Skip to content

Hydration Factor Study for the Stayhealthy BC3

Pilot Study for the Validation of the BC3 Bioelectrical Impedance Hydration Factor

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01209936
Enrollment
360
Registered
2010-09-27
Start date
2009-09-30
Completion date
2010-07-31
Last updated
2023-12-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Body Composition

Keywords

Body Fat, Total body water

Brief summary

This was a pilot study to validate the BC3 hydration factor which will be important for determining how well this value represents hydration status compared to assessment of total body water by D20.

Detailed description

The BC3 body composition analyzer is manufactured by Stayhealthy. It is a handheld, noninvasive bioelectrical impedance device. When held, the BC3 sends an electrical impulse through the body to determine body fat, lean mass, and total body hydration. Bio-impedance devices measure the change in electrical resistance in body tissues by using a detectable electrical signal that passes through the body. The method is based on the principle that lean body mass contains virtually all the water and conducting electrolytes in the body and provides a good electrical pathway. In contrast, fat or fat-containing tissues produce a poor electrical pathway. By inducing a low-energy, high-frequency signal, a measurement of the baseline resistance to flow of electrical current can be assessed. The resistance measurement relates directly to the volume of the conductor. Through repeated measurements and identifying other parameters (i.e. gender, weight, age and height) proprietary equations were developed that determine total body water, lean mass, and % body fat. The purpose of this study is to validate the calculated hydration factor produced by the BC3 bioelectrical impedance device. If this value is valid and comparable to the gold standard (D20) then this hydration factor will offer a very quick and inexpensive method to determine total body hydration status using a hand-held bioelectrical impedance device.

Interventions

DEVICED2O

All participants received deuterium oxide (D2O) and testing with the BC3 for the measurement of total body water.

Sponsors

University of Southern California
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
HEALTH_SERVICES_RESEARCH
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 80 Years
Healthy volunteers
Yes

Inclusion criteria

• Men and women 18-80 years of age

Exclusion criteria

* Acute illness (infection, recent surgery, trauma). * Pregnancy * Pacemaker or defibrillator

Design outcomes

Primary

MeasureTime frameDescription
Estimated Body Fat % for Children Aged 10-17 Years OldOne time point (only one day visit)Stayhealthy BC3 was analyzed against hydrostatic weighing for children aged 10-17 years old
Percent Body Fat in Women Aged 18-80 Years OldOne time point (only one day visit)Body fat percentage was measured via Stayhealthy BC3 as well as a full body DXA
Estimated Percent Body Fat for Men Aged 18-80One time point (only one day visit)The body fat percentage as estimated from BC3 and as calculated by a full body DXA.

Countries

United States

Participant flow

Participants by arm

ArmCount
Adults
All adults underwent a 10-15 minute total body DXA Scan. All adults used the Stayhealthy body composition analyzer BC3 for BIA to measure body composition. All participants received deuterium oxide (D2O) and testing with the BC3 for the measurement of total body water.
245
Children
All children were measured using Hydrostatic Weighing to assess body composition. All children used the Stayhealthy body composition analyzer BC3 for BIA to measure body composition. All participants received deuterium oxide (D2O) and testing with the BC3 for the
115
Total360

Baseline characteristics

CharacteristicChildrenTotalAdults
Age, Continuous14.9 Years
STANDARD_DEVIATION 1.7
30.15 Years
STANDARD_DEVIATION 9.225
45.4 Years
STANDARD_DEVIATION 16.75
Race and Ethnicity Not Collected0 Participants
Region of Enrollment
United States
115 participants360 participants245 participants
Sex: Female, Male
Female
64 Participants192 Participants128 Participants
Sex: Female, Male
Male
51 Participants168 Participants117 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
— / —— / —
other
Total, other adverse events
0 / 2450 / 117
serious
Total, serious adverse events
0 / 2450 / 117

Outcome results

Primary

Estimated Body Fat % for Children Aged 10-17 Years Old

Stayhealthy BC3 was analyzed against hydrostatic weighing for children aged 10-17 years old

Time frame: One time point (only one day visit)

ArmMeasureGroupValue (MEAN)Dispersion
Estimated Body Fat % From BC3 for Children Aged 10-17 Years OldEstimated Body Fat % for Children Aged 10-17 Years OldBC318.4 Percent body fatStandard Deviation 8.4
Estimated Body Fat % From BC3 for Children Aged 10-17 Years OldEstimated Body Fat % for Children Aged 10-17 Years OldHydrostatic Weighing18.8 Percent body fatStandard Deviation 8.3
p-value: 0.87t-test, 2 sided
Primary

Estimated Percent Body Fat for Men Aged 18-80

The body fat percentage as estimated from BC3 and as calculated by a full body DXA.

Time frame: One time point (only one day visit)

ArmMeasureGroupValue (MEAN)Dispersion
Estimated Body Fat % From BC3 for Children Aged 10-17 Years OldEstimated Percent Body Fat for Men Aged 18-80BC320.3 Percent body fatStandard Deviation 5.3
Estimated Body Fat % From BC3 for Children Aged 10-17 Years OldEstimated Percent Body Fat for Men Aged 18-80DXA20.4 Percent body fatStandard Deviation 5.6
p-value: 0.55t-test, 2 sided
Primary

Percent Body Fat in Women Aged 18-80 Years Old

Body fat percentage was measured via Stayhealthy BC3 as well as a full body DXA

Time frame: One time point (only one day visit)

ArmMeasureGroupValue (MEAN)Dispersion
Estimated Body Fat % From BC3 for Children Aged 10-17 Years OldPercent Body Fat in Women Aged 18-80 Years OldBC332.8 Percent body fatStandard Deviation 9.1
Estimated Body Fat % From BC3 for Children Aged 10-17 Years OldPercent Body Fat in Women Aged 18-80 Years OldDXA33.4 Percent body fatStandard Deviation 9.6
p-value: 0.17t-test, 2 sided