Skip to content

Low Cylinder Toric

Evaluation of the Acrysof Toric Model SN60T3 Intraocular Lens in Patients With Predicted Residual Corneal Astigmatism Between 0.75 and 1.03 Diopters (D)

Status
Completed
Phases
Phase 4
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT00732030
Enrollment
26
Registered
2008-08-11
Start date
2007-06-30
Completion date
Unknown
Last updated
2010-04-07

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cataracts

Keywords

Toric IOL

Brief summary

Study evaluates the visual outcomes and patient satisfaction after implantation of a toric intraocular lens (IOL).

Interventions

DEVICEToric IOL (SN60T3)

Each enrolled eye implanted with the AcrySof Toric Model SN60T3 intraocular lens (IOL) for the treatment of cataract.

Sponsors

Alcon Research
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
21 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Diagnosed with Cataracts * Predicted residual corneal astigmatism between 0.75 and 1.03 Diopters (D)

Exclusion criteria

* Preoperative ocular pathology * Planned monovision

Design outcomes

Primary

MeasureTime frameDescription
Uncorrected Distance Visual Acuity6 monthsUncorrected Distance Visual Acuity at month 6 measured in logMAR. LogMAR is the logarithm of the minimum angle of resolution. It is a unit of measure for visual acuity (VA).
Best Corrected Distance Visual Acuity6 MonthsBest Corrected Distance Visual Acuity at month 6 measured in LogMAR. LogMAR is the logarithm of the minimum angle of resolution. It is a unit of measure for visual acuity (VA).
Residual Refractive Cylinder6 MonthResidual Refractive Cylinder at month 6 measured in diopters (D).

Secondary

MeasureTime frameDescription
Patient Satisfaction Survey6 monthsSatisfaction for daytime distance vision, nighttime distance vision, and indoors distance vision on a scale of 1 to 5, with 1 being very dissatisfied and 5 being very satisfied.

Countries

United States

Participant flow

Recruitment details

Patients were \>21 years of age, of any race and either gender. Patients had operable cataracts in at least 1 eye and were able to provide informed consent. Patients were free of systemic diseases affecting ocular health and had a predicted residual corneal astigmatism between 0.75 and 1.03 Diopters (D) by the AcrySof Toric calculator.

Pre-assignment details

During the preoperative exam, subjects were examined to ensure they met the inclusion/exclusion criteria. Only subjects who signed an informed consent and qualified to be in the study by meeting all inclusion/exclusion criteria were enrolled.

Participants by arm

ArmCount
AcrySof Toric T3
Each enrolled eye receives AcrySof Toric SN60T3 IOL
26
Total26

Baseline characteristics

CharacteristicAcrySof Toric T3
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
19 Participants
Age, Categorical
Between 18 and 65 years
7 Participants
Sex: Female, Male
Female
16 Participants
Sex: Female, Male
Male
10 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 24
serious
Total, serious adverse events
0 / 24

Outcome results

Primary

Best Corrected Distance Visual Acuity

Best Corrected Distance Visual Acuity at month 6 measured in LogMAR. LogMAR is the logarithm of the minimum angle of resolution. It is a unit of measure for visual acuity (VA).

Time frame: 6 Months

Population: Data was collected for 29 eyes in 24 patients.

ArmMeasureValue (MEAN)Dispersion
AcrySof Toric T3Best Corrected Distance Visual Acuity0.02 logMARStandard Deviation 0.09
Primary

Residual Refractive Cylinder

Residual Refractive Cylinder at month 6 measured in diopters (D).

Time frame: 6 Month

ArmMeasureValue (MEAN)Dispersion
AcrySof Toric T3Residual Refractive Cylinder0.4 DioptersStandard Deviation 0.4
Primary

Uncorrected Distance Visual Acuity

Uncorrected Distance Visual Acuity at month 6 measured in logMAR. LogMAR is the logarithm of the minimum angle of resolution. It is a unit of measure for visual acuity (VA).

Time frame: 6 months

Population: Data was collected for 29 eyes in 24 patients.

ArmMeasureValue (MEAN)Dispersion
AcrySof Toric T3Uncorrected Distance Visual Acuity0.09 logMARStandard Deviation 0.14
Secondary

Patient Satisfaction Survey

Satisfaction for daytime distance vision, nighttime distance vision, and indoors distance vision on a scale of 1 to 5, with 1 being very dissatisfied and 5 being very satisfied.

Time frame: 6 months

ArmMeasureGroupValue (MEAN)Dispersion
AcrySof Toric T3Patient Satisfaction SurveyDaytime Distance Vision4.6 Units on a ScaleStandard Deviation 0.6
AcrySof Toric T3Patient Satisfaction SurveyNighttime Distance Vision4.4 Units on a ScaleStandard Deviation 0.7
AcrySof Toric T3Patient Satisfaction SurveyIndoors Distance Vision4.6 Units on a ScaleStandard Deviation 0.7

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026