Coronary Artery Disease, Myocardial Ischemia
Conditions
Keywords
Coronary Angioplasty, Coronary Stents, Scoring Balloons, Coronary Artery Bifurcation Lesions
Brief summary
The purpose of this study is to evaluate a new angioplasty catheter, AngioSculpt® for the treatment of bifurcation lesions (blockages occurring at branch points) in coronary arteries.
Detailed description
Background: Bifurcation lesions, which consist of a narrowing occurring at branch points of coronary arteries, typically involve both the main branch (parent vessel) and an adjacent side branch. These lesions pose a particularly challenging situation for angioplasty procedures due to the difficulty of covering both branches with stents and a higher rate of recurrence (restenosis). A new angioplasty scoring balloon catheter (AngioSculpt®) has recently been approved for the treatment of narrowings in coronary arteries. The AngioSculpt® catheter incorporates a nitinol device that consists of spiral wires that wrap around the balloon catheter. As the balloon inflates, the spiral wires score the lesion allowing the balloon to be more stable (avoid slippage) and may enlarge the narrowed sections of the artery with less pressure or risk of dissection (uncontrolled tearing of the inner lining of the artery wall commonly seen with conventional balloons). Study Purpose: To demonstrate the safety and efficacy of the AngioSculpt® used in conjunction with coronary stents (implantable wire mesh tubes for scaffolding blocked arteries) for the treatment of coronary artery bifurcation narrowings and to compare these results with the historical outcomes associated with the use of conventional balloons and stents in the treatment of bifurcation lesions. Study Design: A prospective, multi-center, non-randomized, single-arm study with results compared to a literature search derived historical control for conventional balloon angioplasty (OPC - Objective Performance Criteria). The intent of this study is to enroll and treat 100 patients at 8 U.S. interventional cardiology programs with clinical follow-up planned at 30 days and 9 months following the procedure.
Interventions
Use of a stent for the main branch of a bifurcation lesion and use of a scoring balloon for the side branch of the same bifurcation lesion
Treatment of the side-branch vessel of true bifurcation lesions using the AngioSculpt device
Sponsors
Study design
Eligibility
Inclusion criteria
* At least 18 years of age and able to give informed consent. * Patients with significant (\> 50% diameter stenosis) native coronary artery disease involving a bifurcation and the ostium of the side branch vessel (Medina class (x, x, 1)) including stable or unstable angina and silent ischemia. * Patients with lesions suitable for percutaneous coronary intervention (PCI).
Exclusion criteria
* Concomitant use of Rotablator, Cutting Balloon, or investigational coronary devices. * Additional planned coronary interventions for a non-target lesion within 9 months of the study procedure. * Left ventricular ejection fraction \< 35% * Patients refusing or not candidates for emergency coronary artery bypass grafting (CABG)surgery * Uncontrolled severe hypertension (systolic BP \>180 mm Hg or diastolic BP \>110 mm Hg) * Patients who are not candidates for chronic treatment with aspirin or Clopidogrel/Ticlopidine * Severe renal failure with creatinine \>2.0 mg/dL * Untreated pre-procedural hemoglobin \<10 g/dL * Coagulopathy manifested by platelet count \<100,000 or International Normalized ratio (INR) \>2.0 (INR is only required in patients who have taken warfarin within 2 weeks of enrollment) * Women who are known or suspected to be pregnant * Patients in cardiogenic shock * Acute myocardial infarction (MI) within the past 72 hours, and/or elevated CPK (and abnormal Troponin-I) at the time of enrollment * Patients with a life expectancy of less than 1 year * Target main branch vessel \< 2.5 mm in diameter * Target main branch lesion \> 30 mm in length * Intended use of a bare metal stent (BMS) in the main branch * Target side branch vessel \< 2.0 mm in diameter * Target side branch lesion \> 15 mm in length * Target bifurcation angle \> 90º (distal angle) * Totally obstructed target coronary arteries (TIMI 0 or 1 flow) * Target bifurcation lesion within a previously placed stent (i.e. in-stent restenosis (ISR)) * Target bifurcation lesion within a surgical conduit (e.g. saphenous vein or internal mammary) * Target lesion demonstrating severe dissection prior to planned deployment of the AngioSculpt device * Unprotected Left Main diameter stenosis ≥ 50% * Visible thrombus (by angiography) at target lesion site * Coronary spasm in the absence of a significant stenosis * Patients who are concurrently participating in an investigational study when such participation could confound the treatment or outcomes of this study
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Procedural Success | 1 day | Defined as less than or equal to 30% diameter stenosis in the main branch and less than or equal to 70% diameter stenosis in the side branch at the conclusion of the procedure (including adjunctive stenting) in the absence of in-hospital major adverse cardiac events (MACE) \[cardiac death, myocardial infarction (MI), or target lesion revascularization (TLR\] |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Major Adverse Cardiovascular Events | 30 days | cardiac death, myocardial infarction, or target lesion revascularization |
| Major Adverse Cardiac Events | 9 months | cardiac death, myocardial infarction, or target lesion revascularization |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| AngioSculpt Device | 93 |
| Total | 93 |
Withdrawals & dropouts
| Period | Reason | FG000 |
|---|---|---|
| Overall Study | Death | 1 |
| Overall Study | Lost to Follow-up | 1 |
Baseline characteristics
| Characteristic | AngioSculpt Device |
|---|---|
| Age, Categorical <=18 years | 0 Participants |
| Age, Categorical >=65 years | 36 Participants |
| Age, Categorical Between 18 and 65 years | 57 Participants |
| Age Continuous | 61.54 years STANDARD_DEVIATION 11.6 |
| Region of Enrollment United States | 93 participants |
| Sex: Female, Male Female | 26 Participants |
| Sex: Female, Male Male | 67 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 7 / 93 |
| serious Total, serious adverse events | 2 / 93 |
Outcome results
Procedural Success
Defined as less than or equal to 30% diameter stenosis in the main branch and less than or equal to 70% diameter stenosis in the side branch at the conclusion of the procedure (including adjunctive stenting) in the absence of in-hospital major adverse cardiac events (MACE) \[cardiac death, myocardial infarction (MI), or target lesion revascularization (TLR\]
Time frame: 1 day
Population: intention to treat
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Overall Study | Procedural Success | 91.4 percentage of participants |
Major Adverse Cardiac Events
cardiac death, myocardial infarction, or target lesion revascularization
Time frame: 9 months
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Overall Study | Major Adverse Cardiac Events | 5.4 percentage of participants |
Major Adverse Cardiovascular Events
cardiac death, myocardial infarction, or target lesion revascularization
Time frame: 30 days
Population: intention to treat
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Overall Study | Major Adverse Cardiovascular Events | 1.1 percentage of particpants |