Tetra-NIRS Clinical Study

Diagnostic Utility of Tetra-NIRS in Conjunction With Urodynamic Studies

Status
Terminated
Phases
Phase 4
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT00871975
Enrollment
155
Registered
2009-03-30
Start date
2009-04-30
Completion date
2012-02-29
Last updated
2016-07-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Lower Urinary Tract Symptoms, Overactive Bladder

Keywords

LUTS, OAB

Brief summary

The purpose of this evaluation is to validate the previous clinical study results using a Near-Infrared Spectroscopy (NIRS) device, as compared to the conventional urodynamics (UDS) testing. The experiment will use the commercially available Tetra-NIRS and Laborie UDS equipment. The hypothesis is that the Tetra-NIRS device, in the clinical environment, provides data to enhance the analysis using the standard urodynamics equipment (stand-alone Urodynamic procedure).

Detailed description

This is a study of the commercially available Tetra-NIRS device to compare the results collected against conventional urodynamics. The study will evaluate male patients with lower urinary tract symptoms (LUTS) as well as female subjects with over-active bladder (OAB). The Tetra-NIRS unit uses near infrared light at 3 different wavelengths to measure changes in haemoglobin and cytochrome from the detrusor muscle from the human bladder wall. This process is similar to what is used for cerebral and muscle oxygenation monitoring (e.g., pulse oximetry). Previous clinical feasibility studies have identified the correlation between the changes in haemoglobin and cytochrome of the bladder, and bladder pressure values obtained during urodynamics procedures and uroflow procedures. For this study the Tetra-NIRS results will not be used by the urologist/nurse to guide in patient management and treatment decisions, but rather will only be used to collect measurements from the patient for the purposes of this study. The Tetra-NIRS sensor patch will be adhered externally to the skin surface where the bladder is located. For this study, a minimum of 30 male subjects with LUTS and 30 female subjects with OAB scheduled to undergo a urodynamics procedure are to be enrolled. Up to 20 female subjects with OAB will be enrolled. The investigators will attempt to equally enrol male subjects into each of the two categories including obstructed and non-obstructed groups according to prostate hyperplasia (equivocal subjects will be determined by the investigator).

Interventions

DEVICEUrodynamics + TETRA - NIRS

Urodynamics will be performed as per standard of care for patients enrolled, whereby one catheter is inserted into the bladder and one in the rectum to analyze bladder filling, storage and voiding conditions via pressures measured and urine flow rates detected by a uroflowmeter. Tetra-NIRS will be concurrently performed where Tetra-NIRS unit uses near infrared light at 3 different wavelengths to measure changes in haemoglobin and cytochrome. A patch is applied externally to measure these changes in the detrusor muscle of the human bladder wall. This process is similar to what is used for cerebral and muscle oxygenation monitoring (e.g., pulse oximetry). Previous clinical feasibility studies have identified a correlation between the changes in haemoglobin and cytochrome and the pressure values obtained during urodynamics procedures.

Sponsors

Laborie Medical Technologies Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* The selection of subjects enrolled for testing must meet the following criteria: * Subjects must be 18 years of age or older. * Subjects are patients of one of the institutions and are currently scheduled for UDS * Male subjects must have LUTS * Female subjects must have OAB * Subjects must give their informed consent prior to enrolment.

Exclusion criteria

* The patient has an existing health condition which the investigators feel will not allow for safe or accurate measurements with the Tetra-NIRS device.

Design outcomes

Primary

MeasureTime frameDescription
Number of Participants With Urodynamic Detrusor Overactivity Events or Prostatic Obstruction as Detected by Tetra-NIRS Compared to Urodynamics1 YearThe Tetra-NIRS device provides a linear pattern similar to the pressures obtained during urodynamics. The NIRS output shows relative change in hemoglobin concentrations (oxygenated and deoxygenated) where the numerical value does not actually indicate the concentration, so there is no unit of measure. The numerical output is used to track change over time, or trendline analysis. A qualified interpreter studied tracings for significant changes (+/-2 Hb units) in the NIRS patterning during detrusor overactivity events. As well, under its' approved intended use, Tetra-NIRS trendline analysis was compared against urodynamics during voiding in males, such that a downward trend during voiding indicates urethral obstruction, and an upward trend indicates an unobstructed urethra.

