Skip to content

One-Lung Ventilation in the Morbidly Obese Patient: Comparison of Double Lumen Versus Bronchial Blockers

One-Lung Ventilation in Morbidly Obese Patients:

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT00813176
Enrollment
50
Registered
2008-12-22
Start date
2007-03-31
Completion date
2008-12-31
Last updated
2018-06-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Obesity

Keywords

Obesity, One-Lung Ventilation, Anesthesia, Thoracic, One-lung ventilation in the morbidly obese patient

Brief summary

The purpose of this study is to determine whether the double-lumen endotracheal tube or the standard single-lumen tracheal tube, with an Arndt® blocker, is superior in providing one lung ventilation during thoracic surgery.

Detailed description

The purpose of the study is to compare a double lumen tube versus the standard single-lumen tracheal tube, with an Arndt® blocker, in the morbidly obese patient who requires one-lung ventilation during surgery. The objects is to determine which device is best for intubation and is most effective in terms of optimal positioning of the device, of lung collapse, and quality of lung isolation.

Interventions

DEVICEArndt Bronchial Blocker

Sponsors

Javier H Campos
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
DOUBLE (Subject, Outcomes Assessor)

Intervention model description

prospective randomized study

Eligibility

Sex/Gender
ALL
Age
18 Years to 99 Years
Healthy volunteers
No

Inclusion criteria

1. Subject undergoing thoracic or esophageal surgery requiring one-lung ventilation. 2. Subject is \>18 years of age. 3. Subject's BMI \> 35.

Exclusion criteria

1. Subject \< 18 years of age. 2. Subject's BMI \< 35.

Design outcomes

Primary

MeasureTime frameDescription
Time Taken for Endotracheal Tube Placement Procedure.This measurement occurred after the patient was in the operating room and the randomization group was determined. This measurement began and ended during the intubation procedure.For each group, the time in minutes required for the anesthesiologist to successfully place the endotracheal tube. Intubation time was recorded as the time from when the tracheal tube passed the vocal cords until the anesthetist concluded that the DLT (Arm 1) or the single-lumen tube with the Arndt blocker (Arm 2) was correctly placed and optimal position was confirmed with the fiberoptic bronchoscope.
Number of Participants in Which the Tube Was Successfully Positioned at First AttemptThis data measure was occurred during surgery.Number of participants in each group in which the tube was successfully positioned at first attempt

Secondary

MeasureTime frameDescription
Effectiveness of Lung CollapseThis measure occured during surgery just after intubation.For each group, the number of participants in which lung collapse was evaluated as: (i) good-complete collapse without surgical interference, (ii) fair-total collapse, but with residual air, and (iii) poor-no or partial collapse, lung inflated and interfering with the surgical exposure.
Time Required to Collapse the LungThis measurement occurred during surgery.For each group, the mean time for lung isolation/collapse was measured from the institution of one-lung ventilation to the time of total lung collapse.
Number of Participants With Successful ReinflationThis occurred during surgery.In each group, the number of Participants in which successful lung reinflation occurred.
Mean Intraoperative Oxygenation During One-lung VentilationThis measure occured during surgery.For each group, the mean intraoperative oxygenation (arterial partial pressure of oxygen measured by arterial blood gases) were taken during two-lung ventilation (baseline) and after 30 min of one-lung ventilation.

Countries

United States

Participant flow

Participants by arm

ArmCount
Double Lumen Endotracheal Tube
Double Lumen Endotracheal Tube Device where lung collapse can be achieved by clamping one of the lumens (the operated lung) 25 subjects were enrolled in this group
25
Arndt Bronchial Blocker
Arndt Bronchial Blocker in conjunction with a single lumen endotracheal tube. Device that allows collapse of one lung (the operated lung). 25 subjects were enrolled in this group
25
Total50

Baseline characteristics

CharacteristicArndt Bronchial BlockerDouble Lumen Endotracheal TubeTotal
Age, Continuous53 years52 years52 years
Body Mass Index40.2 kg/m^2
STANDARD_DEVIATION 4.5
39.9 kg/m^2
STANDARD_DEVIATION 4.8
40 kg/m^2
STANDARD_DEVIATION 4.5
Body Weight116.5 Kilograms
STANDARD_DEVIATION 22.4
114.5 Kilograms
STANDARD_DEVIATION 17.1
115.6 Kilograms
STANDARD_DEVIATION 19.9
double lumen tubes versus bronchial blockers25 Participants25 Participants50 Participants
Neck Circumference47 Centimeters
STANDARD_DEVIATION 7
46 Centimeters
STANDARD_DEVIATION 4
46.3 Centimeters
STANDARD_DEVIATION 5.9
Sex: Female, Male
Female
13 Participants14 Participants27 Participants
Sex: Female, Male
Male
12 Participants11 Participants23 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
— / —— / —
other
Total, other adverse events
0 / 250 / 25
serious
Total, serious adverse events
3 / 252 / 25

