Study of In-line Pressure Using Various Techniques of Subcutaneous Administration of Fluids Enabled by Human Recombinant Hyaluronidase (INFUSE-AT1A)

The INcreased Flow Utilizing Subcutaneously-Enabled Fluids Administration Technique 1A Study (INFUSE-AT 1A): A Randomized, Parallel Group Study to Evaluate the In-line Pressure, Flow Rate, Safety and Tolerability of Hylenex-facilitated Subcutaneous Fluid Administration In Healthy Volunteers Using Various Subcutaneous Infusion Techniques

Status
Terminated
Phases
Phase 4
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01116102
Acronym
INFUSE-AT1A
Enrollment
20
Registered
2010-05-04
Start date
2010-05-31
Completion date
2010-06-30
Last updated
2011-10-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Dehydration

Keywords

dehydration, fluid therapy, hyaluronoglucosaminidase, hyaluronidase, hypodermoclysis, clysis, subcutaneous hydration, subcutaneous rehydration, hyaluronan, rHuPH20

Brief summary

The purpose is to characterize the in-line pressure profiles associated with several infusion technique factors during subcutaneous (SC) infusion of Lactated Ringer's solution, preceded by recombinant human hyaluronidase (hylenex). The safety and tolerability of hylenex-augmented SC infusion of Lactated Ringer's solution fluid is also being evaluated.

Interventions

DRUGSubcutaneous HYLENEX (recombinant human hyaluronidase) and Lactated Ringer's solution

Subcutaneous administration of 150 U HYLENEX, followed by 1000 mL Lactated Ringer's solution

Sponsors

Halozyme Therapeutics
CollaboratorINDUSTRY
Baxter Healthcare Corporation
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 60 Years
Healthy volunteers
Yes

Inclusion criteria

* Male or female, aged 18 to 60 years, inclusive. * Body mass index 19.0 to 35.0 kg/m2 * Intact normal skin without potentially obscuring tattoos, pigmentation, or lesions on the upper back in the area intended for infusion * Free from any clinically significant abnormality on the basis of medical/medication history or physical examination * Vital signs and clinical laboratory parameters within normal range or, if outside normal range, deemed not clinically significant * Negative urine drug and alcohol screens.

Exclusion criteria

* Upper back pathology that could interfere with study outcome. * History of congestive heart failure, known coronary heart disease, active or recent pulmonary disease, or renal insufficiency. * Rales on lung auscultation. * Known allergy to hyaluronidase or any other ingredient in the formulation of hylenex recombinant * Treatment with furosemide, benzodiazepines, or phenytoin. * Pregnancy or breastfeeding. * Exposure to any experimental drug within 30 days prior to study admission, or previous participation in this study. * Any other reason which, in the investigator's opinion, would prevent the safe participation in the study.

Design outcomes

Primary

MeasureTime frameDescription
Maximum Measured In-line Fluid Pressureeach minute for the first 15 minutes of fluid infusion, every 5 minutes for the next 45 minutes of infusion, and every 15 minutes thereafter until the end of infusionMaximum fluid pressure measured (15 sec period) in delivery line during subcutaneous fluid infusion at specified time point after start of infusion

Secondary

MeasureTime frameDescription
Number of Attempts Needed for Successful Subcutaneous Catheter/Needle Placementend of catheter/needle placement
Cumulative Fluid Volume Deliveredeach minute for the first 15 minutes of fluid infusion, every 5 minutes for the next 45 minutes of infusion, and every 15 minutes thereafter until the end of infusion
Technical Challenges Encountered During Fluid Infusionat any occurence of a defined challenge or at end of infusion if no challenges occurredObserved challenges, including catheter kinking, catheter/needle dislodgement/pull-out, infusion pump alarm, other technical problems

Countries

United States

Participant flow

Participants by arm

ArmCount
24 ga Catheter, Dose Flush, Single-step Rate Scheme2
24 ga Catheter, No Dose Flush, Single-step Rate Scheme3
24 ga Catheter, Dose Flush, Up-titrated Rate Scheme3
24 ga Catheter, No Dose Flush, Up-titrated Rate Scheme3
25 ga Needle, Dose Flush, Single-step Rate Scheme2
25 ga Needle, No Dose Flush, Single-step Rate Scheme2
25 ga Needle, Dose Flush, Up-titrated Rate Scheme2
25 ga Needle, No Dose Flush, Up-titrated Rate Scheme3
Total20

