Ease of Use and Blood Clotting for the Gambro Polyflux HD-C4 Big Dialyzer

The Gambro Polyflux HD-C4 Big Ease Of Use And Thrombogenicity Study

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT00659724
Enrollment
33
Registered
2008-04-16
Start date
2008-03-31
Completion date
2008-05-31
Last updated
2025-07-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Chronic Kidney Failure

Keywords

Dialyzer

Brief summary

The purpose of this study is to collect information on the ease of use and thrombogenicity of the Polyflux HD-C4 Big and Fresenius Optiflux 180NR or 200NR dialyzers under conditions of routine clinical use for hemodialysis.

Detailed description

This ease of use study involves the use of FDA 510(k) cleared hemodialyzers and will have no impact on the patient's routine dialysis therapy. This study is aimed at obtaining feedback from the nursing and technical staff regarding the use and thrombogenicity of the Polyflux HD-C4 Big, and Fresenius Optiflux 180NR or 200NR dialyzers. The dialysis nursing staff will complete ease of use and thrombogenicity assessments following routine dialysis therapy. Any adverse events will be tracked for the study patients.

Interventions

None listed

Sponsors

Advanced Dialysis Center of Potomac
CollaboratorINDUSTRY
Advanced Dialysis Center of Easton
CollaboratorINDUSTRY
Gambro Renal Products, Inc.
CollaboratorINDUSTRY
Baxter Healthcare Corporation
CollaboratorINDUSTRY
Vantive Health LLC
Lead SponsorINDUSTRY

Study design

Observational model
OTHER
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Hemodialysis patients 18 years of age or older

Exclusion criteria

* n/a

Design outcomes

Primary

MeasureTime frameDescription
Ease of Use: Priming Blood SidePriming at each treatmentFor every study treatment/dialyzer, the nursing staff rated the ease of priming blood side as follows: Very Easy:Air is easily removed without knocking or clamping procedures. Acceptable:Knocking and/or clamping required for efficient air removal. Difficult:Knocking and/or clamping and additional volume of saline or extended recirculation needed to remove air. Very Difficult:Air could not be removed.
Ease of Use: Priming Dialysate SidePriming at each treatmentFor every study treatment/dialyzer, the nursing staff rated the ease of priming dialysate side as follows: Dialysate Side at 500 ml Priming Volume (check one): Perfect: No air visible. Acceptable: Some air visible, but considered insignificant. Not Acceptable: Additional actions needed to sufficiently remove air.
Dialyzer Assessment: Fiber ConditionEach treatment: assessment of the condition of the dialyzer fibers after rinse-backFor every study treatment/dialyzer, the nursing staff rated the condition of the dialyzer fibers after rinse-back as follows: Very Good:All fibers appear white. Good:Few fibers (less than 10) appear PINK / RED. (check one) Poor:Several fibers (more than 10) appear PINK / RED. (check one) Very Poor:Most fibers (more than 75%) appear PINK / RED. (check one)
Dialyzer Assessment: Arterial Header ConditionEach treatment: assessment of the condition of the arterial header after rinse-backFor every study treatment/dialyzer, the nursing staff rated the condition of the arterial header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.
Dialyzer Assessment: Venous Header ConditionEach treatment: the assessment of the condition of the venous header after rinse-backFor every study treatment/dialyzer, the nursing staff rated the condition of the venous header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.

Countries

United States

Participant flow

Participants by arm

ArmCount
Hemodialysis Patients
Hemodialysis patients 18 years of age or older receiving routine dialysis treatment for chronic renal failure.
33
Total33

Baseline characteristics

CharacteristicHemodialysis Patients
Age, Continuous66.0 years
STANDARD_DEVIATION 11.1
Height65.6 inches
STANDARD_DEVIATION 4.3
Sex: Female, Male
Female
18 Participants
Sex: Female, Male
Male
15 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 33
serious
Total, serious adverse events
0 / 33

Outcome results

Primary

Dialyzer Assessment: Arterial Header Condition

For every study treatment/dialyzer, the nursing staff rated the condition of the arterial header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.

