An Inpatient Performance Evaluation of a New Subcutaneous Glucose Sensor

An Inpatient Performance Evaluation of a New Subcutaneous Glucose Sensor

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01112696
Enrollment
100
Registered
2010-04-28
Start date
2010-04-30
Completion date
2010-11-30
Last updated
2018-01-16

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Type 1 Diabetes Mellitus, Type 2 Diabetes Mellitus

Keywords

diabetes, glucose sensor, CGM

Brief summary

The purpose of the study is to assess the performance of a new subcutaneous glucose sensor over a seven-day sensor life when used with currently marketed Medtronic Diabetes devices. In addition performance will be calculated for use of the new sensor with proposed new devices using new calibration algorithms.

Detailed description

The first generation Medtronic MiniMed Subcutaneous Glucose Sensor (Sof-Sensor) was originally approved by the FDA for commercialization as part of the Continuous Glucose Monitoring System (CGMS) on June 15, 1999 (PMA 980022). The Sensor is composed of a microelectrode with a thin coating of glucose oxidase beneath several layers of biocompatible membrane. This same sensor was used as part of subsequent continuous glucose monitoring (CGM) systems, such as the Guardian REAL-Time and Paradigm REAL-Time sensor augmented insulin pump. The current configuration of sensor has undergone in vitro and in vivo testing. A new second-generation glucose sensor (herein referred to as the Comfort Sensor) has been developed. The first-generation glucose sensor was approved with reported sensor accuracy (MAD) of 18%; it was labeled for a maximum use duration of 72 hours, using only the abdomen as an insertion site. The new sensor is shorter and has a smaller diameter, with a smaller gauge introducer needle. The new sensor inserter is designed to be used with the new sensor. The objectives of this study are to 1) Assess performance of the Comfort Sensor when used over a period of seven days with currently available devices, and 2) Assess performance of the Comfort Sensor when used over a period of seven days using new calibration algorithms (post-processed with algorithm for future devices). Accuracy data will be calculated based on comparing calibrated glucose sensor values to a gold standard (YSI plasma glucose values) in adult subjects during in-clinic testing, and on comparing calibrated glucose sensor values to glucose meter values during in-home testing. The devices that will be used for gathering sensor data during this study will be: 1) the Comfort Sensor, 2) the MiniLink (Transmitter), 3) the Guardian REAL-Time Display Device and 4) the CGMS iPro.

Interventions

All subjects to wear sensors

Sponsors

Medtronic Diabetes
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 75 Years
Healthy volunteers
No

Inclusion criteria

1. Males and females aged 18 through 75 inclusive 2. Diagnosed with Type 1 or type 2 Diabetes Mellitus, using insulin to treat their diabetes (Each site's and overall study enrollment of subjects with Type 2 diabetes will target 20% of the total number) 3. Willingness to perform required study and data collection procedures and to adhere to operating requirements of the Guardian REAL-TIME and CGMS iPro Systems 4. Willingness to perform at least 4 capillary blood glucose tests per day while wearing the Guardian REAL-TIME and CGMS iPro Systems 5. Willingness to participate in a 10 hour frequent blood sampling session during the study 6. Subject agrees to comply with the study protocol requirements 7. Informed Consent, HIPAA Authorization, and California Experimental Subject Bill of Rights (if applicable) signed by the subject 8. The Subject is willing to wear both the Guardian REAL-Time and CGMS iPro Systems for 14 days (\ 340 hours)

Exclusion criteria

1. The Subject has a history of tape allergies that have not been resolved 2. The Subject has any skin abnormality (e.g. psoriasis, rash, staphylococcus infection) that has not been resolved and would inhibit them from wearing the sensors 3. Any additional condition(s) that in the Investigator's opinion would warrant exclusion from the study or prevent the Subject from completing the study 4. Subject is currently participating in an investigational study (drug or device)

Design outcomes

Primary

MeasureTime frameDescription
Glucose Sensor Accuracy When Compared to Laboratory Standard (YSI): Proportion of Glucose Sensor Readings That Met Accuracy CriteriaDays one through six of sensor useThe primary accuracy parameter (primary effectiveness endpoint) was the comparative readings of paired sensor and YSI glucose readings, measured on days 1 through 6. Accuracy is defined as within 20% agreement between YSI and paired sensor (within 20 mg/dL if YSI \<80 mg/dL). Accuracy ranges from 0 - 100, with higher number suggests better accuracy.

Secondary

MeasureTime frameDescription
Device Related Moderate or Device Related Severe Adverse Eventsdays one through six of sensor wearDevice related moderate adverse event: low level of inconvenience or concern to the subject and may interfere with daily activities but is usually improved by simple therapeutic remedy. Device related severe adverse event: interrupts a subject's daily activity and typically requires intervening treatment. Note: device related determination is made by the site that there is a reasonable possibility that the adverse event may have been caused by the device.

Countries

United States

Participant flow

Participants by arm

ArmCount
Group 1
Sensor Users (All Subjects)
100
Total100

Baseline characteristics

CharacteristicGroup 1
Age, Categorical
<=18 years
1 Participants
Age, Categorical
>=65 years
7 Participants
Age, Categorical
Between 18 and 65 years
92 Participants
Age, Continuous42.2 years
STANDARD_DEVIATION 15
Region of Enrollment
United States
100 participants
Sex: Female, Male
Female
31 Participants
Sex: Female, Male
Male
69 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 100
other
Total, other adverse events
0 / 100
serious
Total, serious adverse events
0 / 100

Outcome results

Primary

Glucose Sensor Accuracy When Compared to Laboratory Standard (YSI): Proportion of Glucose Sensor Readings That Met Accuracy Criteria

The primary accuracy parameter (primary effectiveness endpoint) was the comparative readings of paired sensor and YSI glucose readings, measured on days 1 through 6. Accuracy is defined as within 20% agreement between YSI and paired sensor (within 20 mg/dL if YSI \<80 mg/dL). Accuracy ranges from 0 - 100, with higher number suggests better accuracy.

Time frame: Days one through six of sensor use

Population: 98 subjects of 100 enrolled subjects (a total of 5857 paired sensor and YSI readings) completed participation in the inpatient frequent blood sampling procedure.

ArmMeasureValue (NUMBER)
All Completed SubjectsGlucose Sensor Accuracy When Compared to Laboratory Standard (YSI): Proportion of Glucose Sensor Readings That Met Accuracy Criteria79.45 paired sensor and YSI glucose readings
Secondary

Device Related Moderate or Device Related Severe Adverse Events

Device related moderate adverse event: low level of inconvenience or concern to the subject and may interfere with daily activities but is usually improved by simple therapeutic remedy. Device related severe adverse event: interrupts a subject's daily activity and typically requires intervening treatment. Note: device related determination is made by the site that there is a reasonable possibility that the adverse event may have been caused by the device.

Time frame: days one through six of sensor wear

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
All Completed SubjectsDevice Related Moderate or Device Related Severe Adverse Events0 Participants