Clinical Evaluation of Therapy Cool Path Duo Cardiac Ablation System for the Treatment of Ischemic Ventricular Tachycardia (VT)

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT00925522
Acronym
Ablate-VT
Enrollment
20
Registered
2009-06-22
Start date
2009-06-30
Completion date
2011-07-31
Last updated
2019-04-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Ischemic Ventricular Tachycardia

Brief summary

To demonstrate that ablation with the Therapy Cool Path Duo cardiac ablation system can eliminate ischemic VT and that its use does not result in an unacceptable risk of serious adverse events.

Detailed description

Ventricular Tachycardia (VT) is a common complication of ischemic heart disease that is often life threatening. Ventricular tachycardia is defined as three or more beats of ventricular origin in succession at a rate greater than 100 beats per minute. There are no normal QRS complexes associated with VT, and the rhythm is usually regular. Patients with VT experience unpleasant heart palpitations, dyspnea, and syncope, all of which decrease their overall quality of life. They also suffer the consequences of impaired hemodynamics, loss of cardiac efficiency, and are at an increased risk of sudden cardiac death (SCD), which accounts for approximately half of all deaths from patients with cardiovascular disease. Historically, antiarrhythmic medications served as the first line therapy for the treatment of VT, in spite of low efficacy and high recurrence rates. But in recent years, internal cardioverter defibrillators (ICDs) have become the treatment of choice. Although, ICDs are effective in terminating most arrhythmias, rarely does so without some episodes requiring poorly tolerated shock therapy. Ablative therapy offers an important option for frequent control with the potential for long-term VT elimination. Radiofrequency catheter ablation offers the ability to provide immediate control of recurrent VT. Since the early 2000's, strategies, tools, and techniques have been improving to identify key ablation sites and to deliver effective lesions. Radiofrequency ablation is successful in treating some but not all arrhythmias. With non irrigated technology it has been difficult to ablate ischemic VT. This may be due to an inadequate lesion size. Irrigated electrodes were developed to create larger lesions. Open irrigation at the catheter tip not only produces larger lesions, but also helps reduce the risk of coagulum (clots) and charring due to high temperatures.

Interventions

DEVICETherapy Cool Path Duo Cardiac Ablation System

Cardiac Ablation Procedure for Ischemic VT using a flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode.

Sponsors

Abbott Medical Devices
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patient is 18 years of age or older * Patient has an ICD (Implantable Cardiac Defibrillator) or will be implanted with one prior to discharge of ablation procedure * Patient has had at least 2 documented spontaneous episode of sustained ischemic VT (Ventricular Tachycardia) within the previous 6 months * Patient is resistant, intolerant or refractory to at least one Class I or III AAD (Anti-Arrhythmic Drug) * Patient has been informed of the nature of the study and has agreed to its provisions and provided written informed consent approved by the Institutional Review Board

Exclusion criteria

* Non-ischemic VT * History of stroke or transient ischemic attack within 6 months prior to enrollment * MI or previous cardiac surgery within 2 months prior to enrollment * Patient is pregnant or nursing * Patient has chronic NYHA (New York Heart Association) class IV heart failure * Limited life expectancy of 6 months or less * Patient is currently participating in another investigational drug or device study * Patient is unable or unwilling to cooperate with the study procedures * Known presence of intracardiac thrombi * Severe aortic stenosis or flailed mitral valve * Major contraindication to anticoagulation therapy or coagulation disorder * Left Ventricular Ejection Fraction \<10%

Design outcomes

Primary

MeasureTime frame
Primary Safety is Defined as the Incidence of Intra-procedural, Acute or Sub-chronic, Serious Cardiac Adverse Events, up to 7 Days Post-procedure.7 days

Countries

United States

Participant flow

Recruitment details

Patients with atrial flutter are evaluated by ablation doctor in the EP clinic and assessed for possibility of meeting other inclusion criteria.

Pre-assignment details

This is a non-randomized study.

Participants by arm

ArmCount
Therapy Cool Path Duo Cardiac Ablation System
Therapy Cool Path Duo Cardiac Ablation System: Consists three system components: A flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. The 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode. The Cool Point ™ Irrigation Pump and Cool Point™ Tubing Set is an external volumetric pump that is intended for use in administration of general I.V. fluids.
20
Total20

Withdrawals & dropouts

PeriodReasonFG000
Overall Studystudy closed -- business decision20

Baseline characteristics

CharacteristicTherapy Cool Path Duo Cardiac Ablation System
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
5 Participants
Age, Categorical
Between 18 and 65 years
15 Participants
Age, Continuous66.8 years
STANDARD_DEVIATION 10.5
Region of Enrollment
United States
20 participants
Sex: Female, Male
Female
0 Participants
Sex: Female, Male
Male
20 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
16 / 20
serious
Total, serious adverse events
7 / 20

Outcome results

Primary

Primary Safety is Defined as the Incidence of Intra-procedural, Acute or Sub-chronic, Serious Cardiac Adverse Events, up to 7 Days Post-procedure.

Time frame: 7 days

ArmMeasureValue (NUMBER)
Therapy Cool Path Duo CatheterPrimary Safety is Defined as the Incidence of Intra-procedural, Acute or Sub-chronic, Serious Cardiac Adverse Events, up to 7 Days Post-procedure.7 participants