Brief summary
The Cirrus HD OCT, a new spectral domain optical coherence tomography (OCT), has better resolution than the previous time domain Stratus OCT. Discriminating ability of these two OCTs for diagnosis of glaucoma and judgement of disease progression will be studied.
Detailed description
The Cirrus HD OCT is a new spectral domain OCT. Even though its resolution and speed are much better than the previous time domain Stratus OCT, its diagnostic power for glaucoma detection and judgement of disease progression should be directly compared with Stratus OCT. In the present study, we will review the data of peripapillary retinal nerve fiber layer thickness measurement using the Cirrus HD OCT and Stratus OCT. Their discriminating ability for diagnosis of glaucoma and judgement of disease progression will be compared using the ROC curves and Cohen's kappa.
Interventions
None listed
Sponsors
Study design
Eligibility
Inclusion criteria
* Retinal nerve fiber layer analysis with Cirrus OCT and Stratus OCT within 1 month
Exclusion criteria
* Intraocular surgeries history * DM * High refractive error ( \< -4.00 diopters, or \> + 4.00 diopters) * Media opacity
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| Retinal nerve fiber layer thickness | When the OCT images are taking |
Secondary
| Measure | Time frame |
|---|---|
| Visual field | When the OCT images are taking |
| Red-free fundus photo | When the OCT images are taking |
| Intraocular pressure | When the OCT images are taking |
| Central corneal thickness | When the OCT images are taking |
Countries
South Korea