Esteem Totally Implantable Hearing System

Esteem Totally Implantable Hearing System Clinical Trial U.S. Pivotal Trial

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01092910
Enrollment
57
Registered
2010-03-25
Start date
2008-01-31
Completion date
2010-08-31
Last updated
2020-01-02

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Mild to Severe Sensorineural Hearing Loss

Keywords

Sensorineural, Hearing Loss

Brief summary

To evaluate the safety and efficacy of the Esteem Totally Implantable Hearing System in subjects suffering from mild to severe hearing loss.

Detailed description

Title: Esteem® Totally Implantable Hearing System Clinical Trial Purpose: To evaluate the safety and efficacy of the Esteem® Totally Implantable Hearing System (hereafter called the Esteem® System) in subjects suffering from mild to severe hearing loss. Trial Design: This pivotal trial is designed as a prospective, multi-center, non-randomized, clinical trial to evaluate the safety and efficacy of the Esteem® System. For this trial the subject will act as his or her own control. This trial has been designed to meet the United States' regulatory requirements. Enrollment Size: It is currently estimated that 50-60 subjects will be enrolled in this trial. Subject Population: Adult subjects who have mild to severe sensorineural hearing loss, have a healthy middle ear, have a speech discrimination score of equal to or better than 40%, and have previously worn a hearing aid for a minimum of 3 months with an average daily use rate of at least 4 hours are eligible for inclusion in the trial. See detailed inclusion/exclusion criteria for specifics. Primary Objectives: 1. To demonstrate that the Esteem® System improves the speech threshold of sensitivity for hearing and identifying speech signals as well as or better than the pre-implant hearing aid (aided condition). Endpoint: Comparison of the speech reception threshold (SRT) using the Esteem® System (4 months post activation) as compared to the pre-implant aided condition. 2. To demonstrate that the Esteem® System at the 4 months post- activation visit is as effective as or better than the hearing aid for improving speech discrimination (intelligibility) as shown by the word recognition score at 50 dB. Endpoint: Comparison of the word recognition score using the Esteem® at 4 months post-activation compared to the pre-implant aided condition. 3. To determine the incidence of Serious Adverse Device Effects (SADE) and the incidence rate of device failures and replacements. Endpoint: The analysis of the incidence of SADEs and device failures and replacements at each follow-up. 4. To demonstrate that the subjects' cochlear function remains unchanged with the Esteem® System as shown by comparison of the subjects' pre-implant baseline bone conduction threshold versus the subjects' 4 month post-activation visit bone conduction threshold. Endpoint: Comparison of bone conduction threshold (BC) using forehead placement at 4 months post activation compared to the pre-implant BC threshold. Secondary Objectives: 1. To demonstrate that the Esteem® System at the 4 months post activation visit improves the 3-frequency (500, 1000, and 2000 Hz) pure tone average (PTA) when compared to the baseline unaided condition. 2. To demonstrate that the Esteem® System at the 4 months post-activation visit is as effective as or better than the hearing aid for improving speech discrimination (intelligibility) as shown by the QuickSIN (speech in noise) test results. 3. To show that the Esteem® System improves Quality-of-Life when compared to the baseline aided condition as shown by APHAB results. 4. To gather subject feedback and comments on the use of the Esteem® System relative to the pre-implant hearing aid (aided condition) as shown by the Esteem® Questionnaire.

Interventions

DEVICEEsteem Totally Implantable Hearing System

The Esteem System is a totally implantable hearing system designed to improve hearing in adult subjects suffering from mild to severe hearing loss that is sensorineural in origin.

Sponsors

Envoy Medical Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Subject is 18 years old * Subject understands the nature of the procedure and has signed the Subject Informed Consent Form prior to the procedure * Subject is willing and able to comply with specified follow-up evaluations and understands the audiological test procedures and use of the Esteem System. * Subject has mild to severe sensorineural hearing loss between 500 and 4000 Hz in the ear to be implanted with pure tone air-conduction threshold levels within the limits of a Hearing Aid (HA) as follow: Freq (Hz) 500 1000 2000 3000 4000 LL\* (dB HL) 30 35 35 35 35 UL\* (dB HL) 100 100 100 100 100 \*LL = Lower Level; UL = Upper Level * Subject's air-bone gap is no greater than 10 dB at 4 of the 5 following frequencies: 500, 1000, 2000, 3000 and 4000 Hz. * Subject has an unaided maximum word recognition score of greater than or equal to 60% with recorded delivery using a phonetically balanced word list at SRT + 40 dB or at maximum tolerable presentation level. * Subject is a current user of a properly functioning and appropriately fit hearing aid. This is defined as the subject has used this aid for at least four (4) hours (average) per day (in the ear to be implanted) for at least three (3) months for a new aid or one (1) month for an adjusted aid. * Subject's hearing aid, in the ear to be implanted, shall appropriately fit optimally. * Subject has normally functioning eustachian tube * Subject has normal tympanic membrane * Subject has a normal middle ear anatomy * Subject has adequate space for Esteem System implant determined via fine cut temporal bone CT scan * Subject is a native speaker of the English language. * Subject is a hearing aid user in the ear to be implanted.