Time frame: Each treatment: assessment of the condition of the arterial header after rinse-back

ArmMeasureGroupValue (NUMBER)
Revaclear MAXDialyzer Assessment: Arterial Header ConditionGood49 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Arterial Header ConditionVery Poor1 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Arterial Header ConditionVery Good155 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Arterial Header ConditionPoor13 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Arterial Header ConditionPoor26 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Arterial Header ConditionVery Good0 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Arterial Header ConditionVery Poor3 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Arterial Header ConditionGood66 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Arterial Header ConditionVery Poor2 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Arterial Header ConditionGood64 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Arterial Header ConditionVery Good0 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Arterial Header ConditionPoor30 Number of Dialyzers
Primary

Dialyzer Assessment: Fiber Condition

For every study treatment/dialyzer, the nursing staff rated the condition of the dialyzer fibers after rinse-back as follows: Very Good:All fibers appear white. Good:Few fibers (less than 10) appear PINK / RED. (check one) Poor:Several fibers (more than 10) appear PINK / RED. (check one) Very Poor:Most fibers (more than 75%) appear PINK / RED. (check one)

Time frame: Each treatment: assessment of the condition of the dialyzer fibers after rinse-back

ArmMeasureGroupValue (NUMBER)
Revaclear MAXDialyzer Assessment: Fiber ConditionVery Good126 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Fiber ConditionGood75 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Fiber ConditionPoor12 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Fiber ConditionVery Poor5 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Fiber ConditionVery Poor11 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Fiber ConditionVery Good23 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Fiber ConditionPoor17 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Fiber ConditionGood44 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Fiber ConditionVery Poor7 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Fiber ConditionGood45 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Fiber ConditionPoor30 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Fiber ConditionVery Good15 Number of Dialyzers
Primary

Dialyzer Assessment: Venous Header Condition

For every study treatment/dialyzer, the nursing staff rated the condition of the venous header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.

Time frame: Each treatment: the assessment of the condition of the venous header after rinse-back

ArmMeasureGroupValue (NUMBER)
Revaclear MAXDialyzer Assessment: Venous Header ConditionPoor7 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Venous Header ConditionVery Poor5 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Venous Header ConditionGood140 Number of Dialyzers
Revaclear MAXDialyzer Assessment: Venous Header ConditionVery Good65 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Venous Header ConditionVery Good0 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Venous Header ConditionGood31 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Venous Header ConditionVery Poor11 Number of Dialyzers
Optiflux F180NRDialyzer Assessment: Venous Header ConditionPoor52 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Venous Header ConditionVery Poor5 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Venous Header ConditionVery Good0 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Venous Header ConditionGood35 Number of Dialyzers
Optiflux F200NRDialyzer Assessment: Venous Header ConditionPoor57 Number of Dialyzers
Primary

Ease of Use: Priming Blood Side

For every study treatment/dialyzer, the nursing staff rated the ease of priming blood side as follows: Very Easy:Air is easily removed without knocking or clamping procedures. Acceptable:Knocking and/or clamping required for efficient air removal. Difficult:Knocking and/or clamping and additional volume of saline or extended recirculation needed to remove air. Very Difficult:Air could not be removed.

Time frame: Priming at each treatment

ArmMeasureGroupValue (NUMBER)
Revaclear MAXEase of Use: Priming Blood SideVery Easy139 Number of Dialyzers
Revaclear MAXEase of Use: Priming Blood SideAcceptable80 Number of Dialyzers
Optiflux F180NREase of Use: Priming Blood SideVery Easy13 Number of Dialyzers
Optiflux F180NREase of Use: Priming Blood SideAcceptable82 Number of Dialyzers
Optiflux F200NREase of Use: Priming Blood SideVery Easy13 Number of Dialyzers
Optiflux F200NREase of Use: Priming Blood SideAcceptable84 Number of Dialyzers
Primary

Ease of Use: Priming Dialysate Side

For every study treatment/dialyzer, the nursing staff rated the ease of priming dialysate side as follows: Dialysate Side at 500 ml Priming Volume (check one): Perfect: No air visible. Acceptable: Some air visible, but considered insignificant. Not Acceptable: Additional actions needed to sufficiently remove air.

Time frame: Priming at each treatment

ArmMeasureGroupValue (NUMBER)
Revaclear MAXEase of Use: Priming Dialysate SidePerfect131 Number of Dialyzers
Revaclear MAXEase of Use: Priming Dialysate SideAcceptable89 Number of Dialyzers
Optiflux F180NREase of Use: Priming Dialysate SidePerfect4 Number of Dialyzers
Optiflux F180NREase of Use: Priming Dialysate SideAcceptable91 Number of Dialyzers
Optiflux F200NREase of Use: Priming Dialysate SidePerfect2 Number of Dialyzers
Optiflux F200NREase of Use: Priming Dialysate SideAcceptable95 Number of Dialyzers