Brief summary

This study is designed to obtain information for design purposes on use of an ECG guided monitoring system to aid in the correct placement of PICC lines. There is no formal study hypothesis.

Detailed description

PICC line placement is a common procedure, made more difficult by the need to ensure that catheter tip placement is in an acceptable location, typically the superior vena cava (SVC), and ideally at the junction of the SVC and right atrium. Tip location is determined at present by chest X-ray, which is performed after the procedure. If the tip location is incorrect, additional procedures and/or X-rays are needed until proper placement is ensured. Use of an electrode placed inside the catheter during insertion has been reported to provide identifiable changes in the p-wave sufficient to guide placement. This study is designed to collect ECG and catheter depth information during routine PICC placement procedures, and to correlate these ECG changes to tip location as determined by contrast enhanced fluoroscopy.

Interventions

DEVICEECG-guided PICC placement

Patients undergoing PICC placement will have ECG data collected as a function of tip location.

Sponsors

C. R. Bard
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
BASIC_SCIENCE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 80 Years
Healthy volunteers
No

Inclusion criteria

* Require Power PICC placement as standard care * Signed informed consent

Exclusion criteria

* Has an intracardiac device * Does not have normal sinus rhythm * Has known abnormality of central venous system * Plasma creatinine \>=1.5 mg/dL * Pregnancy * Allergy or prior reaction to contrast material

Design outcomes

Primary

MeasureTime frameDescription
Distance of Maximum P-wave Amplitude in Relation to Superior Vena Cava (SVC)/Right Atrium(RA) Junctionduring catheter insertionIn order to determine correlation of PICC tip location with the intracatheter ECG, the PICC was advanced at 1cm increments and the p-wave observed for amplitude changes.

Countries

United States

Participant flow

Recruitment details

The study was conducted at the University of Utah between August 2008 and December 2008.

Participants by arm

ArmCount
PICC Insertion
All patients, aged 18 to 80 years, who required PICC insertion for their standard care will be studied.
7
Total7

Baseline characteristics

CharacteristicPICC Insertion
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
4 Participants
Age, Categorical
Between 18 and 65 years
3 Participants
Age, Continuous58.86 years
STANDARD_DEVIATION 15.48
Region of Enrollment
United States
7 participants
Sex: Female, Male
Female
3 Participants
Sex: Female, Male
Male
4 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 7
serious
Total, serious adverse events
0 / 7

Outcome results

Primary

Distance of Maximum P-wave Amplitude in Relation to Superior Vena Cava (SVC)/Right Atrium(RA) Junction

In order to determine correlation of PICC tip location with the intracatheter ECG, the PICC was advanced at 1cm increments and the p-wave observed for amplitude changes.

Time frame: during catheter insertion

ArmMeasureValue (MEAN)Dispersion
PICC PlacementDistance of Maximum P-wave Amplitude in Relation to Superior Vena Cava (SVC)/Right Atrium(RA) Junction0.5 cmStandard Deviation 1.2