Exclusion criteria

* Subject has a history of post-adolescent chronic middle ear infections, inner ear disorders or recurring vertigo requiring treatment, disorders such as mastoiditis, Hydrops or Meniere's syndrome or disease * Subject has a history of otitis externa or eczema for the outer ear canal and the investigator believes this will affect the Esteem System implantation * Subject has cholesteatoma or destructive middle ear disease * Subject has life expectancy of two (2) years due to other medical conditions * Subject has retrocochlear or central auditory disorders * Subject is known to be suffering from any psychological, developmental, physical, or emotional disorder that the investigator feels would interfere with the surgery or follow-up testing * Subject has a known history of fluctuating air conduction and/or bone conduction hearing loss over a one-year period of 15 dB in either direction at 2 or more frequencies (from 500 - 4000 Hz) * Subject has sudden hearing loss due to unknown cause * Subject has a history of disabling tinnitus, defined as tinnitus which required treatment. * Subject is unable to adequately perform audiological testing * Subject has a medical condition or undergoing a treatment that may affect healing and the investigator does not believe the subject is a good candidate for the trial. * Subject has diabetes that is not well controlled with medication or diet and the investigator does not believe in his best medical judgment that the subject would be a good candidate for the trial * Subject is pregnant at the time of device implant * Subject has a history of keloid formation * Subject has known hypersensitivity to silicone rubber, polyurethane, stainless steel, titanium and/or gold

Design outcomes

Primary

MeasureTime frameDescription
Assessment of Cochlear Function at 4-months and 10-months Post-activation, as Evidenced by Bone-conduction (BC) Thresholds.Baseline, 4-, and 10-Months Post-ActivationComparison of bone conduction thresholds at the 4-month and 10-month post-activation follow-up, relative to the pre-implant baseline bone conduction thresholds. Lower thresholds represent better (more sensitive) outcomes.
SRT Change4 Months Post ActivationComparison of the speech reception threshold (SRT) using the Esteem System as compared to the pre-implant aided condition
Word Recognition Score Change4 Months Post ActivationComparison of the word recognition score using the Esteem compared to the pre-implant aided condition
SADEs10 Months Post-ActivationThe incidence of Serious Adverse Device Effects (SADE) and the incidence rate of device failures and replacements

Secondary

MeasureTime frameDescription
Comparisons of Change in Global Score on Abbreviated Profile of Hearing Aid Benefit (APHAB) at 4 Months and at 10 Months Post-activation, Relative to Pre-implant Baseline Score4 and 10 Months Post-ActivationTo assess whether the Esteem System improves Quality-of-Life, when compared to the baseline aided condition (pre-implant) as shown by the global Abbreviated Profile of Hearing Aid Benefit (APHAB) score. The APHAB is a 24-item self-assessment inventory in which patients report the amount of trouble they are having with communication in various everyday situations using a scale ranging from 0 to 100, with higher scores indicating less difficulty (or increased benefit). The Global Score is the average of 3 subscales. Scores reported here are calculated by comparing the patient's reported difficulty in the baseline unaided condition with their amount of difficulty when using amplification (hearing aids or implant) -- i.e. Global Scores at pre-implant baseline subtracted from scores at aided baseline and at 4 and 10 months post-implant. Positive numbers denote increased benefit relative to unaided baseline, and the higher the number, the greater the benefit.
Scores on Esteem Questionnaire4 and 10 Months Post ActivationTo gain subject feedback on the use of the Esteem System relative to their pre-implant hearing aid (aided condition) as shown by the (unvalidated) Esteem Questionnaire. Subjects completed a 7-item questionnaire rating various subjective attributes concerning their experience with Esteem as compared to their previous hearing aid. Ratings were on a scale of 1 to 5, where 1 is much worse, 3 is the same and 5 is much better. Reported here are proportions of participants responding to each score (1 to 5) on the scale, for average scores across all 7 items for each subject.
Change in Pure Tone Average (PTA) at 4 Months, Relative to Baseline Pre-implant4 Months Post ActivationFor each subject, the 4-month post-activation Air Conduction PTA (average of thresholds at three frequencies: 500, 1000, and 2000 Hz) was compared to the baseline unaided PTA. A negative mean difference (in dB) from baseline indicates improved hearing sensitivity with the Esteem, relative to baseline.
Difference in QuickSIN (Quick Speech-In-Noise Test) Score at 4 Months Relative to Baseline Aided Condition4 Months Post ActivationThe Quick-SIN is a test of sentence recognition in varying levels of background noise. The score achieved is termed SNR loss, and a higher score indicates poorer performance on the test. SNR Loss is defined as the dB increase in signal-to-noise ratio required by a hearing-impaired person to understand speech in noise, compared to someone with normal hearing. (The range of possible scores is -4.5 to 25.5, with lower scores indicating better performance.) For this study, baseline scores for each subject are subtracted from 4 month scores, providing a difference score. The mean difference across subjects is reported here, with 0 meaning no change and a negative difference indicating better performance with Esteem, compared to the pre-implant aided condition.