Countries

Canada, United Kingdom, United States

Participant flow

Recruitment details

Each site recruited in a Urology clinic setting. All patients were scheduled for Urodynamics testing. Sunnybrook recruitment period spanned Apr 2010 - Feb 2012. Vanderbilt Sept 2009 - May 24 2010. King's College May 2010 - August 2010.

Pre-assignment details

There were no group assignments.

Participants by arm

ArmCount
Urodynamics + Tetra-NIRS
Single arm - all patients were recruited following a requirement for urodynamic testing. The patient is the invited to the clinic room to perform the urodynamics test concurrently with the Tetra-NIRS device. The clinic staff perform urodynamics per their standard protocol, with the Tetra-NIRS patch applied externally above the location of the bladder. The numerical pressure data resulting from the urodynamics machine plus the hemoglobin levels (oxygenated and deoxygenated) measured by the Tetra-NIRS device are displayed on a computer and saved. The test would take approx. 45 minutes to perform, after which time the patient is free to go home. Interpretation of the Tetra-NIRS had no bearing on the patients' diagnosis, however urodynamic interpretation was performed per their standard protocol.
155
Total155

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyCatheters could not be inserted1
Overall StudyMotion artifact, tracing uninterpretable5
Overall StudyPatient unable to void/insufficient vol13
Overall StudyTechnical difficulty w155-1ith equipment3

Baseline characteristics

CharacteristicUrodynamics + Tetra-NIRS
Age, Customized
Greater than or equal to 18 years
155 participants
Region of Enrollment
Canada
37 participants
Region of Enrollment
United Kingdom
100 participants
Region of Enrollment
United States
18 participants
Sex: Female, Male
Female
121 Participants
Sex: Female, Male
Male
34 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 155
serious
Total, serious adverse events
0 / 155

Outcome results

Primary

Number of Participants With Urodynamic Detrusor Overactivity Events or Prostatic Obstruction as Detected by Tetra-NIRS Compared to Urodynamics

The Tetra-NIRS device provides a linear pattern similar to the pressures obtained during urodynamics. The NIRS output shows relative change in hemoglobin concentrations (oxygenated and deoxygenated) where the numerical value does not actually indicate the concentration, so there is no unit of measure. The numerical output is used to track change over time, or trendline analysis. A qualified interpreter studied tracings for significant changes (+/-2 Hb units) in the NIRS patterning during detrusor overactivity events. As well, under its' approved intended use, Tetra-NIRS trendline analysis was compared against urodynamics during voiding in males, such that a downward trend during voiding indicates urethral obstruction, and an upward trend indicates an unobstructed urethra.

Time frame: 1 Year

Population: Male and female patients were included, where the urodynamics tracings were compared against the Tetra NIRS tracings.

ArmMeasureGroupValue (NUMBER)
Urodynamics + Tetra NIRSNumber of Participants With Urodynamic Detrusor Overactivity Events or Prostatic Obstruction as Detected by Tetra-NIRS Compared to UrodynamicsFalse Positive on Tetra-NIRS + UDS59 participants
Urodynamics + Tetra NIRSNumber of Participants With Urodynamic Detrusor Overactivity Events or Prostatic Obstruction as Detected by Tetra-NIRS Compared to UrodynamicsTrue Positive on Tetra-NIRS + UDS30 participants
Urodynamics + Tetra NIRSNumber of Participants With Urodynamic Detrusor Overactivity Events or Prostatic Obstruction as Detected by Tetra-NIRS Compared to UrodynamicsFalse Negative on Tetra-NIRS + UDS10 participants
Urodynamics + Tetra NIRSNumber of Participants With Urodynamic Detrusor Overactivity Events or Prostatic Obstruction as Detected by Tetra-NIRS Compared to UrodynamicsTrue Negatives on Tetra-NIRS + UDS34 participants
Comparison: A contingency table was used to compare presence or absence of an event (eg. detrusor overactivity) on the urodynamic tracing with interpretation of events on the Tetra-NIRS tracings.
Comparison: A contingency table was used to compare presence or absence of an event (eg. detrusor overactivity) on the urodynamic tracing with interpretation of events on the Tetra-NIRS tracings.

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026