ReSTOR Natural +3.0D Study In Japan

ReSTOR Natural +3.0D (Diopter) Study In Japan

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT00758576
Enrollment
64
Registered
2008-09-25
Start date
2008-02-29
Completion date
2009-12-31
Last updated
2010-12-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cataract

Keywords

Cataract

Brief summary

The objective of this study is to evaluate safety and effectiveness of ACRYSOF Multifocal Single-piece Intraocular Lens (IOL) Model SN6AD1 when implanted to replace the natural lens following cataract removal.

Interventions

DEVICEACRYSOF Multifocal Single-piece Intraocular Lens (IOL) Model SN6AD1

Implantation with the AcrySof ReSTOR Intraocular Lens (IOL) Model SN6AD1 following cataract removal.

Sponsors

Alcon Research
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
20 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Planned bilateral cataract removal by phacoemulsification * Potential postoperative best corrected visual acuity of 0.5 or over * 1.5 Diopter (D) or less of astigmatism preoperatively, measured by keratometry

Exclusion criteria

* Poorly controlled glaucoma * Progressive diabetic retinopathy

Design outcomes

Primary

MeasureTime frameDescription
Uncorrected Intermediate Visual Acuity1 year after surgeryBinocular Uncorrected Intermediate Visual Acuity, measured at 1 meter and 50 centimeters measured in decimal visual acuity. Decimal visual acuity is the normal method to measure visual acuity in Japan. A higher decimal visual acuity value indicates better visual acuity. 0.5 in decimal visual acuity means 20/40 in Snellen fraction.
Uncorrected Distance Visual Acuity1 year after surgeryBinocular Uncorrected Dinstance Visual Acuity measured in decimal visual acuity. Decimal visual acuity is the normal method to measure visual acuity in Japan. A higher decimal visual acuity value indicates better visual acuity. 0.5 in decimal visual acuity means 20/40 in Snellen fraction.
Uncorrected Near Visual Acuity1 year after surgeryBinocular Near Visual Acuity measured in decimal visual acuity. Decimal visual acuity is the normal method to measure visual acuity in Japan. A higher decimal visual acuity value indicates better visual acuity. 0.5 in decimal visual acuity means 20/40 in Snellen fraction.

Countries

Japan

Participant flow

Recruitment details

Only adult subjects of either gender in need of cataract extraction in both eyes were considered for enrollment. 64 subjects were implanted bilaterally with the ACRYSOF® ReSTOR® Aspheric +3.0 D (Diopter) Add Power Intraocular Lens (IOL) Model SN6AD1

Pre-assignment details

Subjects were examined to ensure inclusion/exclusion criteria. Only subjects who signed an informed consent and qualified to be in the study were enrolled. Subject numbers were assigned after obtaining consent meeting inclusion/exclusion criteria. Subjects were considered enrolled once implanted in the first operative eye.

Participants by arm

ArmCount
SN6AD1
ACRYSOF® ReSTOR® Aspheric +3.0 D (Diopter) Add Power Intraocular Lens (IOL) Model SN6AD1
64
Total64

Baseline characteristics

CharacteristicSN6AD1
Age, Customized
60 - 69 years of age
23 Participants
Age, Customized
<60 years of age
6 Participants
Age, Customized
70 - 79 years of age
33 Participants
Age, Customized
>80 years of age
2 Participants
Sex: Female, Male
Female
60 Participants
Sex: Female, Male
Male
4 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 64
serious
Total, serious adverse events
1 / 64

Outcome results

Primary

Uncorrected Distance Visual Acuity

Binocular Uncorrected Dinstance Visual Acuity measured in decimal visual acuity. Decimal visual acuity is the normal method to measure visual acuity in Japan. A higher decimal visual acuity value indicates better visual acuity. 0.5 in decimal visual acuity means 20/40 in Snellen fraction.

Time frame: 1 year after surgery

Population: One patient dropped out after the visit 6 months following surgery but before the 1 year visit.

ArmMeasureValue (MEAN)
SN6AD1Uncorrected Distance Visual Acuity1.122 Decimal Visual Acuity
Primary

Uncorrected Intermediate Visual Acuity

Binocular Uncorrected Intermediate Visual Acuity, measured at 1 meter and 50 centimeters measured in decimal visual acuity. Decimal visual acuity is the normal method to measure visual acuity in Japan. A higher decimal visual acuity value indicates better visual acuity. 0.5 in decimal visual acuity means 20/40 in Snellen fraction.

Time frame: 1 year after surgery

Population: One patient dropped out after the visit 6 months following surgery but before the 1 year visit.

ArmMeasureGroupValue (MEAN)
SN6AD1Uncorrected Intermediate Visual Acuity1 meter0.771 Decimal Visual Acuity
SN6AD1Uncorrected Intermediate Visual Acuity50 centimeters0.861 Decimal Visual Acuity
Primary

Uncorrected Near Visual Acuity

Binocular Near Visual Acuity measured in decimal visual acuity. Decimal visual acuity is the normal method to measure visual acuity in Japan. A higher decimal visual acuity value indicates better visual acuity. 0.5 in decimal visual acuity means 20/40 in Snellen fraction.

Time frame: 1 year after surgery

Population: One patient dropped out after the visit 6 months following surgery but before the 1 year visit.

ArmMeasureValue (MEAN)
SN6AD1Uncorrected Near Visual Acuity1.082 Decimal Visual Acuity

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026