Countries

United States

Participant flow

Recruitment details

Subjects were recruited at three different sites: The Ear Center of Greensboro Greensboro, NC 27401 Shohet Ear Associates Medical Group, Inc. Newport Beach, CA 92663 Lahey Clinic Burlington, MA 01805

Participants by arm

ArmCount
Esteem Implant
Subjects implanted with the Esteem Totally Implantable Hearing System
57
Total57

Baseline characteristics

CharacteristicEsteem Implant
Age, Continuous52.9 years
STANDARD_DEVIATION 15.8
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
57 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
1 Participants
Race (NIH/OMB)
Black or African American
0 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
56 Participants
Sex: Female, Male
Female
19 Participants
Sex: Female, Male
Male
38 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
43 / 57
serious
Total, serious adverse events
6 / 57

Outcome results

Primary

Assessment of Cochlear Function at 4-months and 10-months Post-activation, as Evidenced by Bone-conduction (BC) Thresholds.

Comparison of bone conduction thresholds at the 4-month and 10-month post-activation follow-up, relative to the pre-implant baseline bone conduction thresholds. Lower thresholds represent better (more sensitive) outcomes.

Time frame: Baseline, 4-, and 10-Months Post-Activation

Population: 4-month and 10-month BC PTA (average 500, 1kHz, 2kHz) compared to baseline BC PTA

ArmMeasureGroupValue (MEAN)Dispersion
Esteem ImplantAssessment of Cochlear Function at 4-months and 10-months Post-activation, as Evidenced by Bone-conduction (BC) Thresholds.Baseline bone conduction PTA55.6 dB HLStandard Error 1.3
Esteem ImplantAssessment of Cochlear Function at 4-months and 10-months Post-activation, as Evidenced by Bone-conduction (BC) Thresholds.4 mo. bone conduction PTA55.6 dB HLStandard Error 1.2
Esteem ImplantAssessment of Cochlear Function at 4-months and 10-months Post-activation, as Evidenced by Bone-conduction (BC) Thresholds.10 mo. bone conduction PTA55.3 dB HLStandard Error 1.4
Primary

SADEs

The incidence of Serious Adverse Device Effects (SADE) and the incidence rate of device failures and replacements

Time frame: 10 Months Post-Activation

Population: Includes all events, cumulatively, from enrollment through 10-month followup

ArmMeasureValue (NUMBER)
Esteem ImplantSADEs6 events
Primary

SRT Change

Comparison of the speech reception threshold (SRT) using the Esteem System as compared to the pre-implant aided condition

Time frame: 4 Months Post Activation

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantSRT Change10.6 dBStandard Error 1.6
Primary

SRT Change

Comparison of the speech reception threshold (SRT) using the Esteem System as compared to the pre-implant aided condition

Time frame: 10 Months Post-Activation

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantSRT Change11.4 dBStandard Error 1.8
Primary

Word Recognition Score Change

Comparison of the word recognition score using the Esteem compared to the pre-implant aided condition

Time frame: 10 Months Post Activation

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantWord Recognition Score Change19.8 change in % correctStandard Error 4.3
Primary

Word Recognition Score Change

Comparison of the word recognition score using the Esteem compared to the pre-implant aided condition

Time frame: 4 Months Post Activation

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantWord Recognition Score Change21.7 change in % correctStandard Error 4.2
Secondary

Change in Pure Tone Average (PTA) at 4 Months, Relative to Baseline Pre-implant

For each subject, the 4-month post-activation Air Conduction PTA (average of thresholds at three frequencies: 500, 1000, and 2000 Hz) was compared to the baseline unaided PTA. A negative mean difference (in dB) from baseline indicates improved hearing sensitivity with the Esteem, relative to baseline.