Baseline characteristics

CharacteristicTotal24 ga Catheter, Dose Flush, Single-step Rate Scheme24 ga Catheter, No Dose Flush, Single-step Rate Scheme24 ga Catheter, Dose Flush, Up-titrated Rate Scheme24 ga Catheter, No Dose Flush, Up-titrated Rate Scheme25 ga Needle, Dose Flush, Single-step Rate Scheme25 ga Needle, No Dose Flush, Single-step Rate Scheme25 ga Needle, Dose Flush, Up-titrated Rate Scheme25 ga Needle, No Dose Flush, Up-titrated Rate Scheme
Age, Categorical
<=18 years
0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants
Age, Categorical
>=65 years
0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants0 Participants
Age, Categorical
Between 18 and 65 years
20 Participants2 Participants3 Participants3 Participants3 Participants2 Participants2 Participants2 Participants3 Participants
Age Continuous33.3 years45.9 years32.6 years39.6 years29.3 years35.4 years39.7 years28.2 years20.9 years
Region of Enrollment
United States
20 participants2 participants3 participants3 participants3 participants2 participants2 participants2 participants3 participants
Sex: Female, Male
Female
12 Participants2 Participants2 Participants2 Participants1 Participants1 Participants2 Participants2 Participants0 Participants
Sex: Female, Male
Male
8 Participants0 Participants1 Participants1 Participants2 Participants1 Participants0 Participants0 Participants3 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
EG004
affected / at risk
EG005
affected / at risk
EG006
affected / at risk
EG007
affected / at risk
deaths
Total, all-cause mortality
— / —— / —— / —— / —— / —— / —— / —— / —
other
Total, other adverse events
2 / 23 / 33 / 33 / 32 / 22 / 22 / 23 / 3
serious
Total, serious adverse events
0 / 20 / 30 / 30 / 30 / 20 / 20 / 20 / 3

Outcome results

Primary

Maximum Measured In-line Fluid Pressure

Maximum fluid pressure measured (15 sec period) in delivery line during subcutaneous fluid infusion at specified time point after start of infusion

Time frame: each minute for the first 15 minutes of fluid infusion, every 5 minutes for the next 45 minutes of infusion, and every 15 minutes thereafter until the end of infusion

Population: Per protocol (two subjects excluded due to technical error in in-line fluid pressure measurement)

