Caverject User Study

Representative Users Study Of Operating Characteristics Of The Caverject Delivery System.

Status
Completed
Phases
Phase 3
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT01008605
Enrollment
48
Registered
2009-11-06
Start date
2011-02-28
Completion date
2011-02-28
Last updated
2012-04-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Erectile Dysfunction

Keywords

Erectile Dysfunction

Brief summary

The purpose of this study is to demonstrate the usability of the system.

Detailed description

demonstrate usability

Interventions

syringe

Sponsors

Pfizer
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Masking
NONE

Eligibility

Sex/Gender
MALE
Age
40 Years to 70 Years
Healthy volunteers
Yes

Inclusion criteria

Age 40 to 70

Exclusion criteria

Prior syringe skills

Design outcomes

Primary

MeasureTime frameDescription
Percentage of Participants Who Successfully Operated the Caverject Impulse Delivery SystemDay 1Percentage of participants who were able to successfully expel the selected dose from the Caverject Impulse Delivery System when relying on the modified Instructions for Use. The process was considered successful if the lower bound of the 95% confidence interval (CI) was more than (\>) 80% overall.

Secondary

MeasureTime frameDescription
Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 2Day 1Participant Assessment Tool, Question 2: Instructions provided were clear? Participant responses were reported as follows: Very Clear, Somewhat Clear, Not Very Clear, Not Clear At All.
Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 3Day 1Participant Assessment Tool, Question 3: Most difficult step? Participant responses were reported as follows: No Steps Particularly Difficult, Attaching Needle, Mixing Solution, Getting The Air Out Of Syringe, Dialing Dose, Pushing Plunger, Other.
Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 1Day 1Participant Assessment Tool, Question 1: Instructions provided were useful? Participant responses were reported as follows: Very Useful, Somewhat Useful, Not Very Useful, Not Useful At All.
Time Required to Perform Each Step While Using the Caverject Impulse Delivery SystemDay 1Steps involved while using the Caverject Impulse Delivery System included assembly, mixing the dose, de-aeration, setting the dose, and injecting the dose.
Number of Participants Providing Comments to Any Question on the Participant Assessment ToolDay 1Number of participants providing comments on questions in the Participant Assessment Tool. Questions were as follows: were instructions clear, were instructions useful, which was the most difficult step, and was the syringe easy to use.
Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 4Day 1Participant Assessment Tool, Question 4: Syringe easy to use? Participant responses were reported as follows: Very Easy, Somewhat Easy, Somewhat Difficult, Very Difficult

Countries

United States

Participant flow

Participants by arm

ArmCount
All Participants
All participants who delivered a predefined dose and volume of Alprostadil (prostaglandin E1 \[PGE1\], Caverject) into a receptacle, using the assigned Caverject Impulse delivery system. Participants did not receive study medication and were monitored during the handling of the Caverject Impulse delivery system.
48
Total48

Baseline characteristics

CharacteristicAll Participants
Age Continuous51.6 years
STANDARD_DEVIATION 8.5
Age, Customized
>= 65 years
6 participants
Age, Customized
Between 18 and 44 years
11 participants
Age, Customized
Between 45 and 64 years
31 participants
Sex: Female, Male
Female
0 Participants
Sex: Female, Male
Male
48 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 48
serious
Total, serious adverse events
0 / 48

Outcome results

Primary

Percentage of Participants Who Successfully Operated the Caverject Impulse Delivery System

Percentage of participants who were able to successfully expel the selected dose from the Caverject Impulse Delivery System when relying on the modified Instructions for Use. The process was considered successful if the lower bound of the 95% confidence interval (CI) was more than (\>) 80% overall.

Time frame: Day 1

Population: Full Analysis Set (FAS): All eligible participants who read the instructions and attempted to deliver Alprostadil using the Caverject Impulse Delivery System. Each participant tested one device/dose combination.

ArmMeasureValue (NUMBER)
All ParticipantsPercentage of Participants Who Successfully Operated the Caverject Impulse Delivery System95.83 percentage of participants
Secondary

Number of Participants Providing Comments to Any Question on the Participant Assessment Tool

Number of participants providing comments on questions in the Participant Assessment Tool. Questions were as follows: were instructions clear, were instructions useful, which was the most difficult step, and was the syringe easy to use.

Time frame: Day 1

Population: FAS. Each participant tested one device/dose combination.

ArmMeasureGroupValue (NUMBER)
All ParticipantsNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?1 participants
All ParticipantsNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?1 participants
All ParticipantsNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?3 participants
All ParticipantsNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?1 participants
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?2 participants
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?1 participants
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?0 participants
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?0 participants
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?0 participants
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?2 participants
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?0 participants
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?1 participants
Caverject 10 Mcg Device, 10 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?0 participants
Caverject 10 Mcg Device, 10 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?0 participants
Caverject 10 Mcg Device, 10 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?1 participants
Caverject 10 Mcg Device, 10 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?1 participants
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?2 participants
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?1 participants
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?1 participants
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?1 participants
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?0 participants
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?0 participants
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?0 participants
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?1 participants
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?0 participants
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?1 participants
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?1 participants
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?1 participants
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were clear?1 participants
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolMost difficult step?2 participants
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolSyringe easy to use?2 participants
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingNumber of Participants Providing Comments to Any Question on the Participant Assessment ToolInstructions provided were useful?1 participants
Secondary

Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 1

Participant Assessment Tool, Question 1: Instructions provided were useful? Participant responses were reported as follows: Very Useful, Somewhat Useful, Not Very Useful, Not Useful At All.