Time frame: 4 Months Post Activation

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantChange in Pure Tone Average (PTA) at 4 Months, Relative to Baseline Pre-implant-27 dBStandard Error 1
Secondary

Comparisons of Change in Global Score on Abbreviated Profile of Hearing Aid Benefit (APHAB) at 4 Months and at 10 Months Post-activation, Relative to Pre-implant Baseline Score

To assess whether the Esteem System improves Quality-of-Life, when compared to the baseline aided condition (pre-implant) as shown by the global Abbreviated Profile of Hearing Aid Benefit (APHAB) score. The APHAB is a 24-item self-assessment inventory in which patients report the amount of trouble they are having with communication in various everyday situations using a scale ranging from 0 to 100, with higher scores indicating less difficulty (or increased benefit). The Global Score is the average of 3 subscales. Scores reported here are calculated by comparing the patient's reported difficulty in the baseline unaided condition with their amount of difficulty when using amplification (hearing aids or implant) -- i.e. Global Scores at pre-implant baseline subtracted from scores at aided baseline and at 4 and 10 months post-implant. Positive numbers denote increased benefit relative to unaided baseline, and the higher the number, the greater the benefit.

Time frame: 4 and 10 Months Post-Activation

Population: Data from entire subject cohort not available.

ArmMeasureGroupValue (MEAN)Dispersion
Esteem ImplantComparisons of Change in Global Score on Abbreviated Profile of Hearing Aid Benefit (APHAB) at 4 Months and at 10 Months Post-activation, Relative to Pre-implant Baseline ScoreBaseline Aided Global Score35.0 score on a scaleStandard Error 2.2
Esteem ImplantComparisons of Change in Global Score on Abbreviated Profile of Hearing Aid Benefit (APHAB) at 4 Months and at 10 Months Post-activation, Relative to Pre-implant Baseline Score4 mo.Global Score with Esteem47.9 score on a scaleStandard Error 3.1
Esteem ImplantComparisons of Change in Global Score on Abbreviated Profile of Hearing Aid Benefit (APHAB) at 4 Months and at 10 Months Post-activation, Relative to Pre-implant Baseline Score10 mo. Global Score with Esteem46.6 score on a scaleStandard Error 3
Secondary

Difference in QuickSIN (Quick Speech-In-Noise Test) Score at 4 Months Relative to Baseline Aided Condition

The Quick-SIN is a test of sentence recognition in varying levels of background noise. The score achieved is termed SNR loss, and a higher score indicates poorer performance on the test. SNR Loss is defined as the dB increase in signal-to-noise ratio required by a hearing-impaired person to understand speech in noise, compared to someone with normal hearing. (The range of possible scores is -4.5 to 25.5, with lower scores indicating better performance.) For this study, baseline scores for each subject are subtracted from 4 month scores, providing a difference score. The mean difference across subjects is reported here, with 0 meaning no change and a negative difference indicating better performance with Esteem, compared to the pre-implant aided condition.

Time frame: 4 Months Post Activation

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantDifference in QuickSIN (Quick Speech-In-Noise Test) Score at 4 Months Relative to Baseline Aided Condition-1 score on a scaleStandard Error 1
Secondary

Scores on Esteem Questionnaire

To gain subject feedback on the use of the Esteem System relative to their pre-implant hearing aid (aided condition) as shown by the (unvalidated) Esteem Questionnaire. Subjects completed a 7-item questionnaire rating various subjective attributes concerning their experience with Esteem as compared to their previous hearing aid. Ratings were on a scale of 1 to 5, where 1 is much worse, 3 is the same and 5 is much better. Reported here are proportions of participants responding to each score (1 to 5) on the scale, for average scores across all 7 items for each subject.

Time frame: 4 and 10 Months Post Activation

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Esteem ImplantScores on Esteem QuestionnaireResponse 1 - 4 mo average across items2 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 2 - 4 mo5 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 3 - 4 mo4 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 4 - 4 mo15 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 5 - 4 mo23 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 1 - 10 mo ave across items2 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 2 - 10 mo3 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 3 - 10 mo5 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 4 - 10 mo12 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse 5 - 10 mo24 Participants
Esteem ImplantScores on Esteem QuestionnaireResponse Missing - 10 mo3 Participants

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026