ArmMeasureGroupValue (MEAN)
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure9 min6.5 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure270 min3.3 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure13 min6.7 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure40 min7.0 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure255 min3.1 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure20 min7.3 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure90 min4.1 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure75 min4.6 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure225 min3.4 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure50 min6.4 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure195 min3.8 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure10 min6.5 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure60 min5.3 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure11 min6.5 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure12 min6.7 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure55 min5.3 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure5 min6.1 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure14 min6.6 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure180 min3.1 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure6 min6.2 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure285 min2.9 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure2 min6.8 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure3 min6.7 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure25 min6.8 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure15 min7.3 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure165 min3.1 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure30 min5.8 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure210 min3.6 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure150 min2.9 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure135 min4.0 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure45 min6.6 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure7 min6.2 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure4 min6.6 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure120 min3.4 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure35 min6.5 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure240 min3.2 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure8 min6.4 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure105 min5.0 psi
24 ga Catheter, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure1 min6.9 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure25 min2.4 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure2 min2.4 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure255 min4.0 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure1 min3.0 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure270 min4.4 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure285 min4.9 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure20 min2.2 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure15 min4.0 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure30 min2.2 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure14 min3.0 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure35 min2.7 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure45 min2.3 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure13 min3.6 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure40 min2.2 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure12 min4.2 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure50 min2.2 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure55 min2.4 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure11 min3.7 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure60 min3.1 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure10 min3.9 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure75 min2.6 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure9 min3.8 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure90 min3.6 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure105 min2.7 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure8 min3.6 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure120 min2.6 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure7 min3.2 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure135 min2.2 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure6 min3.3 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure165 min2.6 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure150 min7.8 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure180 min2.1 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure5 min2.6 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure195 min4.0 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure4 min2.1 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure210 min4.0 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure3 min2.1 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure225 min4.3 psi
24 ga Catheter, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure240 min3.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure13 min7.7 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure285 min1.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure15 min3.4 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure5 min9.1 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure7 min8.8 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure3 min7.4 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure180 min2.1 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure14 min3.5 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure150 min2.0 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure255 min8.5 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure90 min2.1 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure8 min9.0 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure135 min2.6 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure35 min2.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure195 min2.5 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure210 min2.2 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure11 min11.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure30 min8.4 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure270 min8.8 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure55 min2.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure40 min2.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure6 min8.0 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure4 min7.3 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure45 min2.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure240 min2.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure105 min2.2 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure2 min7.8 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure60 min2.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure225 min1.9 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure165 min2.6 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure10 min7.7 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure1 min6.8 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure50 min3.0 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure12 min13.6 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure25 min3.0 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure75 min2.3 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure120 min2.5 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure20 min3.3 psi
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure9 min8.2 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure9 min7.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure1 min5.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure12 min10.3 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure90 min4.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure8 min7.3 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure35 min10.9 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure210 min4.8 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure105 min4.5 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure15 min15.8 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure40 min11.5 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure7 min7.1 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure13 min10.5 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure3 min6.1 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure120 min5.2 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure30 min8.6 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure240 min7.4 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure135 min5.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure150 min5.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure6 min7.1 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure14 min17.7 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure165 min5.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure225 min4.9 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure5 min5.7 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure285 min4.4 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure255 min4.8 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure180 min4.8 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure25 min8.7 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure4 min6.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure270 min5.0 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure2 min5.9 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure195 min5.2 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure11 min7.6 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure55 min7.9 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure45 min7.4 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure10 min6.2 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure50 min7.8 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure60 min7.7 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure20 min11.9 psi
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure75 min5.9 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure6 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure45 min4.3 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure135 min3.8 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure25 min4.8 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure30 min4.8 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure35 min4.5 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure50 min4.5 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure55 min4.4 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure60 min3.7 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure75 min3.6 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure90 min3.5 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure105 min4.0 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure120 min4.3 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure150 min3.7 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure165 min4.6 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure180 min3.3 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure195 min3.8 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure210 min3.9 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure225 min3.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure240 min3.4 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure255 min3.8 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure270 min3.5 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure285 min2.8 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure1 min3.8 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure2 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure3 min4.2 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure4 min4.3 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure5 min4.4 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure40 min4.6 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure7 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure8 min4.2 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure9 min4.2 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure10 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure11 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure12 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure13 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure14 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure15 min4.1 psi
25 ga Needle, Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure20 min4.1 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure6 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure210 min2.8 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure14 min3.3 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure225 min2.4 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure120 min2.9 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure3 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure165 min2.6 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure7 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure180 min2.6 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure45 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure285 min2.5 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure105 min2.7 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure25 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure35 min3.3 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure2 min3.1 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure8 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure5 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure40 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure90 min2.7 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure195 min2.6 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure12 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure75 min2.8 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure135 min2.7 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure9 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure270 min2.7 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure30 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure13 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure150 min2.6 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure60 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure4 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure50 min3.3 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure10 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure255 min2.5 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure240 min2.6 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure15 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure55 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure20 min3.2 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure1 min2.8 psi
25 ga Needle, No Dose Flush, Single-step Rate SchemeMaximum Measured In-line Fluid Pressure11 min3.2 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure40 min3.4 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure11 min4.0 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure45 min3.4 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure12 min3.8 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure165 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure20 min3.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure35 min4.2 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure5 min1.1 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure30 min3.8 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure240 min2.5 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure14 min3.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure270 min2.5 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure1 min1.1 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure285 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure210 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure13 min3.9 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure150 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure6 min2.2 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure255 min2.9 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure25 min3.7 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure225 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure120 min2.8 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure15 min3.7 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure7 min2.1 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure3 min1.2 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure105 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure90 min2.7 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure8 min2.1 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure195 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure75 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure9 min2.2 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure2 min1.1 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure135 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure60 min3.4 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure4 min1.0 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure10 min2.2 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure55 min3.4 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure180 min2.6 psi
25 ga Needle, Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure50 min3.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure14 min6.9 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure1 min3.0 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure225 min3.3 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure2 min2.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure210 min6.3 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure135 min7.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure3 min2.5 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure195 min3.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure180 min3.2 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure4 min2.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure165 min3.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure5 min2.5 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure150 min7.0 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure6 min2.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure120 min6.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure105 min6.3 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure7 min3.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure90 min7.5 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure15 min6.5 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure8 min4.6 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure75 min7.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure9 min4.1 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure60 min6.5 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure55 min6.0 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure40 min5.9 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure10 min3.9 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure50 min5.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure11 min6.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure45 min5.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure12 min7.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure35 min6.4 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure30 min5.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure20 min6.2 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure25 min6.1 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure270 min4.8 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure285 min4.3 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure13 min7.0 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure255 min4.9 psi
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeMaximum Measured In-line Fluid Pressure240 min5.3 psi
Secondary

Cumulative Fluid Volume Delivered

Time frame: each minute for the first 15 minutes of fluid infusion, every 5 minutes for the next 45 minutes of infusion, and every 15 minutes thereafter until the end of infusion

Population: Per Protocol (two subjects excluded due to technical errors in fluid pressure measurement)

