Comparison of 50% Tilt and Tuned Waveforms in Single-Coil Active Can Configuration

Prospective Comparison of 50/50% Tilt and Tuned Defibrillation Waveforms in a Single Coil Active Can Configuration

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT00874445
Acronym
PROMISE
Enrollment
77
Registered
2009-04-02
Start date
2009-03-31
Completion date
2010-09-30
Last updated
2019-02-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Sudden Cardiac Death

Brief summary

The purpose of this study is to compare the defibrillation efficacy between the 50/50% tilt biphasic waveform and the Tuned biphasic waveform in a single coil active can configuration.

Detailed description

There has been one prospective study that found some benefit of using tuned waveforms over 50/50% tilt waveforms.11 Other studies have failed to show much change in defibrillation efficacy with different duration biphasic waveforms. A vast majority of patients in these studies were tested with dual coil leads. Although dual coil leads can lower defibrillation thresholds, it is known that dual coil leads can make the lead extraction difficult, lead to venous obstruction and are more prone to failure. Results from a recent study showed that that a single coil configuration is adequate in a vast majority of patients when DFT testing was performed with tuned waveforms. This makes the use of single-coil leads during ICD/ CRT-D implants an attractive option. Theoretical analysis has shown that difference between the tuned and 50% tilt waveforms is amplified at high impedance levels with tuned waveforms being superior (unpublished). Accordingly, this study has been designed to prospectively compare the DFT estimates with tuned and 50% tilt waveforms in left-sided, active pectoral defibrillation lead systems when the SVC coil has been turned OFF or is not part of the shocking circuit.

Interventions

Patients in experimental group will have their ICDs programmed to Tuned waveform and those in the control group will have their ICDs programmed to Fixed Tilt waveform.

Sponsors

Abbott Medical Devices
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Patient meets standard indication for an ICD/ CRT-D and will be implanted with an FDA approved SJM ICD/ CRT-D and compatible defibrillation lead system in the left pectoral region. * Patient is able to tolerate DFT testing.

Exclusion criteria

* Patient has the pulse generator on the right side. * Patient is pregnant. * Patient is less than 18 years old. * SVC coil was turned ON during DFT testing

Design outcomes

Primary

MeasureTime frame
True DFTs in fixed tilt and tuned waveforms obtained in volts (V)Implant

Secondary

MeasureTime frame
True DFTs in fixed tilt and tuned waveforms obtained in joules (J)Implant