Time frame: Day 1

Population: FAS. Each participant tested one device/dose combination.

ArmMeasureGroupValue (NUMBER)
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 1Very useful44 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 1Somewhat useful4 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 1Not very useful0 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 1Not useful at all0 participants
Secondary

Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 2

Participant Assessment Tool, Question 2: Instructions provided were clear? Participant responses were reported as follows: Very Clear, Somewhat Clear, Not Very Clear, Not Clear At All.

Time frame: Day 1

Population: FAS. Each participant tested one device/dose combination.

ArmMeasureGroupValue (NUMBER)
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 2Very clear43 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 2Somewhat clear4 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 2Not very clear1 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 2Not clear at all0 participants
Secondary

Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 3

Participant Assessment Tool, Question 3: Most difficult step? Participant responses were reported as follows: No Steps Particularly Difficult, Attaching Needle, Mixing Solution, Getting The Air Out Of Syringe, Dialing Dose, Pushing Plunger, Other.

Time frame: Day 1

Population: FAS. Each participant tested one device/dose combination.

ArmMeasureGroupValue (NUMBER)
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 3No steps particularly difficult38 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 3Attaching needle2 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 3Mixing solution1 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 3Getting the air out of syringe0 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 3Dialing dose2 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 3Pushing plunger4 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 3Other1 participants
Secondary

Number of Participants With Categorical Responses to the Participant Assessment Tool: Question 4

Participant Assessment Tool, Question 4: Syringe easy to use? Participant responses were reported as follows: Very Easy, Somewhat Easy, Somewhat Difficult, Very Difficult

Time frame: Day 1

Population: FAS. Each participant tested one device/dose combination.

ArmMeasureGroupValue (NUMBER)
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 4Very easy43 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 4Somewhat easy4 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 4Somewhat difficult1 participants
All ParticipantsNumber of Participants With Categorical Responses to the Participant Assessment Tool: Question 4Very difficult0 participants
Secondary

Time Required to Perform Each Step While Using the Caverject Impulse Delivery System

Steps involved while using the Caverject Impulse Delivery System included assembly, mixing the dose, de-aeration, setting the dose, and injecting the dose.

Time frame: Day 1

Population: FAS. Each participant tested one device/dose combination. n = number of participants with available data.

ArmMeasureGroupValue (MEAN)Dispersion
All ParticipantsTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)103.2 secondsStandard Deviation 16.98
All ParticipantsTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)98.7 secondsStandard Deviation 85.41
All ParticipantsTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 621.0 secondsStandard Deviation 9.65
All ParticipantsTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)55.8 secondsStandard Deviation 18.27
All ParticipantsTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)55.5 secondsStandard Deviation 28.28
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)57.4 secondsStandard Deviation 30.94
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 624.8 secondsStandard Deviation 7.09
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)39.2 secondsStandard Deviation 21.21
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)50.8 secondsStandard Deviation 18.2
Caverject 10 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)141.7 secondsStandard Deviation 26.17
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)61.0 secondsStandard Deviation 38.78
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)62.8 secondsStandard Deviation 68.99
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)119.2 secondsStandard Deviation 58.6
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 670.3 secondsStandard Deviation 106.57
Caverject 10 Mcg Device, 7.5 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)62.3 secondsStandard Deviation 41.55
Caverject 10 Mcg Device, 10 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)52.7 secondsStandard Deviation 26.02
Caverject 10 Mcg Device, 10 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)80.0 secondsStandard Deviation 85.65
Caverject 10 Mcg Device, 10 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)115.4 secondsStandard Deviation 33.53
Caverject 10 Mcg Device, 10 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)58.0 secondsStandard Deviation 37.1
Caverject 10 Mcg Device, 10 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 647.3 secondsStandard Deviation 37.8
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)48.8 secondsStandard Deviation 43
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)127.5 secondsStandard Deviation 54.57
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)50.2 secondsStandard Deviation 22.97
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 643.2 secondsStandard Deviation 41.76
Caverject 20 Mcg Device, 5.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)42.0 secondsStandard Deviation 28.55
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)44.0 secondsStandard Deviation 18.12
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 626.2 secondsStandard Deviation 14.6
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)43.8 secondsStandard Deviation 38.21
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)110.2 secondsStandard Deviation 31.4
Caverject 20 Mcg Device, 10.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)51.6 secondsStandard Deviation 42.69
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)48.2 secondsStandard Deviation 18.18
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)109.7 secondsStandard Deviation 28.77
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)61.2 secondsStandard Deviation 16.44
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)31.0 secondsStandard Deviation 14.34
Caverject 20 Mcg Device, 15.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 623.8 secondsStandard Deviation 11.37
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 5: Injecting dose (n=6, 5, 6, 7, 5, 5, 6, 6)41.5 secondsStandard Deviation 18.22
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 4: Setting the dose (n=6, 4, 6, 7, 5, 5, 6, 637.3 secondsStandard Deviation 22.34
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 1: Assembly (n=6, 6, 6, 7, 6, 5, 6, 6)107.2 secondsStandard Deviation 52.7
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 3: De-aerating (n=6, 5, 6, 7, 5, 5, 6, 6)57.2 secondsStandard Deviation 20.18
Caverject 20 Mcg Device, 20.0 Mcg Dose SettingTime Required to Perform Each Step While Using the Caverject Impulse Delivery SystemStep 2: Mixing solution (n=6, 5, 6, 7, 5, 5, 6, 6)51.8 secondsStandard Deviation 35.31