CANadian Clinical Evaluation of SpiderFlash Arrhythmia DETECTion in SpiderFlash™ Device

CANadian Clinical Evaluation of the Automatic Arrhythmia DETECTion in SpiderFlash™ Device

Status
UNKNOWN
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT01218763
Acronym
CAN-DETECT
Enrollment
101
Registered
2010-10-11
Start date
2010-04-30
Completion date
Unknown
Last updated
2011-06-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Arrhythmias, Cardiac

Keywords

Cardiac Event recorder, Sensitivity, Efficacy

Brief summary

This clinical investigation is a prospective, non-randomized, multi-center pivotal trial. This trial aims to validate the automatic detection features provided by the external long-duration SpiderFlash-t™ Event recorder coupled with its analyser EventScope™, and to assess the performances of the Mode Switch function provided by Sorin Group Implantable Cardioverter-Defibrillators (ICDs). This trial is being performed first in order to demonstrate the ventricular tachycardia (VT)/ ventricular fibrillation (VF) sensitivity of the automatic VT/VF detection function of the EventScope™ software. Patients will be implanted with an ICD from Sorin Group, with PARAD+ arrhythmia detection enabled and Mode Switch (MS) algorithms activated in all patients. Arrhythmic events will be documented over a period of 6 to 8 weeks for each patient.

Interventions

None listed

Sponsors

Sorin Group Canada
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

Patients already implanted with an OVATIO DR or CRT-D device may be included in the study if they presented with at least one clinical VT/VF event at the follow-up visit preceding the enrolment. ICD or CRT-D device upgrading or replacement are allowed but under the condition that existing bipolar RV sensing/pacing lead and/or defibrillation lead and/or LV pacing lead (where applicable) are not of poor or questionable integrity. Patients with implanted pacemakers may be included in the study only if the pacemaker is to be explanted or otherwise disabled prior to implanting the ICD or CRT-D system, to avoid any potential pacemaker-ICD interactions. Patients must comply with all the following criteria in order to be included in the CAN-DETECT study: 1. Patient is eligible for implantation of a dual-chamber ICD or a CRT-D device according to current available guidelines for primary or secondary ICD indication . 2. A stable, optimal (as determined by the enrolling physician) drug regimen 3. Scheduled for implant of an OVATIO DR Model 6550 or OVATIO CRT 6750 or an equivalent device from Sorin Group. 4. Signed and dated informed consent.

Exclusion criteria

Any patient complying with any of the following criteria cannot be included in the CAN-DETECT study: 1. Any contraindication for standard cardiac pacing 2. Any contraindication for ICD therapy 3. Permanent high-degree AV Block for patients planned for dual-chamber ICD implant only. 4. Permanent AF for patients planned for dual-chamber ICD implant only. 5. Abdominal implantation site 6. Acute myocarditis 7. Mechanical tricuspid valve 8. Heart transplant recipient 9. Already participating in another clinical study 10. Life expectancy less than 6 months 11. Inability to understand the purpose of the study or refusal to cooperate 12. Inability or refusal to provide informed consent 13. Unavailability for scheduled follow-up at the implanting center 14. Under guardianship 15. Age of less than 18 years.

Design outcomes

Primary

MeasureTime frameDescription
VT/VF sensitivity2 yearsThe primary purpose of this study is to evaluate the performance of the automatic detection of ventricular tachycardia and ventricular fibrillation (VT/VF) by the EventScope™ analyzer found in the Holter monitor SpiderFlash-t™. In order to validate this objective we will use the arrhythmia data collected in the memory storage found in the Sorin Group DR and CRT ICDs.

Secondary

MeasureTime frameDescription
AF sensitivity2 yearsThe main secondary objective of this study is to evaluate the detection of atrial fibrillation by the Mode Switch algorithm found in the Sorin Group defibrillators. However, as SpiderFlash-t has only been validated on patients with spontaneous rhythm, in order to achieve this objective we will use the AF events detected by the EventScope™ in patients presenting with a spontaneous rhythm. SafeR mode, programmed in DR devices, will permit preservation of such rhythm.

Countries

Canada

Outcome results

None listed