Outcome results

Primary

Number of Participants in Which the Tube Was Successfully Positioned at First Attempt

Number of participants in each group in which the tube was successfully positioned at first attempt

Time frame: This data measure was occurred during surgery.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Double Lumen Endotracheal TubeNumber of Participants in Which the Tube Was Successfully Positioned at First Attempt22 Participants
Arndt Bronchial BlockerNumber of Participants in Which the Tube Was Successfully Positioned at First Attempt23 Participants
Primary

Time Taken for Endotracheal Tube Placement Procedure.

For each group, the time in minutes required for the anesthesiologist to successfully place the endotracheal tube. Intubation time was recorded as the time from when the tracheal tube passed the vocal cords until the anesthetist concluded that the DLT (Arm 1) or the single-lumen tube with the Arndt blocker (Arm 2) was correctly placed and optimal position was confirmed with the fiberoptic bronchoscope.

Time frame: This measurement occurred after the patient was in the operating room and the randomization group was determined. This measurement began and ended during the intubation procedure.

ArmMeasureValue (MEAN)Dispersion
Double Lumen Endotracheal TubeTime Taken for Endotracheal Tube Placement Procedure.3.4 minutesStandard Deviation 1.8
Arndt Bronchial BlockerTime Taken for Endotracheal Tube Placement Procedure.4.1 minutesStandard Deviation 5.2
Secondary

Effectiveness of Lung Collapse

For each group, the number of participants in which lung collapse was evaluated as: (i) good-complete collapse without surgical interference, (ii) fair-total collapse, but with residual air, and (iii) poor-no or partial collapse, lung inflated and interfering with the surgical exposure.

Time frame: This measure occured during surgery just after intubation.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Double Lumen Endotracheal TubeEffectiveness of Lung CollapseGood16 Participants
Double Lumen Endotracheal TubeEffectiveness of Lung CollapseFair7 Participants
Double Lumen Endotracheal TubeEffectiveness of Lung CollapsePoor1 Participants
Double Lumen Endotracheal TubeEffectiveness of Lung CollapseLung Colapse not Required1 Participants
Arndt Bronchial BlockerEffectiveness of Lung CollapseLung Colapse not Required0 Participants
Arndt Bronchial BlockerEffectiveness of Lung CollapseGood17 Participants
Arndt Bronchial BlockerEffectiveness of Lung CollapsePoor1 Participants
Arndt Bronchial BlockerEffectiveness of Lung CollapseFair7 Participants
Secondary

Mean Intraoperative Oxygenation During One-lung Ventilation

For each group, the mean intraoperative oxygenation (arterial partial pressure of oxygen measured by arterial blood gases) were taken during two-lung ventilation (baseline) and after 30 min of one-lung ventilation.

Time frame: This measure occured during surgery.

ArmMeasureValue (MEAN)Dispersion
Double Lumen Endotracheal TubeMean Intraoperative Oxygenation During One-lung Ventilation16.8 kPaStandard Deviation 8.8
Arndt Bronchial BlockerMean Intraoperative Oxygenation During One-lung Ventilation18.6 kPaStandard Deviation 10.3
Secondary

Number of Participants With Successful Reinflation

In each group, the number of Participants in which successful lung reinflation occurred.

Time frame: This occurred during surgery.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Double Lumen Endotracheal TubeNumber of Participants With Successful Reinflation24 Participants
Arndt Bronchial BlockerNumber of Participants With Successful Reinflation25 Participants
Secondary

Time Required to Collapse the Lung

For each group, the mean time for lung isolation/collapse was measured from the institution of one-lung ventilation to the time of total lung collapse.

Time frame: This measurement occurred during surgery.

ArmMeasureValue (MEAN)Dispersion
Double Lumen Endotracheal TubeTime Required to Collapse the Lung18 MinutesStandard Deviation 9.3
Arndt Bronchial BlockerTime Required to Collapse the Lung15.8 MinutesStandard Deviation 9.6

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026