Validation of the Stayhealthy BC1 Body Comp Analyzer

Validation of the BC1 Bioelectrical Impedance Body Comp Analyzer

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01209923
Enrollment
362
Registered
2010-09-27
Start date
2007-05-31
Completion date
2009-08-31
Last updated
2022-01-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Body Composition

Keywords

Percent body fat, Body Composition, Stayhealthy

Brief summary

The purpose of this study was to validate the Stayhealthy BC1 bioelectrical impedance analyzer to assess percent body fat in children and adults. The BC1 was compared to dual-energy x-ray absorptiometry assessment of body fat for adults and to hydrostatic weighing assessment of body fat for children. We hypothesized that the BC1 would provide similar measures of body fat between devices with acceptable limits of agreement.

Detailed description

The BC1 body composition analyzer is manufactured by Stayhealthy. It is a handheld, noninvasive bioelectrical impedance device. When held, the BC1 sends an electrical impulse through the body to determine body fat, muscle mass, and body hydration. Bio-impedance devices measure the change in electrical resistance in body tissues by using a detectable electrical signal that passes through the body. The method is based on the principle that lean body mass contains virtually all the water and conducting electrolytes in the body and provides a good electrical pathway. In contrast, fat or fat-containing tissues produce a poor electrical pathway. By inducing a low-energy, high -frequency signal, a measurement of the baseline resistance to flow of electrical current can be assessed. The resistance measurement relates directly to the volume of the conductor. Through repeated measurements and identifying other parameters (i.e. gender, weight, age and height) proprietary equations were developed that determine total body water, fat free body mass, and total body fat.

Interventions

DEVICEBC1

Body composition measures taken twice at the same visit for BIA, DEXA, and hydrostatic weighing.

Sponsors

University of Southern California
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
HEALTH_SERVICES_RESEARCH
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
10 Years to 80 Years
Healthy volunteers
Yes

Inclusion criteria

* boys and girls 10-17 yrs and men and women 18-80 yrs

Exclusion criteria

* acute illness * pregnancy * pacemaker or defibrillator

Design outcomes

Primary

MeasureTime frameDescription
Percent Body Fat2 measures were assessed during a single visitPercent body was compared between devices

Countries

United States

Participant flow

Participants by arm

ArmCount
Body Composition Testing
Body composition tested using bioelectrical impedance and the DEXA or hydrostatic weighing methods. BC1: Body composition measures taken twice at the same visit for BIA, DEXA, and hydrostatic weighing.
362
Total362

Baseline characteristics

CharacteristicBody Composition Testing
Age, Categorical
<=18 years
115 Participants
Age, Categorical
>=65 years
10 Participants
Age, Categorical
Between 18 and 65 years
237 Participants
Age, Continuous45.7 years
STANDARD_DEVIATION 17.4
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
360 participants
Sex: Female, Male
Female
192 Participants
Sex: Female, Male
Male
170 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 362
other
Total, other adverse events
0 / 362
serious
Total, serious adverse events
0 / 362

Outcome results

Primary

Percent Body Fat

Percent body was compared between devices

Time frame: 2 measures were assessed during a single visit

Population: 247 adults and 115 children

ArmMeasureGroupValue (MEAN)Dispersion
Stayhealthy BC1 Bioelectric Impedance AnalyzerPercent Body FatMen20.3 percent body fatStandard Deviation 5.3
Stayhealthy BC1 Bioelectric Impedance AnalyzerPercent Body FatWomen32.8 percent body fatStandard Deviation 9.1
Stayhealthy BC1 Bioelectric Impedance AnalyzerPercent Body FatChildren18.4 percent body fatStandard Deviation 8.4
Reference AssessmentPercent Body FatMen20.4 percent body fatStandard Deviation 5.6
Reference AssessmentPercent Body FatWomen33.4 percent body fatStandard Deviation 9.6
Reference AssessmentPercent Body FatChildren18.8 percent body fatStandard Deviation 8.3
Comparison: Children BIA was compared to hydrostatic weighing; adults were compared to DEXA.p-value: 0.001t-test, 2 sided

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026