Asthma
Conditions
Keywords
Asthma, nitric oxide, eNO, FENO, Apieron INSIGHT, Aerocrine NIOX
Brief summary
Collection of quantitative data to demonstrate the accuracy (bias) and precision (repeatability) of the Apieron INSIGHT(tm) eNO Breath System as compared to the FDA cleared NIOX eNO Breath System when testing is performed by asthma patients.
Detailed description
A non-randomized, prospective study of up to 150 subjects with asthma. Patients will perform eNO testing with the Apieron INSIGHT(tm) eNO system and the NIOX eNO system at the physician's office/clinic after a short demonstration and training by a healthcare professional trained in the use of the devices. Exhaled NO testing involves inhaling to lung capacity, and then performing a controlled exhalation into a mouthpiece connected to each of the Apieron INSIGHT(tm) and NIOX systems. Patients will perform an eNO measurement with the Apieron INSIGHT(tm) system and with the FDA cleared NIOX system for the accuracy portion of the study. Patients will perform 2 sequential eNO measurements on each system for the precision part of the study. Subjects that measure eNO values over 100 ppb will be asked to repeat 2 sequential eNO measurements on each system.
Interventions
Subjects will perform a controlled breath maneuver using both the Apieron and NIOX devices. The devices will measure nitric oxide in the subjects' exhaled breath.
Sponsors
Study design
Eligibility
Inclusion criteria
* Age 5 or older * Pre-existing diagnosis of asthma * Non-smoker
Exclusion criteria
* Chronic inflammatory lung disease other than asthma * Medical conditions that preclude hand-eye coordination
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| eNO using Apieron INSIGHT(tm) System and Aerocrine NIOX at 50 ml/s | Single point in time |
Countries
United States