaths Per Minute Between the BM07 5GHz Sensor and SomnoScreen Devices0.28 breaths per minuteStandard Deviation 1.2