ArmMeasureGroupValue (MEAN)
24 ga Catheter, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered60 min250 mL
24 ga Catheter, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered240 min848 mL
24 ga Catheter, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered10 min41.2 mL
24 ga Catheter, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered285 min998 mL
24 ga Catheter, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered180 min649 mL
24 ga Catheter, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered5 min20.8 mL
24 ga Catheter, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered120 min448 mL
24 ga Catheter, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered10 min41.3 mL
24 ga Catheter, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered240 min850 mL
24 ga Catheter, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered120 min450 mL
24 ga Catheter, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered180 min650 mL
24 ga Catheter, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered285 min1000 mL
24 ga Catheter, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered5 min21.4 mL
24 ga Catheter, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered60 min249 mL
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered5 min4.2 mL
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered180 min649 mL
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered240 min849 mL
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered285 min999 mL
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered120 min449 mL
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered10 min11.7 mL
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered60 min249 mL
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered120 min447 mL
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered5 min3.8 mL
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered285 min997 mL
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered10 min11.5 mL
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered240 min848 mL
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered60 min240 mL
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered180 min647 mL
25 ga Needle, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered120 min449 mL
25 ga Needle, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered5 min21.1 mL
25 ga Needle, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered10 min42.0 mL
25 ga Needle, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered60 min249 mL
25 ga Needle, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered180 min649 mL
25 ga Needle, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered240 min849 mL
25 ga Needle, Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered285 min999 mL
25 ga Needle, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered120 min447 mL
25 ga Needle, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered285 min997 mL
25 ga Needle, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered180 min647 mL
25 ga Needle, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered60 min249 mL
25 ga Needle, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered240 min847 mL
25 ga Needle, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered10 min41.2 mL
25 ga Needle, No Dose Flush, Single-step Rate SchemeCumulative Fluid Volume Delivered5 min20.7 mL
25 ga Needle, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered10 min12.1 mL
25 ga Needle, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered5 min4.1 mL
25 ga Needle, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered240 min847 mL
25 ga Needle, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered180 min650 mL
25 ga Needle, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered285 min997 mL
25 ga Needle, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered60 min248 mL
25 ga Needle, Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered120 min448 mL
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered285 min996 mL
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered120 min446 mL
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered240 min846 mL
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered5 min4.1 mL
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered60 min246 mL
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered180 min646 mL
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeCumulative Fluid Volume Delivered10 min11.9 mL
Secondary

Number of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement

Time frame: end of catheter/needle placement

Population: Per Protocol (two subjects excluded due to technical error in in-line fluid pressure measurements)

ArmMeasureGroupValue (NUMBER)
24 ga Catheter, Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts0 Subjects
24 ga Catheter, Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt2 Subjects
24 ga Catheter, No Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts0 Subjects
24 ga Catheter, No Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt3 Subjects
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts0 Subjects
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt2 Subjects
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts1 Subjects
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt2 Subjects
25 ga Needle, Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt2 Subjects
25 ga Needle, Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts0 Subjects
25 ga Needle, No Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt1 Subjects
25 ga Needle, No Dose Flush, Single-step Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts0 Subjects
25 ga Needle, Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt2 Subjects
25 ga Needle, Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts0 Subjects
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement2 attempts0 Subjects
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeNumber of Attempts Needed for Successful Subcutaneous Catheter/Needle Placement1 attempt3 Subjects
Secondary

Technical Challenges Encountered During Fluid Infusion

Observed challenges, including catheter kinking, catheter/needle dislodgement/pull-out, infusion pump alarm, other technical problems

Time frame: at any occurence of a defined challenge or at end of infusion if no challenges occurred

Population: Per Protocol (two subjects excluded due to technical error in in-line fluid pressure measurement)

ArmMeasureGroupValue (NUMBER)
24 ga Catheter, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge0 participants
24 ga Catheter, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking0 participants
24 ga Catheter, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm0 participants
24 ga Catheter, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion0 participants
24 ga Catheter, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion0 participants
24 ga Catheter, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking0 participants
24 ga Catheter, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge0 participants
24 ga Catheter, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm0 participants
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge0 participants
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion0 participants
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking0 participants
24 ga Catheter, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm0 participants
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge2 participants
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking1 participants
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm2 participants
24 ga Catheter, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion1 participants
25 ga Needle, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion0 participants
25 ga Needle, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm0 participants
25 ga Needle, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking0 participants
25 ga Needle, Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge0 participants
25 ga Needle, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion0 participants
25 ga Needle, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm0 participants
25 ga Needle, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking0 participants
25 ga Needle, No Dose Flush, Single-step Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge0 participants
25 ga Needle, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion0 participants
25 ga Needle, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge0 participants
25 ga Needle, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm0 participants
25 ga Needle, Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking0 participants
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionCatheter Kinking0 participants
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionOcclusion Alarm0 participants
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionIntervention Needed to Secure/Maintain Infusion0 participants
25 ga Needle, No Dose Flush, Up-titrated Rate SchemeTechnical Challenges Encountered During Fluid InfusionAny Technical Challenge0 participants