Induced Hypoxia Study for Validation of SpO2 Accuracy

Pulse Oximeter Responses to Multiple Levels of Stable Hypoxia

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT00788983
Enrollment
12
Registered
2008-11-11
Start date
2008-11-30
Completion date
2008-11-30
Last updated
2013-09-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hypoxia

Keywords

Induced hypoxia, Pulse oximeter, Nonin Medical, Verification of SpO2 accuracy

Brief summary

Validate Non-Motion SpO2 Accuracy from 70%-100% of various pulse oximeter/ sensor combinations and various OEM pulse oximeter modules and sensor combinations during induced hypoxia (desaturation).

Detailed description

The testing is conducted on 12 healthy, consenting, non-smoking subjects in accordance with the IRB approved Protocol. The subjects shall be distributed across both genders and a range of skin tones as equally as practical. An arterial line will be placed in the radial artery of each subject's right arm, and sensors will be attached to each subject. The subjects will be placed in a semi-supine position and allowed to breathe through a mouthpiece while the nose is blocked with a nose-clip. Hypoxia will be induced by on each subject, as levels of oxyhemoglobin saturation (between 70% and 100%) are achieved by breathing mixtures of nitrogen, room air and carbon dioxide. Inspired O2 concentration will be adjusted breath-by-breath using a computed saturation, based on end-tidal PO2 and PCO2, as sampled by a mass spectrometer. Predicted levels of oxyhemoglobin saturations will be attained and held stable. At a minimum of sixty seconds into each plateau, a 0.5cc waste blood draw through the arterial line, followed by the first 1.0cc sample blood draw. After approximately thirty additional seconds, the second 1.0cc sample blood draw will be taken. The samples will immediately be analyzed by a Radiometer OSM-3 multi-wavelength co-oximeter and recorded. The SpO2 data from the oximeters will be collected via a laptop computer. Concurrent with the end of each blood draw, a marker will be generated on the laptop computers to identify the event.

Interventions

None listed

Sponsors

Nonin Medical, Inc
Lead SponsorINDUSTRY

Study design

Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to 40 Years
Healthy volunteers
Yes

Inclusion criteria

* Non-smokers * Healthy; no known blood disorders * Finger sizes within the sensor specifications * Minimum 18 years of age

Exclusion criteria

* Artificial nails or fingernail polish

Design outcomes

Primary

MeasureTime frame
SpO2 Accuracy Verification Study (Arms)of ≤3 between 70 & 100%During Analysis

Countries

United States

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026