Skip to content

Pivotal Study of the Intravenous Blood Glucose (IVBG) System, In-Clinic Setting

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT00925080
Enrollment
50
Registered
2009-06-19
Start date
2009-06-30
Completion date
2009-08-31
Last updated
2010-02-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes, Hyperglycemia

Brief summary

The purpose of this study is to evaluate safety and accuracy of the IVBG System (the System) when used to track blood glucose in insulin treated subjects with diabetes mellitus in an in-clinic setting for up to 72 hours (per subject). Reference blood glucose measurements will be collected across the entire reportable range of the System (e.g., 40 400 mg/dL) with adequate sampling at the upper and lower ends of this range. IVBG System accuracy will primarily be assessed relative to ISO 15197 criteria (i.e., within ±15 mg/dL at YSI glucose levels \< 75 mg/dL, and within ±20% at YSI glucose levels \>75 mg/dL).

Interventions

DEVICEIVBG

Venous blood glucose measurement every 7.5 minutes for 72-hours

Sponsors

DexCom, Inc.
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Age 18 years or older; 2. Willing to participate in one 72-hour in-clinic session during which up to 72 venous blood samples drawn for YSI measurement of glucose concentration (maximum sampling frequency of 4 per hour), a fingerstick blood glucose measurement will also be performed at the time of each YSI blood draw; 3. If instructed, be willing not to perform any activities that would result in submersion of the Sensor/Potentiostat in water or willing to wear a waterproof covering when submerging the Sensor/Potentiostat; 4. Have been diagnosed with insulin-requiring diabetes mellitus and are on multiple daily injections (MDI) or Continuous Subcutaneous Insulin Infusion (CSII) insulin therapy; 5. Able to speak, read, and write English.

Exclusion criteria

1. Have skin conditions or existing (or planned) medical instrumentation and/or dressings that preclude wearing the IVBG Sensor (e.g., extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis, or surgical dressings at the proposed wear site); 2. Are pregnant (as demonstrated by a positive pregnancy test within 72-hours of device insertion); 3. Have a contraindication to placement of a dedicated peripheral IV line; 4. Have a known history of heparin-induced thrombocytopenia; 5. Current participation in another investigational study protocol (if a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study); 6. Have any condition that, in the opinion of the Investigator, would interfere with their participation in the trial or pose an excessive risk to study staff handling venous blood samples (e.g., known history of hepatitis B or C).

Design outcomes

Primary

MeasureTime frame
IVBG System accuracy will primarily be assessed relative to ISO 15197 criteria (i.e., within ±15 mg/dL at YSI glucose levels < 75 mg/dL, and within ±20% at YSI glucose levels >75 mg/dL).72-hours

Countries

United States

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026